K102615 · Ge Medical Systems, LLC · MQB · Nov 18, 2010 · Radiology
Device Facts
Record ID
K102615
Device Name
WIRELESS DR IMAGING OPTION-WDR1
Applicant
Ge Medical Systems, LLC
Product Code
MQB · Radiology
Decision Date
Nov 18, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Wireless DR Imaging Option -WDR1, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy. It is intended for use in all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma, wherever conventional screen-film systems or CR systems maybe used. The device is not intended for mammography or fluoroscopy applications.
Device Story
Wireless DR Imaging Option (WDR1) is a digital radiographic detector system comprising a wireless detector, interface box, computer, and display. It functions as a receptor replacement for existing radiographic X-ray systems, substituting for film/screen or CR systems. The detector uses amorphous silicon with a cesium iodide scintillator to capture X-ray signals, which are transmitted via wireless signal, tethered cable, or direct dock connection to the system computer. Used in clinical settings (e.g., ICU, trauma, routine radiography) by healthcare professionals. The system processes captured data into digital images for clinician review, facilitating diagnostic decision-making. Benefits include improved workflow through wireless portability, reduced weight, and power efficiency compared to traditional detectors.
Clinical Evidence
Bench testing only. No clinical data presented. The device was evaluated for electrical, mechanical, and radiation safety through verification and validation testing, confirming performance to specifications and adherence to international standards (IEC 60601-1) and CDRH requirements.
Technological Characteristics
Amorphous silicon detector with cesium iodide scintillator. Wireless, tethered, or docked connectivity. Complies with IEC 60601-1 and 21 CFR subchapter J. System includes detector, interface box, computer, and display. Hardware/firmware optimized for reduced weight and power consumption.
Indications for Use
Indicated for generating radiographic images of human anatomy in patients of all ages, including pediatric, intensive care, and trauma populations. Not indicated for mammography or fluoroscopy.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K192440 — FDR SE Lite · Fujifilm Corporation · Nov 1, 2019
K183713 — Wireless Digital Flat Detector · Iray Technology Taicang , Ltd. · Jan 9, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
K102615
P. 1 of 3
NOV 1 8 2010
GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDRI
### 510(K) Summary of Safety and Effectiveness
Prepared in accordance with 21 CFR Part 807.92.
| Date Prepared: | September 3, 2010 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare, (GE Medical Systems, LLC)<br>3000 N. Grandview Blvd.<br>Waukesha, WI 53188 |
| Primary Contact : | John L. Schmidt<br>Regulatory Affairs Leader, X-Ray<br>Telephone: 262-548-4964; Fax: 262-548-2032<br>e-mail: John.L.Schmidt@ge.com |
| Secondary Contact : | David Blonski.<br>Regulatory Affairs Director, X-Ray<br>GE Healthcare, (GE Medical Systems, LLC)<br>Telephone: 262-513-4072; Fax: 262-548-2032<br>e-mail: David.Blonski@ge.com |
# DEVICE INDENTIFICATION
| Trade Name: | GE Wireless DR Imaging Option – WDR1 |
|----------------------|---------------------------------------------------------------------------------------|
| Common/Usual Name: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| Classification Name: | Solid State X-ray Imager (flat panel/digital imager)<br>Class II, MQB, 21CFR 892.1650 |
# DEVICE DESCRIPTION:
The Wireless DR Imaging Option - WDR1 consists of a wireless detector, system interface box, computer, display, keyboard and mouse. It is designed to acquire digital radiographic images when used with existing radiographic x-ray systems. Images captured can be communicated to the system via wireless signal, tethered cable or direct connection (docked). The Wireless detector is very similar to the digital detector currently in use as the GE Digital Radiographic Detector (K982196), and the Philips
{1}------------------------------------------------
K102615
P. 2 of 3
GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDRI
Medical Systems Wireless Portable Detector FD-W17 (K090625) comprised of amorphous silicon with a cesium iodide scintillator.
It is used on radiographic X-ray systems as a substitute for film/screen systems or CR systems. The Radiographic X-ray system remains unchanged except for the replacement of the systems receptor with the Wireless WDR 1 detector.
