Swoop Portable MR Imaging System

K221393 · Hyperfine, Inc. · LNH · Jun 10, 2022 · Radiology

Device Facts

Record IDK221393
Device NameSwoop Portable MR Imaging System
ApplicantHyperfine, Inc.
Product CodeLNH · Radiology
Decision DateJun 10, 2022
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesAI/ML, Pediatric

AI Performance

OutputAcceptanceObservedDev DSTest DS
MRI Image Reconstruction (Noise and Blurring Reduction)Image performance with advanced reconstruction must meet all image quality criteria per NEMA MS 1-2008, MS 3-2008, MS 9-2008, MS 12-2016, and ACR Phantom Test Guidance.Passed all testing in accordance with internal requirements and applicable standards.Not specified.Verification and validation testing including software verification (testing to verify advanced reconstruction models do not alter image features or introduce artifacts) and image performance testing using NEMA and ACR phantom standards.

Indications for Use

The Swoop® Portable MR Imaging System is a bedside magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Device Story

Swoop® Portable MR Imaging System is a bedside MRI device for head imaging; used when full diagnostic MRI is not practical. System uses a 63.3 mT permanent magnet to acquire MR signals; data processed via tablet-based interface. Features include modified pulse sequences and enhanced noise correction to remove residual line noise. Operated by clinicians at point-of-care (emergency rooms, critical care, hospital/rehabilitation rooms). System generates T1W, T2W, FLAIR, and DWI image sets. Deep learning-based reconstruction algorithm reduces noise and blurring. Clinicians view images on tablet for quality control and diagnostic assessment; output assists in clinical decision-making by providing internal head structure visualization.

Clinical Evidence

Bench testing only. Performance verified via software verification, image quality assessment using NEMA MS standards and ACR phantom guidance, and cybersecurity testing. Biocompatibility, cleaning/disinfection, and electrical safety/EMC testing leveraged from predicate device.

Technological Characteristics

Permanent magnet (63.3 mT); 24 mT/m gradient strength; 22 T/m/s slew rate. Form-fitting linear volume head coil (TX/RX). Tablet-based interface. Software includes deep learning-based image reconstruction (Advanced Gridding, Conjugate Gradient) and advanced denoising. Standards: IEC 62304, NEMA MS 1/3/9/12, ISO 10993, ANSI/AAMI ES 60601-1, IEC 60601-1-2.

