K982563 · Ramsoft, Inc. · LLZ · Oct 6, 1998 · Radiology
Device Facts
Record ID
K982563
Device Name
ULTRAPRO MODEL 98
Applicant
Ramsoft, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 6, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use by medical professionals to acquire data from FDA listed imaging equipment such as ultrasound, CT and MRJ scanners, x-ray imagers and video cassette recorders that provide a video output or a digital output. Archival is to be performed on media that is FDA listed. Ultrapro can be configured to use JPEG lossy compression. When lossy compression is used, the compression ratio will be imprinted in the non-diagnostic area of the image. The compression ratio can be adjusted during acquisition.
Device Story
Ultrapro is a PACS workstation for medical imaging practices. It acquires images from standard medical video sources or digital inputs (DICOM, JPEG, TIFF) via video frame grabber or scanner. The system stores images on PC hardware; supports lossless RLE and lossy JPEG/Motion JPEG compression (ratios up to 60:1). Radiologists use the workstation to review images, adjust display parameters (contrast, brightness, zoom), perform annotations, and make distance measurements. Reports are created within the system. Data is communicated to other workstations or third-party DICOM 3.0 components via network/modem and archived to optical media. The system is intended for use by medical professionals in clinical settings to facilitate diagnostic review and clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence determination is not based on an assessment of performance tests.
Technological Characteristics
PC-based PACS workstation. Inputs: video frame grabber, TWAIN scanner, DICOM/PC formats. Outputs: display, network communication, optical media archive. Features: NTSC/PAL video support, RLE/JPEG/Motion JPEG compression, DICOM 3.0 compliance. Software-based image processing for contrast, brightness, zoom, annotation, and measurement. Standalone or networked configuration.
Indications for Use
Indicated for use by medical professionals to acquire, display, communicate, and archive diagnostic images from FDA-listed imaging equipment, including ultrasound, CT, MRI, x-ray, and video cassette recorders.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
ALI UltraPACS (K925965/A) with ALI comPACS data compression module (K963610)
CEMAX-ICON Archive Manager 2.0 and Autorad modules (K955092)
K041935 — DR SYSTEMS PACS, RELEASE 6.1 · Dr Systems, Inc. · Aug 16, 2004
K072972 — MEDSTREAMING MEDICAL OFFICE SOFTWARE · Medstreaming, LLC · Nov 6, 2007
K022970 — AMICAS LIGHT BEAM WORKSTATION, MODEL 1.0 · Amicas, Inc. · Nov 22, 2002
K061035 — TIGERVIEW PROFESSIONAL · Televere Systems · Jun 5, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo on the left and a date on the right. The logo is circular and contains the text "RamSoft". The date is "OCT 6 1998". The text is in a bold, sans-serif font.
K982 56.3
# 510(k) Summary for Ultrapro 98 1
# 1.1 General Information
# 1.1.1 Name and Address of Manufacturer
RamSoft Inc. 37 Bankview Circle Toronto. ON M9W 6S6, CANADA Tel: (416) 674-1347 Fax: (416) 674-7147 Email: ram@ramsoft.on.ca Contact Person: Naganatha S Ramanathan
# 1.1.2 Establishment Registration Number
9006410
1.1.3 Common and Proprietary names of Device
Trade or Proprietary Name: Ultrapro™ Common Name: Picture Archiving and Communication System
### 1.1.4 Class
Systems, digital image communications - Class | 892.2020 LMD
### 12 Substantially Equivalent Devices
ALI UltraPACS (K925965/A) with ALI comPACS data compression module (K963610) Manufactured by: ALI Technologies, Inc
CEMAX-ICON Archive Manager 2.0 and Autorad modules (K955092) Manufactured by: CEMAX-ICON Inc.
### 1.3 Device Description
Ultrapro is a tool for a medical imaging practice. It can acquire medical images from any standard medical video source. Digital data in the DICOM format or other popular PC formats such as JPEG, TIFF, etc. can also be acquired.
Images may be acquired in uncompressed or compressed format (lossless or lossy). The lossy compression ratio is selected by viewing the original image and the lossy compressed image side by side and increasing the ratio while retaining clinically significant characteristics. Images that have been lossy compressed have the text "LOSSY xx:1" where xx is the compression ratio placed in the visible, non-diagnostic area of the image.
