Mimics inPrint

K173619 · Materialise NV · LLZ · Mar 21, 2018 · Radiology

Device Facts

Record IDK173619
Device NameMimics inPrint
ApplicantMaterialise NV
Product CodeLLZ · Radiology
Decision DateMar 21, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Mimics inPrint is intended for use as a software interface and image segmentation system for the transfer of DICOM imaging information from a medical scanner to an output file. It is also used as pre-operative software for treatment planning. For this purpose, the Mimics inPrint output file can be used for the fabrication of the output file using traditional or additive manufacturing methods. The physical replica can be used for diagnostic purposes in the field of orthopedic, maxillofacial and cardiovascular applications. Mimics inPrint should be used in conjunction with other diagnostic tools and expert clinical judgement.

Device Story

Mimics inPrint is image processing/segmentation software; imports DICOM data from medical scanners (CT/MRI); transforms imaging data into output files for physical model fabrication. Used by clinicians for pre-operative treatment planning in orthopedic, maxillofacial, and cardiovascular fields. Software organizes functionality into user-guided workflows to improve experience; does not simulate treatment options. Output files are used to create physical replicas via traditional or additive manufacturing. Clinicians use these physical models as diagnostic aids alongside other tools and expert judgment to assist in surgical planning.

Clinical Evidence

Bench testing only. Validation confirmed measurement accuracy and 3D calculation capabilities met specifications. Validation of physical replica fabrication demonstrated that anatomical models for cardiovascular, orthopedic, and maxillofacial cases can be printed accurately using compatible 3D printers.

Technological Characteristics

Software-based image processing and segmentation system. Operates on DICOM imaging data. Features user-guided workflow for pre-operative planning and 3D model generation. Built on the same code base as the predicate device. Connectivity via DICOM standard. No hardware components; software-only device.

Indications for Use

Indicated for use as a software interface and image segmentation system for DICOM imaging data to facilitate pre-operative treatment planning and the fabrication of physical anatomical models via traditional or additive manufacturing. Intended for diagnostic use in orthopedic, maxillofacial, and cardiovascular applications. Must be used with other diagnostic tools and expert clinical judgment.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ March 21, 2018 Image /page/0/Picture/1 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue. Materialise N.V. % Mr. Oliver Clemens Regulatory Affairs Officer Technologielaan 15 3001 Leuven BELGIUM Re: K173619 Trade/Device Name: Mimics inPrint Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 21, 2018 Received: February 21, 2018 Dear Mr. Clemens: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs. Ph.L Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K173619 Device Name Mimics inPrint #### Indications for Use (Describe) Mimics inPrint is intended for use as a software interface and image segmentation system for the transfer of DICOM imaging information from a medical scanner to an output file. It is also used as pre-operative software for treatment planning. For this purpose, the Mimics inPrint output file can be used for the fabrication of the output file using traditional or additive manufacturing methods. The physical replica can be used for diagnostic purposes in the field of orthopedic, maxillofacial and cardiovascular applications. Mimics inPrint should be used in conjunction with other diagnostic tools and expert clinical judgement. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary The following section is included as required by the Safe Medical Devices Act (SMDA) of 1990 and 21CFR 807.92 | Company name | Materialise N.V. | |-----------------------------------|-----------------------------------| | Establishment registration number | 3003998208 | | Street Address | Technologielaan 15 | | City | Leuven | | Postal code | 3001 | | Country | Belgium | | Phone number | +32 16 39 62 80 | | Fax number | +32 16 39 66 06 | | Principal Contact person | Oliver Clemens | | Contact title | Regulatory Affairs Officer | | Contact e-mail address | Regulatory.Affairs@materialise.be | | Additional contact person | Mieke Janssen | | Contact title | Senior Regulatory Officer | | Contact e-mail address | Regulatory.Affairs@materialise.be | #### Submission date The date of the Traditional 510(k) submission is November 20, 2017 # Submission information | Trade Name | Mimics inPrint | |-----------------------------|----------------------------------------| | Common Name | Image processing system | | Classification Name | System, Image processing, Radiological | | Classification product code | LLZ (892.2050) | # Predicate Devices The primary predicate device to which substantial equivalence is claimed: | Trade or proprietary or model name | Mimics | |------------------------------------|-----------------| | 510(k) number | K073468 | | Decision date | April 2nd, 2008 | {4}------------------------------------------------ | Classification product code | LLZ (892.2050) | |-----------------------------|------------------| | Manufacturer | Materialise N.V. | The reference device: | Trade or proprietary or model name | D2P | |------------------------------------|-------------------| | 510(k) number | K161841 | | Decision date | January 9th, 2017 | | Classification product code | LLZ (892.2050) | | Manufacturer | 3D Systems, Inc | ## Description and functioning of the device Mimics inPrint is an image processing and segmentation software that was built on top of the Mimics application framework, an image processing and segmentation framework for the transfer of imaging information to an output file. #### Intended Use Mimics inPrint is intended for use as a software interface and image segmentation system for the transfer of DICOM imaging information from a medical scanner to an output file. It is also used as pre-operative software for treatment planning. For this purpose, the Mimics inPrint output file can be used for the fabrication of physical replicas of the output file using traditional or additive manufacturing methods. The physical replica can be used for diagnostic purposes in the field of orthopedic, maxillofacial and cardiovascular applications. Mimics inPrint should be used in conjunction with other diagnostic tools and expert clinical judgement. #### Technological Characteristics A detailed comparison shows the subject device is substantially equivalent in intended use, design, functionality, operating principles, materials and performance characteristics to the predicate device. Both the predicate and subject device are intended for use as a software interface and image segmentation system for the transfer of imaging information from a medical scanner to an output file, and can both be used as preoperative software. Both devices use the same image segmentation functionalities. Both devices generate an output file. Both devices have functionalities to perform pre-surgical planning. The subject device originates from the same code base as the predicate device; and follows the same development cycle, and testing procedures as the predicate device. The verification and validation of both predicates and subject device has been done following the same procedures and workflows. {5}------------------------------------------------ The following technological differences exist between the subject and predicate device: - . The subject device's intended use explicitly reflects 3D printing of the which can be used for diagnostic purposes in orthopedic, maxillofacial and cardiovascular applications, whereas this was only implicitly included for the primary predicate device - . The subject device only intends to be a software for treatment planning and does not include the simulation of treatment options. - . Whereas the primary predicate device was intended to import imaging information from a medical scanner such as CT or MRI scanner, the subject device is intended to import DICOM compliant types of imaging information. - The subject device shares the same technology and functionality as the primary predicate device, however, functionality is organized into a user guided workflow for better user experience. # Performance Data ## Non-clinical tests The Mimics inPrint application has been validated for its intended use to determine substantial equivalence to the predicate device. Measurement accuracy and calculate 3D study were performed and confirmed to be within specification. Validation of printing of physical replicas was performed and demonstrated that anatomical models for cardiovascular, orthopedic and maxillofacial cases can be printed accurately when using any of the compatible 3D printers. ## Summary The characteristics that determine the functionality and performance of the Mimics inPrint are substantially equivalent to the devices cleared under K073468. The non-clinical testing indicates that the subject device is as safe, as effective, and performs as well as the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...