K180336 · Siemens Medical Solutions USA, Inc. · LLZ · Apr 19, 2018 · Radiology
Device Facts
Record ID
K180336
Device Name
syngo.MR Applications
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Apr 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
syngo.MR Applications is a syngo based post-acquisition image processing software for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra.
Device Story
syngo.MR Applications is a post-acquisition image processing software suite for viewing, manipulating, and analyzing MR, MR-PET, CT, PET, and MR spectra data. It consolidates eight previously cleared Siemens MR software applications into a single platform. The device operates as a workstation-based software tool used by radiologists and clinicians to perform routine and advanced image evaluations, including cardiac function analysis, neuro-perfusion, fMRI, tractography, oncology lesion segmentation, and spectroscopy. Inputs consist of DICOM-compliant medical images; outputs include processed images, quantitative metrics (e.g., blood flow dynamics, lesion volumes, perfusion mismatch ratios), and structured reports. The software assists clinicians in diagnostic decision-making and surgical planning. It is intended for use in clinical environments by trained healthcare professionals. The system supports DICOM standards for interoperability and utilizes established image processing algorithms for segmentation, registration, and quantification.
Clinical Evidence
Bench testing only. Software verification and validation were performed in accordance with ISO 14971:2007 and IEC 62304:2006. No clinical data was required or provided to support the substantial equivalence of the consolidated software applications.
Technological Characteristics
Software-based post-processing application. Conforms to DICOM PS 3.1-3.20 (2016), ISO 14971:2007 (risk management), and IEC 62304:2006 (software life cycle). Operates on syngo.via platform. Features include image segmentation, registration, histogram analysis, and quantitative measurement tools for MR, CT, and PET data.
Indications for Use
Indicated for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra in a clinical setting by healthcare professionals.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
syngo.MR Post-Processing Software Version SMRVB20A (K163294)
syngo.MR Post-Processing Software Version SMRVB10A (K151353)
syngo.MR Post-Processing Software Version SMRVA16A (K130749)
syngo.MR Post-Processing Software Version SMRVA16B (K133401)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Siemens Medical Solutions USA, Inc. % Cordell L. Fields, Esq. Regulatory Affairs Technical Specialist 40 Liberty Boulevard, Mailcode 65-1A MALVERN PA 19355
April 19, 2018
Re: K180336
Trade/Device Name: syngo.MR Applications Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ LNH Dated: February 6, 2018 Received: February 7, 2018
Dear Mr. Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
(1-800-638-2041 or 301-796-7100).
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a light blue or teal.
## DEPARTMENT OF HEALTH AND HUMAN SERVICES
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
| 510(k) Number (if known) | K180336 |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | syngo.MR Applications |
| Indications for Use (Describe) | syngo.MR Applications is a syngo based post-acquisition image processing software for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra. |
| Type of Use (Select one or both, as applicable) | <div> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Traditional 510(k) Premarket Notification for: syngo.MR Applications VB30
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a light blue or cyan. The background is plain white, which makes the text stand out.
# Section 5 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
## Date of Summary Preparation: February 06, 2018
## I. General Information
| Importer / Distributor: | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Mailcode: 65-1A<br>Malvern, PA 19355 |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number: | 2240869 |
| Manufacturer: | Siemens Healthcare GmbH<br>Henkestr. 127<br>D-91052 Erlangen, Germany |
| Registration Number: | 3002808157 |
| Contact Person: | Cordell L. Fields, Esq.<br>Regulatory Affairs Technical Specialist<br>Siemens Medical Solutions USA, Inc.<br>65 Valley Stream Pkwy<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Phone: (610) 448-6469<br>Fax: (610) 427-3465<br>Email: cordell.fields@siemens-healthineers.com |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a light blue or turquoise color.
