Lung Nodule Assessment and Comparison Option (LNA)

K162484 · Philips Medical Systems Nederland B.V. · LLZ · Feb 23, 2017 · Radiology

Device Facts

Record IDK162484
Device NameLung Nodule Assessment and Comparison Option (LNA)
ApplicantPhilips Medical Systems Nederland B.V.
Product CodeLLZ · Radiology
Decision DateFeb 23, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K162484 · Feb 23, 2017Lung Nodule Assessment and Comparison Option (LNA)Philips Medical Systems Nederland B.V.Real recorded clinical data cases (for software validation); Pan-Canadian Early Detection of Lung Cancer Study (PanCan) cohort; British Columbia Cancer Agency (BCCA) chemoprevention trial cohortReal-world clinical data cases were used to validate the software's clinical workflow. Additionally, the device's integrated risk calculator is based on a model derived from and validated against large-scale population-based clinical study cohorts.Software validation; Risk calculator; Clinical workflow simulation; Population-based study

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Pan-Canadian Early Detection of Lung Cancer Study (PanCan) and BCCA chemoprevention trials; Population-based prospective clinical studyParticipants undergoing low-dose CT screening for lung nodulesNot applicable for this studyProbability of malignancy in pulmonary nodules detected on first screening CT
Software Validation using Clinical Data; Workflow simulationReal clinical patients (recorded cases)Not applicable for this studyClinical workflow adequacy

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Lung Nodule Characterization
Lung Nodule Malignancy Risk EstimationMultivariable logistic regression modelAUC > 0.90Pan-Canadian Early Detection of Lung Cancer Study (PanCan)Chemoprevention trials at the British Columbia Cancer Agency (BCCA)

Indications for Use

The Philips Medical Systems Lung Nodule Assessment and Comparison Option is intended for use as a diagnostic patient-imaging tool. It is intended for the review and analysis of thoracic CT images, providing quantitative and characterizing information about nodules in the lung in a single study, or over the time course of several thoracic studies. Characterizations include diameter, volume and volume over time. The system automatically performs the measurements, allowing lung nodules and measurements to be displayed.

Device Story

Diagnostic imaging software for thoracic CT analysis; processes CT images to identify, segment, and characterize lung nodules. Features include automatic lung/lobe segmentation, single-click nodule segmentation, and manual editing tools. Supports temporal comparison by loading up to 8 concurrent studies, synchronizing time-points, and calculating growth metrics (doubling time, volume/diameter change). Provides risk estimation for malignancy using the McWilliams et al. (2013) model based on patient and nodule characteristics. Used by radiologists and technologists in clinical settings. Output includes quantitative measurements (diameter, volume, mean HU), nodule characterization (type, location, shape, spiculation), and reports based on Lung-RADS guidelines. Assists clinicians in determining nodule growth patterns and malignancy risk, facilitating longitudinal monitoring and screening workflows.

Clinical Evidence

No new clinical trials were conducted for the device. The device relies on the validated McWilliams et al. (2013) risk prediction model, which was developed and validated using the Pan-Canadian Early Detection of Lung Cancer Study (PanCan) and British Columbia Cancer Agency (BCCA) cohorts. The model demonstrated AUC > 0.90. Device performance was validated using real recorded clinical data cases to simulate clinical workflow.

Technological Characteristics

Radiological image processing software; operates on thoracic CT images. Features automated segmentation (lungs, lobes, nodules), 3D/MIP visualization, and quantitative measurement algorithms. Connectivity includes integration with PACS/imaging systems. Complies with ISO 14971 (risk management), IEC 62304 (software lifecycle), and IEC 62366-1 (usability).