Indications for Use

Indicated for adult and pediatric patients (≥ 0 years) requiring bedside magnetic resonance imaging of the head where full diagnostic examination is not clinically practical.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. June 10, 2022 Hyperfine, Inc. % Christine Kupchick Sr. Regulatory Specialist 351 New Whitfield Street GUILFORD CT 06437 Re: K221393 Trade/Device Name: Swoop® Portable MR Imaging System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, MOS Dated: May 23, 2022 Received: May 24, 2022 Dear Christine Kupchick: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Michael D. O'Hara, Ph.D. Deputy Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221393 #### Device Name Swoop® Portable MR Imaging System #### Indications for Use (Describe) The Swoop® Portable MR Imaging System is a bedside magnetic resonance imaging device for producing images that display the internal structure of the bead where full diagnostic examination is not climically practical. When interpreted by a trained physician, these images provide information that can be useful in letermining a diagnois. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) PSC Publishing Services (301) 443-6740 PP {3}------------------------------------------------ # HYPERFINE # 510(k) Summary Swoop® Portable MR Imaging System K221393 # 510(k) Submitter | Company Name: | Hyperfine, Inc. | |------------------|----------------------| | Company Address: | 351 New Whitfield St | | | Guilford, CT 06437 | # Contact | Name: | Christine Kupchick | |------------|------------------------| | Telephone: | (203) 343-3404 | | Email: | ckupchick@hyperfine.io | Date Prepared: June 10, 2022 #### Device Identification | Trade Name: | Swoop® Portable MR Imaging System | |----------------------|-----------------------------------------------------------------------------------| | Common Name: | Magnetic Resonance Imaging | | Regulation Number: | 21 CFR 892.1000 | | Classification Name: | System, Nuclear Magnetic Resonance Imaging Coil, Magnetic Resonance,<br>Specialty | | Product Code: | LNH; MOS | | Regulatory Class: | Class II | #### Predicate Device Information The subject Swoop® Portable MR Imaging System is substantially equivalent to the predicate Swoop® System (K212456). #### Device Description The Swoop® system is a portable MRI device that allows for patient bedside imaging. The system enables visualization of the internal structures of the head using standard magnetic resonance imaging contrasts. The main interface is a commercial off-the-shelf device that is used for operating the system, providing access to patient data, exam setup, exam execution, viewing MRI image data for quality control purposes, and cloud storage interactions. The system can generate MRI data sets with a broad range of contrasts. The Swoop® system user interface includes touch screen menus, controls, indicators, and navigation icons that allow the operator to control the system and to view imagery. The Swoop® {4}------------------------------------------------ System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, and FLAIR sequences, specifically in terms of reductions in image noise and blurring. This subject device in this submission includes modified pulse sequence options and an enhancement to the existing noise correction feature to remove residual line noise. #### Indications for Use The Swoop® Portable MR Imaging System is a bedside magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. #### Substantial Equivalence Discussion The table below compares the subject device to the predicate. | Specification | Subject Swoop® System | Predicate Swoop® System (K212456) | |-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------| | Intended Use/Indications for Use: | The Swoop® Portable MR Imaging<br>System is a bedside magnetic<br>resonance imaging device for<br>producing images that display the<br>internal structure of the head where<br>full diagnostic examination is not<br>clinically practical. When interpreted<br>by a trained physician, these images<br>provide information that can be<br>useful in determining a diagnosis. | Same | | Patient Population: | Adult and pediatric patients (≥ 0<br>years) | Same | | Anatomical Sites: | Head | Same | | Environment of Use: | At the point of care in medical<br>facilities, including emergency rooms,<br>critical care units, hospital or<br>rehabilitation rooms. | Same | | Energy Used and/or delivered: | Magnetic Resonance | Same | | Magnet: | | | | Physical Dimensions | 835 mm x 630 mm x 652 mm | Same | | Bore Opening | 610 mm x 315 mm | Same | | Weight | 320 kg | Same | | Field Strength | 63.3 mT permanent magnet | Same | | Gradient: | | | | Strength | 24 mT/m | Same | | Rise Time | 1.1 ms | Same | | Slew Rate | 22 T/m/s | Same | | Computer Display | Hyperfine-supplied tablet | Same | | RF Coils: | | | | Number of Coils | 1 head coil | Same | | Coil Type | TX/RX | Same | | Coil Geometry | Form-fitting | Same | {5}------------------------------------------------ | Specification | Subject Swoop® System | Predicate Swoop® System (K212456) | |--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------| | Inner Dimensions (mm) | 205 mm x 240 mm | Same | | Coil Design | Linear Volume | Same | | Patient Weight Capacity | 200 kg | Same | | Operation Temperature | 15-30 C | Same | | Warm Up Time | <3 minutes | Same | | Temperature Control | No | Same | | Humidity Control | No | Same | | Image Reconstruction Algorithm | | | | T1W<br>- T1-Standard<br>- T1-Gray/White Contrast | Advanced Gridding | Advanced Gridding<br>(T1-Gray/White Contrast only) | | T2W<br>- T2<br>- T2-Fast | Advanced Gridding | Advanced Gridding (T2 only) | | FLAIR | Advanced Gridding | Same | | DWI | Conjugate Gradient | Same | | Image Post-Processing | - Advanced Denoising (applies to T1W, T2W, and FLAIR only)<br>- image orientation transform<br>- Geometric distortion correction<br>- Receive coil intensity correction<br>- DICOM output | Same | The subject device and the predicate device have intended use, operating principles, and similar technological characteristics. The subject device differs from the predicate in pulse sequence options and noise correction. These differences do not raise new questions of safety and effectiveness as compared to the predicate. # Non-Clinical Performance As part of demonstrating substantial equivalence to the predicate, a risk-based assessment was completed to identify the risks associated with the modifications. Based on the risk assessment, the following verification and validation testing was performed. The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence. | Test | Test Description | Applicable Standard(s) | |--------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Software<br>Verification | Testing to verify that the advanced reconstruction<br>models do not alter image features or introduce<br>artifacts. | ● IEC 62304:2006<br>● FDA Guidance, “Guidance for the<br>Content of Premarket Submissions<br>for Software Contained in Medical<br>Devices” | | | Testing to verify that image quality with advanced<br>reconstruction is acceptable. | | | | Testing to verify basic software functionality is<br>unchanged between releases. | | | | NESSUS scan test to verify any vulnerabilities and<br>serve as a security baseline. | | {6}------------------------------------------------ | Image<br>Performance | Testing to verify image performance with advanced<br>reconstruction meets all image quality criteria. | NEMA MS 1-2008 (R2020) NEMA MS 3-2008 (R2020) NEMA MS 9-2008 (R2020) NEMA MS 12-2016 American College of Radiology (ACR)<br>Phantom Test Guidance for Use of<br>the Large MRI Phantom for the ACR<br>MRI Accreditation Program American College of Radiology<br>standards for named sequences | |------------------------|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Software<br>Validation | Validation studies to ensure the device meets user<br>needs and performs as intended. | FDA Guidance, “Guidance for the<br>Content of Premarket Submissions<br>for Software Contained in Medical<br>Devices” | | Cybersecurity | Testing to verify cybersecurity controls and<br>management. | Cybersecurity as recommended in<br>FDA guidance, “Content of<br>Premarket Submissions for<br>Management of Cybersecurity in<br>Medical Devices” | The following testing was leveraged from the predicate device. Test results from the predicate were used to support the subject device because the conditions were identical or the subject device modifications did not introduce a new worst-case configuration or scenario for testing. | Test | Test Description | Applicable Standard(s) | |---------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------| | Biocompatibility | Biocompatibility testing of patient-contacting materials. | ISO 10993-1:2018 ISO 10993-5:2009 ISO 10993-10:2010 | | Cleaning/<br>Disinfection | Cleaning and disinfection validation of patient-contacting materials. | FDA Guidance, "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" ISO 17664:2017 ASTM F3208-17 | | Safety | Electrical Safety, EMC, and Essential Performance testing. | ANSI/AAMI ES 60601-1:2005/(R)2012 IEC 60601-1-2:2014 IEC 60601-1-6:2013 | | Performance | Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems. | NEMA MS 8-2016 | # Conclusion Based on the indications for use, technological characteristics, performance results, and comparison to the predicate, the subject Swoop® Portable MR Imaging System has been shown to be substantially equivalent to the predicate device identified in this submission and does not present any new issues of safety or effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...