AI-Rad Companion Organs RT

K232899 · Siemens Medical Solutions U.S.A. · QKB · Apr 3, 2024 · Radiology

Device Facts

Record IDK232899
Device NameAI-Rad Companion Organs RT
ApplicantSiemens Medical Solutions U.S.A.
Product CodeQKB · Radiology
Decision DateApr 3, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K232899 · Apr 3, 2024AI-Rad Companion Organs RTSiemens Medical Solutions U.S.A.Retrospective clinical CT and MR image datasets from multiple clinical sitesRetrospective performance studies were conducted using previously acquired clinical CT and MR images to validate the accuracy of the AI-Rad Companion Organs RT software in contouring organs-at-risk compared to manual ground truth annotations.Retrospective performance study; Clinical image datasets; Algorithm validation; Organs-at-risk contouring

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective MR performance study; Retrospective performance studyAdult patients (22 years and older) undergoing MR imaging; Sample Size: 66 cases; Number of Sites: Multiple clinical sites across North America & EuropeNot applicable for this studyDice coefficient, absolute symmetric surface distance (ASSD), and failure rate
Retrospective CT performance study; Retrospective performance studyPatients undergoing CT for RT treatment planning; Sample Size: 414 cases; Number of Sites: 25 clinical sites (North America, South America, Asia, Australia, and Europe)Not applicable for this studyDice coefficient and absolute symmetric surface distance (ASSD)

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Automatic organ segmentation (CT)Deep learning algorithmAverage Dice score difference < 3% compared to predicateDice coefficients ranging from 47.6% to 92.6% depending on organTraining: 1,312 datasets (Cochlea: 215, Thyroid: 293, Constrictor Muscles: 335, Chest Wall: 48, LN Supraclavicular/Axilla/Internal Mammaries: 228, Duodenum/Bowels/Sigmoid: 332, Stomach: 371, Pancreas: 369, Pulmonary Artery/Vena Cava/Trachea/Spinal Canal/Proximal Bronchus: 113, Ventricles & Atriums: 706, Descending Coronary Artery: 252, Penile Bulb: 854, Uterus: 381)>1 (experienced annotators) + 1 (board-certified radiation oncologist)Retrospective performance study: 414 cases (Cohort A: 73, Cohort B: 40, Cohort C: 301)>1 (experienced annotators) + 1 (board-certified radiation oncologist)
Automatic organ segmentation (MR)Deep learning algorithmAverage Dice value equivalent or better than predicate; Fail rate < 15%Average Dice 85.75%; Fail rate 2.75%Training: 1,404 datasets (T1 VIBE/Dixon W: 219, T2 W TSE: 225, Prostate (T2W): 960)>1 (experienced annotators) + 1 (board-certified radiation oncologist)Retrospective performance study: 66 cases>1 (experienced annotators) + 1 (board-certified radiation oncologist)

Indications for Use

AI-Rad Companion Organs RT is a post-processing software intended to automatically contour DICOM CT and MR predefined structures using deep-leaming-based algorithms. Contours that are generated by AI-Rad Companion Organs RT may be used as input for clinical workflows including external beam radiation therapy treatment planning. AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by AI-Rad Companion Organs RT. The output of AI-Rad Companion Organs RT are intended to be used by trained medical professionals. The software is not intended to automatically detect or contour lesions.

Device Story

Post-processing software for automatic segmentation of organs-at-risk (OAR) in CT or MR images; intended for radiation therapy treatment planning. Input: DICOM CT/MR series. Operation: Deep-learning-based algorithms generate contours (DICOM-RTSTRUCT format). Workflow: Software processes images; results sent to confirmation UI for review/rejection or directly to configurable DICOM node (e.g., Treatment Planning System). Usage: Clinical environment; operated by trained medical professionals. Benefit: Streamlines contouring workflow; provides initial segmentation for manual review/editing by clinicians. Not a diagnostic device; not for clinical decision-making.

Clinical Evidence

No clinical trials conducted. Performance validated via retrospective bench testing on 414 CT cases and 66 MR cases. Metrics: Dice coefficient, absolute symmetric surface distance (ASSD), and fail rate. CT performance: Dice scores for new organs comparable to literature/cleared devices; existing organs within 3% of predicate. MR performance: Dice scores equivalent or superior to predicate/reference devices.

Technological Characteristics

Software-based medical image management and processing system. Deep-learning-based segmentation algorithms. Inputs: DICOM CT/MR. Outputs: DICOM-RTSTRUCT. Deployment: Edge and cloud-based. Web-based UI. Locked algorithms. Compliant with DICOM PS 3.1-3.20, IEC 62304, ISO 14971, IEC 62366-1.