Indications for Use

Indicated for use as a diagnostic patient-imaging tool for the review and analysis of thoracic CT images to provide quantitative and characterizing information about lung nodules in a single study or over time. Intended for use by radiologists and technologists in diagnostic and screening evaluations, including Low Dose CT Lung Cancer Screening.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 23, 2017 Philips Medical Systems Nederland B.V. % Yoram Levy General Manager Osite 31 Haavoda Street Binyamina, 30500 ISRAEL Re: K162484 Trade/Device Name: Lung Nodule Assessment and Comparison Option (LNA) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ, JAK Dated: January 15, 2017 Received: January 23, 2017 Dear Yoram Levy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K162484 Device Name Lung Nodule Assessment and Comparison Option (LNA) Indications for Use (Describe) The Lung Nodule Assessment and Comparison Option is intended for use as a diagnostic patient-imaging tool. It is intended for the review and analysis of thoracic CT images, providing quantitative and characterizing information about nodules in the lung in a single study, or over the time course of several thoracic studies. Characterizations include diameter, volume and volume over time. The system automatically performs the measurements, allowing lung nodules and measurements to be displayed. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------|-----------------------------------------------------------| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. # 510(k) Summary # K162484 # Lung Nodule Assessment and Comparison Option This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. Date prepared: February 16, 2017 # I. Submitter's name and address | <b>I. Submitter's name and address</b> | | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | Establishment name: | Philips Medical Systems Nederland B.V. | | Establishment address: | Veenpluis 4-6<br>5684 PC Best<br>The Netherlands | | Establishment registration: | 3003768277 | | Primary Contact person: | Yoram Levy, Qsite<br>QA/RA Consultant<br>31 Haavoda Street<br>Binyamina, Israel 30500<br>Tel +(972)4-638-8837;<br>Fax (972)4-638-0510<br>Yoram@qsitemed.com | | Alternative contact person | Mr. Yonel Braunstein<br>Head of Regulatory and Clinical Affairs<br>Philips Medical Systems Nederland B.V<br>E-mail: Yonel.Braunstein@philips.com | | <b>II. Device information</b> | | | Trade name: | Lung Nodule Assessment and Comparison Option (LNA) | | Device Classification Name | System, Image processing, Radiological | | Device Class | Class II | | Classification Panel | LLZ, JAK | | Product Code | Radiological Image Processing Software | | Regulation Description | 21 CFR 892.2050, 21CFR 892.1750 | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. # III. Predicate device information # Predicate devices: | Device trade name | 510(k) number | Date of clearance | Classification name | Product code | Regulation | Class | Classification panel | |----------------------------------|---------------|---------------------|----------------------------------------------|--------------|-------------------|-------------|----------------------| | Lung<br>Analysis<br>Software | K151283 | October<br>30, 2015 | System, Image<br>processing,<br>Radiological | JAK,<br>LLZ | 21CFR<br>892.1750 | Class<br>II | Radiology | | Lung Nodule<br>Assessment<br>and | K023785 | Feb 10<br>2003 | System, Image<br>processing,<br>Radiological | 90<br>JAK | 21CFR<br>892.1750 | Class<br>II | Radiology | # Reference devices: | Device trade<br>name | 510(k)<br>number | Date of<br>Clearance | Classification<br>name | Product code | Regulation | Class | Classification<br>panel | |----------------------------------------------------------------------------------------|------------------|----------------------|-----------------------------------------------|--------------|--------------------|-------|-------------------------| | Philips<br>Multi-slice<br>CT system<br>with Low<br>Dose CT<br>lung cancer<br>screening | K153444 | April 8,<br>2016 | Computed<br>Tomography<br>X- ray system | JAK | 21 CFR<br>892.1750 | II | Radiology | | Brilliance<br>iCT<br>(Brilliance<br>Volume) | K060937 | June 5,<br>2006 | System, X-<br>Ray,<br>Tomography,<br>Computed | JAK | 21 CFR<br>892.1750 | II | Radiology | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white. | I4 (Integrated<br>Intelligent<br>Imaging<br>Informatics)<br>system | K160315 | February<br>19, 2016 | Picture<br>archiving and<br>communication<br>system | LLZ,<br>JAK | 21 CFR<br>892.2050 | II | Radiology | |--------------------------------------------------------------------|---------|----------------------|-----------------------------------------------------|-------------|--------------------|----|-----------| | EBW NM2.0 | K111336 | February<br>3, 2011 | Picture<br>archiving and<br>communication<br>system | LLZ | 21 CFR<br>892.2050 | II | Radiology | # IV. Device description The Lung Nodule Assessment and Comparison Option application is intended for use as a diagnostic patient-imaging tool. It is intended for the review and analysis of thoracic CT images, providing quantitative and characterizing information about nodules in the lung in a single study, or over the time course of several thoracic studies. The system automatically performs the measurements, allowing lung nodules and measurements to be displayed. The user interface and automated tools help to determine growth patterns and compose comparative reviews. The Lung Nodule Assessment and Comparison Option application requires the user to identify a nodule and to determine the type of nodule in order to use the appropriate characterization tool. Lung Nodule Assessment and Comparison Option may be utilized in both diagnostic and screening evaluations supporting Low Dose CT Lung Cancer Screening*. *The screening must be performed within the established inclusion criteria of programs/ protocols that have been approved and published by either a governmental body or professional medical society. -Please refer to clinical literature that includes the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information. # Key Features - 1. Automatic Lungs and Lobes segmentation {6}------------------------------------------------ - 2. Single-click lung nodule segmentation - 3. Manual nodule segmentation editing tools - 4. Load up to 8 concurrent studies for temporal measurements - 5. Restore previously segmented nodules from prior studies for comparison - 6. Enhanced comparison feature that allows matching and establishing correspondence between pre identified nodules in two studies - 7. Synchronization between studies from different time-points - 8. 3D or MIP visualization of segmented nodules - 9. Reporting based on Lung-RADS™ guidelines and option to export to Clipboard - 10. Risk Calculator tool based on patient and nodule characteristics for estimation of the probability that lung nodules detected on baseline screening low-dose CT scans are malignant, based on McWilliams, Annette, et al. "Probability of cancer in pulmonary nodules detected on first screening CT." New England Journal of Medicine 369.10 (2013): 910-919. - 11. Automatic software calculation of the following measurements for each segmented nodule: - Quantification of nodule parameters: A - Long Axis- Longest diameter on an axial slice (mm) - · Short axis- Longest diameter perpendicular to the long axis on the same - slice (mm) . - Average \ Max 3D \ Effective diameter (mm) - Volume (mm3) - Mean HU (HU) - Manual edit of the nodule segmentation contour lines with automatic recalculation of . geometric measurements post-editing - > Specification of the following characteristics for each nodule in configurable pre-sets - Nodule type (solid, part-solid, ground-glass, calcified) - Lobe Location - Nodule shape (Circular/Oval/Triangular) - Nodule Spiculation - > Comparison and matching feature automatic calculations of the following measurements between each follow-up scan and the previous scan: {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the only element in the image and it is centered. - I Doubling time in days - I Percent (%) and absolute change of all numerical parameters (growth in nodule long axis, short axis, average diameter, max 3D diameter, effective diameter, volume, mean HU). ### Additional Information for Risk Calculator The Risk Calculator feature in LNA is based on the full model with spiculation developed by Brock University as described in McWilliams, et al (2013). This model allows estimating the probability that lung nodules detected on baseline screening low-dose CT scans are malignant. The model's performance was validated using two large population-based prospective studies: the Pan-Canadian Early Detection of Lung Cancer Study (PanCan) and the chemoprevention trials at the British Columbia Cancer Agency (BCCA), sponsored by the U.S. National Cancer Institute. Further details can be found in McWilliams, A., Tammemagi, M.C., Mayo, J.R., Roberts, H., Liu, G., Soghrati, K., Yasufuku, K., Martel, S., Laberge, F., Gingras, M. and Atkar-Khattra, S., (2013). Probability of cancer in pulmonary nodules detected on first screening CT. New England Journal of Medicine, 369(10), pp.910-919 # V. Intended use The Philips Medical Systems Lung Nodule Assessment and Comparison Option is intended for use as a diagnostic patient-imaging tool. It is intended for the review and analysis of thoracic CT images, providing quantitative and characterizing information about nodules in the lung in a single study, or over the time course of several thoracic studies. Characterizations include diameter, volume and volume over time. The system automatically performs the measurements, allowing lung nodules and measurements to be displayed. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is plain white. # VI. Substantial Equivalence Comparison Philips believes that Lung Nodule Assessment and Comparison Option product is substantially equivalent to the identified predicate devices , primary- Toshiba's Lung Analysis Software (K151283) and predicate- Philips' Lung Nodule Assessment and Comparison option (K023785), with additional/enhanced functionality deriving from the reference devices: Philips Multi-slice CT system with Low Dose CT lung cancer screening (K153444), Brilliance iCT (Brilliance Volume) (K060937), 14 (Integrated Intelligent Imaging Informatics) system (K160315) and EBW NM2.0 (K111336) Below is the comparison table to summarize the similarities and differences of the subject, predicates, and reference devices: | Feature | The proposed<br>device:<br>Lung Nodule<br>Assessment and<br>Comparison<br>Option (LNA<br>application) | Primary Predicate:<br>Toshiba's Lung<br>Analysis Software<br>(K151283) | Predicate:<br>Philips' Lung<br>Nodule<br>Assessment<br>and<br>Comparison<br>option<br>(K023785) | Reference<br>device | |-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------| | Device Classification<br>Name | System, Image<br>processing,<br>Radiological | System, Image<br>processing,<br>Radiological | System,<br>Image<br>processing,<br>Radiological | Not relevant<br>for this<br>subject | | Device Class | Class II | Class II | Class II | Not relevant for<br>this subject | | Classification<br>Panel | Radiology | Radiology | Radiology | Not relevant for<br>this subject | | Product Code | LLZ, JAK | JAK, LLZ | 90 JAK | Not relevant for<br>this subject | | Regulation<br>Description | Radiological<br>Image Processing<br>Software | Radiological Image<br>Processing Software | Radiological<br>Image<br>Processing<br>Software | Not relevant for<br>this subject | | Regulation<br>Number | 21 CFR 892.2050<br>21CFR 892.1750 | 21CFR 892.1750 | 21CFR<br>892.1750 | Not relevant for<br>this subject | | Indication For<br>Use | The Lung Nodule<br>Assessment and<br>Comparison<br>Option is intended<br>for use as a<br>diagnostic patient-<br>imaging tool.<br>It is intended for<br>the review and<br>analysis of<br>thoracic CT<br>images, providing<br>quantitative and<br>characterizing<br>information about<br>nodules in the<br>lung in a single<br>study, or over the<br>time course of<br>several thoracic<br>studies.<br>Characterizations<br>include diameter,<br>volume and<br>volume over time.<br>The system<br>automatically<br>performs the<br>measurements,<br>allowing lung<br>nodules and<br>measurements to<br>be displayed. | The separately licensed<br>Lung Analysis option is<br>intended for the review<br>and analysis of thoracic<br>CT images for the<br>purposes of<br>characterizing nodules<br>in the lung in a single<br>study, or over the time<br>course of several<br>thoracic studies.<br>Characterizations<br>include diameter,<br>volume and volume<br>over time. The system<br>automatically<br>performs the<br>measurements, allowing<br>lung nodules and<br>measurements to be<br>displayed. | The Lung<br>Nodule<br>Assessment<br>and<br>Comparison<br>Option is<br>intended for<br>use as a<br>diagnostic<br>patient-<br>imaging tool.<br>It is intended<br>to provide<br>quantitative<br>information<br>about<br>physician-<br>indicated lung<br>nodules that<br>are identified<br>on high-<br>resolution<br>computed<br>tomography<br>images of the<br>lung. | Not relevant<br>for this subject | | Intended users | Radiologists and<br>Technologist | Radiologists,<br>Technologist and<br>Oncologists | Radiologists<br>and<br>Technologist | Not relevant<br>for this<br>subject | | Intended body<br>part | Chest | Chest | Chest | Not relevant for<br>this subject | | Type of scans | Thoracic CT<br>images | Thoracic CT images | Thoracic CT<br>images | Not relevant for<br>this subject | | 2D image review | Yes | Yes | Yes | Not relevant for<br>this subject | | 3D image review | Yes | Yes | Yes | Not relevant for<br>this subject | | 2D and 3D<br>comparative<br>review | Yes | Yes | No | Not relevant for<br>this subject | | | | | | | | 2D<br>measurements | Yes | Yes | Yes | Not relevant for<br>this subject | | Segmentation