K222036 · Medipixel, Inc. · QIH · Mar 22, 2023 · Radiology
Device Facts
Record ID
K222036
Device Name
MPXA-2000
Applicant
Medipixel, Inc.
Product Code
QIH · Radiology
Decision Date
Mar 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K222036 · Mar 22, 2023
MPXA-2000
Medipixel, Inc.
Retrospective coronary angiograms (305 images) from Philips Medical Systems
Retrospective clinical angiographic images were used to conduct standalone performance testing to validate the device's vessel segmentation, vessel classification, and ROI analysis algorithms.
MPXA-2000 is software intended to be used for performing calculations in X-ray angiographic images of the coronary arteries. These calculations are based on vessel contours which are automatically detected by the software and subsequently presented for review and manual editing. The analysis results obtained with MPXA-2000 are intended for use by cardiologists and radiologists: - to support clinical decisions concerning the coronary arteries. - to support the evaluation of intervention or drug therapy applied for conditions of the coronary arteries.
Device Story
MPXA-2000 is a standalone cardiovascular image analysis software for viewing and quantifying X-ray angiographic images of coronary arteries. It accepts DICOM-format X-ray angiographic images as input. The device utilizes deep-learning (for vessel classification and segmentation) and computer vision algorithms to automatically detect vessel contours, classify vessel types (LAD, LCX, RCA), and calculate vessel dimensions (e.g., minimum lumen diameter, % diameter stenosis, reference diameter) in real-time. The software allows for manual editing of automated results. It is intended for use by cardiologists and radiologists in clinical settings. Output is displayed on-screen and can be exported as PDF or Excel files. By providing reproducible quantitative data, the device supports clinical decision-making and treatment evaluation for patients with coronary artery disease.
Clinical Evidence
No clinical data. Bench testing only. Standalone performance testing used 305 retrospective coronary angiograms from the US population. Primary endpoints included vessel segmentation (Dice similarity coefficient 0.9200), vessel classification (accuracy 0.9934), and ROI analysis (minimum luminal diameter estimation). Results met pre-specified pass criteria.
Technological Characteristics
Standalone software; DICOM input; deep-learning (segmentation/classification) and computer vision (measurement) algorithms. Features: automatic vessel segmentation, vessel type classification, ROI quantification, manual editing tools. Complies with IEC 62304:2015. Cybersecurity controls implemented. Vendor-independent compatibility with FDA-approved X-ray angiography systems (Philips, Siemens, GE).
Indications for Use
Indicated for use in clinical settings where validated and reproducible quantified results are needed to support calculations in X-ray angiographic images of coronary arteries, for adult patients with coronary artery disease (CAD). Results are intended for use by clinicians to support clinical decision-making or treatment evaluation and are not the sole basis for diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Medipixel, Inc. % Hye-ri Choi RA Manager Dooam Building 7F, 61, Yangwha-ro, Mapo-gu Seoul. 04037 REPUBLIC OF KOREA
Re: K222036
March 22, 2023
Trade/Device Name: MPXA-2000 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QIH Dated: February 21, 2023 Received: February 23, 2023
Dear Hye-ri Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222036
Device Name MPXA-2000
#### Indications for Use (Describe)
MPXA-2000 is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the calculations in X-ray angiographic images of the coronary arteries, for use on individual patients with coronary artery disease (CAD). MPXA-2000 is indicated for use in adult patient only.
When the quantified results provided by MPXA-2000 are used in a clinical setting on X-ray images of an individual patient, they can be used to support the clinical decisions making for the patient or the evaluation of the treatment applied. In this case, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical diagnosis, and they are only intended for use by the responsible clinicians.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Medipixel. The logo consists of a colorful icon to the left of the word "medipixel" in gray. The icon is made up of four different colored sections: purple, light blue, dark blue, and orange.