Indications for Use

Indicated for adult patients requiring radiation therapy treatment planning. Used for automatic contouring of predefined organs-at-risk (OAR) on CT or MR images. Not intended for lesion detection or contouring. Must be used by trained medical professionals.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ April 3, 2024 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Siemens Medical Solutions U.S.A. % Kira Morales Regulatory Affairs Manager 40 Liberty Blvd. MALVERN, PA 19355 Re: K232899 Trade/Device Name: AI-Rad Companion Organs RT Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QKB Dated: March 1, 2024 Received: March 4, 2024 Dear Kira Morales: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming {1}------------------------------------------------ product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Loran Weidner Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K232899 Device Name AI-Rad Companion Organs RT #### Indications for Use (Describe) AI-Rad Companion Organs RT is a post-processing software intended to automatically contour DICOM CT and MR predefined structures using deep-leaming-based algorithms. Contours that are generated by AI-Rad Companion Organs RT may be used as input for clinical workflows including external beam radiation therapy treatment planning. AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by AI-Rad Companion Organs RT. The output of AI-Rad Companion Organs RT are intended to be used by trained medical professionals. The software is not intended to automatically detect or contour lesions. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots. # 510(k) SUMMARY FOR AI-Rad Companion Organs RT Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Date Prepared: April 2, 2024 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990 and 21 CFR §807.92. ### 1. Submitter | Importer/Distributor | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Malvern, PA 19355<br>Mail Code: 65-3<br>Registration Number: 2240869 | |----------------------|-------------------------------------------------------------------------------------------------------------------------------------| | Manufacturing Site | Siemens Healthcare GmbH<br>Henkestrasse 127<br>Erlangen, Germany 91052<br>Registration Number: 3002808157 | # 2. Contact Person Kira Morales Senior Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19335 Phone: +1 (484) 901 - 9471 Email: kira.morales@siemens-healthineers.com ### 3. Device Name and Classification | Product Name: | AI-Rad Companion Organs RT | |----------------------|------------------------------------------------| | Common Name: | Medical Imaging Software | | Classification Name: | Medical Image Management and Processing System | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots. Classification Panel: CFR Section: Device Class: Product Code: ### 4. Predicate Device Product Name: Common Name: 510(k) Number: Clearance Date: Classification Name: Classification Panel: CFR Section: Device Class: Primary Product Code: Recall Information: # 5. Reference Devices Product Name: Common Name: 510(k) Number: Clearance Date: Classification Name: Classification Panel: CFR Section: Device Class: Primary Product Code: #### Product Name: Common Name: 510(k) Number: Clearance Date: Classification Name: Classification Panel: CFR Section: Device Class: Primary Product Code: #### Product Name: Common Name: 510(k) Number: Clearance Date: Classification Name: Radiology 21 CFR §892.2050 Class II OKB AI-Rad Companion Organs RT Medical Imaging Software K221305 October 14, 2022 Medical Image Management and Processing System Radiology 21 CFR §892.2050 Class II OKB N/A #### MRCAT Pelvis Medical Imaging Software K182888 April 30, 2019 Medical charged-particle radiation therapy system Radiology 21 CFR §892.2050 Class II MUJ #### RT Image Suite Medical Imaging Software K220783 September 7, 2022 Medical charged-particle radiation therapy system Radiology 21 CFR §892.2050 Class II MUJ #### Contour Protégé AI Medical Imaging Software K231765 November 8, 2023 Medical image management and processing system {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. Classification Panel: Radiology 21 CFR §892.2050 CFR Section: Device Class: Class II Primary Product Code: OKB Product Name: AI Segmentation Common Name: Medical Image Segmentation Software 510(k) Number: K211881 Clearance Date: Classification Name: Classification Panel: CFR Section: Device Class: Primary Product Code: September 2, 2021 Medical charged-particle radiation therapy system Radiology 21 CFR §892.2050 Class II MUJ # 6. Indications for Use AI-Rad Companion Organs RT is a post-processing software intended to automatically contour DICOM CT and MR pre-defined structures using deep-learning-based algorithms. Contours that are generated by AI-Rad Companion Organs RT may be used as input for clinical workflows including external beam radiation therapy treatment planning. AI-Rad Companion Organs RT