K203312 · Canon Medical Systems Corporation · LLZ · Feb 16, 2021 · Radiology
Device Facts
Record ID
K203312
Device Name
Vitrea Software Package, VSTP-001A
Applicant
Canon Medical Systems Corporation
Product Code
LLZ · Radiology
Decision Date
Feb 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
Vitrea Software Package is an application package developed for use on Vitrea®, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea has the following additional indications: The Cerebral Aneurysm Analysis application is intended to facilitate the extraction and segmentation of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio. The MR Wall Motion Tracking application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and regional myocardial function that is used for patients with suspected heart disease. The MR Coronary Tracking application is intended to assist physicians with performing coronary artery analysis for MR heart images which are intended for the qualitative and quantitative analysis of coronary arteries. The SUREVolume Synthesis application is intended to load volume images acquired by whole-body X-ray CT scanners, X-ray angiography systems, and MRI systems and displays fusion images. The Angio Viewer application displays image data acquired using an X-ray angiography system. It supports cine display, subtraction, and distance measurement. The US Cardiac Fusion application enables fusion display of the analysis results obtained using the US 3D Wall Motion Tracking application and the CT Coronary Artery Analysis application. The Ultrasound Clinical Applications are indicated for the visualization of structures, and dynamic processes with the human body using saved ultrasound DICOM images to provide image information for diagnosis. The Spectral Stone Analysis application is intended to serve as an adjunct visualization tool for the differentiation between uric acid and non-uric acid stones greater than 3 mm with Spectral CT studies acquired on the Canon Medical Systems scanner. The Spectral Composition Analysis application is intended to assist a physician in visualizing the presence of monosodium urate in anatomical structures. The clinical syndrome of gout is characterized by the presence of monosodium urate crystals in joints or soft tissue. The Embolization Plan application is a post processing software that is intended to assist physicians in the visualization of the liver arterial tree using 3D images of CT or 3D images of Cone Beam CT acquired by Toshiba or Canon Medical Systems. It provides tools to assist the user in analysis of these images. The output is intended to be an adjunct means that allows automatic and manual planning of the liver arterial vessels for guidance of the embolization procedure. The output is a 3D visualization of the hepatic arteries to high dense lesion in the liver. The Spectral Analysis application is a CT, non-invasive image analysis software package, which may be used to aid in the visualization of anatomical and pathological materials. The software provides quantification of Hounsfield units of iodine attenuation differences and iodine concentration and display by color. Effective Z and Electron Density maps may aid in the differentiation and characterization of different tissues in the human body.
Device Story
Vitrea Software Package (VSTP-001A) is a suite of post-processing applications for medical imaging workstations. It ingests multi-dimensional DICOM images from CT, X-ray angiography (XA), MRI, and ultrasound systems. Applications perform specialized tasks: segmenting cerebral aneurysms; tracking cardiac wall motion; analyzing coronary arteries; fusing multi-modality volumes; visualizing liver arterial trees for embolization planning; and processing spectral CT data (e.g., iodine maps, effective Z, electron density). Physicians use these tools to perform measurements, visualize anatomical structures, and plan interventions. The software provides quantitative and qualitative outputs, including 3D visualizations, color-coded maps, and distance measurements, which serve as adjuncts to clinical diagnosis and procedural guidance. The system operates on dedicated workstations within clinical environments; it does not connect to external networks, ensuring data remains within the local domain.
Clinical Evidence
Bench testing only. Verification and validation activities included phantom studies to assess feature improvements and a physician review of clinical images to support promotional claims regarding spectral iodine maps in lung visualization. No clinical trials were conducted.
Technological Characteristics
Software-only package for Vitrea workstations. Implements image processing algorithms for spectral CT analysis (monochromatic images 35-200keV, iodine maps, effective Z, electron density, basis material decomposition). Complies with IEC 62304 and IEC 60601-1-6. Standalone deployment; no network connectivity.
Indications for Use
Indicated for patients undergoing diagnostic imaging (CT, XA, MR, US) requiring post-processing analysis of cerebral arteries, cardiac function, coronary arteries, liver arterial trees, or spectral CT tissue characterization (gout, stone differentiation).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K192923 — Vitrea Software Package · Canon Medical Systems Corporation · Feb 14, 2020
K180323 — Vitrea Software Package; VSTP-001A · Canon Medical Systems Corporation · Mar 30, 2018
K182601 — Vitrea Software Package, VSTP-001A · Canon Medical Systems Corporation · Dec 14, 2018
K173088 — Vitrea Software Toshiba Package VSTP-001A · Toshibamedical Systems Corporation · Jan 31, 2018
K183013 — Vitrea Software Package, VSTP-001A · Canon Medical Systems Corporation · Nov 26, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
February 16, 2021
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Canon Medical Systems Corporation % Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K203312
Trade/Device Name: Vitrea Software Package, VSTP-001A Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: January 27, 2021 Received: January 28, 2021
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K203312
Device Name Vitrea Software Package, VSTP-001A
# Indications for Use (Describe)
Vitrea Software Package is an application package developed for use on Vitrea®, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimens acquired from a variety of imaging devices. Vitrea has the following additional indications:
The Cerebral Aneurysm Analysis application is intended to facilitate the extraction of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio.
