K994210 · Philips Medical Systems, Inc. · LLZ · Feb 4, 2000 · Radiology
Device Facts
Record ID
K994210
Device Name
INTURIS SUITE
Applicant
Philips Medical Systems, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 4, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Philips Inturis Suite is intended to be used for the display and reading of cardiac angiographic images stored on a compact disk (CD) or an image server. Inturis Suite also allows the user to change the appearance of the images and export selected frames to other programs for printing and further manipulation.
Device Story
Inturis Suite is modular software for import, storage, archive, review, analysis, and database management of digital cardiovascular angiographic images. Inputs include DICOM 3.0 formatted images from CDs or image servers. Software operates on Windows NT PCs or UNIX servers; used in clinical environments (e.g., Cath Lab) by healthcare professionals. Features include image display, digital zoom, window/level adjustment, annotation, and export of frames (bmp, jpg, avi). Output allows clinicians to visualize and manipulate angiographic data to support cardiovascular diagnostic and procedural decision-making.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based image review system. Operates on Windows NT or UNIX platforms. Supports DICOM 3.0 standard for image import/export. Features include digital zoom, window/level adjustment, and annotation. Connectivity via network or CD-ROM. Modular architecture.
Indications for Use
Indicated for the display and reading of cardiac angiographic images for clinical review and analysis. No specific patient population or contraindications defined.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
ImageView Coronary Angiography Display and Review System (K972229)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo for Philips. The logo is in black and white and features the word "PHILIPS" at the top. Below the word is a shield-like shape containing a design with two wavy lines and two four-pointed stars. The overall design is simple and recognizable.
# Philips Medical Systems
# 510(K) Summary
K994210
In accordance with the requirements of the Safe Medical Device Act, Philips Medical Systems North America Company herewith submits a 510(K) summary of safety and effectiveness for the following device.
| SUBMITTER NAME / ADDRESS: | Philips Medical Systems North America Company<br>710 Bridgeport Avenue<br>Shelton, CT 06484-0917 |
|---------------------------|--------------------------------------------------------------------------------------------------|
| CONTACT PERSON / TEL NO: | Frank Gianelli<br>Tel No: (203) 926-7729 |
| DATE SUMMARY PREPARED: | December 8, 1999 |
| ESTABLISHMENT NO.: | 1217116 |
| TRADE/PROPRIETARY NAME: | Inturis Suite |
| COMMON/USUAL NAME: | Image Review Software for Angiographic X-ray System |
| CLASSIFICATION NAME: | Angiographic X-ray System<br>(21 CFR 892.1600; Class II; Tier 2; 90-IZI) |
| PREDICATE DEVICE(S): | ImageView Coronary Angiography Display and Review System; Heartware, Inc. |
# DEVICE DESCRIPTION:
Philips Inturis Suite is software comprised of modular software programs designed to perform the necessary functions required for the import, storage, archive, review, analysis, reporting and database management of digital cardiovascular angiographic images in DICOM 3.0 formats within and beyond the Cath Lab domain. Inturis Suite allows multiple users to quickly access and transfer multiple angiographic exams. Inturis Suite software can be installed and run on Windows NT based personnel computers for small applications and on UNIX based servers for high performance applications. The modular software design of Inturis Suite, allows users to tailor their image archive, transfer and communication requirements based on their particular performance needs and the number of cath labs importing images to an Inturis Suite via DICOM. Inturis Suite software will be provided on a CD-Rom which contains the installation program, the applicable executable files, DLL files and a readme file.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the Philips logo. The logo consists of the word "PHILIPS" in bold, uppercase letters, positioned above a shield-like shape. Inside the shield, there is a stylized globe with wavy lines representing water and star-like shapes representing land masses. The logo is black and white.
#### INTENDED USE:
Philips Inturis Suite is intended to be used for the display and reading of cardiac angiographic images stored on a compact disk (CD) or an image server. Inturis Suite also allows the user to change the appearance of the images and export selected frames to other programs for printing and further manipulation.
## SUBSTANTIAL EQUIVALENCE INFORMATION:
Inturis Suite is considered comparable and substantially equivalent to the following predicate device which has been cleared for commercial distribution via its referenced 510(k) submission.
· Device: ImageView Coronary Angiography Display and Review System Manufacturer: HeartWare, Inc. 510(k) No.: K972229
# TECHNOLOGICAL CHARACTERISTICS:
Predicate Device: ImageView Coronary Angiography Display and Review System, HeartWare, Inc. Comparison matrix of main features: Inturis Suite vs. predicate device
| Feature | Philips<br>Inturis Suite | HeartWare<br>ImageView |
|--------------------|--------------------------------------|-------------------------------------|
| Operating system | Windows NT; UNIX | Windows 95; Windows NT |
| Memory requirement | 128 MB (minimum)<br>256 MB (maximum) | 32 MB (minimum)<br>128 MB (maximum) |
| Image source | DICOM 3.0 SCU<br>DICOM 3.0 (CD-rom) | DICOM 3.0 SCU<br>DICOM 3.0 (CD-rom) |
| Display rate | up to 30 frames per sec | up to 30 frames per sec |
| Multiple windows | Yes | Yes |
| Digital zoom | Yes | Yes |
| Window (contrast) | Yes | Yes |
| Level (brightness) | Yes | Yes |
| Annotation | Yes | Yes |
| Image export | bmp, jpg, avi | bmp, gif |
| Network access | Yes | Yes |
## SAFETY INFORMATION:
Inturis Suite introduces no new safety and efficacy issues other than those already identified with the predicate device. The results of the hazard analysis, combined with the appropriate preventive measures taken indicate that the device is of minor level of concern.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings and body. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 2000 FEB
Frank Gianelli Senior Regulatory Affairs Specialist Philips Medical Systems 710 Bridgeport Avenue P.O. Box 860 Shelton, CT 06484-0917
Re:
K994210 Inturis Suite Dated: December 8, 1999 Received: December 14, 1999 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Mr. Gianelli:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act indude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either dass III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verty sych assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdch/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name :
Indications For Use :
Philips Inturis Suite is intended to be used for the display and reading of cardiac angiographic images stored on a compact disk (CD) or an image server. Inturis Suite also allows the user to change the appearance of the images and export selected frames to other programs for printing and further manipulation.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel le. Lyman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devices 510(k) Number
Prescription Use
( Per 21 CFR 801.109 )
OR
Over-The-Counter Use
(Optional Format 1-2-96)