must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by AI-Rad Companion Organs RT. The outputs of AI-Rad Companion Organs RT are intended to be used by trained medical professionals. The software is not intended to automatically detect or contour lesions. # 7. Device Description AI-Rad Companion Organs RT provides automatic segmentation of pre-defined structures such as Organs-at-risk (OAR) from CT or MR medical series, prior to dosimetry planning in radiation therapy. AI-Rad Companion Organs RT is not intended to be used as a standalone diagnostic device and is not a clinical decision-making software. CT or MR series of images serve as input for AI-Rad Companion Organs RT and are acquired as part of a typical scanner acquisition. Once processed by the AI algorithms, generated contours in DICOM-RTSTRUCT format are reviewed in a confirmation window, allowing clinical user to confirm or reject the contours before sending to the target system. Optionally, the user may select to directly transfer the contours to a configurable DICOM node (e.g., the TPS, which is the standard location for the planning of radiation therapy). {6}------------------------------------------------ Image /page/6/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots arranged in a circular pattern. The output of AI-Rad Companion Organs RT must be reviewed and, where necessary, edited with appropriate software before accepting generated contours as input to treatment planning steps. The output of AI-Rad Companion Organs RT is intended to be used by qualified medical professionals. The qualified medical professional can perform a complementary manual editing of the contours or add any new contours in the TPS (or any other interactive contouring application supporting DICOM-RT objects) as part of the routine clinical workflow. # 8. Substantially Equivalent (SE) and Technological Characteristics The indented use of the predicate device and the subject device are equivalent. The two main difference compared to the predicate, AI-Rad Companion Organs RT (K221305): - MR Contouring Algorithms ● - o T1 Model for 3 target organs - o T2 Model for 6 target organs - . Update CT contouring algorithm for 38 new organs AI-Rad Companion Organs RT VA50 and AI-Rad Companion Organs RT VA40 both use a deep learning algorithm to support their AI claims. Additionally, they both process CT data in DICOM format, making them vendor agnostic and create outputs which can be used by any TPS system. The MR contouring has only been validated for Siemens Healthineers' scanner data. The deep learning CT algorithm within AI-Rad Companion Organs RT VA50 has been enhanced from the algorithm in AI-Rad Companion Organs RT VA40 (K221305). All models contained within AI-Rad Companion Organs RT VA50 and AI-Rad Companion Organs RT VA40 (K221305) are locked and cannot be modified by the user. The subject device, AI-Rad Companion Organs RT, is substantially equivalent with regards to the software features, functionalities, and core algorithms. The performance of the new MR contouring algorithm within AI-Rad Companion Organs RT VA50 is comparable to the algorithm in MRCAT Pelvis (K182888). The performance of the new CT contouring algorithms have been validated against FDA/CE cleared devices or from literature. The risk analysis and non-clinical data support that both devices perform equivalently and do not raise different questions of the safety and effectiveness. | | Subject Device | Predicate Device | Reference Device | |-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device<br>Manufacturer | Siemens | Siemens | Philips Medical<br>Systems MR Finland | | Device Name | AI-Rad Companion<br>Organs RT | AI-Rad Companion<br>Organs RT | MRCAT Pelvis | | 510(k) Number | TBD | K221305 | K182888 | | processing software | processing software | Ingenia 1.5T and 3.0T | | | intended to | intended to | MR systems. Intended | | | automatically contour | automatically contour | Use: MRCAT imaging | | | DICOM CT and MR | DICOM CT imaging | is intended to provide | | | pre-defined structures | data using deep- | the operator with | | | using deep-learning- | learning-based | information of tissue | | | based algorithms. | algorithms. | properties for radiation | | | Contours that are | Contours that are | attenuation estimation | | | generated by AI-Rad | generated by AI-Rad | purposes in photon | | | Companion Organs RT | Companion Organs RT | external beam | | | may be used as input | may be used as input | radiotherapy treatment | | | for clinical workflows | for clinical workflows | planning. Indications | | | including external | including external | for use: MRCAT | | | beam radiation therapy | beam radiation therapy | Pelvis is indicated for | | | treatment planning. | treatment planning. AI- | radiotherapy treatment | | | AI-Rad Companion | Rad Companion | planning of soft tissue | | | Organs RT must be | Organs RT must be | cancers in the pelvic | | | used in conjunction | used in conjunction | region. | | | with