Swoop Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System
Applicant
Hyperfine, Inc.
Product Code
LNH · Radiology
Decision Date
Jul 7, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Swoop Point-of-Care Magnetic Resonance Imaging Device is a bedside magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Device Story
Portable MRI scanner for bedside head imaging; operates via 63.3 mT permanent magnet. System inputs include patient data and scan parameters via touchscreen interface; transforms magnetic resonance signals into head imagery. Features automatic alignment and motion correction software. Used in clinical settings by trained operators; images viewed by physicians for diagnostic support. Benefits include accessibility to MRI in environments where traditional scanners are impractical. System includes cloud storage interactions and user-supplied tablet interface.
Clinical Evidence
Bench testing only. Performance evaluated via software verification (IEC 62304), cybersecurity assessment, and leveraged testing from the predicate device (K201722) including biocompatibility (ISO 10993-1), electrical safety/EMC (ANSI/AAMI ES 60601-1, IEC 60601-2-33, IEC 60601-1-2), usability (IEC 60601-1-6), and NEMA MS standards for SNR, image uniformity, SAR, phased array coils, and geometric distortion.
Technological Characteristics
Portable MRI system; 63.3 mT permanent magnet; 24 mT/m gradient strength; 22 T/m/s slew rate. Form-fitting linear volume head coil (TX/RX). Dimensions: 835mm x 630mm x 652mm; Weight: 320kg. Connectivity: Cloud storage, user-supplied tablet interface. Software includes automatic alignment and motion correction. Standards: IEC 62304, ANSI/AAMI ES 60601-1, IEC 60601-2-33, IEC 60601-1-2, ISO 10993-1, NEMA MS series.
Indications for Use
Indicated for bedside MRI imaging of the head in adult and pediatric patients (≥ 0 years) where full diagnostic examination is not clinically practical.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K201722 — Hyperfine Point-Of-Care Magnetic Resonance Imaging (POC MRI) Scanner System · Hyperfine Research, Inc. · Aug 11, 2020
K192002 — Lucy Point-of-Care Magnetic Resonance Imaging Device · Hyperfine Research, Inc. · Feb 6, 2020
K230208 — Swoop® Portable MR Imaging System · Hyperfine, Inc. · Feb 22, 2023
K221923 — Swoop Portable MR Imaging System · Hyperfine, Inc. · Jul 28, 2022
K232644 — neuro42 MRI System · Neuro42, Inc. · Feb 8, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
July 7, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Hyperfine, Inc. % Ms. Christine Kupchick Sr. Regulatory Specialist 530 Old Whitfield Street Guilford, Connecticut 06437
## Re: K211818
Trade/Device Name: Swoop™ Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, MOS Dated: June 10, 2021 Received: June 11, 2021
## Dear Ms. Kupchick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K211818
Device Name
Swoop™ Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System
Indications for Use (Describe)
The Swoop Point-of-Care Magnetic Resonance Imaging Device is a bedside magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
PSC Publishing Services (301) 443-6740 FF
{3}------------------------------------------------
# HYPERFINE
# 510(k) SUMMARY K211818
#### 510(k) Submitter
| Company Name: | Hyperfine, Inc. |
|------------------|--------------------------------------------|
| Company Address: | 530 Old Whitfield St<br>Guilford, CT 06437 |
#### Contact
| Name: | Christine Kupchick |
|------------|------------------------|
| Telephone: | (203) 343-3404 |
| Fax: | (203) 458-2514 |
| Email: | ckupchick@hyperfine.io |
July 2, 2021 Date Prepared:
#### Device Identification
| Trade Name: | Swoop™ Point-of-Care Magnetic Resonance Imaging (POC MRI) Scanner System |
|----------------------|-----------------------------------------------------------------------------------|
| Common Name: | Magnetic Resonance Imaging |
| Regulation Number: | 21 CFR 892.1000 |
| Classification Name: | System, Nuclear Magnetic Resonance Imaging<br>Coil, Magnetic Resonance, Specialty |
| Product Code: | LNH; MOS |
| Regulatory Class: | Class II |
#### Predicate Device Information
The subject Swoop POC MRI Scanner System is substantially equivalent to the predicate POC MRI Scanner System (K201722).
The predicate device has not been subject to a design-related recall.
#### Device Description
The Swoop POC MRI Scanner System is a portable MRI device that allows for patient bedside imaging. The system enables visualization of the internal structures of the head using standard magnetic resonance imaging contrasts. The main interface is a commercial off-the-shelf device that is used for operating the system, providing access to patient data, exam execution, viewing MRI image data for quality control purposes, and cloud storage interactions. The system can generate MRI data sets with a broad range of contrasts. The Swoop POC MRI Scanner System user interface includes touchscreen menus, controls, indicators and navigation icons that allow the operator to control the system and to view imagery.