The MR Wall Motion Tracking application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and regional myocardial function that is used for patients with suspected heart disease.
The MR Coronary Tracking application is intended to assist physicians with performing coronary artery analysis for MR heart images which are intended for the qualitative and quantitative analysis of coronary arteries.
The SUREVolume Synthesis application is intended to load volume images acquired by whole-body X-ray CT scanners, X-ray angiography systems, and MRI systems and displays fusion images.
The Angio Viewer application displays image data acquired using an X-ray angiography system. It supports cine display, subtraction, and distance measurement.
The US Cardiac Fusion application enables fusion display of the analysis results obtained using the US 3D Wall Motion Tracking application and the CT Coronary Artery Analysis application.
The Ultrasound Clinical Applications are indication of structures, and dynamic processes with the human body using saved ultrasound DICOM images to provide image information for diagnosis.
The Spectral Stone Analysis application is intended to serve as an adjunct visualization tool for the differentiation between uric acid and non-uric acid stones greater than 3 mm with Spectral CT studies acquired on the Canon Medical Systems scanner.
The Spectral Composition Analysis application is intended to assist a physician in visualizing the presence of monosodium urate in anatomical structures. The clinical syndrome of gout is characterized by the presence of monosodium urate crystals in joints or soft tissue.
The Embolization Plan application is a post processing software that is intended to assist physicians in the visualization of the liver arterial tree using 3D images of CT or 3D images of Cone Beam CT acquired by Toshiba or Canon Medical Systems. It provides tools to assist the user in analysis of these images. The output is intended to be an adjunct means that allows automatic and manual planning of the liver arterial vessels for guidance of the embolization procedure. The output is a 3D visualization of the hepatic arteries to high dense lesion in the liver.
The Spectral Analysis application is a CT, non-invasive image analysis software package, which may be used to aid in the visualization of anatomical and pathological materials. The software provides quantification of Hounsfield units of iodine
{3}------------------------------------------------
attenuation differences and iodine concentration and display by color.
Effective Z and Electron Density maps may aid in the differentiation of different tissues in the human body.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K203312
### 510(k) Summary
#### 1. SUBMITTER'S NAME
Fumiaki Teshima Sr. Manager, Quality Assurance Department Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
- 2. ESTABLISHMENT REGISTRATION 9614698
#### OFFICIAL CORRESPONDENT/CONTACT PERSON 3. Orlando Tadeo, Jr. Sr. Manager, Requlatory Affairs Department Canon Medical Systems USA, Inc. 2441 Michelle Dr. Tustin, CA 92780 (714) 669-7459
- DATE PREPARED 4. November 06, 2020
- 5. DEVICE NAME(S) Vitrea Software Package, VSTP-001A
- 6. TRADE NAME(S) Vitrea Software Package, VSTP-001A
- 7. COMMON NAME Radiological Image Processing Software
#### 8. DEVICE CLASSIFICATION
Class II (per 21 CFR 892.2050, Picture Archiving and Communications System) Picture Archiving and Communications Systems – Product Code: 90LLZ [per 21 CFR 892.20501
# 9. PREDICATE DEVICE
| Product | Marketed<br>by | Regulation<br>Number | Regulation Name | Product<br>Code | 510(k)<br>Number | Clearance<br>Date |
|-------------------------------------------|-------------------------------------|----------------------|------------------------------------------------------|-----------------|------------------|-------------------|
| Vitrea Software<br>Package,<br>VSTP-001A* | Canon<br>Medical<br>Systems,<br>USA | 21 CFR<br>892.2050 | Picture Archiving<br>and<br>Communications<br>System | LLZ | K192923 | 10/15/2019 |
* Includes: MR Wall Motion Tracking, Cerebral Aneurysm Analysis, (CT/XA Cerebral Artery Morphological Analysis), MR Coronary Tracking, Synthesis, Angio Viewer, US Cardiac Fusion, Ultrasound Applications Package, Dual Energy Stone Analysis, Dual Energy Composition Analysis, Embolization Planning Tool
{5}------------------------------------------------
# 10. REASON FOR SUBMISSION
Modification of an existing medical device.
