K160632 · Toshibamedical Systems Corporation · LNH · Aug 23, 2016 · Radiology
Device Facts
Record ID
K160632
Device Name
Vantage Titan 1.5T, MRT-1510, V3.6 Software
Applicant
Toshibamedical Systems Corporation
Product Code
LNH · Radiology
Decision Date
Aug 23, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Vantage Titan system is indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: - Proton density (PD) (also called hydrogen density) - Spin-lattice relaxation time (T1) - Spin-spin relaxation time (T2) - Flow dynamics - Chemical Shift Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Device Story
Vantage Titan 1.5T is a magnetic resonance imaging (MRI) system. It acquires cross-sectional images of head/body anatomy using nuclear magnetic resonance properties (proton density, T1, T2, flow dynamics, chemical shift). The system uses pulse sequences (UTE, MP-RAGE, T2:mEcho, T2*mEcho) and automated positioning/planning tools (NeuroLine+, CardioLine+, SureVOI) to assist clinicians. Operated by trained professionals in clinical settings, the system provides images for physician interpretation to aid in diagnosis. Hardware remains unchanged from predicate; software updates include new pulse sequences, MRS improvements, multi b-value DWI, and image filters (eFSBB, temporal filter). Benefits include enhanced diagnostic imaging capabilities and workflow efficiency through automated positioning.
Clinical Evidence
Bench testing, phantom imaging, and volunteer clinical imaging were performed to validate new software functionalities. No clinical trials on patients were required; safety and effectiveness were demonstrated through software validation, risk management, and design controls.
Technological Characteristics
1.5T MRI system. Materials and hardware unchanged from K150427. Software-based pulse sequences and automated planning tools. Conforms to IEC 60601-2-33 (2010) for safety parameters (SAR, dB/dt). Software development follows IEC 62304:2006 and IEC 62366:2007.
Indications for Use
Indicated for diagnostic imaging of head or body anatomic structures in patients requiring cross-sectional transaxial, coronal, sagittal, or oblique MRI, including non-contrast MRA. Contraindications include patients where MRI is unsafe; contrast agent use restricted to approved indications.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 23, 2016
Toshiba Medical Systems Corporation % Janine Reves Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K160632
Trade/Device Name: Vantage Titan 1.5T, MRT-1510, V3.6 Software Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: July 27, 2016 Received: July 29, 2016
Dear Ms. Reyes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jeff Bollyea
For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
K160632
510(k) Number (if known)
Device Name
Vantage Titan 1.5T, MRT-1510, V3.6 Software
Indications for Use (Describe)
The Vantage Titan system is indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- · Proton density (PD) (also called hydrogen density)
- · Spin-lattice relaxation time (T1)
- · Spin-spin relaxation time (T2)
- · Flow dynamics
- · Chemical Shift
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
|------------------------------------------------------------------------------------------------|---------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### TOSHIBA AMERICA MEDICAL SYSTEMS, INC.
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
#### 1. CLASSIFICATION AND DEVICE NAME:
| Classification Name: | Magnetic Resonance Diagnostic Device |
|-------------------------|--------------------------------------|
| Regulation Number: | 90-LNH (Per 21 CFR § 892.1000) |
| Trade Proprietary Name: | Vantage Titan 1.5T |
| Model Number: | MRT-1510 |
#### 2. SUBMITTER'S NAME:
Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
#### 3. OFFICIAL CORRESPONDENT:
Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
## 4. CONTACT PERSON, U.S. AGENT AND ADDRESS:
- Contact Person:
Janine Faith Reyes Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive, Tustin, CA 92780 (714) 669-7853
## 5. MANUFACTURING SITE:
Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
- 6. ESTABLISHMENT REGISTRATION: 9614698
- 7. Date Prepared:
July 27, 2016
## U.S. Agent:
Paul Biggins Director, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive, Tustin, CA 92780 (714) 669-7808
{4}------------------------------------------------
#### HIBA AMERICA MEDICAL
### 8. DEVICE NAME:
Vantage Titan 1.5T, MRT-1510, V3.6 Software
## 9. TRADE NAME(S):
Vantage Titan 1.5T
#### 10. CLASSIFICATION NAME:
Magnetic Resonance Diagnostic Device (MRDD)
### 11. CLASSIFICATION PANEL:
Radiology
#### 12. DEVICE CLASSIFICATION:
Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)
#### 13. PRODUCT CODE:
90-LNH
#### 14. PREDICATE DEVICE:
Primary Predicate Device (system): Vantage Titan 1.5T, MRT-1510, V3.1 Reference Predicate Device (added software functionalities): Vantage Titan 3T, MRT-3010, V3.5
| | Subject | Primary Predicate | Reference Predicate |
|----------------|---------------------------------------|---------------------------------------|-------------------------------------|
| System | Vantage Titan 1.5T,<br>MRT-1510, V3.6 | Vantage Titan 1.5T,<br>MRT-1510, V3.1 | Vantage Titan 3T,<br>MRT-3010, V3.5 |
| Marketed By | Toshiba America<br>Medical Systems | Toshiba America<br>Medical Systems | Toshiba America<br>Medical Systems |
| 510(k) Number | This Submission | K150427 | K152371 |
| Clearance Date | | April 17, 2015 | October 23, 2015 |
#### 15. REASON FOR SUBMISSION
Modification of a cleared device.