510(k) Summary MPXA-2000 K222036
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
# 1) DATE PREPARED [21 CFR 807.92(a)(1)]
March 15, 2023
# 2) SUBMITTER'S INFORMATION [21 CFR 807.92(a)(1)]
| 510(k) Owner (Manufacturer) | Medipixel, Inc.<br>Dooam Building 7F, 61, Yanghwa-ro, Mapo-gu, Seoul, Republic of Korea,<br>04037<br>Phone: +82-70-4699-0460<br>Fax: +82-50-7534-4375<br>E-mail: contact@medipixel.io |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Correspondence Person | Seokjin Ham<br>Quality Management Representative of Medipixel, Inc.<br>Phone: +82-10-3664-3896<br>E-mail: harry.ham@medipixel.io<br><br>Hye-Ri Choi<br>RA Manager of Medipixel, Inc.<br>Phone: +82-70-4699-0460<br>E-mail: contact@medipixel.io |
# 3) SUBJECT DEVICE [21 CFR 807.92(a)(2)]
| Proprietary name | Medipixel XA |
|----------------------|--------------------------------------------------|
| Model Name | MPXA-2000 |
| Common/Usual Name | Cardiovascular Image Analysis Software |
| Product Code | QIH |
| Regulation Number | 21 CFR 892.2050 |
| Classification Names | Automated Radiological Image Processing Software |
| Device Class | II |
| Classification Panel | Radiology |
# 4) PREDICATE DEVICE [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follows:
| Proprietary Name | QAngio XA (X-RAY VVA, K112807) |
|--------------------|----------------------------------------|
| Common/Usual Name: | Radiological Image Processing Software |
| Product Code | LLZ |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Medipixel. The logo consists of a colorful icon to the left of the word "medipixel" in gray and black. The icon is made up of four different colored shapes: purple, light blue, dark blue, and orange.
| Regulation Number | 21 CFR 892.2050 |
|----------------------|----------------------------------------|
| Classification Names | System, Image Processing, Radiological |
| Device Class | II |
| Classification Panel | Radiology |
#### DEVICE DESCRIPTION [21 CFR 807.92(a)(4)] ર)
MPXA-2000, the stand-alone application, is cardiovascular image analysis software for the viewing and quantification of X-ray angiographic images of the coronary arteries. Automatic vessel contour detection (i.e. vessel segmentation) forms the basis for the analyses. MPXA-2000 provides automatic analysis through segmentation of vessels and calculation of the segmented vessel's dimensions. MPXA-2000 is developed based on Deep-learning (Classification and Segmentation of vessels) and Computer vision algorithms (Measurements of vessels) to analyze the images and provide quantification of the vessels in real-time. It allows accurate and reproducible quantification of the coronary arteries from a set of those images in DICOM format.
MPXA-2000 is composed of the following key analysis workflow and functionalities:
- Image Uploading, Frame Selection -
- Vessel Segmentation -
- Classification of Vessel types -
- ROIs (regions of Interest) Quantification -
- -Visualization & Export of the Analysis Results.
The analysis results are available on the screen and can be exported in PDF or Excel file format. MPXA-2000 has been validated for the images produced by FDA-approved X-ray angiography systems from Philips Medical Systems, Siemens Healthineers, GE Healthcare. The input data should be X-ray angiographic images that comply with the DICOM standard.
# 6) INTENDED FOR USE/INDICATIONS FOR USE [21 CFR 807.92(a)(5)]
#### Intended Use
MPXA-2000 is software intended to be used for performing calculations in X-ray angiographic images of the coronary arteries. These calculations are based on vessel contours which are automatically detected by the software and subsequently presented for review and manual editing.
The analysis results obtained with MPXA-2000 are intended for use by cardiologists and radiologists:
- to support clinical decisions concerning the coronary arteries.
- to support the evaluation of intervention or drug therapy applied for conditions of the coronary arteries.
# Indications for Use
MPXA-2000 is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the calculations in X-ray angiographic images of the coronary arteries, for use on individual patients with coronary artery disease (CAD). MPXA-2000 is indicated for use in adult patients only.
