MAGNETOM Lumina and MAGNETOM Vida Fit with syngo MR XA50A

K220939 · Siemens Medical Solutions USA, Inc. · LNH · Apr 29, 2022 · Radiology

Device Facts

Record IDK220939
Device NameMAGNETOM Lumina and MAGNETOM Vida Fit with syngo MR XA50A
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeLNH · Radiology
Decision DateApr 29, 2022
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesAI/ML

Intended Use

The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/ or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.

Device Story

MAGNETOM Lumina and Vida Fit are MR diagnostic systems using magnetic resonance to generate cross-sectional images/spectra of head, body, and extremities. Operated by trained physicians/technicians in clinical settings. Inputs include RF signals and magnetic field gradients; processed via syngo MR XA50A software to produce diagnostic images. Features include deep learning-based reconstruction (Deep Resolve Boost/Swift Brain), automated workflows (myExam Prostate/Angio Assist), and advanced pulse sequences (GRE_WAVE, HASTE, BLADE). Vida Fit includes Nexaris hardware for interventional/intraoperative use. Output is visualized by clinicians to assist in diagnosis. Benefits include improved image quality, faster acquisition, and streamlined workflows via automated slice positioning and motion correction.

Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including verification and validation of transferred hardware and software features in accordance with IEC 62304 and FDA guidance.

Technological Characteristics

MR diagnostic system; 21 CFR 892.1000. Features deep learning-based image reconstruction, automated workflow software, and specialized pulse sequences. Hardware includes Nexaris dockable tables and coils. Connectivity via DICOM. Conforms to IEC 60601-1, IEC 60601-2-33, ISO 14971, and NEMA MS standards. Software lifecycle per IEC 62304.