## Device Name and Classification
| Data | Details |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name /<br>Device<br>Proprietary<br>Name: | syngo.MR Applications<br>It covers single and engine applications:<br>• syngo.MR General<br>○ syngo.MR Reading<br>○ syngo.MR General Routine<br>○ syngo.MR Cardiac Reader<br>○ syngo.MR Composing<br>○ syngo.MR General Engine2<br>• syngo.MR Cardiology<br>○ syngo.MR Cardiac 4D Ventricular<br>Function<br>○ syngo.MR Cardiac Flow<br>○ syngo.MR Cardio Engine<br>• syngo.MR Neurology<br>○ syngo.MR Neuro Perfusion<br>○ syngo.MR Neuro Perfusion Mismatch<br>○ syngo.MR Neuro fMRI<br>○ syngo.MR Tractography<br>○ syngo.MR Neuro Perfusion Engine<br>○ syngo.MR Neuro 3D Engine<br>• syngo.MR Oncology<br>○ syngo.MR Onco<br>○ syngo.MR 3D Lesion Segmentation<br>○ syngo.MR Tissue4D<br>○ syngo.MR Onco Engine<br>○ syngo.MR Onco Engine Pro<br>○ syngo.MR OncoCare<br>• syngo.BreVis<br>• syngo.mMR General<br>• syngo.MR Spectroscopy<br>○ syngo.MR Spectro SVS<br>○ syngo.MR Spectro CSI<br>○ syngo.MR Spectro Extension<br>• syngo.MR Vascular<br>○ syngo.MR Vascular Analysis |
| Classification<br>Name: | Regulation Description:<br>Picture Archiving and Communication System (PACS) |
<sup>2</sup> syngo.MR General Engine is the precondition for all other, advanced MR post-processing applications and engines
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the word "SIEMENS" in all capital letters. The text is a light blue color. The background is white.
| Data | Details |
|---------------------------|------------------------------|
| Classification<br>Panel: | Radiology |
| Device<br>Classification: | Class II devices |
| Regulation<br>number: | 21 CFR § 892.2050 |
| Product Code: | Primary: LLZ, Secondary: LNH |
| Data | Details |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name /<br>Device<br>Proprietary<br>Name: | syngo.MR Applications<br>It covers single and engine applications:<br>syngo.MR General syngo.MR Reading syngo.MR General Routine syngo.MR Cardiac Reader syngo.MR Composing syngo.MR General Engine syngo.MR Cardiology syngo.MR Cardiac 4D Ventricular Function syngo.MR Cardiac Flow syngo.MR Cardio Engine syngo.MR Neurology syngo.MR Neuro Perfusion syngo.MR Neuro Perfusion Mismatch syngo.MR Neuro fMRI syngo.MR Tractography syngo.MR Neuro Perfusion Engine syngo.MR Neuro 3D Engine syngo.MR Oncology syngo.MR Onco syngo.MR 3D Lesion Segmentation syngo.MR Tissue4D syngo.MR Onco Engine syngo.MR Onco Engine Pro syngo.MR OncoCare syngo.BreVis syngo.mMR General syngo.MR Spectroscopy |
Traditional 510(k) Premarket Notification for: syngo.MR Applications VB30
February 06, 2018
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the word "SIEMENS" in a sans-serif font. The word is written in a light blue color. The letters are all capitalized and evenly spaced.