When the quantified results provided by MPXA-2000 are used in a clinical setting on X-ray images of an individual patient, they can be used to support the clinical decisions making for the diagnosis of the patient or
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Medipixel. The logo consists of a colorful geometric shape on the left and the word "medipixel" in gray on the right. The geometric shape is made up of four different colored sections: purple, light blue, dark blue, and orange.
the evaluation of the treatment applied. In this case, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical diagnosis, and they are only intended for use by the responsible clinicians.
# 7) TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE [21 CFR 807.92(a)(6)]
The comparison of the technological characteristics of the predicate and subject device is given in the table below.
# General Information
| Item | New device | Predicate Device |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | MPXA-2000 | QAngio XA (X-RAY VVA) |
| Manufacturer | Medipixel, Inc. | Medis Medical Imaging Systems, B.V. |
| 510(k) Number | - | K112807 |
| Product code | QIH | LLZ |
| Regulation No. | 21 CFR 892.2050 | 21 CFR 892.2050 |
| Class | II | II |
| Indications for Use | MPXA-2000 is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the calculations in X-ray angiographic images of the coronary arteries, for use on individual patients with coronary artery disease (CAD). MPXA-2000 is indicated for use in adult patients only.<br>When the quantified results provided by MPXA-2000 are used in a clinical setting on X-ray images of an individual patient, they can be used to support the clinical decisions making for the diagnosis of the patient or the evaluation of the treatment applied. In this case, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical diagnosis, and they are only intended for use by the responsible clinicians. | X-RAY VVA is indicated for use in clinical settings where validated and reproducible quantified results are needed to support the calculations in X-ray angiographic images of the chambers of the and heart of blood vessels, for use on individual patients with cardiovascular disease.<br>When the quantified results provided by X-RAY VVA are used in a clinical setting on X-ray images of an individual patient, they can be used to support the clinical decisions making for the diagnosis of the patient or the evaluation of the treatment applied. In this case, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical diagnosis, and they are only intended for use by the responsible clinicians. |
| Patient Population | Patient with cardiovascular disease. | Patient with cardiovascular disease. |
| Intended Users | Cardiologists, Radiologists | Cardiologists, Radiologists |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Medipixel. The logo consists of a geometric shape made up of four colored squares: purple, light blue, orange, and white. To the right of the shape is the word "medipixel" in gray and black lowercase letters.
| (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1<br>Software Version | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
# Technological characteristics
| | Input data type | X-ray angiographic data in<br>DICOM format<br>(vendor-independent) | X-ray angiographic data in<br>DICOM format<br>(vendor-independent) |
|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manual editing of automatic results by user | | Yes | Yes |
| Calibration | | Automatic calibration<br>based on isocenter<br>calibration factor in<br>DICOM file<br>Other manual calibration<br>options (catheter<br>calibration) | Automatic calibration<br>based on isocenter<br>calibration factor in<br>DICOM file<br>Other manual calibration<br>options (catheter, marker<br>catheter, sphere, grid,<br>circle manual and line<br>calibration) |
| Automatic angiographic series<br>loading into the software from the<br>angiography equipment | | Yes | No |
| Visualization/ Edit Tools | | ✓ Zooming, panning<br>✓ Editing vessel<br>contours<br>✓ Angle, Length, Area,<br>Text Annotation | ✓ Zooming, panning<br>✓ Editing vessel<br>contours |
| Quantitative<br>Analysis | Automated 2D<br>arterial contour<br>segmentation | Yes | Yes |
| | Classification of<br>vessel types | Yes<br>LAD, LCX, and RCA | No |
| | Optional Stent<br>analysis including<br>stent edges | No | Yes |
