TT1 GENERAL PURPOSE COIL

K083434 · Tursiop Technologies, LLC · MOS · Dec 23, 2008 · Radiology

Device Facts

Record IDK083434
Device NameTT1 GENERAL PURPOSE COIL
ApplicantTursiop Technologies, LLC
Product CodeMOS · Radiology
Decision DateDec 23, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The TT1 General Purpose Coil is a receive-only diagnostic MR imaging coil designed for use at a static magnetic field strength of 1.5T with the GE Signa® 1.5T system. It is used for obtaining diagnostic images of a variety of small-to-medium sized anatomical regions, such as jaw, spine, neck, shoulder, thigh, foot, ankle and joints.

Device Story

TT1 General Purpose Coil is a linear, receive-only RF coil; 3-inch diameter loop; designed for 1.5T GE Signa MR systems. Device captures RF signals from hydrogen nuclei excitation; provides high-resolution imaging for small-to-medium anatomical regions (jaw, spine, neck, shoulder, thigh, foot, ankle, joints) with ~10 cm field of view. Active/PIN diode-enabled blocking circuitry prevents coupling with transmitted fields. Used in clinical MR imaging environments; operated by trained MR technologists/radiologists. Output is raw MR signal processed by host system into diagnostic images; assists clinicians in anatomical visualization and diagnosis. Benefits include optimized signal-to-noise ratio (SNR) and targeted coverage for specific body parts.

Clinical Evidence

No clinical data. Substantial equivalence established via bench testing and side-by-side performance comparison with the predicate GE 3-inch General Purpose Coil, evaluating SNR and image quality.

Technological Characteristics

Linear, single-channel, receive-only RF coil; 3-inch diameter loop; ABS plastic housing. Operates at 1.5T. Active/PIN diode-enabled blocking circuitry for RF decoupling. Complies with IEC 60601-1 (safety), UL 94 (flammability), and NEMA MS-6 (coil characterization).

