Volpara Imaging Software

K182310 · Volpara Health Technologies Limited · LLZ · Sep 21, 2018 · Radiology

Device Facts

Record IDK182310
Device NameVolpara Imaging Software
ApplicantVolpara Health Technologies Limited
Product CodeLLZ · Radiology
Decision DateSep 21, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

VolparaDensity is a software application intended for use with the raw data from digital breast x-ray systems, including tomosynthesis. VolparaDensity calculates and quantifies a density map and from that determines volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Volpara provides these numerical values along with a BI-RADS breast density 4th or 5th Edition category to aid health care professionals in the assessment of breast tissue composition. VolparaDensity is not an interpretive or diagnostic aid and should be used only as adjunctive information when the final assessment of breast density category is made by an MQSA-qualified interpreting physician.

Device Story

Volpara Imaging Software processes raw digital mammography and tomosynthesis projection data to generate pixel-based density maps representing fibroglandular tissue thickness. Software computes volumetric statistics: fibroglandular tissue volume, total breast volume, volumetric breast density, average/maximum dense tissue thickness, and focal density (maximum dense tissue volume in 1cm² regions). It assigns BI-RADS 4th/5th Edition density categories using software-defined thresholds, incorporating focal density assessments. Output includes numerical metrics, density maps with marked maxima, and image quality metrics. Operates on Windows servers outside the patient environment; managed by healthcare professionals. Provides adjunctive information to assist MQSA-qualified interpreting physicians in breast density assessment. Benefits include standardized, objective quantification of breast tissue composition to support clinical decision-making.

Clinical Evidence

Bench testing only. Verification and validation performed per ISO 14971:2012 and ISO 62304:2006. Testing included unit-level and integrated system-level verification, with complete predicate validation data repeated to ensure integration and backwards compatibility.

Technological Characteristics

Software-based volumetric breast density assessment. Operates on Windows server. Inputs: raw digital mammography/tomosynthesis data. Outputs: DICOM SCI format density maps and quantitative metrics. Standards: ISO 14971:2012, ISO 62304:2006, DICOM 2015. Software level of concern: Moderate.

