K130724 · Straumann USA · LLZ · Jun 28, 2013 · Radiology
Device Facts
Record ID
K130724
Device Name
CODIAGNOSTIX IMPLANT PLANNING SOFTWARE
Applicant
Straumann USA
Product Code
LLZ · Radiology
Decision Date
Jun 28, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
coDiagnostiX is an implant planning and surgery planning software tool intended for use by dental professionals who have appropriate knowledge in dental implantology and surgical dentistry. This software reads imaging information output from medical scanners such as CT or DVT scanners. It allows pre-operative simulation and evaluation of patient anatomy and dental implant placement. For automated manufacturing of drill guides in the dental laboratory environment, the coDiagnostiX software allows for export of data to 3D manufacturing systems. Alternatively, coDiagnostiX can provide printouts of template plans for the creation of surgical templates using a manually operated gonyX table.
Device Story
coDiagnostiX is a dental implant planning software used by dental professionals in clinical or laboratory settings. It ingests 3D imaging data (CT/DVT) via DICOM. The software enables visualization, measurement, and virtual implant placement. Key features include nerve canal identification, 3D jaw reconstruction, bone densitometry, and segmentation of anatomical structures. It supports surgical planning by allowing users to refine implant positions and generate surgical templates. Outputs include STL files for 3D manufacturing or printed plans for use with manual gonyX tables. The software aids clinicians in pre-operative evaluation and surgical guidance, potentially improving implant placement accuracy.
Clinical Evidence
Bench testing only. Software verification and validation were conducted in accordance with FDA guidance 'General Principles of Software Validation' to ensure design inputs met design outputs. No clinical data was required or provided.
Technological Characteristics
Dental image diagnosis and implant planning software. Operates on Windows or Mac (via virtualization). Supports DICOM, JPEG, BMP, and STL formats. Features include nerve module, 3D segmentation, bone densitometry, and surgical template planning. Connectivity via network or local database. Software-based image processing.
Indications for Use
Indicated for dental professionals with knowledge in implantology and surgical dentistry to perform pre-operative simulation, evaluation of patient anatomy, and dental implant placement planning using CT or DVT scan data. Not for direct patient contact or use with life-sustaining devices.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Straumann company logo and the text "Page 1 of 4". The Straumann logo is a stylized design with the company name in bold, black font. The text "Page 1 of 4" is handwritten in black ink and is located to the right of the logo. The text indicates that this is the first page of a four-page document.
### 510(k) Summary - K130724
JUN 2 8 2013
#### 1. Applicant's Name and Address
Straumann US (on behalf of Institut Straumann AG) 60 Minuteman Rd. Andover, MA 01810 Telephone Number: 800-448-8168, ext 2513 978-747-0023 Fax Number: Contact Person: Elaine Alan Regulatory Project Manager
- March 15, 2013 2. Date of Submission:
#### 3. Name of the Device
Trade Name: Common Name: Classification Name: Regulation Number:
coDiagnostiX Implant Planning Software Dental 3D diagnosis and implant planning software Image Processing System 8892.2050
### Legally Marketed Device to which Equivalence is Claimed (Predicate 4. Device)
K071636, IVS coDiagnostiX
#### 5. Description of the Device
coDiagnostiX Implant Planning Software is designed for diagnosis of 3dimensional datasets and precise image-guided and reproducible preoperative planning of dental implants. The software is provided in three configurations: station, client or server. Patient image data is received from various sources such as CT or DVT scanning. The scanned data will be read with the coDiagnostiX DICOM transfer module using the DICOM format, converted into 3dimensional datasets and stored in a database. Surgical planning is performed through the calculation of special views, analysis of graphic data and the placement of virtual dental implants. Additional functions are available to the user for refinement of the surgical planning, such as:
- Active measurement tools, length and angle, for individual measuring of implant positions
- · Nerve module to assist in distinguishing the nervus mandibularis channel
- . 3D cut for a 3dimensional cut through the jaw for fine adjustment of the implant position
{1}------------------------------------------------
- Segmentation module for coloring several areas inside the slice . dataset, e.g., jaw bone, natural tooth series, or types of tissue, e.g., bone, skin, and creating a 3D reconstruction for the dataset
- Parallelizing function for adjustment of adjacent images ●
- Bone densitometry with a density statistic for density measuring in the area around the positioned implant; a density allocation along, and transverse to, the implant cover area is displayed
All working steps are saved automatically to the patient file; one patient file may contain multiple surgical plan proposals which allow the user to choose the ideal surgical plan. The guided surgical template plan, STL file and/or guided surgical protocol is generated from the final surgical plan.
#### 6. Intended Use of the Device
coDiagnostiX is an implant planning and surgery planning software tool intended for use by dental professionals who have appropriate knowledge in dental implantology and surgical dentistry. This software reads imaging information output from medical scanners such as CT or DVT scanners. It allows pre-operative simulation and evaluation of patient anatomy and dental implant placement.
For automated manufacturing of drill guides in the dental laboratory environment, the coDiagnostiX software allows for export of data to 3D manufacturing systems. Alternatively, coDiagnostiX can provide printouts of template plans for the creation of surgical templates using a manually operated gonyX table.
#### 7. Technological Characteristics
coDiagnostiX Implant Planning Software is a dental image diagnosis and implant planning software. A comparison of the technological characteristics of predicate and subject device is below.