{4}------------------------------------------------
The purpose of this submission is to gain clearance for updates to the software to include automatic alignment and motion correction features.
#### Indications for Use
The Swoop Point-of-Care Magnetic Resonance Imaging Device is a bedside magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
#### Substantial Equivalence Discussion
The table below compares the subject device to the predicate.
| Specification | Subject Swoop Portable MRI Device | Predicate POC MRI Device (K201722) |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Swoop Point-of-Care Magnetic<br>Resonance Imaging Device is a bedside<br>magnetic resonance imaging device<br>for producing images that display the<br>internal structure of the head where<br>full diagnostic examination is not<br>clinically practical. When interpreted<br>by a trained physician, these images<br>provide information that can be useful<br>in determining a diagnosis. | The Point-of-Care Magnetic<br>Resonance Imaging Device is a bedside<br>magnetic resonance imaging device<br>for producing images that display the<br>internal structure of the head where<br>full diagnostic examination is not<br>clinically practical. When interpreted<br>by a trained physician, these images<br>provide information that can be useful<br>in determining a diagnosis. |
| Patient Population | Adult and pediatric (≥ 0 years) | Adult and pediatric (≥ 0 years) |
| Anatomical Site | Head | Head |
| Patient Weight | 200 kg | 200 kg |
| Capacity | | |
| Energy Type | Magnetic Resonance | Magnetic Resonance |
| Operation<br>Temperature | 15-30C | 15-30C |
| Warm Up Time | <3 min | <3 min |
| MAGNET | | |
| Physical Dimensions | 835 mm x 630 mm x 652 mm | 835 mm x 630 mm x 652 mm |
| Bore Opening | 610 mm x 315 mm | 610 mm x 315 mm |
| Weight | 320 kg | 320 kg |
| Field Strength | 63.3 mT permanent magnet | 63.3 mT permanent magnet |
| GRADIENT | | |
| Strength | 24 mT/m | 24 mT/m |
| Rise Time | 0.4 ms | 0.4 ms |
| Slew Rate | 22 T/m/s permanent magnet | 22 T/m/s permanent magnet |
| COMPUTER | | |
| Display | User supplied tablet | User supplied tablet |
| RF COIL | | |
| RF Coils | 1 Head Coil | 1 Head Coil |
| Coil Type | TX/RX | TX/RX |
| Coil Geometry | Form-Fitting | Form-Fitting |
{5}------------------------------------------------
| Inner Dimensions | 205 mm x 240 mm | 205 mm x 240 mm |
|------------------|-----------------|-----------------|
| Coil Design | Linear Volume | Linear Volume |
The subject device and the predicate device have intended use, operating principles and similar technological characteristics. The subject device differs from the predicate in software features, which include automatic alignment and motion correction. Additionally, minor changes were made through letter-to-file to both the software and hardware of the minor changes to the software included updates to the user interface and enhanced security features, and the minor changes to the hardware included modifications to the RF shield and screen, gauss guard, and battery charger through. These differences do not raise new questions of safety and effectiveness as compared to the predicate.
#### Non-Clinical Performance
The subject device has similar technological characteristics as the predicate (K201722) and differs only in software, which includes automatic alignment and motion correction features, updates to the user interface, and enhanced security features. As part of demonstrating substantial equivalence to the predicate, a risk analysis was completed to identify the risks associated with the software modifications. Performance testing was conducted to evaluate the modifications. The subject device passed all the testing in accordance with internal requirements and applicable standards to support substantial equivalence.
- Software Verification per IEC 62304:2006 and as recommended in the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- Cybersecurity Information provided as recommended in FDA Guidance, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices"
The following testing was leveraged from the predicate device (K201722). Test results from the predicate were used to support the subject device because the conditions were identical or the subject device modifications did not introduce a new worst-case configuration or scenario for testing.
- Biocompatibility per ISO 10993-1:2018
- . Electrical Safety, EMC and Essential Performance per ANSI/AAMI ES 60601-1:2005/(R)2012, IEC 60601-2-33:2015, and IEC 60601-1-2:2014
- Electrical Safety Collateral Standard: Usability per IEC 60601-1-6:2013
- NEMA MS 1-2008 (R2014) - Determination of Signal-to-Noise Ratio (SNR) in Diagnostic Magnetic Resonance Imaging
- NEMA MS 3-2008 (R2014) - Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
{6}------------------------------------------------
- . NEMA MS 8-2016 - Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems
- NEMA MS 9-2008 (R2014) Characterization of Phased Array Coils for Diagnostic Magnetic ● Resonance Images
- NEMA MS 12-2016 Quantification and Mapping of Geometric Distortion for Special . Applications
#### Conclusion
The results of the testing described above demonstrate that the subject Swoop POC MRI Scanner System is as safe and effective as the predicate and supports a determination of substantial equivalence.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.