appropriate | with appropriate | | | | software such as | software such as | | | | Treatment Planning | Treatment Planning | | | | Systems and | Systems and | | | | Interactive Contouring | Interactive Contouring | | | | applications, to review, | applications, to review, | | | | edit, and accept | edit, and accept | | | | contours generated by | contours generated by | | | | AI-Rad Companion | AI-Rad Companion | | | | Organs RT. | Organs RT. | | | | The outputs of AI-Rad | The output of AI-Rad | | | | Companion Organs RT | Companion Organs RT | | | | are intended to be used | in the format of | | | | by trained medical | RTSTRUCT objects | | | | professionals. | are intended to be used | | | | The software is not | by trained medical | | | | intended to | professionals. | | | | automatically detect or | The software is not | | | | contour lesions | intended to | | | | | automatically detect or | | | | | contour lesions. Only | | | | | DICOM images of | | | | Algorithm | | | | | Segmentation of Organ at Risk in the Anatomic Regions | Deep Learning | Deep Learning | Machine-learning | | Compatible Modality | CT & MR Images | CT Images | MR | | Compatible Scanner Models | CT: Head & Neck,<br>Thorax, Abdomen & Pelvis<br>Head & Neck lymph nodes<br>(166 OAR)<br>MR: Pelvis (9 OAR) | CT: Head & Neck,<br>Thorax, Abdomen & Pelvis<br>Head & Neck lymph nodes<br>(128 OAR) | Male and female pelvis, with soft-tissue cancer in the anatomical pelvic region below L1 vertebra, including (post-operative) prostate, rectum, anus, bladder and cervix | | Compatible Treatment Planning System | No Limitation on scanner model for CT.<br>Siemens Healthineers' data only for MR.<br>DICOM compliance required. | No Limitation on scanner model,<br>DICOM compliance required. | Ingenia 1.5T and 3.0T MR-RT, Ingenia Ambition 1.5T MR-RT and Ingenia Elition 3.0T MR-RT | | Contraindications | No Limitation on TPS model, DICOM compliance required. | No Limitation on TPS model, DICOM compliance required. | No Limitation on TPS model, DICOM compliance required. | | Target Population | Adult use only | Adult use only | Adult use only | | | AI-Rad Companion Organs RT is designed for use only in adult populations.<br>AI-Rad Companion Organs RT is designed for any patient for whom relevant modality scans are available. | AI-Rad Companion Organs RT is designed for use only in adult populations.<br>AI-Rad Companion Organs RT is designed for any patient for whom relevant modality scans are available. More specifically, the software is validated on previously acquired CT DICOM volumes for radiation therapy treatment planning, including, head and | No information publicly available | | | adult patients are considered to be valid input. | | | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots. {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots. | | neck, thorax, abdomen, and pelvis. | | | |---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------| | Clinical<br>condition the<br>device is<br>intended to<br>diagnose, treat or<br>manage | Limited to patients<br>previously selected for<br>Radiation Therapy. | Limited to patients<br>previously selected for<br>Radiation Therapy. | No information<br>publicly available | | Software<br>Architecture | AI-Rad Companion<br>(Engine) architecture<br>enabling the<br>deployment of AI Rad<br>Companion Organs RT<br>using Edge and in the<br>Cloud. The UI is<br>provided using a web-<br>based interface. | AI-Rad Companion<br>(Engine) architecture<br>enabling the<br>deployment of AI Rad<br>Companion Organs RT<br>using Edge and in the<br>Cloud. The UI is<br>provided using a web-<br>based interface. | No information<br>publicly available | | Deployment<br>Feature | Edge & Cloud<br>Deployment | Edge & Cloud<br>Deployment | No information<br>publicly available | | Organ Templates | Creating, editing and<br>deletion of organ<br>templates. Customize<br>predefined structure<br>database with mapping<br>to international<br>nomenclature schemes. | Creating, editing and<br>deletion of organ<br>templates. Customize<br>predefined structure<br>database with mapping<br>to international<br>nomenclature schemes. | No information<br>publicly available | | Automated<br>workflow | AI-Rad Companion<br>Organs RT<br>automatically<br>processes input image<br>data and sends the<br>results as DICOM-RT<br>Structure Sets to a<br>user-configurable<br>target node. | AI-Rad Companion<br>Organs RT<br>automatically<br>processes input image<br>data and sends the<br>results as DICOM-RT<br>Structure Sets to a<br>user-configurable<br>target node. | Automatic contouring | | Contour<br>visualization and<br>editing feature | AI-Rad Companion<br>Organs RT provides<br>basic result preview of<br>automatic<br>segmentation results,<br>and no editing feature<br>of the automatic<br>segmented contour. | AI-Rad Companion<br>Organs RT provides<br>basic result preview of<br>automatic<br>segmentation results,<br>and no editing feature<br>of the automatic<br>segmented contour. | No information<br>publicly available | | Segmentation<br>Performance | MR: The target<br>performance was<br>validated using 66<br>cases to validate the<br>overall performance of<br>the MR contouring.