GEMSS Medical System Surgical Mobile Fluoroscopic X-ray System
K160065 · Gemss Medical Systems Co., Ltd. · OXO · May 20, 2016 · Radiology
Device Facts
Record ID
K160065
Device Name
GEMSS Medical System Surgical Mobile Fluoroscopic X-ray System
Applicant
Gemss Medical Systems Co., Ltd.
Product Code
OXO · Radiology
Decision Date
May 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Intended Use
KMC-650 Surgical Mobile Fluoroscopic X-ray System is indicated for use in generating fluoroscopic / radioscopic images of human anatomy. This device is not intended for interventional guided procedure & mammographic applications.
Device Story
KMC-650 is a mobile surgical fluoroscopic X-ray system used in hospitals (OR, ER, cast room) by clinicians to visualize anatomical structures. It consists of a high-voltage inverter generator, X-ray tube, beam-limiting device, image intensifier, and mobile cart. The system generates X-rays that pass through the patient; the image intensifier and camera capture the resulting signals, which are processed and displayed on a monitor. The device provides real-time fluoroscopic or static radioscopic images to assist in diagnostic procedures. It features automatic brightness control (ABS) and pulse fluoroscopy modes. Clinicians use the visual output to assess skeletal, respiratory, or urinary conditions. The system benefits patients by providing immediate, mobile diagnostic imaging during medical operations or emergency assessments without requiring patient transport to a radiology suite.
Clinical Evidence
Bench testing only. The device underwent electrical, mechanical, and EMC safety testing according to FDA-recognized standards (IEC 60601 series, ISO 14971). Performance testing confirmed the generator output is sufficient for fluoroscopy and radiography requirements. No clinical data was required or provided.
Indicated for generating fluoroscopic/radioscopic images of human anatomy, including skeletal, respiratory, and urinary systems (skull, spine, chest, abdomen, extremities). Not for interventional guided procedures or mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K032761 — KMC-950 C-ARM MOBILE SYSTEM · United Radiology Systems, Inc. · May 14, 2004
K051754 — ORTHOSCAN, ORTHOSCAN HD · Orthoscan, Inc. · Aug 9, 2005
K181943 — OSCAR (OSCAR Prime, OSCAR Classic) · Genoray Co., Ltd. · Aug 17, 2018
K183144 — uMC 560i · Shanghai United Imaging Healthcare Co., Ltd. · Dec 19, 2019
K250241 — Cios Select · Siemens Medical Solutions USA, Inc. · Nov 4, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 20, 2016
GEMSS MEDICAL SYSTEMS CO., LTD % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025
Re: K160065
Trade/Device Name: KMC-650 GEMSS Medical System Surgical Mobile Fluoroscopic X-ray System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OXO, JAA Dated: April 13, 2016 Received: April 18, 2016
Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K160065
Device Name KMC-650 GEMSS Medical System Surgical Mobile Fluoroscopic X-ray System
Indications for Use (Describe)
KMC-650 Surgical Mobile Fluoroscopic X-ray System is indicated for use in generating fluoroscopic / radioscopic images of human anatomy. This device is not intended for interventional guided procedure & mammographic applications.
Type of Use (Select one or both, as applicable)
| <input checked="" type="checkbox"/> | Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------|----------------------------------------------|
| <input type="checkbox"/> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION 07
# 510(k) SUMMARY
{4}------------------------------------------------
## 510(k) SUMMARY For KMC-650
#### Submitted by:
GEMSS MEDICAL SYSTEMS CO., LTD. Address - 2nd Floor, 29, Dunchon-daero 541 beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea Telephone -- +82-31-764-7321~3 Fax - +82-31-764-7324 Contact - Mr. Sangwoo Lee Internet - http://www.gemss-medical.com
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR Part 807.92.
Date: April 12, 2016
1. General Information:
Establishment: GEMSS MEDICAL SYSTEMS CO., LTD. 2nd Floor, 29, Dunchon-daero 541 beon-gil, Jungwon-gu, seongnam-si, KOREA, REPUBLIC OF 13216 Registration Number: 3003384390
GMESS MEDICAL SYSTEMS CO., LTD. is the holder/owner for this 510(k).