#### 11. DEVICE DESCRIPTION
Vitrea Software Package, VSTP-001A, is an application package developed for use on Vitrea, a medical image processing software, marketed by Vital Images, Inc. Vitrea Software Package, VSTP-001A, currently includes ten post processing applications, MR Wall Motion Tracking, Cerebral Aneurysm Analysis, MR Coronary Tracking, SUREVolume Synthesis, Angio Viewer, US Cardiac Fusion, Ultrasound Applications Package, Dual Energy Stone Analysis, Dual Energy Composition Analysis and Embolization Planning Tool which use brain, body or cardiac image data, obtained from CT/XA/MR/US systems, to assist physicians in performing specialized measurements and analysis.
#### 12. INDICATIONS FOR USE
Vitrea Software Package is an application package developed for use on Vitrea®, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea has the following additional indications:
The Cerebral Aneurysm Analysis application is intended to facilitate the extraction and segmentation of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio.
The MR Wall Motion Tracking application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and regional myocardial function that is used for patients with suspected heart disease.
The MR Coronary Tracking application is intended to assist physicians with performing coronary artery analysis for MR heart images which are intended for the qualitative and quantitative analysis of coronary arteries.
The SUREVolume Synthesis application is intended to load volume images acquired by whole-body X-ray CT scanners, X-ray angiography systems, and MRI systems and displays fusion images.
The Angio Viewer application displays image data acquired using an X-ray angiography system. It supports cine display, subtraction, and distance measurement.
The US Cardiac Fusion application enables fusion display of the analysis results obtained using the US 3D Wall Motion Tracking application and the CT Coronary Artery Analysis application.
The Ultrasound Clinical Applications are indicated for the visualization of structures, and dynamic processes with the human body using saved ultrasound DICOM images to provide image information for diagnosis.
The Spectral Stone Analysis application is intended to serve as an adjunct visualization tool for the differentiation between uric acid and non-uric acid stones greater than 3 mm with Spectral CT studies acquired on the Canon Medical Systems scanner.
The Spectral Composition Analysis application is intended to assist a physician in visualizing the presence of monosodium urate in anatomical structures. TThe clinical syndrome of gout is characterized by the presence of monosodium urate crystals in joints or soft tissue.
<sup>2441</sup> Michelle Drive, Tustin, CA 92780 PHONE: 800-421-1968 https://us.medical.canon
{6}------------------------------------------------
The Embolization Plan application is a post processing software that is intended to assist physicians in the visualization of the liver arterial tree using 3D images of CT or 3D images of Cone Beam CT acquired by Toshiba or Canon Medical Systems. It provides tools to assist the user in analysis of these images. The output is intended to be an adjunct means that allows automatic and manual planning of the liver arterial vessels for quidance of the embolization procedure. The output is a 3D visualization of the hepatic arteries to high dense lesion in the liver.
The Spectral Analysis application is a CT, non-invasive image analysis software package, which may be used to aid in the visualization of anatomical and pathological materials. The software provides quantification of Hounsfield units of iodine attenuation differences and iodine concentration and display by color.
Effective Z and Electron Density maps may aid in the differentiation and characterization of different tissues in the human body.
#### 13. SUBSTANTIAL EQUIVALANCE
The Vitrea Software Package, VSTP-001A, is substantially equivalent to and has the same intended use as the primary predicate device, which received premarket clearance under K192923, and is marketed by Canon Medical Systems USA. The changes being made to the existing device are the modfiication of three applications, Spectral Stone Analysis, Spectral Composition Analysis and Spectral Analysis. The modifications to these three applications were evaluated by risk analysis and a regression review, and it was determined that they were implemented with all risks reduced to an acceptable level.