#### 16. DEVICE DESCRIPTION
The Vantage Titan (Model MRT-1510) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System, previously cleared under K150427. This system is based upon the technology and materials of previously marketed Toshiba MRI systems and is intended to acquire and display cross-sectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body.
#### 17. SUMMARY OF HARDWARE CHANGES
No change from the previous predicate submission, K150427.
{5}------------------------------------------------
#### TOSHIBA AMERICA MEDICAL SYSTEMS, INC.
## 18. SUMMARY OF SOFWARE CHANGES
This submission is to report the following software functionalities have been added:
- UTE (Ultra Short TE) pulse sequence ●
- MP-RAGE pulse sequence
- T2:mEcho and T2*mEcho pulse sequence ●
- . Improvements in MRS
- Multi b-value DWI
- NeuroLine+ (automatic positioning assistance for neuro)
- CardioLine+ (automatic positioning assistance for cardiac)
- SureVOI (automatic cardiac planning assistance)
- eFSBB (phase-enhanced FSBB filter) ●
- Temporal Filter
#### 19. SAFETY PARAMETERS
| Item | Subject Device: Vantage Titan<br>1.5T, V3.6 | Predicate Device: Vantage<br>Titan 1.5T, V3.1 (K150427) | Notes |
|---------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------|
| Static field strength | 1.5T | 1.5T | Same |
| Operational Modes | Normal and 1st Operating Mode | Normal and 1st Operating Mode | Same |
| i. Safety parameter<br>display | SAR dB/dt | SAR dB/dt | Same |
| ii. Operating mode<br>access requirements | Allows screen access to 1st level<br>operating mode | Allows screen access to 1st level<br>operating mode | Same |
| Maximum SAR | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33(2010)) | 4W/kg for whole body (1st<br>operating mode specified in<br>IEC 60601-2-33(2010)) | Same |
| Maximum dB/dt | <1st operating mode specified<br>in IEC 60601-2-33 (2010) | <1st operating mode specified<br>in IEC 60601-2-33 (2010) | Same |
| Potential emergency<br>condition and means<br>provided for shutdown | Shut down by Emergency Ramp<br>Down Unit for collision hazard<br>for ferromagnetic objects | Shut down by Emergency Ramp<br>Down Unit for collision hazard<br>for ferromagnetic objects | Same |
#### 20. IMAGING PERFORMANCE PARAMETERS
No change from the previous predicate submission, K150427.
{6}------------------------------------------------
#### A AMERICA MEDICAL SYSTEMS. INC.
## 21. INDICATIONS FOR USE
The Vantage Titan system indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1)
- Spin-spin relaxation time (T2) ●
- Flow dynamics
- Chemical Shift
Contrast agent use is restricted to the approved drug indications. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
No changes to the previously cleared indication, K150427.
## 22. SUMMARY OF DESIGN CONTROL ACTIVITIES
PS Risk List for new software functionalities and pulse sequences are included in this submission. The test methods used are the same as those submitted in the previously cleared submissions of the primary predicate device, Vantage Titan 1.5T, MRT-1510, V3.1 (K150427), and the reference predicate device, Vantage Titan 3T, MRT-3010, V3.5 (K152371). A declaration of conformity with design controls is included in this submission.
## 23. SAFTEY
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. The changes to this device are limited to the application software that is being presented for review. As the hardware and related safety considerations remain unchanged from the previous submission no information related to these aspects are presented for the determination of safety. Performance and/or clinical testing are being provided to demonstrate the safety and effectiveness of the application software.
This device is in conformance with the applicable parts of the following consensus standards published by the International Electrotechnical Commission (IEC) for medical devices and the National Electrical Manufacturers Association (NEMA).
## LIST OF APPLICABLE STANDARDS
- . IEC62304:2006
- IEC62366:2007 ●
{7}------------------------------------------------
#### AMERICA MEDICAL SYSTEMS, INC.
## 24. TESTING
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
## 25. SUBSTANTIAL EQUIVALENCE
Toshiba Medical Systems Corporation believes that the Vantage Titan 1.5T, MRT-1510, V3.6 Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate devices referenced in this submission. Toshiba Medical Systems Corporation believes that the changes incorporated into the Vantage Titan 1.5T, MRT-1510, V3.6 software are substantially equivalent to the previously cleared predicate devices.
### 26. CONCLUSION
The modifications incorporated into the Vantage Titan 1.5T, MRT-1510, V3.6 software do not change the indications for use or the intended use of the device. Based upon bench testing, phantom imaging, volunteer clinical imaging, successful completion of software validation and application of risk management and design controls, it is concluded that the subject device is safe and effective for its intended use.