INFX-8000V with Wireless Footswitch

K143225 · Toshibamedical Systems Corporation · OWB · Jan 9, 2015 · Radiology

Device Facts

Record IDK143225
Device NameINFX-8000V with Wireless Footswitch
ApplicantToshibamedical Systems Corporation
Product CodeOWB · Radiology
Decision DateJan 9, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1650
Device ClassClass 2

Intended Use

This device is a digital radiography/fluoroscopy system use in a diagnostic interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.

Device Story

Digital radiography/fluoroscopy system for interventional angiography; includes C-arm, x-ray tube, beam limiter, receptor, controller, processing computers, and patient table. Modification adds interchangeable Bluetooth wireless footswitch for system control. Used in clinical interventional settings by physicians. Wireless footswitch maintains existing system functionality; no changes to hardware or software requirements. Benefits include improved operational flexibility during procedures while maintaining established diagnostic capabilities.

Clinical Evidence

No clinical data. Safety and effectiveness verified via bench testing, risk management, and design controls.

Technological Characteristics

Digital radiography/fluoroscopy system. Components: C-arm, x-ray tube, beam limiter, receptor, controller, computers. Connectivity: Bluetooth wireless footswitch. Standards: 21 CFR Subchapter J, 21 CFR 1020, ISO 13485, IEC 60601-1, IEC 60601-2-43, IEC 60601-2-28.

Indications for Use

Indicated for diagnostic and angiographic procedures of blood vessels in the heart, brain, abdomen, and lower extremities.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a flowing, wave-like design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 9, 2015 Toshiba Medical Systems Corporation % Mr. Paul Biggins U.S. Agent/Director Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780 Re: K143225 Trade/Device Name: INFX-8000V with Wireless Footswitch Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: II Product Code: OWB, JAA Dated: December 9, 2014 Received: December 10, 2014 Dear Mr. Biggins: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert A Ochs Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K143225 Device Name INFX-8000V with Wireless Footswitch Indications for Use (Describe) This device is a digital radiography/fluoroscopy system use in a diagnostic interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) - SUMMARY OF SAFETY AND EFFECTIVENESS #### 1. CLASSIFICATION and DEVICE NAME: | Classification Name: | Image Intensified Fluoroscopic X-ray System | |-------------------------|--------------------------------------------------------------------------------------------------------| | Regulation Number: | 21 CFR 892.1650 (Class II) | | Product Code | OWB – Interventional Fluoroscopic X-ray System<br>JAA – System, x-ray, fluoroscopic, image-intensified | | Trade Proprietary Name: | Infinix | | Model Number: | INFX-8000V, with wireless footswitch | ### 2. ESTABLISHMENT REGISTRATION: 9614698 ### 3. CONTACT PERSON, U.S. AGENT and ADDRESS: ### Contact Person/U.S. Agent: Paul Biggins Director, Regulatory Affairs (714) 730-5000 Fax: (714) 730-1310 pbiggins@tams.com ### Establishment Name and Address: Toshiba America Medical Systems, Inc. (TAMS) 2441 Michelle Drive Tustin, Ca. 92780 ### 4. MANUFACTURING SITE Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan CONTACT: AKINORI HATANAKA # 5. Date OF SUBMISSION: November 12, 2014 ## 6. PERFORMANCE STANDARD: 21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard # 7. PREDICATE DEVICE: INFX-8000V,v5.30 (K133535) # 8. REASON FOR SUBMISSION: Modification of a cleared device {4}------------------------------------------------ ### 9. SUBMISSION TYPE: Special 510(k) ### 10. DEVICE DESCRIPTION: This device in an x-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm, which is equipped with an x-ray tube, beam limiter and x-ray receptor, x-ray controller, computers with system and processing software, and a patient radiographic table. #### 11. Indication for Use: This device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities. ### 12. SUMMARY OF CHANGE(S) This device has been modified to use an interchangeable BlueTooth wireless footswitch. The introduction of this footswitch does not change any hardware or software requirements of the cleared device. ### 13. SUBSTANTIAL EQUIVALENCE: This device is substantially equivalent to the INFX-8000V, v5.30 K133535, marketed by Toshiba America Medical Systems. The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device. There are no new indications for use or intended use of the device. ### 14. SAFETY: The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards; IEC 60601-2-43 and IEC60601-2-28. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report. ### 15. CONCLUSION Based upon the results of the verification testing the performance of the modified device is equal to or better than the predicate device. The addition of the wireless footswitch does not change the indications for use or the intended use of the device. Safety and effectiveness have been verified via risk management and application of design controls to the modification. Toshiba concludes that it has demonstrated that the modified INFX-8000V is substantially equivalent to the previous version.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...