MODIFICATION TO: SENS-A-RAY SYSTEM WITH PROIMAGE SOFTWARE

K013176 · Afp Imaging Corp. · EHD · Oct 16, 2001 · Dental

Device Facts

Record IDK013176
Device NameMODIFICATION TO: SENS-A-RAY SYSTEM WITH PROIMAGE SOFTWARE
ApplicantAfp Imaging Corp.
Product CodeEHD · Dental
Decision DateOct 16, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.1800
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The ProImage Software is intended to be used with the Sens-A-Ray system and standard digital dental X-ray systems and computer stations for system operation, archive data storage, image capture and enhancement, and patient data and support.

Device Story

Sens-A-Ray System with ProImage Software is a digital dental imaging software package. It interfaces with standard X-ray systems, sensor units, and image grabbers to capture, process, store, and retrieve dental X-ray images. Operated by dental professionals in a clinical setting, the software provides tools for image enhancement, including auto-contrast, colorization, and histogram viewing. It also manages patient databases and supports endodontic imaging modes. The system facilitates the transition from analog to digital workflows by allowing clinicians to view, manipulate, and archive high-quality dental images on computer stations, aiding in diagnostic evaluation and patient record management.

Clinical Evidence

Bench testing only. Verification and validation testing was performed to confirm that the modified software functions as intended.

Technological Characteristics

Operational software package for digital dental systems. Functions include image capture, enhancement, and database management. Compatible with standard X-ray equipment, sensor units, and intraoral cameras. Software features include auto-contrast, colorization, histogram viewing, and endodontic mode. No specific hardware materials or connectivity standards (e.g., DICOM) are detailed beyond general compatibility with computer stations.

Indications for Use

Indicated for patients receiving routine dental radiography using digital dental X-ray systems.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## Special 510(k) Summary ### for Koi 3176 # Sens-A-Ray System with ProImage Software - SPONSOR 1. AFP Imaging Corporation 250 Clearbrook Road Elmsford, NY 10523 Contact Person: Tzipora Halevi 914-592-6100 Telephone: Date Prepared: September 21, 2000 #### 2. DEVICE NAME | Proprietary Name: | Sens-A-Ray Digital Dental X-Ray System with<br>ProImage Software | |----------------------|------------------------------------------------------------------| | Common/Usual Name: | Accessory to Extraoral X-Ray System | | Classification Name: | Extraoral Source X-Ray System Accessory | #### 3. PREDICATE DEVICES AFP Imaging Corporation Sens-A-Ray System K923067 #### 4. DEVICE DESCRIPTION The ProImage Software is essentially identical to the Sens-A-Ray software described in K923067. (K923067 was submitted to FDA by the former owner of AFP Imaging, Regam Medical Systems AB.) The AFP Imaging ProImage Software and Sens-A-Ray System are intended to be used in conjunction with standard X-Ray systems. X-Ray images are sent by a digitized processor to a computer screen for image presentation. As with the Sens-A-Ray Software, the ProImage Software provides and controls the following functions: {1}------------------------------------------------ - A patient/image database, which logs a patient's statistical and image data ● - Image exposure . - Image storage ● - Image processing ● - Image retrieval and post-processing ● - Image hardcopy . The modifications to the Sens-A-Ray software involve enabling video images and I ho mountread on a manipulated within the same software package. The minor modifications involve better database search and structure to allow the user to catalog patient images in an easier and more intuitive format. The other abor to cannoly feature improvements resulting from user input. The feature improvements include: - Colorizing image, auto-contrast, re-expose, and "view histogram" functions . - Importing multiple images and adjusting the size and resolution of images . - Adding an endodontic mode which allows full-screen X-rays . The modifications above were made to accommodate more advanced digital X-ray systems and to provide the user with better imaging capabilities within a single software package. These modifications remain within the scope of the intended use and fundamental technology of the original Sens-A-Ray system. The ProImage Software is intended for use with a sensor unit, image unit, and image grabber that were previously cleared under K923067. The ProImage Software is also compatible with the AFP IntraOral camera, subject of K922942 and other legally marketed intraoral cameras. #### INTENDED USE પં The ProImage Software is intended to be used with the Sens-A-Ray system and standard digital dental X-ray systems and computer stations for system operation, archive data storage, image capture and enhancement, and patient data and support. {2}------------------------------------------------ #### TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6. The AFP Imaging ProImage Digital Dental X-Ray System Software is similar in intended use and fundamental technology to the Sens-A-Ray Software, subject of K923067. The AFP ProImage Software and the predicate Sens-A-Ray Software have the same intended use. They are both intended to be used with digital dental imaging systems for the capture, evaluation, and storage of high quality digital dental X-rays using existing X-ray equipment. Both devices are indicated for patients receiving routine dental radiography. The AFP ProImage Software and the predicate device are operational software packages that support system operation, image capture, and enhancement, patient data, and support. They are both installed into digital dental systems intended for use with standard X-ray equipment. The AFP Imaging ProImage Software and the predicate devices have essentially identical performance characteristics. These devices use an extraoral source of X-rays for intraoral images in dental radiography. They provide a means for capturing, enhancing, and delivering dental images for viewing and storage #### TESTING 7. Verification and validation testing was successfully performed to confirm that the modified software functions as intended. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract symbol resembling an eagle or bird in flight. The words "DEPARTMENT OF HEALTH &" are vertically oriented and curve along the left side of the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 1 6 2001 Re: K013176 AFP Imaging Corporation % Mary-McNamara-Culline, RAC Staff Consultant Medical Device Consultants, Inc. 49 Plain Street NORTH ATTLEBORO MA 02760 Trade/Device Name: AFP Imaging Sens-A-Ray Digital Dental X-Ray System Software Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: 90 EHD Dated: September 21, 2001 Received: September 24, 2001 ### Dear Ms. McNamara-Cullinane: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): KO/3/7 Device Name: AFP Imaging Sens-A-Ray System with ProImage Software Indications for Use: The AFP Imaging Sens-A-Ray Digital Dental X-Ray System with ProImage Software is intended for use with standard digital dental X-ray systems and computer stations for intended for use with standard digital Genea. It is 'systems and enhancement, and patient data and support. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY) Concurrence of CDRH, Office of Device Evaluation (ODE) David A. Leagson (Division Sign-OM) Division of Reproductive, Abdominal, and Radiological Deviens K013176 510(k) Numbe Prescription Use ✓ OR Over-The-Counter Use (Optional Format 1-2-96) AFP Imaging Special 510(k) ProImage Software September 21, 2000
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...