Indications for Use

Indicated for patients requiring diagnostic MR imaging of the head, body, or extremities, including interventional procedures. Applicable to all patient populations where MR imaging is not contraindicated.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Siemens Medical Solutions USA, Inc. % Alina Goodman Regulatory Affairs Professional 40 Liberty Blvd. MALVERN PA 19355 Re: K220939 Trade/Device Name: MAGNETOM Lumina and MAGNETOM Vida Fit with syngo MR XA50A Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LNI, MOS Dated: March 30, 2022 Received: March 31, 2022 Dear Alina Goodman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ 2 devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for CAPT Patrick Hintz, MSIH, CIH, USPHS Assistant Director Eletronic Products Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220939 Device Name MAGNETOM Lumina and MAGNETOM Vida Fit with syngo MR XA50A #### Indications for Use (Describe) The MAGNETOM system is indicated for use as a magnetic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These inages and/ or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician vield information that may assist in diagnosis. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a pattern of orange dots. # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR § 807.92. #### 1. General Information | Establishment: | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Malvern, PA 19355, USA<br>Registration Number: 2240869 | |----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date Prepared: | March 30, 2022 | | Manufacturer: | Siemens Healthcare GmbH<br>Henkestr. 127<br>91052 Erlangen<br>Germany<br>Registration Number: 3002808157<br><br>Siemens Shenzhen Magnetic Resonance LTD<br>Siemens MRI Center<br>Hi-Tech Industrial park (middle)<br>Gaoxin C. Ave., 2nd<br>Shenzhen 518057<br>P.R. CHINA | #### 2. Contact Information Alina Goodman Regulatory Affairs Professional Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355, USA +1(224)526-1404 Phone: E-mail: alina.goodman@siemens-healthineers.com Registration Number: 3004754211 {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots. #### 3. Device Name and Classification | Device/ Trade name: | MAGNETOM Lumina and MAGNETOM Vida Fit with<br>syngo MR XA50A | |-----------------------|--------------------------------------------------------------| | Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) | | Classification Panel: | Radiology | | CFR Code: | 21 CFR § 892.1000 | | Classification: | II | | Product Code: | Primary: LNH<br>Secondary: LNI, MOS | #### 4. Legally Marketed Predicate Devices | Trade name: | MAGNETOM Lumina with syngo MR XA31A | |-----------------------|---------------------------------------------| | 510(k) Number: | K203443 | | Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) | | Classification Panel: | Radiology | | CFR Code: | 21 CFR § 892.1000 | | Classification: | II | | Product Code: | Primary: LNH<br>Secondary: LNI, MOS | | Trade name: | MAGNETOM Vida Fit with syngo MR XA20A | | Trade name: | MAGNETOM Vida Fit with syngo MR XA20A | |-----------------------|---------------------------------------------| | 510(k) Number: | K192924 | | Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) | | Classification Panel: | Radiology | | CFR Code: | 21 CFR § 892.1000 | | Classification: | II | | Product Code: | Primary: LNH<br>Secondary: LNI, MOS | #### 5. Indications for use The indications for use for the subject devices are the same as the predicate device: The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional imaqes, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots in a circular pattern. The MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles. # 6. Device Description MAGNETOM Lumina and MAGNETOM Vida Fit with software syngo MR XA50A include new software compared to the predicate devices, MAGNETOM Vida Fit with software syngo MR XA20A (K192924) and MAGNETOM Lumina with syngo MR XA31A (K203443). This software and some hardware components are transferred from the reference device MAGNETOM Vida with software syngo MR XA50A (K213693) as well as an imaging feature from MAGNETOM Vida with software syngo MR XA11A (K181433). A high-level summary of the transferred hardware and software is provided below: #### Hardware (Vida Fit only) #### Transferred Hardware: - The Nexaris Dockable Table is a new variant of the MR patient table which is used for intraoperative or interventional imaging. It enables the patient transfer between OR tables and the MR system without repositioning on the MR patient table and vice versa during interventional procedures and surgeries. Additionally, it can be used for diagnostic imaging. - The Nexaris Head Frame holds up to two Ultra Flex Large 18 coils. It can be used for head imaging in combination with the Nexaris Dockable Table when the patient is positioned on the transfer board but not pinned in a head clamp. - Transferred MaRS Computer ### Transferred Coil: - The Nexaris Spine 36 is used in combination with and without transfer board for body imaging on the Nexaris Dockable Table. #### Transferred modifications for hardware: - The Beat Sensor is a contact less method for generating cardiac triggers as an alternative to the already existing ECG or pulse triggers. It is based on a measurement of the modulation of a weak magnetic Pilot Tone, caused by conformation changes in conductive tissues. #### Software Transferred Features and Applications: Vida Fit only: {6}------------------------------------------------ Image /page/6/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern. - SVS EDIT is a special variant of the SVS SE pulse sequence type, which acquires two different spectra (one with editing pulses on resonance, one with editing pulses off resonance) within a single sequence. - BEAT FQ nav allows the user to make use of navigator echo based respiratory gating for flow imaging to acquire 4D flow