#### INDICATIONS FOR USE:
The Wireless DR Imaging Option -WDRI, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy. It is intended for use in all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma, wherever conventional screen-film systems or CR systems maybe used.
The device is not intended for mammography or fluoroscopy applications.
### COMPARISON WITH PREDICATE DEVICES:
The Wireless DR Imaging Option - WDR1 is an enhanced wireless version of and substantially equivalent to, the GE Digital Radiographic Detector cleared under 510(k) number K982196 and Philips Medical Systems wireless detector FD-W17 (K090625). It involves changes from the predicate device (K982196) to add wireless functionality, reduced weight and power, and modifications in design that involve hardware, software, and firmware. The Wireless DR Imaging Option WDR1 and the predicate devices are both for use in acquiring digital radiographic images and have similar indications for use.
Changes in electronics, hardware and firmware have been made to affect these power and weight reductions, but there is no fundamental change in X-ray detection technology, or image creation, and uses virtually identical operating principles to the GE Digital Radiographic Detector currently marketed system (K982196). It has the same technological characteristics related to safety and effectiveness as the predicate devices. A review of all bench and standards testing indicate that the new device provides no new safety concerns and is as safe an effective as the predicate devices.
The Wireless DR Imaging Option - WDR1 is certified to comply with the X-ray requirements of 21 CFR, as well as safety requirements of IEC 60601-1 and associated collateral and particular standards.
{2}------------------------------------------------
K102615
P. 3 of 3
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The logo is black and white and has a vintage look.
GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDR 1
#### ADVERSE EFFECTS ON HEALTH:
The device has been evaluated for electrical, mechanical, and radiation safety, and conforms to applicable medical device safety standards, as confirmed by a Nationally Recognized Test Laboratory.
The potential hazards of electrical and mechanical are identified in a risk management summary (hazard analysis) and controlled by:
- . System verification and validation to ensure performance to specifications, Federal Regulations, and user requirements.
- Adherence and certification to industry and international standards. .
- Compliance to applicable CDRH 21 CFR subchapter J requirements. t
The device is designed and manufactured under the Quality System Regulations of 21 CFR 820.
# CONCLUSION:
The Wireless DR Imaging Option - WDR1 is an evolutionary modification to the GE Digital Radiographic Detector. It does not result in any new potential safety risks, has the same technological characteristics, and performs as well as the devices currently on the market. The Wireless DR Imaging Option WDR1 will be certified to comply with the X-ray requirements of 21 CFR, as well as safety requirements of IEC 60601-1 and associated collateral and particular standards.
After analyzing standards testing and bench data, it is the conclusion of GE Healthcare that the Wireless DR Imaging Option - WDR1 is substantially equivalent to other marketed devices with similar indications for use and meeting the same standards.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around it. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. John J. Schmidt Regulatory Affairs Leader, X-Ray GE Medical Systems, LLC 3000 N. Grandview Blvd. WAUKESHO WI 53188
AUG 2 3 2013
Re: K102615
Trade/Device Name: Wireless DR Imaging Option-WDR1 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: September 2, 2010 Received: September 9, 2010
Dear Mr. Schmidt:
This letter corrects our substantially equivalent letter of November 18, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 rase surved in the encreativent of the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, arones provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing pactice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and If you desire specific acritic for your as not is not is not is Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tod inay of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K102615
P. 1 of 1
GE Medical Systems, LLC 510(k) Premarket Notification Wireless DR Imaging Option WDR I
# Indications for Use
NOV 1 8 2010
510(k) Number (if Known): K | 0 26 15 Device Name: Wireless DR Imaging Option - WDR I
Indications for Use:
The Wireless DR Imaging Option -WDR1, when used with a radiographic imaging system. is indicated for use in generating radiographic images of human anatomy. It is intended for use in all routine radiography exams, and specialized areas including pediatric work, intensive care and trauma, wherever conventional screen-film systems or CR systems maybe used.
The device is not intended for mammography or fluoroscopy applications.
Prescription Use_X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Usc (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRA, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K102615
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.