1W/KG SAR(PART OF VERSION 6-05 OPERATING SYSTEM SOFTWARE)
K962048 · Hitachi Medical Systems America, Inc. · LNH · Aug 14, 1996 · Radiology
Device Facts
Record ID
K962048
Device Name
1W/KG SAR(PART OF VERSION 6-05 OPERATING SYSTEM SOFTWARE)
Applicant
Hitachi Medical Systems America, Inc.
Product Code
LNH · Radiology
Decision Date
Aug 14, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Intended Use
The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Device Story
Device is a software modification for Hitachi STRATIS and MRH-1500 MRI systems. Input: RF signals from proton nuclei in magnetic fields. Transformation: System software controls RF excitation and magnetic field gradients to generate spatial images based on T1, T2, proton density, and flow. Modification increases maximum Specific Absorption Rate (SAR) limit from 0.4 W/kg to 1.0 W/kg, aligning with IEC-601-2-33 standard for normal operating mode. Used in clinical settings by physicians/technicians. Output: Cross-sectional anatomical images (transverse, coronal, sagittal, oblique, curved). Clinical utility: Non-invasive diagnostic visualization of internal structures; aids in diagnosis determination. Benefit: Allows higher SAR levels for imaging while maintaining safety, consistent with international standards.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance is based on adherence to the IEC-601-2-33 standard for SAR limits.
Technological Characteristics
Magnetic Resonance Imaging system. Operates at fields up to 1.5T. RF frequencies up to 65 MHz. Software-controlled SAR limit of 1.0 W/kg. Imaging sequences include SE, FSE, IR, FIR, GFE, and MRA. Connectivity/materials not specified beyond standard MRI hardware components.
Indications for Use
Indicated for diagnostic imaging of the head, body, spine, and extremities in the general patient population to provide physiological and clinical information via T1/T2-weighted imaging, proton density measurements, and MR angiography.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Hitachi STRATIS with Version 3 Operating System Software
Hitachi MRH-1500 with Version 3 Operating System Software
Related Devices
K971279 — VERSION 4 OPERATING SYSTEM SOFTWARE · Hitachi Medical Systems America, Inc. · Jul 3, 1997
K961969 — HITACHI STRATIS, MRH-1500 MAGNETIC RESONANCE DIAGNOSTIC DEVICES VERSION 6 OPERATING SYSTEM SOFTWARE · Hitachi Medical Systems America, Inc. · Sep 5, 1996
K024042 — PANORAMA ENHANCEMENTS · Philips Medical Systems, Inc. · Feb 4, 2003
K981475 — OPART (MRT-600) VERSIONS 2 SOFTWARE UPGRADE · Toshiba America Mri, Inc. · Jul 14, 1998
K974498 — STRATIS II · Hitachi Medical Systems America, Inc. · Feb 26, 1998
Submission Summary (Full Text)
{0}
SIO(k) Summary
AUG 14 1996
K962048
## 1.0 SUBMITTER INFORMATION:
1.1 Submitter: Hitachi Medical Systems America
1963 Case Parkway
Twinsburg, OH 44087
PH: 216 425-1313
FX: 216 425-1410
1.2 Contact: James Jochen Rogers
1.3 Date: May 24, 1996
## 2.0 DEVICE NAME:
2.1 Magnetic Resonance Diagnostic Device
2.2 Classification Name: System, Nuclear Magnetic Resonance Imaging
2.3 Classification Number: 90LNH
2.4 Trade/Proprietary Name: 1W/kg SAR (Part of Version 6-05 Operatin System Software)
2.5 PREDICATE DEVICE(s):
Hitachi STRATIS with Version 3 Operating System Software
Hitachi MRH-1500 with Version 3 Operating System Software
## 3.0 DEVICE DESCRIPTION:
3.1 FUNCTION
The STRATIS / MRH-1500 Operating System Software is modified to change the maximum SAR limit from 0.4 W/kg to 1.0 W/kg as permitted under the International Electrotechnical Commission (IEC) standard Part 2: Particular Requirements for the Safety of Magnetic Resonance Equipment for Medical Diagnosis.
Because of the recent approval of the IEC-601-2-33 standard on MR safety, Hitachi seeks to take advantage of the acceptance of the higher SAR level permitted under that standard for the general patient population. We believe that the FDA has accepted the provisions of the IEC standard with respect to its SAR requirements, in compliance, we interpret the FDA's position with regard to the IEC standard for SAR as superceding its previous limit for SAR (up to 0.4 W/kg whole body is of no concern), and no further evidence is necessary beyond that given in the IEC standard and its rationale.