| Data | Details |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | syngo.MR Spectro SVS syngo.MR Spectro CSI syngo.MR Spectro Extension syngo.MR Vascular syngo.MR Vascular Analysis |
| Single functionalities<br>and engines | Description |
| syngo.MR General | |
| syngo.MR Reading | Enables reading of 2D, 3D, and 4D MR data.<br>Includes:<br>• Basic workflow<br>• Workflow customization<br>• Follow-up support<br>• Rescan handling<br>Context-specific reporting |
| syngo.MR General<br>Routine | It extends syngo.via by adding dedicated worflows and tools for routine<br>and advanced reading of MR examinations. Several workflows are<br>provided like MR Neurology, MR Prostate Reading, MR Breast<br>Reading, and MR Cardio-Vascular Reading workflows.<br>syngo.MR General Routine provides several MR Evaluation tools:<br>Subtraction, MeanCurve, Image Filter, 2D/3D Distortion Correction.<br>ADC and b-value tool (for extrapolated b-values), Multiplication,<br>Division, Addition, Elastic Motion Correction.Workflow optimized report<br>templates. |
| syngo.MR Cardiac<br>Reader | With syngo.MR Cardiac Reader the Cardiac Reading step within the<br>Cardiac Analysis Workflow is provided. The Cardiac Reading enables<br>viewing and evaluating different kinds of cardiac images using the<br>common tools. |
| syngo.MR Composing | syngo.MR Composing is a dedicated offline application for creation of<br>full-format images from overlapping MR volume data sets acquired at<br>multiple stages. It can be used to compose images in any of the other<br>workflows. |
| syngo.MR General<br>Engine4 | syngo.MR General Engine contains:<br>• syngo.MR General Routine<br>• syngo.MR Cardiac Reader<br>syngo.MR General Engine is the precondition for all other, advanced<br>MR post-processing applications and Engines. |
| syngo.MR Cardiology | |
| syngo.MR Cardiac 4D<br>Ventricular Function | syngo.MR Cardiac 4D Ventricular Function processes MR cine images<br>of the heart which enables users to generates quantitative results for<br>physicians in the diagnostic process. The MR cardiac interactive<br>reporting template is included. |
| Single functionalities<br>and engines | Description |
| syngo.MR Cardiac<br>Flow | Processes velocity-encoded MR images to evaluate blood flow<br>dynamics e.g. in the heart and the great vessels. The application<br>generates quantitative results for physicians in the diagnostic process.<br>The MR cardiac interactive reporting template is included. |
| syngo.MR Cardio<br>Engine | <i>syngo.MR Cardio Engine</i> contains:<br>syngo.MR Cardiac 4D Ventricular Function syngo.MR Cardiac Flow |
| <b>syngo.MR Neurology</b> | |
| syngo.MR Neuro<br>Perfusion5 | Extends the MR Neurology workflow with advanced processing tools<br>for the analysis of brain perfusion datasets and allows the<br>assessment and evaluation of various spots of cerebral lesions. |
| syngo.MR Neuro<br>Perfusion Mismatch | Extends the assessment of brain perfusion datasets with calculation of<br>mismatch area and mismatch ratio between two different contrasts (for<br>example diffusion and perfusion) to support therapy decisions. |
| syngo.MR Neuro fMRI | A workflow oriented visualization package for BOLD fMRI. It enables<br>the visualization of task-related areas of activation overlayed onto 2D<br>or 3D anatomical datasets, providing the spatial correspondence of<br>BOLD results with cortical landmarks or brain lesions. Image-based<br>registration and multi-contrast evaluation are also available. Functional<br>and anatomical image data can be exported for surgical planning as<br>DICOM datasets. |
| syngo.MR<br>Tractography | Enables the representation of diffusion paths of the human brain<br>based on diffusion tensor imaging. syngo.MR Tractography supports<br>surgery planning and is suitable for neurophysiological research in<br>relation to cortical networking and pathologies of the white matter. |
| syngo.MR Neuro<br>Perfusion Engine | <i>syngo.MR Neuro Perfusion Engine</i> contains:<br>syngo.MR Neuro Perfusion syngo.MR Neuro Perfusion Mismatch |
| syngo.MR Neuro 3D<br>Engine | <i>syngo.MR Neuro 3D Engine</i> contains:<br>syngo.MR Neuro fMRI syngo.MR Tractography |
| <b>syngo.MR Oncology</b> | |
| syngo.MR Onco | An image viewing, processing and reading software that allows for<br>oncological MR image evaluation in a structured way. |
| Single functionalities<br>and engines | Description |
| syngo.MR 3D Lesion<br>Segmentation | Provides convenient volumetric evaluation of lesions and/or other<br>structure of interest as well as particularly useful tools for oncology<br>applications. |