data. Both navigator echo based respiratory gating as well as flow imaging are part of the predicate device already. New is merely the combination of both. - The HASTE interactive pulse sequence type extends the existing HASTE pulse sequence type by offering the possibility to interactively change imaging parameters. - GRE_WAVE is a special variant of the GRE pulse sequence type which allows larger acceleration factors, measuring one or two contrasts. GRE Wave results in higher signal-to-noise ratio for larger acceleration factors which can be leveraged to allow fast high-resolution 3D susceptibility-weighted imaging. - The myExam Prostate Assist provides an assisted and quided workflow for prostate imaging. This automated workflow leads to higher reproducibility of slice angulation and coverage; this may support exams not having to be repeated. - Iniector coupling is a software application that allows the connection of certain contrast agent injectors to the MR system for simplified, synchronized contrast injection and examination start. ### Lumina onlv: - Compressed Sensing GRASP-VIBE is intended to be used in dynamic and/or non-contrast liver examinations to support patients who cannot reliably hold their breath for a conventional breath-hold measurement. ### Lumina and Vida Fit: - Deep Resolve Swift Brain is a protocol for fast routine brain imaging primarily based on echo planar imaging (EPI) pulse sequences. Its main enablers are multi-shot (ms) EPI pulse sequence types and a deep learning-based image reconstruction. - Deep Resolve Boost is a novel deep learning-based image reconstruction alqorithm for 2D TSE data, which reconstructs images from k-space raw-data. - BLADE diffusion is a multi-shot imaging method based on TSE or TGSE (when EPI factor > 1) readout and a BLADE trajectory with diffusion preparation to enable diffusion weighted imaging with reduced sensitivity to B0 inhomogeneity and reduced T2 decay caused image blurring. - HASTE diffusion (HASTE DIFF) is a single-shot imaging method based on TSE readout with diffusion preparation to enable diffusion weighted imaging with reduced sensitivity to B0 inhomogeneity. Transferred Modifications for Features and Applications: #### Vida Fit only: {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. - The AbsoluteShim mode is a shimming procedure based on a 3-echo gradient echo protocol. - The 3D ASL sequence (tgse_asl) now provides relCBF maps, by implementing an additional M0 scan and performing the corresponding reconstruction method. It also provides BAT maps in multiple inversion time(multi-TI) imaging. #### Lumina and Vida Fit: - Fast GRE RefScan: A speed-optimized reference scan for GRAPPA and SMS kernel calibration for echo planar imaging pulse sequence types. - Static Field Correction is a reconstruction option reducing susceptibilityinduced distortions and intensity variations. - Deep Resolve Sharp is an interpolation algorithm which increases the perceived sharpness of the interpolated images. Functionality is available for different pulse sequence types. (Newly transferred to Vida Fit) - Deep Resolve Gain is a reconstruction option which improves the SNR of the scanned imaqes. Functionality is available for different pulse sequence types. (Newly transferred to Vida Fit) - The myExam Angio Advanced Assist provides an assisted and quided workflow for peripheral angiography examination using care bolus. The main advantage of this new workflow is a simplified and improved planning procedure of multi-station peripherical angiography measurements. # Other transferred Modifications and / or Minor Changes Vida Fit only: - Elastography-AddIn synchronizes settings between the Elastography sequence and the active driver. - HASTE MoCo is an image-based motion correction in the average-dimension for the HASTE pulse sequence type. - Coil independent pulse sequences remove the coil information from the pulse sequences and generate this information during run-time from automatic coil detection and localization. - The Needle Intervention AddIn provides a user interface for workflow improvement of MR-quided needle interventions under real-time imaging conditions. It supports planning a needle trajectory, laser-based localization of the entry point as well as automatic slice positioning. - The PhaseRev Dot Addin/Component supports the measurement workflow of the user by automatically flipping the direction of the phase encoding gradient. - The adjustment mode "offcenter" triggers a transmitter adjustment method that is specialized for offcenter imaging. The transmitter adjustment determines the RF voltage that is required to excite a certain B1 field. ### Lumina and Vida Fit: {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a cluster of orange dots. - TSE MoCo is an image-based motion correction in the average-dimension for the TSE pulse sequence type. - MR Breast Biopsy is improved with an automatic fiducial detection. ### 7. Substantial Equivalence MAGNETOM Lumina and MAGNETOM Vida Fit with software syngo MR XA50A are substantially equivalent to the following predicate devices: | Predicate Device | FDA Clearance Number<br>and Date | Product<br>Code | Manufacturer | |------------------------------------------|---------------------------------------|-----------------|----------------------------| | MAGNETOM Lumina with<br>syngo MR XA31A | K203443, cleared March<br>31, 2021 | LNH<br>LNI, MOS | Siemens Healthcare<br>GmbH | | MAGNETOM Vida Fit with<br>syngo MR XA20A | K192924, cleared on March<br>11, 2020 | LNH<br>LNI, MOS | Siemens Healthcare<br>GmbH | MAGNETOM Lumina and MAGNETOM Vida Fit with software syngo MR XA50A consist of hardware and software features previously cleared with the following reference device: | Reference Device | FDA Clearance Number<br>and Date | Product<br>Code | Manufacturer | |--------------------------------------|---------------------------------------|-----------------|----------------------------| | MAGNETOM