Indications for Use

Indicated for use by healthcare professionals to aid in the assessment of breast tissue composition using raw digital mammography or tomosynthesis images. Not for diagnostic or interpretive use; adjunctive information only.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 21, 2018 Volpara Health Technologies Limited Ralph Highnam CEO Level 7, 44 Victoria Street, Wellington Central WELLINGTON, 6011 NEW ZEALAND Re: K182310 Trade/Device Name: Volpara Imaging Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: August 3, 2018 Received: August 24, 2018 Dear Ralph Highnam: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Hse 2. Nils for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182310 Device Name Volpara Imaging Software #### Indications for Use (Describe) Volpara is a software application intended for use with the raw data from digital breast x-ray systems, including tomosynthesis. Volpara calculates and quantifies a density map and from that determines volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Volpara provides these numerical values along with a BI-RADS breast density 4th or 5th Edition category to aid health care professionals in the assessment of breast tissue composition. Volpara is not an interpretive or diagnostic aid and should be used only as adjunctive information when the final assessment of breast density category is made by an MQSA qualified interpreting physician. Type of Use (Select one or both, as applicable) | <span style="font-size:10px"><b></b></span> | <span style="font-size:10px">Prescription Use (Part 21 CFR 801 Subpart D)</span> | |---------------------------------------------|----------------------------------------------------------------------------------| | | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Volpara Health Technologies. The word "volpara" is written in a bold, sans-serif font in blue. Below the word "volpara" is the phrase "health technologies" in a smaller, sans-serif font, also in blue. To the right of the word "volpara" is a stylized graphic of three blue shapes that resemble leaves or petals. # Special 510(k) Summary Prepared 15th August 2018 - Sponsor: Volpara Health Technologies - Company Address: Level 7, 44 Victoria Street, Wellington, 6011, New Zealand - Contact Person: Ralph Highnam - Telephone: +64 - 4 - 499 - 6029 - +64 4 499 6029 Fax: - ralph.highnam@volparasolutions.com Email: - Submission Date: 15th August 2018 - 510(k) submission number: K182310 - Device Name: Volpara Imaging Software - Common Name: Imaging Software #### Classification: Requlatory Class: = Review Category: Class II Classification Panel: Radiology System, Imaging Processing; 21 CFR 892.2050; LLZ #### Legally Marketed Predicate Devices: The modified software, Volpara Imaging Software ("Volpara") 1.5.6 is substantially equivalent to the original Volpara Imaging Software 1.5.2 cleared pursuant to K153427 (Volpara Health Technologies Limited), which serves as the primary predicate. ### Predicate Device Description: The Volpara Imaging Software ("Volpara") 1.5.2 software provides volumetric assessment of digital x-ray images of the breast, including in that definition both raw digital {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Volpara Health Technologies. The word "volpara" is written in a bold, sans-serif font in a dark blue color. Below the word "volpara" is the phrase "health technologies" in a smaller, lighter font. To the right of the word "volpara" are three stylized shapes in a light blue color, resembling a bird in flight. mammograms and raw tomosynthesis projections (the central projections of which are just raw digital mammograms). The assessment takes the form of generating and validating density maps wherein the value at each pixel represents the thickness of fibroglandular tissue between that pixel and the x-ray source. From the density maps various quantitative density-map based statistics are computed as follows: - . volume of fibroglandular tissue in cm³ - volume of breast in cm3, . - the volumetric breast density (the percentage of fibroglandular tissue in breast) . - average thickness of dense tissue in cm . - maximum thickness of dense tissue in cm . - maximum volume of dense tissue above any 1cm² square region (or "focal ● density") From the volumetric breast density, a BI-RADS 4th Edition and 5th Edition breast density category can be attained by applying thresholds set by the software. The device outputs those metrics along with the density maps themselves marked with the location of the various maxima. Volpara Imaging Software 1.5.2 operates on a Windows server that meets Volpara data input and output requirements and generally is located outside the patient environment. The device does not contact the patient, nor does it control any life-sustaining devices. ### Comparison with Predicate Device: Volpara Imaging Software 1.5.6 is the same core software as the predicated device Volpara Imaging Software 1.5.2 with the addition of the following software and labeling enhancements. - . The modified software uses the focal density information when generating a Bl-RADS breast density category. - The modified software supports data acquired using Hologic's SmartCurve paddle. ● - . The modified software provides output image quality metrics which were validated but not output by 1.5.2. - . The User Manual has been expanded to cover the modifications described above, and with an updated bibliography. - The software provides an option to select highlighting of density scores at a specified threshold. