K061961 · Pie Medical · LLZ · Sep 6, 2006 · Radiology
Device Facts
Record ID
K061961
Device Name
ART LAB SOFTWARE
Applicant
Pie Medical
Product Code
LLZ · Radiology
Decision Date
Sep 6, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
ART.LAB is a software package to be used on an external personal computer for the real time automatic measurement of the Intima Media Thickness, diameter and distension of the carotid artery from RF based signal processing obtained from Esaote or Pie Medical ultrasound systems. The Arterial Stiffness Indicator is calculated from distension and blood pressure (systolic and diastolic).
Device Story
ART.LAB is an optional software package installed on an external PC; processes RF signals acquired from Esaote or Pie Medical ultrasound systems via a data acquisition card. Performs real-time automatic measurement of carotid artery Intima Media Thickness (IMT), diameter, and distension. Calculates Arterial Stiffness Indicator using distension and patient blood pressure (systolic/diastolic) inputs. Used in clinical settings by healthcare professionals to assist in vascular assessment. Output provides quantitative vascular measurements to aid clinicians in evaluating arterial health and stiffness.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and bench testing of the software's measurement capabilities.
Technological Characteristics
Software-based image processing system. Operates on external PC (P3 1 GHz, 256 MB RAM, Windows 98/XP). Requires data acquisition card for RF signal input from ultrasound systems. Uses RF signal processing for automated vascular measurements. Connectivity via connection cable for RF, Z, and trigger signals.
Indications for Use
Indicated for the automatic measurement of carotid artery Intima Media Thickness, diameter, and distension, and calculation of Arterial Stiffness Indicator using RF signals from ultrasound systems. For prescription use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K06 196 1
510(k) Summary ART.LAB Pie Medical
SEP - 6 2006
# 510(k) Summary
The following safety and effectiveness summary has been prepared pursuant to requirement for 510(k) summaries specified in 21CFR¶807.92(a).
807.92(a)(1)
### Submitter Information
| Carri Graham, Official Correspondent | |
|--------------------------------------|----------------|
| 11460 N. Meridian Street, Suite 150 | |
| Carmel, IN 46032 | |
| Phone: | (317) 569-9500 |
| Facsimile: | (317) 569-9520 |
| Contact Person: | Carri Graham |
|-----------------|--------------|
| Date: | July 7, 2006 |
807.92(a)(2)
| Trade Name: | ART.LAB | |
|-------------------------|--------------------------------------------|----------|
| Common Name: | Picture Archiving and Communication System | |
| Classification Name(s): | Image Processing System | 892.2050 |
| Classification Number: | 90IYO<br>90IYN<br>90ITX<br>90LLZ | |
807.92(a)(3)
Predicate Device(s)
| K043360 | IMT.LAB | Pie Medical |
|---------|-----------------------|-------------------------------|
| K961144 | Cardiovision MS. 2000 | Touritu Engineering Co., Inc. |
| K032875 | SSD-5500 | Aloka |
Additional Substantial Equivalence Information is provided in the following substantial Equivalence Comparison Table.
{1}------------------------------------------------
510(k) Summary ART.LAB Pie Medical
#### Device Description 807.92(a)(4)
The ART.LAB software product is an optional software package to be used on an external personal computer for the real time automatic measurement of the Intima Media Thickness, diameter and distension of the carotid artery. The measurements are based upon RF signal processing obtained from Esaote or Pie Medical ultrasound systems. The Arterial Stiffness Indicator is calculated from the distension and blood pressure (systolic and diastolic). The ART.LAB software operates in conjunction with an ultrasound system, a linear probe and is installed on an external PC, used for signal processing.
Specifications: Ultrasound scanner with Fast-B-mode scanning capabilities and a linear probe PC P3 1 GHz with 256 MG memory Microsoft Windows 98 or XP Data acquisition card (for acquisition of RF-signals obtained from ultrasound system) Connection cable (from ultrasound system to PC, for RF, Z & trigger signals) ART.LAB software
807.92(a)(5) ART.LAB is a software package to be used on an external personal computer for the real time automatic measurement of the Intima Media Thickness, diameter and distension of the carotid artery from RF based signal processing obtained from Esaote or Pie Medical ultrasound systems. The Arterial Stiffness Indicator is calculated from distension and
Intended Use(s)
blood pressure (systolic and diastolic).
{2}------------------------------------------------
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the top half of the circle. In the center of the seal is a stylized eagle-like symbol with three curved lines representing its wings or feathers. The seal appears to be a government or official emblem.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Pie Medical % Ms. Carri Graham Consultant Anson Group, LLC 11460 N Meridian St., Ste 150 CARMEL IN 46032
SEP - 6 2006
Re: K061961
Trade/Device Name: ART.LAB Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 7, 2006 Received: July 11, 2006
### Dear Ms. Graham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a black and white circular logo. The logo contains the text "1906-2006" at the top, "PDA" in the center, and "Centennial" below that. There are three stars below the word "Centennial". The logo has a dotted border.
Protecting and Promoting Public Health
{4}------------------------------------------------
Page 2 -
ﺎ،
Please be advised that FDA's issuance of a substantial equivalence determination does not mean r rease of actives that i Driving that your device complies with other requirements of the Act that I Dri has made a and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI K Pat 601); facemig (2) CFR Paulation (21 CFR Part 820); and if applicable, the electronic form in all quality bystelle (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) Phils letter with and To Jour finding of substantial equivalence of your device to a legally premaince noticated device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific as Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under (21 CF From the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): __ < 0 6 / 9 / 1 6 0 / 1 / 6 / 1 / 6 /
Device Name: ART.LAB
Indications for Use:
ART.LAB is a software package to be used on an external personal computer for the real ANY LAD is a software paonage to otima Media Thickness, diameter and distension of thire automatic incasurement of the nal processing obtained from Esaote or Pie Medical the carond artery from Its base sigiffiess Indicator is calculated from distension and blood pressure (systolic and diastolic).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Nancy C. Bridger
(Division Sigh-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.