CEDARA I-SOFTVIEW AND CEDARA I-READMAMMO

K040468 · Cedara Software Corp. · LLZ · Apr 23, 2004 · Radiology

Device Facts

Record IDK040468
Device NameCEDARA I-SOFTVIEW AND CEDARA I-READMAMMO
ApplicantCedara Software Corp.
Product CodeLLZ · Radiology
Decision DateApr 23, 2004
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

Two and three dimensional image review, manipulation, analysis and therapy planning capabilities that support image management display needs in the medical environment from multiple locations within and outside the hospital. Productivity-Enhancing Second Console Workstations - Workstations designed to perform automated, routine tasks such as image review, printing and archiving as well as post processing capabilities that enable special services for referring physicians. Diagnostic Review Workstations - Workstations designed to assist radiologists and surgeons in conducting primary diagnostic review and planning through flexible and interactive manipulation of multi-modality softcopy images including the use of prosthetic template overlays, and including mammography. Physician's Review Workstations - Workstations designed to give easy and economic access to multi-modality softcopy images in multiple locations within and outside the hospital. (e.g. teleconferencing, teleradiology etc.) Lossy compressed mammographic images must not be used for primary diagnostic interpretation unless approved for use in digital mammography. Display monitors used for primary diagnostic interpretation of mammographic images must be approved for use in digital mammography.

Device Story

Cedara I-ReadMammo is medical image review workstation software; provides tools for viewing, manipulating, and analyzing multi-modality medical images. Input: multi-modality medical images. Operation: software enables viewing protocols, workflow management, and diagnostic measurements (e.g., prosthetic template overlays) for surgical planning. Output: processed medical images and measurements displayed on workstation monitors. Used by radiologists and surgeons in clinical environments for primary diagnostic review and planning. Benefits: optimizes workflow efficiency and provides convenient access to images for diagnostic and surgical planning purposes.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use compared to the predicate device.

Technological Characteristics

PACS medical imaging software; multi-modality image processing; supports 2D/3D visualization, measurement tools, and prosthetic template overlays. Software-based workstation application.

Indications for Use

Indicated for radiologists, surgeons, and referring physicians for multi-modality medical image review, manipulation, analysis, and therapy planning. Includes mammography. Contraindicated for primary diagnostic interpretation of mammographic images using lossy compressed images or non-approved display monitors.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for CEDARA. The logo consists of a stylized globe made up of horizontal lines. Below the globe is the word "CEDARA" in a bold, sans-serif font. Special 510(k) Premarket Notification Cedara I-SoftViews I-SoftViews KO40468 # 510 (k) Summary | Submitter: | Cedara Software Corp. | |----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | 6509 Airport Road<br>Mississauga, Ontario<br>Canada L4V 1S7 | | Contact: | Carol Nakagawa. | | Telephone: | (905) 672-2100. | | Date: | February 2004. | | Trade Names: | Cedara I-SoftView; Cedara I-ReadMammo | | Common Name: | PACS medical imaging software | | Classification Name: | Picture archiving and communications system. | | Cleared Devices: | Cedara Software Corp., Cedara I SoftView, 510(k) No. K022881. | | Device Description: | Cedara I-ReadMammo is medical image review workstation<br>software that is comprised of features that are previously cleared in<br>Cedara I-SoftView. The differences lie mainly in workflow and<br>user interface that make mammography image viewing more<br>convenient for the user. The product consists of features that allow<br>the qualified medical professional to view patient medical images<br>with the desired viewing protocol and workflow in order to<br>optimize the efficient use of their time. Additionally, as with all<br>Cedara I-SoftView offerings, measurements that are commonly<br>required for diagnosis and surgical planning are available to the<br>user. | Page 1 of 2 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a logo with the word "CEDARA" in bold, sans-serif font at the bottom. Above the text is a circular graphic made up of many horizontal lines. The lines are closely spaced together, creating a textured effect within the circle. #### Indications for Use: "Two and three dimensional image review, manipulation, analysis and therapy planning capabilities that support image management display needs in the medical environment from multiple locations within and outside the hospital. Productivity-Enhancing Second Console Workstations - Workstations designed to perform automated, routine tasks such as image review, printing and archiving as well as post processing capabilities that enable special services for referring physicians. Diagnostic Review Workstations - Workstations designed to assist radiologists and surgeons in conducting primary diagnostic review and planning through flexible and interactive manipulation of multi-modality softcopy images including the use of prosthetic template overlays, and including mammography. Physician's Review Workstations - Workstations designed to give easy and economic access to multi-modality softcopy images in multiple locations within and outside the hospital. (e.g. teleconferencing, teleradiology etc.) Lossy compressed mammographic images must not be used for primary diagnostic interpretation unless approved for use in digital mammography. Display monitors used for primary diagnostic interpretation of mammographic images must be approved for use in digital mammography." Comparison to Previously Cleared Device: The intended use and technological characteristics of Cedara I-ReadMammo Workstation software are substantially equivalent, in the opinion of Cedara Software Corp. to those of the previously cleared device and do not pose any new issues of safety and effectiveness. Page 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## APR 2 3 2004 Ms. Carol Nakagawa Manager of Regulatory Affairs Cedara Software Corp. 6509 Airport Road Mississauga Ontario, L4V 1S7 CANADA Re: K040468 Trade/Device Name: Cedara I-SoftView Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: April 6, 2004 Received: April 7, 2004 Dear Ms. Nakagawa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050 {3}------------------------------------------------ Page 2 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement | 510(k) Number | K040468 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Cedara I-SoftView | | Indications for Use | "Two and three dimensional image review, manipulation, analysis and<br>therapy planning capabilities that support image management display<br>needs in the medical environment from multiple locations within and<br>outside the hospital.<br><br>Productivity-Enhancing Second Console Workstations - Workstations<br>designed to perform automated, routine tasks such as image review,<br>printing and archiving as well as post processing capabilities that enable<br>special services for referring physicians.<br><br>Diagnostic Review Workstations - Workstations designed to assist<br>radiologists and surgeons in conducting primary diagnostic review and<br>planning through flexible and interactive manipulation of multi-modality<br>softcopy images including the use of prosthetic template overlays, and<br>including mammography.<br><br>Physician's Review Workstations - Workstations designed to give easy and<br>economic access to multi-modality softcopy images in multiple locations<br>within and outside the hospital. (e.g. teleconferencing, teleradiology etc.)<br><br>Lossy compressed mammographic images must not be used for primary<br>diagnostic interpretation unless approved for use in digital<br>mammography. Display monitors used for primary diagnostic<br>interpretation of mammographic images must be approved for use in<br>digital mammography." | #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED ._.___________________________________________________________________________________________________________________________________________________________________________ ... .. Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use (Per 21 CFR 801. 109) | <div style="display:inline-block;"></div> | |----------------------------------------|-------------------------------------------| |----------------------------------------|-------------------------------------------| OR | Over-The-Counter Use | | |----------------------|--| |----------------------|--| | (Division Sign-Off) | | |---------------------------------------------------------------|---------| | Division of Reproductive, Abdominal, and Radiological Devices | | | 510(k) Number | K040468 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...