of<br>lung airway,<br>lungs and lung<br>lobes | Yes | Yes | No | Brilliance iCT<br>(Brilliance<br>Volume),<br>(K060937) | | Single click lung<br>nodule<br>segmentation | Yes | Yes | Yes | Not relevant for<br>this subject | | Nodule<br>Characteristics | Yes | Yes | No | Not relevant for<br>this subject | | Automatic<br>calculation of<br>measurements<br>for each<br>segmented<br>nodule | Yes<br>■ Short<br>axis- Longest<br>diameter<br>perpendicular to<br>the long axis on<br>the slice(mm)<br>■ Long<br>Axis- Longest<br>diameter on an<br>axial slice (mm)<br>■ Average<br>\ Max 3D \<br>Effective diameter<br>(mm)<br>■ Volume<br>(mm3)<br>■ Mean densities<br>(HU) | Yes<br>■ Volume (mm³)<br>■ Mean diameter<br>(mm)<br>■ Maximum<br>diameter (mm)<br>■ Short axis<br>diameter (mm)<br>■ Average/mini<br>mum/maximum<br>densities (HU) | No | Not relevant for<br>this subject | | Comparison and<br>Matching | Yes | Yes | Yes | EBW NM<br>2.0<br>(K111336)<br>■ I4<br>(Integrated<br>Intelligent<br>Imaging<br>Informatics)<br>system<br>■ (K160315) | | Comparison and<br>matching<br>automatic<br>calculations<br>between each<br>follow-up scan<br>and the previous<br>scan | Yes<br>■ Doubling<br>time in days<br>■ Percent<br>(%) and absolute<br>change of all<br>numerical<br>parameters | Yes<br>■ Elapsed time<br>in days<br>■ Doubling time<br>in days<br>Percent (%) growth in<br>nodule volume | Yes | Not relevant for<br>this subject | | | long axis, short<br>axis, average<br>diameter, max 3D<br>diameter,<br>effective<br>diameter, volume,<br>mean HU). | | | Not relevant for | | Workflow | ■ Detect and<br>Segment<br>■ Comparison and<br>Matching<br>■ Results | ■ Point-and-click<br>detection<br>■ Automated contouring<br>■ Automated<br>measurements<br>■ Manual correction | ■ Detect and<br>Segment<br>■ Comparison<br>and Matching<br>■ Results | this subject | | Loading multiple<br>studies | Yes<br>Up to 8 studies | Yes<br>Up to 3 studies | Yes<br>Up to 3 studies | Not relevant for<br>this subject | | Supporting Low-<br>dose CT | Yes | Yes | No | Not relevant for<br>this subject | | Reporting results | Yes<br>The results<br>includes the<br>following:<br>• Patient<br>related<br>information<br>• Dictation<br>• Table with<br>Nodule result<br>table and<br>additional<br>findings<br>• Lung-<br>RADS<br>• Risk<br>Calculator | Yes<br>The results includes the<br>following:<br>• Dictation<br>• Table with Nodule<br>result table<br>• Lung-RADS<br>• Fleischer<br>• Criteria | Yes<br>The results<br>includes the<br>following:<br>• Dictation Table<br>with Nodule<br>result table | Not relevant for<br>this subject | | Printing Option | Yes | Yes | Yes | Not relevant for<br>this subject | {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is plain white, which makes the blue letters stand out. {10}------------------------------------------------ Image /page/10/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is plain white, which makes the blue letters stand out. {11}------------------------------------------------ Image /page/11/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is plain white, which makes the blue letters stand out. Philips Medical Systems Lung Nodule Assessment and Comparison Option (LNA application) and the identified predicate devices, primary- Toshiba's Lung Analysis Software (K151283) and predicate- Philips' Lung Nodule Assessment and Comparison option (K023785) are substantial equivalent in terms of indication for use and intended {12}------------------------------------------------ users, design features, principle of operation and fundamental scientific technology, and safety and/or effectiveness. The above listed technological differences are considered low risk, providing further support to clinicians in accessing current technologies or features. These functionalities are derived from reference predicate devices and were verified and validated, and do not raise new questions on safety and/or effectiveness. These features have not changed the intended use and operational principles of the device. Therefore, the Lung Nodule Assessment and Comparison Option is substantially equivalent to the identified predicate devices , primary- Toshiba's Lung Analysis Software (K151283) and predicate- Philips' Lung Nodule Assessment and Comparison option (K023785)and the reference devices in terms of technological characteristics. # VII. Brief discussion of the nonclinical tests submitted, referenced or relied on No performance standards for PACS systems or components have been issued under the authority of Section 514. Non-clinical performance testing has been performed on ISPP and demonstrates compliance with the following International and FDA-recognized