| syngo.MR Tissue4D | A post-processing workflow which supports the physician in reading of<br>dynamic contrast-enhanced MR data sets. |
| syngo.MR OncoCare | Enables the physician to evaluate signal intensities in segmented<br>regions of interest with the help of histograms and color maps as well<br>as evaluation the change of time in typical lesion parameters like<br>diameter or volume (trending). |
| syngo.MR Onco<br>Engine | syngo.MR Onco Engine contains:<br>• syngo.MR Onco<br>• syngo.MR 3D Lesion Segmentation |
| syngo.MR Onco<br>Engine Pro | syngo.MR Onco Engine Pro contains:<br>• syngo.MR Onco Engine<br>• syngo.MR OncoCare |
| syngo.MR BreVis | A software package intended for use in viewing and analyzing<br>magnetic resonance imaging (MRI) studies. |
| syngo.mMR General | A syngo based post-processing software for viewing, manipulating and<br>evaluating MR, PET and MR-PET images. |
| syngo.MR Spectroscopy | |
| syngo.MR Spectro<br>SVS | Provides evaluation of MR Single Voxel Spectroscopy (SVS) data with<br>comprehensive workflow guidance. |
| syngo.MR Spectro CSI | Provides evaluation of MR Chemical Shift Imaging (CSI) data with<br>comprehensive workflow guidance. syngo.MR Spectro CSI includes<br>the possibility of an integrated reading of MR images and CSI<br>spectroscopy data for prostate exams. |
| syngo.MR Spectro<br>Extension | Provides access to advanced parameters, which allow the advanced<br>user to configure the post processing and display of spectro results<br>according to his / her personal needs. Both Single Voxel Spectroscopy<br>(SVS) and Chemical Shift Imaging (CSI) data are supported. |
| syngo.MR Vascular | |
| syngo.MR Vascular<br>Analysis | Enables assessment / quantification of general vascular pathologies. |
#### II. Safety and Effectiveness Information Supporting Substantial Equivalence
#### Intended Use
"syngo.MR Applications is a syngo based post-acquisition image processing software for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra."
#### Device Description
Applications are syngo svnao.MR based post-processing The software/applications to be used for viewing and evaluating MR images provided by a magnetic resonance diagnostic device and enabling structured evaluation of MR images.
The syngo.MR Applications is a combination of eight (8) former separately cleared medical devices which are now handled as features / functionalities within syngo.MR Applications.
These functionalities are combined unchanged compared to their former cleared descriptions; of course some minor enhancements and improvements are made. The syngo.MR Applications are syngo.via based MR data viewing, processing and reading software allowing MR image evaluation in a structured way and supporting convenient reading and / or evaluation of MR images and data.
<sup>3</sup> While viewing (i.e. assessing) of images from other vendors is always possible; for advanced post-processing applications, some of the post-processing steps may depend on information contained in private DICOM tags ,therefore evaluation and processing of images can't be guaranteed for other vendors.
{7}------------------------------------------------
## The table below shows the functionalities of the syngo.MR Applications:
<sup>4</sup> syngo.MR General Engine is the precondition for all other, advanced MR post-processing applications and engines
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a light blue or cyan color.
<sup>§</sup> The application syngo.MR Neuro Dynamics described in K151353 (and cleared August 07, 2015) is only available within syngo.MR Neuro Perfusion (neurospecific Mean Curve Tool), but no longer as single application. The syngo.MR Neuro Perfusion Engine Pro, described in K151353 (and cleared August 07, 2015), is therefore obsolete within SMRVB20, as all applications are already part of syngo.MR Neuro Perfusion Engine.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the word "SIEMENS" in a sans-serif font. The word is written in all capital letters and is a light blue color. The background is white, which makes the word stand out.
Traditional 510(k) Premarket Notification for: syngo.MR Applications VB30
February 06, 2018
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a light blue or turquoise.
Except for the new product composition, there are some minor improvements and enhancements of the existing functionalities within the syngo.MR Applications.