The Hitachi MRH-1500 and STRATIS MR Devices were originally cleared for marketing with an SAR limit of 0.4 W/kg (K943798, K922846B, and K945386), in compliance with the FDA's August 2, 1988 "Guidance for the Content and Review of a Magnetic Resonance Diagnostic Device 510(k) Application". In the FDA guidance, Safety Parameter Action Levels limit SAR to ≤ 3.2 W/kg averaged over the head, to ≤ 0.4 W/kg whole body, or to demonstrate that exposure to RF fields is insufficient to produce a core temperature increase in excess of 1°C and localized heating greater than 38°C in the head, 39°C in the trunk, and 40°C in the extremities.
{1}
The IEC standard, in defining the NORMAL OPERATING MODE, allows SAR values up to $1.5\mathrm{W / kg}$ under conditions of favorable environmental conditions (a scan room temperature $\leq 24^{\circ}\mathrm{C}$, scan room relative humidity $\leq 60\%$). However, the IEC standard dictates that maximum SAR values be derated up to a floor value of $1.0\mathrm{W / kg}$ for scan room temperature and humidity above these baseline environmental conditions, provided these ambient conditions are consistent with the overall device operating specifications. Since the MRH-1500 and STRATIS do not presently have bore temperature or humidity sensors, maximum SAR is derated to a maximum of $1.0\mathrm{W / kg}$. In order to comply with the NORMAL OPERATING MODE defined in the IEC standard, we at Hitachi propose that SAR control for the MRH-1500 and STRATIS MRI systems be limited to a maximum of $1.0\mathrm{W / kg}$, from the current limit of $0.4\mathrm{W / kg}$
No marketing claims will be made for the MRH-1500 or STRATIS stating compliance with the IEC standard. A separate future 510(k) premarket notification will describe full implementation of the IEC standard with respect to control of SAR, including, 1) the three operating modes (normal, first-level, and second-level, operating modes) as defined in the IEC standard, 2) modified clinical user interface through visual screens, and 3) control of access to the upper operating modes. However, the IEC standard permits operation of an MR device entirely within the NORMAL OPERATING MODE without any of these features.
## 3.2 SCIENTIFIC CONCEPTS
Magnetic Resonance (MR) is based on the fact that certain atomic nuclei have electromagnetic properties which cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nucleus used in current imaging experiments in magnetic resonance. When placed in a magnetic field, there is a slight net orientation or alignment of these atomic nuclei with the magnetic field. The introduction of a short burst of radiofrequency (RF) excitation of wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a reorientation of the proton's magnetization vector. When the RF excitation is removed, the proton relaxes and returns to its original orientation. The rate of relaxation is exponential, and varies with the character of the proton and its adjacent molecular environment. This reorientation process is characterized by two exponential relaxation times called T1 and T2 which can be measured.
These relaxation events are accompanied by an RF emission or echo which can be measured and used to develop a representation of these emissions on a three dimensional matrix. Spatial localization is encoded into the echo by varying the RF excitation and by appropriately applying magnetic field gradients in x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of NMR characteristics of the nuclei under consideration can be constructed by using image processing techniques similar to those used in CT.
For magnetic fields up to 1.5T, the RF frequencies commonly used range up to $65\mathrm{MHz}$. The RF fields have pulse powers from several watts to greater than 10 kilowatts, and repeat at rates from once every few seconds to greater than fifty per second. The time-varying magnetic gradient fields have a typical duration of sub-millisecond to several milliseconds.
{2}
## 3.3 PHYSICAL AND PERFORMANCE CHARACTERISTICS
MR is currently of great interest because it is capable of producing high quality anatomical images without the associated risks of ionizing radiation. In addition, the biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In x-ray imaging, differences in x-ray attenuation, largely based on differences in electro density are responsible for the contrast observed in x-ray images. In MR imaging, differences in proton density, blood flow, and relaxation times T1 and T2 all may contribute to image contrast. In addition, by varying the duration and spacing of the RF pulses, images may be produced in which the contrast is primarily dependent on T1 relaxation, T2 relaxation, proton density, or a combination of all three.
## 4.0 DEVICE INTENDED USE:
The MR system is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
- Anatomical Region: Head, Body, Spine, Extremities
- Nucleus excited: Proton
- Diagnostic uses:
- 2D T1- / T2-weighted imaging
- T1, T2, proton density measurements
- MR Angiography
- image processing
- Imaging capabilities:
- 2D, 3D Spin Echo (SE)
- 2D Short Spin Echo (SES)
- 2D, 3D Fast Spin Echo (FSE)
- 2D Inversion Recovery (IR)
- 2D, 3D Fast Inversion Recovery (FIR)
- 2D,3D Gradient Field Echo (GFE); also with rephasing
- 2D, 3D Rapid Scan (RS)
- MTC, RF Spoiling
- MR Angiography, (2D INFA, 3D INFA, 2D GFEA, 3D GFEA, Sloped Slab Profile (SSP))
- RF Coil Uniformity
- Adaptive Image post-processing
## 5.0 DEVICE TECHNOLOGICAL CHARACTERISTICS:
Identical to the Predicate Device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.