| Item | Vitrea Software Package, VSTP-001A (V8.8) | Vitrea Software Package, VSTP-001A (V8.5) |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | (Subject Device) | K192923 |
| Anatomical region | Whole body | Whole body |
| Type of Input Data | Acquired by spectral scan (TSX-306A/3, V10.3 or earlier and V10.4 and later) | Acquired by spectral scan (TSX-306A/3, V10.3 or earlier) |
| | Spectral Stone Analysis<br>• Monochromatic images | Spectral Stone Analysis<br>• Monochromatic images |
| | Spectral Composition Analysis<br>• Basis material images and monochromatic image | Spectral Composition Analysis<br>• Basis material images and monochromatic image |
| | Spectral Analysis<br>• Basis material images and monochromatic image | Spectral Analysis<br>• Basis material images and monochromatic image |
| Type of Output Data | Spectral Stone Analysis<br>• Analysis result display<br>• Secondary capture (RGB image)<br><br>Spectral Composition Analysis<br>• Analysis result display<br>• Secondary capture (RGB image)<br>• Batch MPR and DICOM volume save | Spectral Stone Analysis<br>• Analysis result display<br>• Secondary capture (RGB image)<br><br>Spectral Composition Analysis<br>• Analysis result display<br>• Secondary capture (RGB image) |
| Item | Vitrea Software Package, VSTP-001A (V8.8) | Vitrea Software Package, VSTP-001A (V8.5) |
| 510(k) Number | (Subject Device) | K192923 |
| | Spectral Analysis<br>• Analysis result display<br>• Secondary capture (RGB image)<br>• Batch MPR and DICOM volume save | Spectral Analysis<br>• Analysis result display<br>• Secondary capture (RGB image) |
| Image Processing | Spectral Stone Analysis<br>• Visualization of the differentiation<br>between uric acid and non-uric acid<br>stones | Spectral Stone Analysis<br>• Visualization of the differentiation<br>between uric acid and non-uric acid<br>stones |
| | Spectral Composition Analysis<br>• Visualization of monosodium urate<br>presence within surrounding anatomical<br>structures | Spectral Composition Analysis<br>• Visualization of monosodium urate<br>presence within surrounding anatomical<br>structures |
| | Spectral Analysis<br>• Generation of monochromatic images<br>(ranging from 35keV – 200keV)<br>o Smoothing filter<br>• Generation of lodine map<br>• VNC image<br>• Generation of Electron Density Image<br>• Generation of Effective Z Image<br>• Generation of Basis material image (bone/water) | Spectral Analysis<br>• Generation of monochromatic images<br>(ranging from 35keV - 135keV)<br>• Generation of lodine map<br>• VNC image<br>• Generation of Electron Density Image<br>• Generation of Effective Z Image |
| Analysis | Segmentation | |
| | Spectral Stone Analysis<br>• The region included in the range of set<br>CT numbers is extracted. | Spectral Stone Analysis<br>• The region included in the range of set<br>CT numbers is extracted. |
| | Spectral Composition Analysis<br>• The compositions are extracted based<br>on the composition setting. | Spectral Composition Analysis<br>• The compositions are extracted based<br>on the composition setting. |
| | Spectral Analysis<br>• Monochromatic image<br>o Basis material dual energy analysis<br>• lodine map<br>o Three material decomposition<br>• Electron density<br>o Basis material dual energy analysis<br>• Effective Z<br>o Basis material dual energy analysis<br>• Basis material image (bone/water)<br>o Basis material dual energy analysis | Spectral Analysis<br>• Monochromatic image<br>o Basis material dual energy analysis<br>• lodine map<br>o Three material decomposition<br>• Electron density<br>o Basis material dual energy analysis<br>• Effective Z<br>o Basis material dual energy analysis |
| Item | Vitrea Software Package, VSTP-001A (V8.8) | Vitrea Software Package, VSTP-001A (V8.5) |
| 510(k) Number | (Subject Device) | K192923 |
| Display | Spectral Stone Analysis*<br>• MPR | Spectral Stone Analysis<br>• MPR |
| | Spectral Composition Analysis<br>• MPR, 3D | Spectral Composition Analysis<br>• MPR, 3D |
| | Spectral Analysis<br>• MPR, Fusion image, MIP image, MinIP image, Displaying three MPR planes | Spectral Analysis<br>• MPR, Fusion image |
{7}------------------------------------------------
{8}------------------------------------------------
*Color Display Function in Spectral Stone Analysis was updated, for the Material Line Preset label to list only "Uric Acid" and "Non Uric Acid
#### 14. SAFETY
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC62304 and IEC60601-1-6.
## 15. TESTING
Risk analysis and verification/validation activities conducted through bench testing which is included in this submission demonstrate that the established specifications for the device have been met. Additional performance testing, using phantom studies, were conducted to assess the improvements to existing features. A physician review of clinical images was also performed, in order to support a new promotional claim for the previously cleared Spectral Analysis feature, specifically, that spectral iodine maps are designed to assist physicians in visualizing iodine differences in the lung. Results of all these studies demonstrate that the features included in this submission meet specifications and perform as intended. Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also referenced for this submission.
#### 16. CYBERSECURITY
The Vitrea Software Package, VSTP-001A is a portfolio of software applications developed by Canon that are integrated into the Vital Images Vitrea workstations. As such this software has no connections to the internet, wired or wireless networks, etc. All import and export of data is within the domain of the Vital product. Based upon this information it should be noted that CyberSecurity requirements do not apply to this device.
## 17. CONCLUSION
The software applications modified in the Vitrea Software Package. VSTP-001A, perform in a manner similar to and are intended for the same use as the predicate device. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device, Vitrea Software Package, VSTP-001A, 510(k) cleared per K192923.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.