2. Official Correspondent Contact:
Dave Kim Mtech Group 8310 Buffalo Speedway Houston, TX 77025 TEL : 713-467-2607
FAX : 713-583-8988
E-mail : davekim@mtech-inc.net
3. Device name and Classification
Trade Name : KMC-650 Classification Name : Image-intensified Fluoroscopic Mobile X-ray System Classification Panel : Radiology Classification Regulation : 21 CFR 892. 1650
{5}------------------------------------------------
| Device Classification | : Class II |
|------------------------|------------|
| Primary product code | : OXO |
| Secondary product code | : JAA |
4. Legally Marketed Predicated Device
| Trade Name | : KMC-950 |
|----------------------|------------------------------------------------------|
| 510(k) Clearance # | : K032761 |
| Clearance date | : 05/14/2004 |
| Classification Name | : Image Intensified Fluoroscopic Mobile X-ray System |
| Classification Panel | : Radiology |
| CFR Section | : 21CFR 892.1650 |
| Device Class | : Class II |
| Product Code | : JAA, OWB, OXO |
#### 5. Device Description
The KMC-650 Surgical Mobile Fluoroscopic X-rav System consists of a high voltage (HV) inverter generator. a tube support unit, an X-ray beam limiting device, mobile cart, a detector, operating software, and a tube, and is primarily used in a hospital for diagnosis of diseases in skeletal, respiratory and urinary systems such as the skull, spinal column, chest, abdomen, extremities, and other body parts. Not to be used for mammography.
KMC-650 is a solution to produce radiological images of patient during medical operations. This inverter control X-ray unit visualizes the anatomical structure on screen, which is obtained by X-ray fluoroscopy and the image intensifier. This system can be applied in emergency room, operation room, cast room or etc. of hospital.
#### 6. Indications for Use
KMC-650 Surgical Mobile Fluoroscopic X-ray System is indicated for use in generating fluoroscopic / radioscopic images of human anatomy. This device is not interventional guided procedure & mammographic applications
#### 7. Substantial Equivalence
KMC-650 is substantially equivalent to the commercially available KMC-950 of GEMSS Medical System (K032761). The detailed SE discussion is provided in SECTION 12.
| | | Predicate device | Proposed device |
|---------------|----------------------|--------------------------------|--------------------------------|
| Model Name | | KMC-950 | KMC-650 |
| 510(k) Number | | K032761 | |
| Manufacturer | | GEMSS Medical System Co., Ltd | GEMSS Medical System Co., Ltd |
| X-ray tube | Manufacturer | Varian (RAD-99) | Toshiba (DF-151SBR) |
| | Anode Type | Rotating | Stationary |
| | Heat Capacity | 300,000 HU | 45,000 HU |
| | Anode Heat Cooling | 70kHU/min | 47kHU/min |
| | Focal size | 0.3mm / 0.6mm | 0.5mm/1.5mm |
| X-ray | Manufacturer | POSKOM (HTC-120) | POSKOM (C-650) |
| generator | X-ray Generator Type | High frequency / inverter type | High frequency / inverter type |
{6}------------------------------------------------
| | Power Output | 12.5 kW | 2.2 kW |
|--------------|---------------------------------------|------------------------------------------------------|------------------------------------------------------|
| Fluoroscopic | kV range | 40 to 125 kV See below<br>discussion for differences | 40 to 110 kV See below<br>discussion for differences |
| Mode | mA range | 0.5 to 5 mA. See below<br>discussion for differences | 0.5 to 4 mA. See below<br>discussion for differences |
| | Pulse Fluoro | Yes | Yes |
| | ABS function | Yes | Yes |
| | Snap Shot | 8.0 mA shot available | 8.0 mA shot available |
| | Boost Shot | 20.0 mA shot available | 20.0 mA shot available |
| Image | Manufacturer | Thales / Toshiba | Thales / Toshiba |
| Intensifier | Size | 9" | 9" |
| | Magnification | 9" / 6" / 4.5" | 9" / 6" / 4.5" |
| Camera | Manufacturer | RAYSIS | GEMSS |
| | Type | 1/2" CCD | 1/2" CCD |
| | Active pixels | 512 x 512 | 1,024 x 1,024 |
| C-arm | Manufacturer | GEMSS medical Systems Co.,<br>Ltd | GEMSS medical Systems Co.,<br>Ltd |
| | SID | 950mm | 1000mm |
| | Range of C-ram Rail | 115° (90° / 25°) | 135° (90° / 45°) |
| | Rotation | | |
| | Range of the Liner<br>FR-arm Movement | 200 mm | 200 mm |
| | Range of the Linear<br>T-arm Movement | 500 mm | 500 mm |
| | Range of Swing-arm<br>Rotation | $\pm$ 12.5° | $\pm$ 12.5° |
| | Range of Stay-arm<br>Rotation | 360° | 360° |
| Image | Storage Capacity | Digital | Digital |
| Processing | Image Matrix Size | 5,000 images | 5,000 images |
