K972990 · GE Medical Systems · LNH · Feb 5, 1998 · Radiology
Device Facts
Record ID
K972990
Device Name
MAGNETIC RESONANCE DIAGNOSTIC SYSTEM
Applicant
GE Medical Systems
Product Code
LNH · Radiology
Decision Date
Feb 5, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Intended Use
Diffusion Weighted EPI imaging produces magnetic resonance (MR) images whose contrast is dependent on the local diffusion coefficient of water. Diffusion Weighted EPI can be useful in visualizing the apparent loss of diffusion (mobility) by water molecules in brain tissues affected by acute stroke. Diffusion Weighted EPI is more accurate than conventional MRI pulse sequence (ie Fast FLAIR and Fast Spin Echo) in identifying the occurrence of acute stroke.
Device Story
Diffusion Weighted EPI Imaging Option is an add-on software feature for existing MRI systems; utilizes single-shot Echo Planar Imaging (EPI) pulse sequence. Input: MR signals; Output: MR images with contrast dependent on local water diffusion coefficients. Used in clinical radiology settings by trained MRI technologists/radiologists. Provides visualization of restricted water diffusion in brain tissue; assists clinicians in identifying acute stroke more accurately than conventional Fast FLAIR or Fast Spin Echo sequences. Benefits patient by enabling earlier or more precise stroke detection.
Clinical Evidence
Bench testing only. Evaluation performed against IEC 601-2-33 international medical equipment safety standard for Magnetic Resonance Systems. Accuracy of imaging performance verified against user manual specifications.
Technological Characteristics
Software-based imaging option for MRI systems. Utilizes single-shot Echo Planar Imaging (EPI) pulse sequence. Operates on existing MR hardware. Complies with IEC 601-2-33 safety standards.
Indications for Use
Indicated for patients requiring visualization of water molecule diffusion in brain tissue to identify acute stroke.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Siemens Medical System Diffusion Weighted MR Imaging Option (K971055)
Related Devices
K974530 — DUFFUSION-WEIGHTED MR IMAGING PACKAGE · Philips Medical Systems (Cleveland), Inc. · Feb 6, 1998
K971055 — DIFFUSION WEIGHTED MR IMAGING /MAGNETOM VISION · Siemens Medical Solutions USA, Inc. · Jun 20, 1997
K973658 — DIFFUSION PACKAGE / GYROSCAN NT · Philips Medical Systems North America, Inc. · May 6, 1998
K230016 — Cercare Medical Neurosuite · Cercare Medical A/S · Jul 14, 2023
Submission Summary (Full Text)
{0}------------------------------------------------
K972990
Image /page/0/Picture/1 description: The image shows a black and white logo. The logo is circular and contains the letters 'GE' in a stylized font. The letters are intertwined and appear to be handwritten. The logo is simple and recognizable.
## GE Medical Systems
PO Box 414. W-709 Milwaukee, WI 53201 USA
## SUMMARY OF SAFETY AND EFFECTIVENESS
FEB - 5 1998
- o This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
- o Identification of Submitter Larry A. Kroger, Ph.D., 414-544-3894, August 13, 1997
- o Identification of the Product Diffusion Weighted EPI Imaging Option
Manufactured by:
GE Medical Systems 3200 N. Grandview Blvd. Waukesha. Wi 53188
- o Device Description
The Diffusion Weighted EPI Imaging Option provides an additional imaging option to the Echo Planar Imaging pulse sequence. The DW-EPI option is a single shot EPI pulse designed to create images that differentiates tissues with restricted diffusion from tissues with normal diffusion.
o Indications for Use
> Diffusion Weighted EPI imaging produces magnetic resonance (MR) images whose contrast is dependent on the local diffusion coefficient of water. Diffusion Weighted EPI can be useful in visualizing the apparent loss of diffusion (mobility) by water molecules in brain tissues affected by acute stroke.
Diffusion Weighted EPI is more accurate than conventional MRI pulse sequence (ie Fast FLAIR and Fast Spin Echo) in identifying the occurrence of acute stroke.
- 0 Comparison with Predicate
The Diffusion Weighted EPI Imaging Option is substantially equivalent to the currently marketed Siemens Medical System Diffusion Weighted MR Imaging Option (510k #K971055).
- o Summary of Studies
The Diffusion Weighted EPI Imaging Option was evaluated to the IEC 601-2-33 International medical equipment safety standard for Magnetic Resonance Systems. Evaluation testing confirmed accuracy statements in the User Manual.
- o Conclusions
It is the opinion of GE that the Diffusion Weighted EPI Imaging Option does not result in any new potential hazards.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. The logo is simple and monochromatic.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Larry Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical System PO Box 414, W-709 Milwaukee, WI 53201
Re:
K972990 Diffusion Weighted EPI Imaging Option Dated: November 20, 1997 Received: November 21, 1997 Regulatory class: II 21 CFR 892.1000/Procode: 90 LNH
FEB - 5 1998
Dear Dr. Kroger:
We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. ................................................................................................................................
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (30) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html":. -------------------------------------------------------------------------------------------
Sincerely vours,
h. liau Yi
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Device Name: Diffusion Weighted EPI Imaging Option
Indications For Use:
- ﺬ-
Diffusion Weighted EPI imaging produces magnetic resonance (MR) images whose contrast is dependent on the local diffusion coefficient of water. Diffusion Weighted EPI can be useful in visualizing the apparent loss of diffusion (mobility) by water molecules in brain tissues affected by acute stroke.
Diffusion Weighted EPI is more accurate than conventional MRI pulse sequence (ie Fast FLAIR and Fast Spin Echo) in identifying the occurrence of acute stroke.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Wint le Seymon
on of Reproductive, Abdominal, E d Radiological De
510(k) Number K972990
OR
**Prescription Use**
(Per 21 CFR 801.109) *✓*
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.