| Analysis<br>Results | Results for multiple<br>lesions and additional<br>user-defined ROIa) | Results for multiple lesions and<br>additional user-defined ROIa) | Results for multiple lesions and<br>additional user-defined ROIa) |
| | Vessel analysis | main and side branch | Straight (main), ostial, and<br>bifurcation (side branch) |
| | Vessel quantifications | - % Diameter Stenosis<br>- Minimum lumen diameter<br>(MLD)<br>- Proximal and distal<br>diameters (at P- and D-<br>marker positions)<br>ROIa) length<br>- Reference Diameter | - % Diameter Stenosis<br>- Minimum lumen diameter<br>(MLD)<br>- Proximal and distal<br>diameters (at P- and D-<br>marker positions)<br>- Lesion length<br>- Reference Diameter<br>- Length of stent and stent |
| | Stent related statistics | No | |
| | | edges<br>- MLD and its position<br>- In-stent Mean diameter | |
| Ventricle analysis | No | Yes | |
| Automatically load and visualize ECG data acquired from the DICOM file | Yes | No | |
| Automatic EoD (End of Diastole) phase detection in DICOM file based on ECG. | Yes | No | |
| Data Reporting | Patient and study details, calibration, annotation, measurement, and analysis details. | Patient and study details, calibration, annotation, measurement, and analysis details. | |
| Export file formats | PDF, Excel | DICOM, BMP, JPEG, AVI and the raw file format | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Medipixel. The logo consists of a stylized square made up of four different colored sections: purple, light blue, orange, and dark blue. To the right of the square is the word "medipixel" in gray, lowercase letters. The logo is simple and modern, and the colors are bright and eye-catching.
a) The term ROI (Region of Interest) refers to a narrowed area in a segmented blood vessel
MPXA-2000 has virtually the same intended use and indications for use as the predicate devices QAngio XA (K112807). Both devices are software intended to be used for the quantitative analysis in X-ray angiographic images. The qualified results provided by both devices can be used to support the clinical decisions making for the diagnosis of the patient or the evaluation of the treatment applied. The basic features as shown below in technological characteristics of both MPXA-2000 and the predicate device (QAngio XA) are the same:
- . Input images provided in a DICOM format
- . Automatic vessel segmentation (contour detection) of the coronary vessels
- Manual editing of the automated result by the user
- . Quantification of coronary arteries based on the analysis of angiographic image
- . Reporting of quantitative analysis results
The differences in technical characteristics between MPXA-2000 and the predicate OAngio XA are as follows:
- -MPXA-2000 provides the automatic classification of vessel types such as LAD, LCX, and RCA. QAngio XA does not provide it.
- -Both MPXA-2000 and the predicate device, QAngio XA, have the same feature to automatically detect the contour of coronary vessels. However, this feature differs slightly between the two products in terms of technical characteristics. The difference is that MPXA-2000 automatically finds the main and bifurcation vessels in the image, and also automatically detects their contours, but in QAngio XA, a user must designate the vessel to be segmented in advance.
- -The predicate device, QAngio XA has the following additional features that MPXA-2000 does not provide.
- Calculations in X-ray angiographic images of the left ventricle of the heart ●
- . Segmentation and quantification of the ostium part of vessels
- Analysis of vessels treated with stents
Among the differences mentioned above, the features of the automatic vessel type classification and vessel segmentation that MPXA-2000 provides are based on Deep-learning algorithm. The Deep-learning algorithm used in MPXA-2000 was validated by the standalone performance test (Doc. No. TD-XA2-SPTR) using the method of comparing the analysis results obtained from MPXA-2000 with the ground truth annotated by experienced experts. The testing results are as follows:
- The vessel classification accuracy: 0.9934 .
- The vessel segmentation Dice similarity Coefficient (DSC): 0.9200(Pass criteria of > 0.8)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Medipixel. The logo consists of a colorful geometric shape on the left and the word "medipixel" in gray on the right. The geometric shape is made up of four different colored sections: purple, light blue, dark blue, and orange. The word "medipixel" is written in a sans-serif font.