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Volpara Health Technologies. The logo features the word "volpara" in a bold, blue sans-serif font. Above the right side of the word is a stylized graphic of three blue teardrop shapes. Below the word "volpara" are the words "health technologies" in a smaller, sans-serif font. # Substantial Equivalence Comparison Table | | Predicate Device<br>Volpara 1.5.2 (K153427) | Submission Device,<br>Volpara 1.5.6 | |-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | VolparaDensity is a software application intended for use with the raw data from digital breast x-ray systems, including tomosynthesis. VolparaDensity calculates and quantifies a density map and from that determines volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Volpara provides these numerical values along with a BI-RADS breast density 4th or 5th Edition category to aid health care professionals in the assessment of breast tissue composition. VolparaDensity is not an interpretive or diagnostic aid and should be used only as adjunctive information when the final assessment of breast density category is made by an MQSA-qualified interpreting physician. | Same as predicate. | | Intended Users | Health Care Professionals. | Health Care Professionals. | | Image Source | Digital mammography images. | Digital mammography images. | | Image Sources | Digital mammograms from mammography or tomosynthesis systems. | Digital mammograms from mammography or tomosynthesis systems, including those obtained using with curved paddles. | | Anatomical Area | Breast | Breast | | Assessment Scope | Volumetric | Volumetric | | Operating Environment | Windows | Windows | | Image Storage and Report Generation | Yes | Yes | | | Output to the console. | Output to the console. | | Numeric Output | Volume of Fibroglandular tissue<br>Volume of Breast<br>Volumetric Breast Density<br>BIRADS 4th or 5th Edition Breast Density Category<br>Average thickness of dense tissue<br>Maximum thickness of dense tissue (and location). | Volume of Fibroglandular tissue<br>Volume of Breast<br>Volumetric Breast Density<br>BIRADS 4th or 5th Edition Breast Density Category, with highlighting above a certain volumetric threshold or if focal density present<br>Average thickness of dense tissue | | | Maximum volume of dense tissue<br>above any 1cm² square region<br>(and location). | Maximum thickness of dense<br>tissue (and location).<br>Maximum volume of dense tissue<br>above any 1cm² square region<br>(and location).<br>Image quality assessment metrics | | Image Output | Density map in DICOM SCI format,<br>for visualization as user specifies. | Density map in DICOM SCI format,<br>for visualization as user specifies | | Classification | 21 CFR 892.2050; LLZ | 21 CFR 892.2050; LLZ | | Software Level of Concern | Moderate | Moderate | {6}------------------------------------------------ # Intended Use: VolparaDensity is a software application intended for use with the raw data from digital breast x-ray systems, including tomosynthesis. VolparaDensity calculates and quantifies a density map and from that determines volumetric breast density as a ratio of fibroglandular tissue and total breast volume estimates. Volpara provides these numerical values along with a BI-RADS breast density 4th or 5th Edition category to aid health care professionals in the assessment of breast tissue composition. VolparaDensity is not an interpretive or diagnostic aid and should be used only as adjunctive information when the final assessment of breast density category is made by an MQSA-qualified interpreting physician. ### Device Description The Volpara Imaging Software ("Volpara") 1.5.6 software provides volumetric assessment of digital x-ray images of the breast, including in that definition both raw digital mammograms and raw tomosynthesis projections (the central projections of which are just raw digital mammograms). The assessment takes the form of generating and validating density maps wherein the value at each pixel represents the thickness of fibroglandular tissue between that pixel and the x-ray source. From the density maps various quantitative density-map based statistics are computed as follows: - . volume of fibroglandular tissue in cm3 - volume of breast in cm3. ● - the volumetric breast density (the percentage of fibroglandular tissue in breast) ● - average thickness of dense tissue in cm ● - maximum thickness of dense tissue in cm ● - maximum volume of dense tissue above any 1cm² square region (or "focal ● density") {7}------------------------------------------------ Image /page/7/Picture/0 description: This image shows the Volpara health technologies logo on the left side of the image. On the right side of the image, the document type is listed as a report. The DMS number is Mtk3252-007-1, the revision date is August 3, 2018, and the DMS reference is Eng/Ultron. The title of the document is Volpara Imaging Software 510(k) Summary, and it is page 5 of 5. - image quality assessment metrics ● From the volumetric breast density, a BI-RADS 4th Edition and 5th Edition breast density category can be attained by applying thresholds set by the software, with an increase in density category if a focal density appears to be present. The device outputs those metrics along with the density maps themselves marked with the location of the various maxima. Volpara operates on a Windows server that meets Volpara data input and output requirements and generally is located outside the patient environment. The device does not contact the patient, nor does it control any life-sustaining devices. ### Performance Data The Volpara software has been verified and validated according to the company's design control process. All of the documents specified in the FDA's various software quidance documents have been submitted in this Special 510(k) Notification. An ISO 14971 compliant risk analysis has been provided and incorporated into the development effort. Software testing included both unit level and integrated system level testing. A report of outstanding anomalies was included in the software information. The modified device was tested and determined to be compliant to the following standards: ISO 14971:2012 Medical devices – Application of risk management to medical devices: ISO 62304-2006 Software Life Cycle Processes, DICOM 2015. In addition to the verification and validation testing conducted for the specific modification to the software detailed in this Special 510(k) submission, complete verification and validation data testing conducted for the predicate was repeated in order to ensure integration and backwards compatibility. ### General Safety and Effectiveness Concerns: The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device. Risk management is ensured via a risk analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. # Conclusion: The Special 510(k) Premarket Notification for VolparaDensity 1.5.6 contains adequate information and data to demonstrate substantial equivalence to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...