| | Monitor Size | 17" | 19" |
| Collimator | Model Name | KMC-950CM | KMC-650CMR |
| | Manufacturer | GEMSS medical Systems Co.,<br>Ltd | GEMSS medical Systems Co.,<br>Ltd |
| | Collimator | Motor control / rotation | Motor control / rotation |
8. Summary of technological characteristics of the device compared to the predicate device
※ Performance Differences in the Generator
| Items | | KMC-950 | KMC-650 |
|--------------------------|-----------|----------------------|----------------------|
| Generator | MODEL | HTC-120 | C-650 |
| | Max Power | 12.5 kW | 2.2 kW |
| Device<br>Specifications | kV | Fluoro : 40 - 125 kV | Fluoro : 40 - 110 kV |
| | mA | Fluoro : 0.5~ 5 mA | Fluoro : 0.5 - 4 mA, |
→ The generator capacity must be sufficient to meet the fluoroscopy and Radiography exposure requirements. The power requirement for a generator depends on the size, length and thickness of X-ray Tube (Anode) Filament. The generator's mA output also depends on the focal spot sizes.
{7}------------------------------------------------
A typical fluoroscopy mode for KMC-950 and KMC-650 requires the power of 1 kW or less. Therefore, the generator outputs for KMC-650 and KMC-950 in fluoroscopy mode is not significantly different.
#### KMC-650/950 Firmware Comparison Table
| Items | KMC-650 | KMC-950 | Remarks |
|------------------|---------|---------|--------------------------------------------------------------------------------------------------------------------------|
| Camera Board | 650CAMB | 950CAMB | Camera specifications are different.(active<br>pixels) |
| Main Board | 650MCB | 950MCB | Generator Control, Key Input function<br>operates differently. |
| OP Control Board | 650OPCB | - | A separate OP Board exists for KMC-650<br>whereas KMC-950 has the main control<br>board only without a separate OP board |
KMC-650 is designed as a set of components (X-ray tube and housing, imaging intensifier, collimator, generator). KMC-650 and components conform to the FDA recognized standards. Based on the recognized standard conformity evidences related to electro-, mechanical-, software-, clinical-, and risk management, it is confirmed that the all technological characteristics of KMC-650 is substantially equivalent to the predicate device.
#### 9. General Safety and Effectiveness Concern
To minimize electrical, mechanical, and radiation hazards, GEMSS MEDICAL SYSTEMS CO., LTD. adheres to recognized and established industry practice, and all equipment complies with the relevant FDA and international standards.
Through verification and validation activities, the safety and effectiveness of KMC-650 is verified and validated. Engineering testing and standards compliance testing were successfully conducted and did not raise any new safety questions or concerns or identify new risks. And instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Electrical, EMC, mechanical safety and performance testing according to FDA recognized standards were performed. All test results were satisfactory. Applied standards are as follows:
- > IEC60601-1:2005
- A IEC60601-1-2:2007
- A IEC60601-1-3:2008
- A IEC60601-2-28:2010
- A IEC60601-2-43:2010
- A IEC60601-2-54:2009
- A NEMA PS 3.1-3.20
- A ISO14971:2012
And, EPRC regulation was satisfactory.
- 21CFR1020.32 A
In addition, FDA quidance were satisfactorily considered.
- > Guidance for the Submissions of 510(k)s for Solid State X-ray Imaging Devices
{8}------------------------------------------------
Based on the robust set of results from the non-clinical and clinical studies that have been performed, KMC-650 has been found to be substantially equivalent to the predicate device as well as found to be a safe and effective surgical mobile fluoroscopic X-ray imaging system.
#### 10. Conclusion as to Substantial Equivalence
KMC-650 is substantially equivalent to the predicate device KMC-950 (K032761). These devices are very similar in the intended use, the design principle, the performance and the applicable standards. Some characteristics, for example, their appearance, the user interfaces are different. However the compliance reports, demonstrations and descriptions in this submission STED provide demonstration that these small and minor differences do not raise any new questions of safety and effectiveness. So GEMSS MEDICAL SYSTEMS CO., LTD. concludes KMC-650 is substantially equivalent to the predicate device KMC-950 (K0327618).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.