K153612 · Custom Orthopaedic Solutions, Inc. · LLZ · Mar 15, 2016 · Radiology
Device Facts
Record ID
K153612
Device Name
ArthrexVIP Web Portal
Applicant
Custom Orthopaedic Solutions, Inc.
Product Code
LLZ · Radiology
Decision Date
Mar 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The Arthrex VIP Web Portal is intended for use as a software interface of imaging information from a medical scanner such as a CT scanner. It is also intended as software for displaying/editing implant placement and surgical treatment options that were generated in the OrthoVis Desktop Software by trained COS technicians. The Arthrex VIP Web Portal is intended for use with the Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS) and with the Arthrex OrthoVis Preoperative Plan. It is indicated for use with the following glenoid implant lines: Arthrex Univers Apex, Arthrex Univers II, and Arthrex Univers Revers.
Device Story
ArthrexVIP Web Portal is a software interface for preoperative planning of total shoulder joint replacement. Surgeons upload medical images (e.g., CT scans) to the portal; COS technicians download these images to create preoperative plans using OrthoVis Desktop Software. Surgeons then access the portal to review, edit, or approve the proposed implant placement and orientation. Approved plans are downloaded by technicians to produce patient-specific instruments (Arthrex Glenoid IRIS). The device facilitates communication between surgeons and technicians to finalize surgical treatment options. It is used in clinical settings by surgeons to improve surgical planning accuracy and patient outcomes.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical software verification and validation, including regression testing, unit testing, code reviews, and integration testing.
Technological Characteristics
Web-based software interface for image transfer and preoperative planning. Features include viewer controls, implant placement editing, and measurement tools. Operates as a client-server system for surgeon-technician collaboration. Software-only device.
Indications for Use
Indicated for use by surgeons to review, edit, and approve preoperative plans for shoulder arthroplasty using Arthrex Univers Apex, Univers II, and Univers Revers glenoid implant lines.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS) (K151500)
Related Devices
K193523 — ArthrexVIP Web Portal · Arthrex, Inc. · Apr 30, 2020
K230904 — Arthrex Virtual Implant Positioning (VIP) System · Arthrex, Inc. · Jun 22, 2023
K241097 — Arthrex Virtual Implant Positioning (VIP) System Software · Arthrex, Inc. · Oct 2, 2024
K162697 — Arthrex Glenoid Intelligent Reusable Instrument System, Arthrex OrthoVis Preoperative Plan, SmartBase for Arthrex Glenoid IRIS, ArthrexVIP Web Portal · Custom Orthopaedic Solutions, Inc. · Feb 10, 2017
K151568 — Arthrex OrthoVis Preoperative Plan · Custom Orthopaedic Solutions, Inc. · Jul 31, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are depicted in a simple, flowing line style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 15, 2016
Custom Orthopaedic Solutions, Incorporated Mr. Keith Grafmever Project Manager 10000 Cedar Avenue Cleveland, Ohio 44106
Re: K153612
Trade/Device Name: ArthrexVIP Web Portal Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ, KWS, PHX Dated: December 7, 2015 Received: December 17, 2015
Dear Mr. Grafmeyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) NA K153612
Device Name ArthrexVIP Web Portal
#### Indications for Use (Describe)
The Arthrex VIP Web Portal is intended for use as a software interface of imaging information from a medical scanner such as a CT scanner. It is also intended as software for displaying/editing implant placement and surgical treatment options that were generated in the OrthoVis Desktop Software by trained COS technicians. The Arthrex VIP Web Portal is intended for use with the Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS) and with the Arthrex OrthoVis Preoperative Plan. It is indicated for use with the following glenoid implant lines: Arthrex Univers Apex, Arthrex Univers II, and Arthrex Univers Revers.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k) SPONSOR / MANUFACTURER: | Custom Orthopaedic Solutions, Inc.<br>10000 Cedar Avenue<br>Cleveland, Ohio 44106 |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON: | Keith Grafmeyer<br>Project Manager<br>Custom Orthopaedic Solutions<br>10000 Cedar Ave.<br>Cleveland, OH 44106<br>Tele: (216) 445 -3403<br>Email: kgrafmeyer@customorthopaedics.com |
| TRADE NAME: | ArthrexVIP Web Portal |
| DATE PREPARED: | 07-Dec-2015 |
| COMMON NAMES: | Image processing system and preoperative software for<br>simulating / evaluating implant placement and surgical<br>treatment options |
| Product | Product Code | Regulation and<br>Classification Name | Device<br>Class |
|-----------------------|--------------|------------------------------------------------------------------------------------------|-----------------|
| ArthrexVIP Web Portal | LLZ | 21 CFR 892.2050<br>Picture Archiving and<br>Communications System | II |
| ArthrexVIP Web Portal | KWS | 21 CFR 888.3660<br>Prosthesis, Shoulder, Semi-<br>Constrained, Metal/Polymer<br>Cemented | II |
| ArthrexVIP Web Portal | PHX | 21 CFR 888.3660<br>Shoulder Prosthesis, Reverse<br>Configuration | II |
#### Predicate Devices:
K151501: OrthoVis Web Portal K151568: Arthrex OrthoVis Preoperative Plan K151500: Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS)
# Device Description:
The ArthrexVIP Web Portal is composed of software intended for use to facilitate upload of medical
{4}------------------------------------------------
images, preoperative planning, and plan approval of placement and orientation of total shoulder joint replacement components. Each surgeon user's uploaded images are associated with specific cases and associated with that surgeon's profile. Uploaded images can be downloaded from the portal by COS technicians and used to create preoperative plans (see 510(k) K151568) in the OrthoVis Desktop Software. The surgeon user is then able to login to the ArthrexVIP Web Portal to review the plan and either approve or modify the location and/or orientation of the joint replacement component. The approved plan is then downloaded by COS technicians for production (see 510(k) K151500 and K151568) as part of the Arthrex Glenoid IRIS device.
# Intended Use and Indications:
The ArthrexVIP Web Portal is intended for use as a software interface and for the transfer of imaging information from a medical scanner such as a CT scanner. It is also intended as software for displaying/editing implant placement and surgical treatment options that were generated in the OrthoVis Desktop Software by trained COS technicians. The ArthrexVIP Web Portal is intended for use with the Arthrex Glenoid Intelligent Reusable Instrument System (Arthrex Glenoid IRIS) and with the Arthrex OrthoVis Preoperative Plan. It is indicated for use with the following glenoid implant lines: Arthrex Univers Apex, Arthrex Univers II, and Arthrex Univers Revers.
# Basis of Substantial Equivalence:
The ArthrexVIP Web Portal has the same intended use and indications for use as the OrthoVis Web Portal (K151501), with the exception of ArthrexVIP being indicated for use solely with the Arthrex line of shoulder arthroplasty implants.
The subject device is a version of the predicate device with changes to the following:
- Measurement Features
- Viewer Controls
- Implant Controls and Options
- Organization of information
- User Interface (UI)
# Non-Clinical Testing
The following testing was performed to demonstrate substantial equivalency of the ArthrexVIP Web Portal to the OrthoVis Web Portal:
- Software verification and validation
- Regression Testing
- Unit Testing
- Code reviews and checks
- Integration Testing
- Dimensional Validation (performed on predicate device and code has not changed for the subject device)
# Clinical Testing
Clinical testing was not necessary to determine substantial equivalence between to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.