Indications for Use

Indicated for patients requiring diagnostic MR imaging of small-to-medium anatomical regions including jaw, spine, neck, shoulder, thigh, foot, ankle, and joints using a 1.5T GE Signa system.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ### A. GENERAL INFORMATION | Classification Name | Magnetic Resonance Diagnostic Device (21 CFR 892.1000) | | |-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------| | Device Trade Name: | TTI General Purpose Coil | | | Applicant's Name and Address: | Tursiop Technologies, LLC<br>11000 Cedar Ave. Suite 280<br>Cleveland OH 44106 | DEC 23 2008 | | Submitter: | Raju Viswanathan<br>Chief Technology Officer<br>Tursiop Technologies, LLC<br>11000 Cedar Ave. Suite 280<br>Telephone: (216) 658-4521<br>Email: raju@tursiop.com | | #### B. INTENDED USE The TT1 General Purpose Coil is a receive-only diagnostic MR imaging coil designed for use at a static magnetic field strength of 1.5T with the GE Signa® 1.5T system. It is used for obtaining diagnostic images of a variety of small-to-medium sized anatomical regions, such as jaw, spine, neck, shoulder, thigh, foot, ankle and joints. #### C. DEVICE DESCRIPTION The TT1 General Purpose Coil is a linear receive-only coil for operation at a magnetic field strength of 1.5T with the GE Signa® 1.5T system. The coil is a single element coil consisting of a 3-inch diameter loop. Coupling of the coil to the transmitted field is prevented through an active blocking circuit. The TT1 General Purpose Coil is specially designed for MR imaging of small-to-medium sized anatomical regions, such as jaw, spine, neck, shoulder, thigh, foot, ankle and joints. The coil provides optimum signal to noise ratio and coverage, allowing high-resolution imaging, while the sensitive region of the coil covers an approximately a 10 cm Field of View. #### D. PERFORMANCE TESTING & STANDARDS - IEC 60601-1 (1988): Medical electrical equipment Part 1: General requirements for safety, . including Amendment 1 (1991) and Amendment 2 (1995). - UL 94; Tests for Flammability of Plastic Materials for parts in Devices and Appliance . - . NEMA: MS-6 (2008) Characterization of Special Purpose Coils for Diagnostic Magnetic Resonance Images - NEMA Standardized 510(k) Pre-Market Notification Submission Template for RF Coil t Accessories Intended for Use with Magnetic Resonance Diagnostic Imaging Devices {1}------------------------------------------------ # E. PREDICATE DEVICES The TT1 General Purpose Coil is substantially equivalent to the predicate devices. The table below identifies similarities between the devices. Although the GE 3-inch General Purpose Coil is in commercial distribution it is not included in the FDA 510(k) database and/or has been purged from the FDA PMA database. Tursiop Technologies LLC has conducted side by side testing between the TT1 General Purpose Coil and the GE 3-inch General Purpose Coil to verify substantial equivalence. | Para<br>mete<br>r | TT1 General Purpose Coil | Predicate Coil: GE 3-inch General<br>Purpose Coil | Predicate Coil: MRgFUS General<br>Purpose and Breast Coil (K061715) | |---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended/Indications for<br>Use | The TT1 General Purpose Coil is<br>a receive-only diagnostic MR<br>imaging coil designed for use at a<br>static magnetic field strength of<br>1.5T with the GE Signa® 1.5T<br>system. It is used for obtaining<br>diagnostic images of a variety of<br>small-to-medium sized<br>anatomical regions, such as jaw,<br>spine, neck, shoulder, thigh, foot,<br>ankle and joints. | The GE 3-inch General Purpose Coil<br>is a receive-only RF coil designed for<br>1.5T MR imaging of various<br>anatomical regions, such as jaw, spine,<br>neck, shoulder, thigh, foot, ankle and<br>joints. | The MRgFUS General Purpose and<br>Breast Coil is a receive-only RF coil<br>designed for MR imaging of breast<br>and auxiliary tissue and various<br>medium sized anatomical regions such<br>as spine, neck, shoulder, thigh, foot,<br>ankle and joints. MRgFUS is designed<br>for use with GE Signa (1.5T or 3.0T)<br>MR Systems | | Dime<br>ns | Coil length - 3 inches<br>Coil width - 3 inches<br>Similar in weight | Coil length - 3 inches<br>Coil width - 3 inches<br>Similar in weight | Coil length - 6.5 inch<br>Similar in size and weight | | Coil Architecture | Linear/Single channel receive-<br>only for 1.5T operation | Linear/Single channel receive-only for<br>1.5T operation | Single channel receive-only for 1.5T<br>or 3.0T operation | | | Housing Type - ABS plastic | Housing Type - Fiberglass | n/a | | | Primary Decoupling - Active/PIN<br>diode-enabled blocking circuitry | Primary Decoupling - Active/PIN<br>diode-enabled blocking circuitry | Active and Passive RF Decoupling<br>circuits | | Performance<br>& Safety | IEC 60601-1<br>Flammability UL 94<br>NEMA MS6 | IEC 60601-1<br>Flammability UL 94<br>NEMA MS6 | IEC 60601-1<br>Flammability UL 94<br>NEMA MS6 | | Principles of<br>Operation | Hydrogen nuclei excitation for<br>imaging of the scanned organ | Hydrogen nuclei excitation for<br>imaging of the scanned organ | Hydrogen nuclei excitation for<br>imaging of the scanned organ | {2}------------------------------------------------ fechnologic haracteristic Optimal Signal to Noise Ratio (SNR) and coverage allowing high-resolution imaging, while the sensitive region of the coil covers an approximately 10 cm Field of View. Optimal Signal to Noise Ratio (SNR) and coverage and high resolution imaging Optimal Signal to Noise Ratio (SNR) and coverage, allowing highresolution imaging, while the sensitive region of the coil covers an approximately 15 cm Field of View. ## F. SAFETY AND EFFECTIVENESS Tursiop Technologies, L.L.C will comply with FDA's GMPs and the TT1 General Purpose Coil does comply with voluntary standards for safety/effectiveness (IEC 60601, UL 94) all of which mandate that components are tested to minimize hazards (electrical, mechanical, and flammability). A risk assessment (FMEA) has been conducted. Risk management practices will be utilized to assess potential risks throughout the device life cycle and mitigate unacceptable levels of risk. Tursiop Technologies, L.L.C. has conducted testing to establish the safety and effectiveness concerning the TT1 General Purpose Coil. Side by side comparison testing was conducted with the predicate device, GE 3-inch General Purpose Coil which demonstrated substantially equivalent performance. The TT1 General Purpose Coil does not introduce any new potential safety risks and operates in a manner similar to the predicate devices. #### G. SUBSTANTIAL EQUIVALENCE STATEMENT The TT1 General Purpose Coil is substantially equivalent to the GE 3-inch General Purpose Coil and the MRgFUS General Purpose and Breast Coil (K061715). This opinion is based on the fact that comparing the TT1 General Purpose Coil technological characteristics, coil architecture and operating principles with that of the predicate devices reveals that the devices comply with the same or equivalent standards and have the same or equivalent intended uses. {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular fashion around the symbol. #### Public Health Service Food and Drug Administration 9200 Corporate Bonlevard Rockville MD 20850 # DEC 2 3 2008 Tursion Technologies LLC % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 53313 Re: K083434 Trade/Device Name: TT1 General Purpose Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: December 6, 2008 Received: December 8, 2008 #### Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ #### Page 2 Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 | |----------------|---------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, hoque In Whang Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use 510(k) Number (if known): Koog 3434 Device Name: TT1 General Purpose Coil Indications for Use: The TT1 General Purpose Coil is a receive-only diagnostic MR imaging coil designed for use at a static magnetic field strength of 1.5T with the GE Signa® 1.5T system. It is used for obtaining diagnostic images of a variety of small-to-medium sized anatomical regions, such as jaw, spine, neck, shoulder, thigh, foot, ankle and joints. PrescriptionUse X (Part 21 CFR 801 Subpart D) AND/OR - Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...