K032555 · Totoku Electric Co., Ltd. · LLZ · Feb 26, 2004 · Radiology
Device Facts
Record ID
K032555
Device Name
3M COLOR MEDICAL FLAT PANEL DISPLAY, CCL316
Applicant
Totoku Electric Co., Ltd.
Product Code
LLZ · Radiology
Decision Date
Feb 26, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Intended Use
3M Color Medical Flat Panel Display, CCL316, is intended for use with Picture Archiving Communication Systems (PACS) for medical imaging application by physicians. It is not meant to be used for digital mammography.
Device Story
3M Color Medical Flat Panel Display, CCL316, functions as a visual display unit for medical images within a PACS environment. Operated by physicians in clinical settings to view diagnostic images; device receives video signals from a workstation; displays images for clinical interpretation and decision-making. Intended to assist in the review of medical imaging data; excludes digital mammography applications.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Color medical flat panel display; intended for PACS integration; 21 CFR 892.2050; Class II; Product Code 90 LLZ.
Indications for Use
Indicated for use by physicians for medical imaging applications within Picture Archiving Communication Systems (PACS). Not indicated for digital mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Related Devices
K073292 — MM/MMA/COLOR SERIES DIGITAL FLAT PANEL DISPLAY SYSTEM · Heeyoung Co., Ltd. · Jan 18, 2008
K032638 — DOME CX DIGITAL FLAT-PANEL DISPLAY SYSTEM MODEL, C3 COLOR AND C3 GRAY · Planar Systems, Inc. · Sep 11, 2003
K092613 — COLOR LCD MONITOR, MODEL RADIFORCE RS210 · Eizo Nanao Corporation · Mar 3, 2010
K073358 — LUMIMED SERIES 5MP DIGITAL FLAT PANEL DISPLAY SYSTEM, MODEL MM50/MM50A · Heeyoung Co., Ltd. · Apr 25, 2008
K042634 — WIDE 1MP LCD MONITOR SYSTEM · Wide Corporation · Oct 8, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight. The symbol is composed of three curved lines that create a sense of movement and dynamism.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## FEB 26 2004
Mr. Mikio Hasegawa General Manager TOTOKU ELECTRIC CO., LTD. MM Company, Design Group 300 Oya, Ueda, Nagano 386-0192 JAPAN
Re: K032555 Trade/Device Name: 3M Color Medical Flat Panel Display, CCL316 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: February 6, 2004 Received: February 5, 2004
Dear Mr. Hasegawa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May.28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brydon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{2}------------------------------------------------
510(k) Number: K032555
3M Color Medical Flat Panel Display, CCL316 Device Name:
Indications for Use:
3M Color Medical Flat Panel Display, CCL316, is intended for use with Picture Archiving Communication Systems (PACS) for medical imaging application by physicians. It is not meant to be used for digital mammography.
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NECESSARY)
## Concurrence of CDRH, Office of Device Evaluation
Prescription Use
✓
OR Over-The-Counter Use
(Optional Format 1-2-96)
David A. Lemon
(Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devic 510(k) Number _
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.