{2}------------------------------------------------
| FEATURE | PROPOSED DEVICE<br>coDiagnostiX | PREDICATE DEVICE<br>IVS coDiagnostiX |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Subject Submission | K071636 |
| Intended Use | Straumann coDiagnostiX is an implant<br>planning and surgery planning software tool<br>intended for use by dental practitioners with<br>appropriate training in dental implantology.<br>This software reads imaging information<br>output from medical scanners such as CT or<br>DVT scanners. It allows pre-operative<br>simulation and evaluation of patient anatomy,<br>dental implant placement and surgical<br>treatment options.<br><br>For automated manufacturing of drill guides in<br>the dental laboratory environment, the<br>coDiagnostiX software allows for export of<br>data to 3D manufacturing systems, or<br>coDiagnostiX can provide printouts of<br>templant plans for the creation of surgical<br>templates using a manually operated gonyX<br>table.<br><br>The target patient population is the general<br>public. coDiagnostiX is not intended to be<br>used in direct contact with the patient nor is it<br>intended to be used with life sustaining<br>devices. | coDiagnostiX is intended for use by<br>medical trained people as a Windows-<br>based diagnosis and implant planning<br>software. This software is an interface for<br>the transfer of imaging information from<br>medical scanners such as CT or DVT<br>scanners and also a pre-operative software<br>for simulation and evaluation dental implant<br>placement and surgical treatment options.<br><br>The patient population will be the general<br>public. coDiagnostiX is not intended to be<br>used in direct contact with the patient nor<br>it intended to be used with life sustaining<br>devices.<br><br>It is possible to use coDiagnostiX as a<br>pattern for manufacturing drilling template<br>with conventional rotary tables or gonyX<br>table from IVS Solutions AG in laboratory<br>environment. This drilling template is then<br>used in direct contact with the patient to<br>realize the implant planning with<br>coDiagnostiX. |
| Minimum<br>System<br>Requirements | OS: Microsoft Windows 7, Windows SP,<br>Vista<br>Processor: Intel Core or AMD Athlon 64<br>X2<br>Main memory: 2 GB<br>Hard disk: 1GB free, plus 50 MB per<br>plan<br>Screen resolution: 1024 x 768 pixels<br>DVD drive<br>One free USB port<br>Optional: Microsoft SQL Server Desktop<br>Engine (MSDE) if user selects to use<br>the SQL Server Module<br>The software can also be run on the<br>Apple operating system with minimum<br>system requirements of:<br>Macintosh computer with Intel<br>processor<br>Mac OSX 10.6 or newer<br>"Parallels Desktop" for Macintosh<br>including valid Windows license<br>(software purchased separately) | WIN 98/NT/2000/XP<br>Pentium compatible processor with<br>800MHz<br>128 MB RAM<br>500 MB hard disc<br>17" color display with 1024x768<br>resolution and 16.7 million colors<br>CD ROM device<br>USB interface<br>Optional: Microsoft SQL Server<br>Desktop Engine (MSDE) if user selects<br>to use the SQL Server Module<br>The software can also be run on the<br>Apple operating system with minimum<br>system requirements of:<br>Macintosh computer with Intel<br>processor<br>Mac OSX 10.6 or newer<br>"Parallels Desktop" for Macintosh<br>including valid Windows license<br>(software purchased separately) |
| Target Population | Same | General public |
| Supported image<br>formats | Same | JPEG, BMP |
| DICOM compliant | Yes | yes |
| Image import | Same, plus STL file format | DICOM data from CT/DVT scanner<br>saved on CDROM, PC or network |
| Image export | STL file format | unknown |
| Printouts | Same | Report/plan with patient information,<br>screenshot, implant list, template |
and the comments of the comments of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the contribution of the co
{3}------------------------------------------------
K130724
Page 4 of 4
#### 8. Performance Testing
Software verification and validation testing were performed to ensure that the device subject to this 510(k) Premarket Notification functions as intended and that design input matches design output. Testing has been carried out in accordance with the FDA guidance document, "General Principles of Software Validation: Final Guidance for Industry and FDA Staff," issued on January 11, 2002. The proposed software met the acceptance criteria.
#### க் Conclusion
coDiagnostiX Implant Planning Software has identical intended use, and operational and functional features, and level of concern as the currently cleared software. Results of the software validation indicate that the software performs as intended and that the proposed changes do not raise any new issues of safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three diagonal lines representing the branches of government. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" is arranged in a circular pattern around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Cantral Center - WO66-G609 Silver Spring, MD 20093-0002
June 28. 2013
Straumann USA. LLC % Ms. Elaine Alan Regulatory Project Manager 60 Minuteman Road ANDOVER MA 01830
Re: K130724
Trade/Device Name: coDiagnostiX Implant Planning Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 17, 2013 Received: May 20, 2013
Dear Ms. Alan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practive requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Alan
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K130724
Device Name: coDiagnostiX Implant Planning Software
Indications for Use:
coDiagnostiX is an implant planning and surgery planning software tool intended for use by dental professionals who have appropriate knowledge in dental implantology and surgical dentistry. This software reads imaging information output from medical scanners such as CT or DVT scanners. It allows pre-operative simulation and evaluation of patient anatomy and dental implant placement.
For automated manufacturing of drill guides in the dental laboratory environment, the coDiagnostiX software allows for export of data to 3D manufacturing systems. Alternatively, coDiagnostiX can provide printouts of template plans for the creation of surgical templates using a manually operated gonyX table.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-the-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiologic Health (OIR)
Jamingh. Mesa
Division Sign-Off Office of In Vitro Diagnostics and Radiologic Health
510(k) K130724
Page 1 of 1