<br>CT: The target<br>performance was<br>validated using 414<br>cases distributed to<br>three cohorts.<br>Both: To objectively<br>evaluate the target<br>performance, the DICE<br>coefficient, the<br>absolute symmetric<br>surface distance<br>(ASSD) and the fail<br>rate was evaluated.<br>The segmentation<br>performance of the<br>subject was equivalent<br>to the overall<br>performance compared<br>to the predicate,<br>reference device and<br>comparable literature<br>& devices. | The target performance<br>was validated using<br>157 cases distributed to<br>two cohorts. Cohort A<br>is clinical routine<br>treatment planning CT<br>and it is split into two<br>sub-cohort and Cohort<br>B is PET-CT data. To<br>objectively evaluate<br>the target performance,<br>the DICE coefficient,<br>the absolute symmetric<br>surface distance<br>(ASSD) and the fail<br>rate was evaluated.<br>The segmentation<br>performance of the<br>subject and reference<br>device were equivalent<br>as well as the overall<br>performance compared<br>to the predicate device. | The mean and standard<br>deviation Dice<br>coefficients, along with<br>the lower 95th<br>percentile confidence<br>bound were calculated. | | User Interface -<br>Results Preview<br>(Confirmation) | Basic visualization<br>functionality of<br>original data and<br>generated contours | Basic visualization<br>functionality of<br>original data and<br>generated contours | No information<br>publicly available | | User Interface<br>Configuration | Configuration UI | Configuration UI | No information<br>publicly available | | Automated<br>Workflow to TPS | Results send to<br>Confirmation UI &<br>Optional bypassing of<br>Confirmation UI to<br>TPS | Results send to<br>Confirmation UI &<br>Optional bypassing of<br>Confirmation UI to<br>TPS | No information<br>publicly available | | Human Factors | Design to be used by<br>trained clinicians. | Design to be used by<br>trained clinicians. | Designed to be used by<br>trained clinicians | {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is an orange graphic of a series of dots. Table 1: Indications for Use and Segmentation Feature Comparison {11}------------------------------------------------ #### SIEME Healthineers The conclusions from all verification and validation data suggests that these enhancements are equivalent with respect to safety and effectiveness of the predicate device. These modifications do not change the intended use of the product. Siemens is of opinion that AI-Rad Companion Organs RT VA50 is substantially equivalent to the currently marketed device, AI-Rad Companion Organs RT (K221305). # 9. Nonclinical Tests Non-clinical tests were conducted to test the functionality of AI-Rad Companion Organs RT. Software validation and bench testing have been conducted to assess the performance claims as well as the claim of substantial equivalence to the predicate device. AI-Rad Companion has been tested to meet the requirements of conformity to multiple industry standards. Non-clinical performance testing demonstrates that AI-Rad Companion Organs RT complies with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005) as well as with the following voluntary FDA recognized Consensus Standards listed in Table 2. | Recognition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>Date | Standards<br>Development<br>Organization | |-----------------------|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|------------------------------------------| | 5-129 | General | Medical Devices –<br>Application of usability<br>engineering to medical<br>devices | 62366-1 Ed<br>1.1 2020-06<br>CV | IEC | | 5-125 | General | Medical Devices –<br>application of risk<br>management to medical<br>devices | 14971:2019-<br>12 | ISO | | 13-79 | Software/<br>Informatics | Medical device software –<br>software life cycle<br>processes [Including<br>Amendment 1 (2016)] | 62304 Ed 1.1<br>2015-06 CV | AAMI<br>ANSI<br>IEC | | 12-349 | Radiology | Digital Imaging and<br>Communications in<br>Medicine (DICOM) Set | PS 3.1 – 3.20<br>2022d | NEMA | | 5-134 | General | Medical devices – symbols<br>to be used with information<br>to be supplied by the<br>manufacturer - Part 1:<br>General Requirements | 15223-1<br>Fourth edition<br>2021-07 | ISO<br>IEC | | 13-97 | Software/<br>Informatics | Health software – Part 1:<br>General requirements for<br>product safety | 82304-1<br>Edition 1.0<br>2016-10 | IEC | Table 2: List of recognized standards ### Verification and Validation {12}------------------------------------------------ Image /page/12/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a pattern. Software documentation level, per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on June 14, 2023, is also included as part of this…
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...