Cardiac / Angio improvements:
- Improved functional analysis in the MR Cardiac Analysis workflow: ●
- o Mode for manual segmentation of the LV and RV
- o Pen tool for improved editing of LV and RV contours
- o Result values can be exported as text file
- o LV and RV contours can be saved and restored
- The workflows MR Angio SingleStation, MultiStation, TWIST and . TimCT are consolidated into the new workflow MR Angiography
Additional minor improvements:
- 4D Navigation Toolbar ●
- o Mini-Toolbar for 4D navigation
- o Display of clinical phase labels
- o Easy navigation in mulit-phase data
- Freehand ROI for seed selection in Tractography ●
- Series Navigator: series numbers now displayed ●
- ADC / computed b-values: now as pre-processing ●
- Tissue4D Maps: color-bar added to image segment ●
- Median added to histogram statistics ●
- Relative frequency added to histogram ●
- Color-bar added to image segment (LUT) .
- Wild card for series description ●
- Organ specific workflow .
#### General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk Management is ensured via a Risk Analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards in a Risk Analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in software development. SW testing and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards. Furthermore, the operators are healthcare professionals familiar with and responsible for the evaluating and post-processing of MR images.
syngo.MR Applications conform to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. The background is white.
| Rec.- No. | Product Area | Title of Standard | Ref.- No. & Date | Standards Development Organization |
|-----------|--------------|----------------------------------------------------------------------------------------|-------------------------|------------------------------------|
| 5-96 | General | Medical devices -<br>Application of usability<br>engineering to medical<br>devices | 62366-1:2015 | AAMI ANSI<br>IEC |
| 5-40 | General | Medical devices - application of<br>risk management to medical<br>devices | 14971:2007 | ISO |
| 13-32 | Software | Medical device software -<br>Software life cycle<br>processes | 62304:2006 | AAMI ANSI<br>IEC |
| 12-300 | Radiology | Digital Imaging and<br>Communications in<br>Medicine (DICOM) Set PS<br>3.1-3.20 (2011) | PS 3.1 -<br>3.20 (2016) | NEMA |
The standards conformed to are the following:
## Substantial Equivalence
The syngo.MR Applications VB30 has a different Intended Use as compared to the predicate devices. This is the result of the new product setup of combining the former eight (8) medical devices into one (1) device. However, the main functionalities of the former separate medical devices are used unchanged for syngo.MR Applications VB30, except for the aforementioned minor enhancements and improvements. The conclusions from the non-clinical data suggest that the additional features bear an equivalent safety and performance profile as that of the predicate devices and also does not affect the Indications for use of it. Therefore syngo.MR Applications are considered to be substantially equivalent to their primary predicate devices, which are current legally marketed devices.
| Table 1: Predicate devices for syngo.MR Applications | |
|------------------------------------------------------|--|
|------------------------------------------------------|--|
| Predicate Device | FDA Clearance | Product Code |
|-------------------------------------------------------|-------------------------------------|--------------|
| syngo.MR Post-Processing<br>Software Version SMRVB20A | K163294<br>cleared February 6, 2017 | LLZ, LNH |
| syngo.MR Post-Processing<br>Software Version SMRVB10A | K151353<br>cleared August 1, 2015 | LLZ,LNH |
| syngo.MR Spectroscopy | K120315<br>cleared April 13, 2012 | LLZ |
| syngo.MR Post-Processing<br>Software Version SMRVA16A | K130749<br>cleared August 20, 2013 | LLZ,LNH |
| syngo.MR Post-Processing<br>Software Version SMRVA16B | K133401<br>cleared March 11, 2014 | LLZ,LNH |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. The word is displayed horizontally and is centered in the image.
### Conclusion as to Substantial Equivalence
The syngo.MR Applications are intended for similar indications as cleared in their according primary predicate devices.
In summary, Siemens is of the opinion that the syngo.MR Applications do not raise new questions of safety or effectiveness and are substantially equivalent to the currently marketed primary predicate devices as shown in Table 1.
Resulting from the new product setup of combining the eight (8) formerly cleared separate medical devices to one (1) medical device, the Intended Use changed. However, the main functionalities of the former devices will remain unchanged except for minor improvements and enhancements to give the device greater capabilities.
Therefore, Siemens believes that the subject device, the syngo.MR Applications is substantially equivalent to their primary predicate devices listed above in Table 1.