Vida with<br>syngo MR XA50A | K213693, cleared February<br>25, 2022 | LNH<br>LNI, MOS | Siemens Healthcare<br>GmbH | | MAGNETOM Vida with<br>syngo MR XA11A | K181433, cleared October<br>19, 2018 | LNH<br>LNI, MOS | Siemens Healthcare<br>GmbH | ### 8. Comparison of technological characteristics with the predicate devices The subject devices, MAGNETOM Lumina and MAGNETOM Vida Fit with software syngo MR XA50A, are substantially equivalent to the predicate devices with regard to the operational environment, programming language, operating system and performance. The subject devices conform to the standard for medical device software (IEC 62304) and other relevant IEC and NEMA standards. There are some differences in technological characteristics between the subject devices and predicate devices, including transferred hardware and software from the reference device. The transferred features and components have been tested and the conclusion from the non-clinical data suggests that the features bear an equivalent safety and performance profile to that of the predicate devices. ### 9. Nonclinical Tests The following performance testing was conducted on the subject devices. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern. | Performance Test | Tested Hardware or Software | Source/Rationale for test | |-----------------------------|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | Verification and validation | Transferred hardware and<br>software features | Guidance for the Content of<br>Premarket Submissions for<br>Software Contained in Medical<br>Devices / 21 CFR §820.30 | The results from each set of tests demonstrate that the devices perform as intended and are thus substantially equivalent to the predicate devices to which they have been compared. ### 10.Clinical Tests No additional clinical tests were conducted to support substantial equivalence for the subject devices. ### 11.Safety and Effectiveness The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device. Risk Management is ensured via a risk analysis in compliance with ISO 14971, to identify and provide mitigation of potential hazards early in the design cycle and continuously throughout the development of the product. Siemens Healthcare GmbH adheres to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images. MAGNETOM Lumina and MAGNETOM Vida Fit with software syngo MR XA50A conform to the following FDA recognized and international IEC, ISO and NEMA standards: | Recognition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and date | Standards<br>Development<br>Organization | |-----------------------|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|------------------------------------------| | 19-4 | General II<br>(ES/<br>EMC) | C1:2009/(R)2012 and<br>A2:2010/(R)2012 (Consolidated<br>Text) Medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance (IEC<br>60601-1:2005, MOD) | ES60601-<br>1:2005/(R)2012<br>and A1:2012 | ANSI AAMI | {10}------------------------------------------------ Image /page/10/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the word "Healthineers" is a graphic of orange dots. | 12-295 | Radiology | Medical electrical equipment -<br>Part 2-33: Particular<br>requirements for the basic<br>safety and essential<br>performance of magnetic<br>resonance equipment for<br>medical diagnosis | 60601-2-33 Ed. 3.2<br>b:2015 | IEC | |--------|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|------------------| | 5-40 | General I<br>(QS/<br>RM) | Medical devices - Application of<br>risk management to medical<br>devices | 14971 Second<br>edition 2007-03-01 | ISO | | 5-114 | General I<br>(QS/<br>RM) | Medical devices - Part 1:<br>Application of usability<br>engineering to medical devices | 62366-1:2015 | ANSI AAMI<br>IEC | | 13-79 | Software/<br>Informatics | Medical device software -<br>Software life cycle processes<br>[Including Amendment 1 (2016)] | 62304:2006/A1:2016 | ANSI AAMI<br>IEC | | 12-300 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set 03/16/2012<br>Radiology | PS 3.1 - 3.20<br>(2016) | NEMA | | 19-8 | General II<br>(ES/<br>EMC) | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>Standard: Electromagnetic<br>disturbances - Requirements<br>and tests | 60601-1-2 Edition<br>4.0 2014-02 | IEC | | 12-331 | Radiology | Characterization of<br>Radiofrequency (RF) Coil<br>Heating in Magnetic Resonance<br>Imaging Systems | Standards<br>Publication MS 14-<br>2019 | NEMA | | 12-232 | Radiology | Acoustic Noise Measurement<br>Procedure for Diagnosing<br>Magnetic Resonance Imaging<br>Devices | MS 4-2010 | NEMA | | 12-288 | Radiology | Standards Publication<br>Characterization of Phased<br>Array Coils for Diagnostic<br>Magnetic Resonance Images | MS 9-2008 (R2014) | NEMA | ### 12.Conclusion as to Substantial Equivalence MAGNETOM Lumina and MAGNETOM Vida Fit with software syngo MR XA50A have the same intended use and same basic technological characteristics than the predicate device systems, MAGNETOM Lumina with syngo MR XA31A and MAGNETOM Vita Fit with sygno XA20A, with respect to the magnetic resonance features and functionalities. While there are some differences in technical features compared to the predicate device, the transfer of software features has been tested and the conclusion from all verification and validation data suggest {11}------------------------------------------------ Image /page/11/Picture/1 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a cluster of orange dots. that the features bear an equivalent safety and performance profile to that of the predicate and reference devices. Siemens believes that MAGNETOM Lumina and MAGNETOM Vida Fit with software syngo MR XA50A are substantially equivalent to the currently marketed device MAGNETOM Lumina with syngo MR XA31A (K203443, cleared on March 31, 2021) and MAGNETOM Vida Fit with software syngo MR XA20A (K192924, cleared on March 11, 2020).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...