Related patient information can also be entered or scanned in with a TWAIN compliant scanner. Once acquired, Ultrapro serves as a digital storage for these images. Images are stored on the computer. Images may be communicated to other Ultrapro workstations or to third-party DICOM 3.0 compliant PACS components. The radiologist may review the images with Ultrapro and create reports. The display of images (contrast, brightness, zoom, etc.) may be adjusted at any time. Annotations may be drawn on images. Distance measurements may be made on images. These image parameters may be saved along with the image while still preserving the original image. This information can be communicated to other workstations through a network or modem. Finally, all of the information may be archived onto optical media for permanent storage.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo for a company called "RalliSoft". The logo features the company name in a stylized font inside of an oval shape. The oval is positioned in the center of a globe graphic. The globe is depicted with curved lines to represent the continents and oceans.
# 1.4 Intended Use Statement
Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use with FDA approved imaging equipment that provides a video output or a digital output.
| | ALI UltraPACS with ALI comPACS | RamSoft Ultrapro |
|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Image Acquisition | Video frame grabber or DICOM<br>input | Video frame grabber, FDA<br>approved film scanner or DICOM<br>input |
| Compatible Video Signals | NTSC (up to 30 frames/s,<br>640x480, 256 grayscale, 16.7<br>million colors)<br>PAL (up to 25 frames/s, 768x576.<br>256 grayscale, 16.7 million colors) | NTSC (up to 30 frames/s,<br>640x480, 256 grayscale, 16.7<br>million colors)<br>PAL (up to 25 frames/s, 768x576,<br>256 grayscale, 16.7 million colors) |
| Cine Capability (motion capture) | Yes | Yes |
| Image Compression | Lossless RLE and Lossy JPEG | Lossless RLE and Lossy JPEG |
| Lossy Compression Ratio<br>Range | 4:1 to 30:1 | 4:1 to 60:1 |
| Uses JPEG lossy compression<br>in viewing, transmitting and<br>archiving still images | Yes | Yes |
| Uses Motion JPEG lossy<br>compression for viewing,<br>transmitting and archiving cine<br>images | Yes | Yes |
| DICOM 3.0 compliant | Yes | Yes |
| Imaging Modalities | Multi-modality | Multi-modality |
| Image Display Capability | Yes | Yes |
| Remote and Local Area Network<br>Communication | Yes | Yes |
| Image Annotation Capability | Yes | Yes |
| Archives Image on Optical<br>Media | Yes | Yes |
# 1.5 Companson of Technological Characteristics
# 1.6 Conclusion
In conclusion, Ultrapro is substantially equivalentitititi to the ALI product. The determination of substantial equivalence is not based on an assessment of performance tests.
ﺎ
55555 : 35453 :
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with its wings spread, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 6 1998
Naganatha S. Ramanathan, Ph.D. President Ramsoft, Inc. 37 Bankview Circle Toronto, Ontario Canada M9W 6S6
Re:
K982563 "Ultrapro 98, Digital Image Communications System Dated: July 22, 1998 Received: July 23, 1998 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Dr. Ramanathan:
We have reviewed your Section 5100% notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments on to devices that have been reclassified in accordance with the Feceral Food Drues and Cosmetic Act (Act). You may, therefore, narket the device, subject to the general controls of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical-Devices: Generaliation (2) CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition; FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odch/dsmamain.html"
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and -Radiological Health
Enclosure
{3}------------------------------------------------
Page of
t
| 510(k) Number (if known): | K982563 |
|---------------------------|-------------|
| Device Name: | ULTRAPRO 98 |
Indicatic is For Use:
Ultrapro is a system that utilizes standard personal computer hardware to acquire, display, communicate and archive diagnostic images. The system is designed for use by medical professionals to acquire data from FDA listed imaging equipment such as ultrasound, CT and MRJ scanners, x-ray imagers and video cassette recorders that provide a video output or a digital output. Archival is to be performed on media that is FDA listed.
Ultrapro can be configured to use JPEG lossy compression. When lossy compression is used, the compression ratio will be imprinted in the non-diagnostic area of the image. The compression ratio can be adjusted during acquisition.
(PL) ASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Gimel li. Syson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Device 510(k) Number
OR
Prescription Use
(Per 21 CFR 801.109)
Over-The-Counter Use
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.