SECURVIEW DX SOFTCOPY WORKSTATION

K041555 · Hologic, Inc. · LLZ · Aug 9, 2004 · Radiology

Device Facts

Record IDK041555
Device NameSECURVIEW DX SOFTCOPY WORKSTATION
ApplicantHologic, Inc.
Product CodeLLZ · Radiology
Decision DateAug 9, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

The Hologic SecurView DX device is intended for selection, display, manipulation, filming and media interchange of multi-modality images from a variety of different storage and printing devices using DICOM or similar interface standards. The device used with FDA cleared monitors may be used by a trained physician for display, manipulation and interpretation of lossless compressed or non-compressed mammographic images for screening and diagnostic mammography, as well as any other DICOM multi-modality image. The SecurView DX is typically used by trained professionals, including, but not limited to, radiologists, technicians and assistants.

Device Story

SecurView DX is a multi-modality review workstation software; accepts DICOM-compliant medical images from external storage/printing devices; provides visualization, manipulation, and interpretation tools for radiologists and technicians. Operates on Windows 2000-based PC hardware with high-resolution monitors. System facilitates clinical review of mammography and other radiological images; supports lossless compressed or non-compressed image display. Output allows clinicians to interpret diagnostic images to support clinical decision-making. Used in clinical environments by trained medical staff.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Software-based workstation; runs on Windows 2000 OS; requires Intel CPU (>=2.0 GHz), >=2.0 GB RAM, >=140 GB HDD. Supports DICOM interface standards for image input/output. Utilizes high-resolution monitors. No specific algorithm architecture or hardware standards (ASTM) detailed.

Indications for Use

Indicated for trained professionals, including radiologists, technicians, and assistants, for the display, manipulation, and interpretation of DICOM multi-modality images, including screening and diagnostic mammography images.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K041555 ## SecurView DX AUG - 9 2004 510(k) Summary Product Name: SecurView DX Product Classification Name: Picture archiving and communication system CFR Section: 892.2050 Product Classification Code: LLZ Classification Panel: Radiology Class II | Manufacturer: | Hologic, Inc.<br>36 Apple Ridge Road<br>Danbury, CT 06810 USA | | | | | |-----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--| | Contact Person: | Gail Yaeker-Daunis<br>Telephone Number: (203) 731-8337<br>(203) 731-8440<br>Fax Number: | | | | | | Date Prepared: | June 4, 2004 | | | | | | | R. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1. 1.<br>1 2 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 | | | | | K033400 Seno Advantage, General Electric Medical Systems Predicate Device: Device Description: The Hologic SecurView DX is mainly a software product. It is Device Description. The Hologie Docurity hadiology images. The Secur View DX is a used for visualization and processing of argitus two high-resolution gray start on horned multi-modality review workstation: It metades we keyboard and workstation keypad. (FDA cleared for Mammography), 3-button mouse, keyboard and workstation keypad. (I'DA created for mailinegraphy); . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The soltware accepts mages which have occarded. Image processing is external to the Presentation Specification in the DIOON Station and and of the system or no compression at all when displaying mammography images. The minimum computer requirements are: Windows 2000® Operating System Intel CPU with a clock rate of 2.0 GHz or greater 2.0 GB RAM or greater 140 GB Hard drive or greater, operating at 160 Mbs or greater CD-ROM/R/RW, DVD+/-RW 10/100 Base TX Network Interface {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes extending from its head, representing health, services, and human needs. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG - 9 2004 Mr. Richard Follett Vice President, RA/QA Hologic, Inc. 35 Crosby Drive BEDFORD MA 01730 Re: K041555 Trade/Device Name: SecurView DX Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: June 4, 2004 Received: June 16, 2004 Dear Mr. Follett: We have reviewed your Section 510(k) premarket notification of intent to market the device WE have reviewed your Scotton 910(t) pe device is substantially equivalent (for the indications felerenced above and nave decembined ly marketed predicate devices marketed in interstate for use stated in the encrosure) to regally manating to the Medical Device Americal Device of Dove commerce prior to way 26, 1776, the classified in accordance with the provisions of the Federal Food, DVA devices that have been recuasified in accordains approval of a premarket approval application (PMA). and Cosment Act (Rea) that do not require approvince approvisions of the Act. The Y ou may, dierelore, market the detires, baojet requirements for annual registration, listing of general controls provisions of the tice, labeling, and prohibitions against misbranding, and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classince (see a0010) ins. Existing major regulations affecting your device can may be subject to such additional controls. Existing major regulations of t may be subject to such additional controlly - Lines 800 to 898. In addition, FDA may be found in the Ood of I outstan ongerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not i Picase oe advised hat I DA s issualloo of a device complies with other requirements of the Act that FDA has made a deceimnation that your istered by other Federal agencies. You must comply or any Pederal Statures and regalations administered of the registration and listing (2 CFR Parth in the with an the Act STCquirements, metading, curfacturing practice requirements as set forth in the 807), laocing (21 CFR Part 601), goods and 1820), and if application of the electronic product quality systems (QD) regulation (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will allow you to begin marketing your antial equivalence of your device to a legally premarket notification. The PDA inding of substantial equive and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please (1), please If you desire specific advice for your dones on one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the promotion and and the regulation entitled, "Misbranding Othce of Compliance at (301) 574-1057. Floor, production of obtain. Other general by relerence to prematic nonitation - (21 OF Act may be obtained from the Division of Small Information on your responsionalisms and its toll-free number (800) 638-2041 or Manufacturers, International and Consumer Assistance at its toll-free in the l Manufacturers, International and Collean.br ---------------------------------------------------------------------------------------------------------------------------------- Sincerely yours, Nancy C. Boyden Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## 510(k) No. K041555 ## Indications For Use ## Device Name: SecurView DX The Hologic SecurView DX device is intended for selection, display, manipulation, of different The Hologic Secur View DX device is imended 10 images from a variety of different filming and media interchange of multi-modality images and printing devices u filming and media merchange of indit-modality mages storage and printing devices using DICOM or similar interface standards. The device used with FDA cleared monitors may be used by a trained physician for The device used with FDA cleared inomitive into of lossless and on non-compressed or non-compressed or non-compressed display, manipulation and interpretation of Nossiess compressed on the country. mammographic images for screening and diagnostic mammography, as well as any other DICOM multi-modality image. The SecurView DX is typically used by trained professionals, including, but not limited The Secur View DX is typically dood of isl technicians and assistants. Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | | |----------------------|--| | OR | | | Over-the-Counter Use | | Nancy C Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices | 510(k) Number | K041555 | |---------------|---------| |---------------|---------|
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...