In addition to that, the results of ROI analysis in the standalone performance test show that the estimation of minimum luminal diameter by MPXA-2000, which is based upon this automatic segmentation feature, is comparable to the reported performance of the predicate device QAngio XA. (Doc. No. TD-XA2-SPTR) All those results above demonstrate that those differences in technical characteristics do not raise any different questions of safety and effectiveness.
And the additional features above of the predicate device do not affect or alter the intended use and indication for use. Although MPXA-2000 does not have these features, it does not raise different questions of safety and effectiveness.
Furthermore, the safety and effectiveness of MPXA-2000 were evaluated and validated in accordance with software specifications and applicable performance standards through software V&V testing and standalone performance testing. The standalone performance test consists of 1) vessel segmentation assessment, 2) vessel classification assessment, and 3) ROI analysis assessment. The software validation activities were performed in accordance with IEC 62304: 2015 Medical device software life cycle processes, in addition to the FDA Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submission for Management of Cybersecurity in Medical Devices."
In conclusion, MPXA-2000's new features added to increase the convenience of use and functionality do not alter the intended use of the device to be used for performing calculations in X-ray angiographic images of the coronary arteries. And the new device does not raise different questions of safety and effectiveness.
#### 8) Summary of Non-Clinical Testing Performed [21 CFR 807.92(b)(1)]
The product has been evaluated for software safety and performance and has been found to conform to applicable medical device safety standards and guidance documents.
MPXA-2000 complies with the following standards:
- IEC 62304 Edition 1.1 2015-06 Medical device software Software life cycle processes -
MPXA-2000 complies with the following guidance:
- General Principles of Software Validation (Jan 11, 2002) -
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices: -Guidance for Industry and FDA Staff (Nov 05, 2005)
- -Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, (Oct 2, 2014)
#### Performance testing
MPXA-2000 was developed and tested in accordance with Medipixel' Design Control processes and has been subject to extensive safety and performance testing. Non-clinical verification test results established that the device meets its design requirements and intended use. Specifically, software verification was conducted at unit (module) and system integration levels. Risk management analysis generated multiple risk mitigation measures and verification activities. A cybersecurity verification testing was conducted to demonstrate the integrity, confidentiality, and availability of MPXA-2000 through testing for identified vulnerabilities of product.
Stand-alone performance testing, a quantitative assessment of MPXA-2000's performance, was carried out by using 305 coronary angiograms collected from the US population. The angiograms for this testing were collected retrospectively from Philips Medical Systems Nederland B.V. under the data-sharing agreement. A
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the word "medipixel" in a sans-serif font. The word is written in gray and black. To the left of the word is a logo that is made up of four different colored squares. The colors of the squares are purple, light blue, dark blue, and orange.
quantitative assessment of MPXA-2000's performance was carried out by using 305 coronary anglograms collected from the US population. Test datasets were strictly segregated from algorithm training datasets. Three features of MPXA-2000 were tested: vessel segmentation, vessel classification, and ROI analysis. For all tests, the results met the Pass criteria which were pre-specified based on the results of related studies found in literature. It demonstrates that MPXA-2000 performs reliably its functions for US data.
Based on the results of non-clinical verification and validation tests, it is concluded that MPXA-2000 is safe and effective in the assessment of vessel segmentation, vessel classification, and ROI analysis
# 9) Summary of Clinical Tests [21 CFR 807.92(b)(2)]
This premarket submission, MPXA-2000, did not require clinical studies to support substantial equivalence.
# 10) Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
There are no significant differences between MPXA-2000 and the predicate devices, QAngio XA that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.
# 11) Conclusions [21 CFR 807.92(b)(3)]
Based on the information submitted in this premarket notification, the intended use/indications for use. technological characteristics, and performance testing, MPXA-2000 raises no new questions of safety and effectiveness. It is substantially equivalent to the predicative devices in safety, effectiveness, and performance.
END of 510(k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.