consensus standards and FDA guidance document: - . ISO 14971 Medical devices – Application of risk management to medical devices - IEC 62304 Medical device software Software life cycle processes - IEC 62366-1 Medical devices Part 1:Application of usability engineering to medical ■ devices - . Guidance for Industry and FDA Staff – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Lung Nodule Assessment and Comparison Option was tested in accordance with Philips verification and validation processes. Verification and Validation tests have been performed to address intended use, the technological characteristics claims, requirement specifications and the risk management results. For the Lung Nodule Assessment and Comparison Option (LNA), validation and verification {13}------------------------------------------------ activities were conducted to provide objective evidence that the design meets user needs and intended use and application specification. These validation activities assure that the lung and lobe segmentation, the comparison, as well as the nodule matching and propagation functionality, and the Prefill functionality for the Lung RADS score and the risk prediction are adequate from an overall product perspective. The LNA application was validated using real recorded clinical data cases in order to simulate the actual use of the software .Each test case was evaluated for the complete clinical workflow. In addition, a usability study was conducted according to the standards The test results in this 510(k) premarket notification demonstrate that Lung Nodule Assessment and Comparison Option: - . Complies with the aforementioned international and FDA-recognized consensus standards and FDA guidance document, and - . Meets the acceptance criteria and is adequate for its intended use and specifications. # VIII. Brief discussion of clinical tests submitted, referenced or relied on McWilliams, A et al ' conducted a population-based prospective clinical study to determine factors predicting the probability that lung nodules detected on the first screening low-dose CT scans are malignant or will be found to be malignant on follow-up. The authors indicated that the final parsimonious and full models showed excellent discrimination and calibration, with areas under the receiver-operating-characteristic curve of more than 0.90, even for nodules that were 10 mm or smaller in the validation set. McWilliams A et al concluded that Predictive tools based on patient and nodule characteristics can be used to accurately estimate the probability that lung nodules detected on baseline screening low-dose CT scans are malignant. The investigators analyzed data from two cohorts of participants undergoing low-dose CT screening. The development data set included participants in the Pan-Canadian Early Detection of <sup>1</sup> McWilliams, A., Tammemagi, M.C., Mayo, J.R., Roberts, H., Liu, G., Soghrati, K., Yasufuku, K., Martel, S., Laberge, F., Gingras, M. and Atkar-Khattra, S.,. Probability of cancer in pulmonary nodules detected on first screening CT. New England Journal of Medicine, 2013: 369(10), 910-919. {14}------------------------------------------------ Lung Cancer Study (PanCan). The validation data set included participants involved in chemoprevention trials at the British Columbia Cancer Agency (BCCA), sponsored by the U.S. National Cancer Institute. The final outcomes of all nodules of any size that were detected on baseline low-dose CT scans were tracked. Parsimonious and fuller multivariable logisticregression models were prepared to estimate the probability of lung cancer. # IX. The conclusions drawn from the nonclinical and clinical tests Verification and Validation (V&V) activities required to establish performance and functionality of Lung Nodule Assessment and Comparison Option were performed. Testing performed demonstrated the Lung Nodule Assessment and Comparison Option meets all defined functionality requirements and performance claims. - Overall The Lung Nodule Assessment and Comparison Option is X. substantially equivalent to the identified predicate devices , conclusion: primary- Toshiba's Lung Analysis Software (K151283) and predicate- Philips'Lung Nodule Assessment and Comparison option (K023785), in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, verification and validation testing demonstrate the safety and efficacy of the device to meet its intended use and specification. Philips Medical believes that the proposed device, Lung Nodule Assessment and Comparison Option, is substantially equivalence to its identified predicate device and is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...