PIXI BONE DENSITOMETER
Device Facts
| Record ID | K970224 |
|---|---|
| Device Name | PIXI BONE DENSITOMETER |
| Applicant | Lunar Corp. |
| Product Code | KGI · Radiology |
| Decision Date | Apr 16, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1170 |
| Device Class | Class 2 |
Intended Use
The PIXI Bone provides an estimation of Bone Mineral Density (BMD in g/cm²) for the regions of the forearm and heel (os calcis).
Device Story
PIXI Bone Densitometer; estimates bone mineral density (BMD) in g/cm² for forearm and heel (os calcis). Device utilizes 5-second radiation exposure (20 mrem) to scan anatomical sites. System processes attenuation data to calculate BMD values. Used in clinical settings by trained operators. Output provides quantitative BMD measurements to healthcare providers for assessment of bone health. High correlation (r=0.998) with calcium hydroxyapatite standards; precision metrics (in vivo %CV 1.5% forearm, 1.97% os calcis) support clinical utility for bone density monitoring.
Clinical Evidence
Bench testing only. BMD estimations correlated with calcium hydroxyapatite pellets (r=0.998). In vitro short-term precision (%CV) was 0.68%. In vivo short-term precision (%CV) was 1.5% for forearm BMD and 1.97% for Os Calcis BMD. Results demonstrate comparability to predicate devices.
Technological Characteristics
Bone densitometer; utilizes ionizing radiation (20 mrem dose per 5-second exposure). Measures BMD in g/cm² via attenuation of radiation through bone tissue. Form factor designed for forearm and heel scanning. No specific materials, software architecture, or connectivity standards disclosed.
Indications for Use
Indicated for the estimation of bone mineral density (BMD) in the forearm and heel (os calcis) for patients requiring bone density assessment.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Related Devices
- K983724 — ORCA-DX BONE DENSITOMETER · Lunar Corp. · Dec 8, 1998
- K982267 — DPX-RX BONE DENSITOMETER · Lunar Corp. · Aug 18, 1998
- K033550 — DXL CALSCAN · Demetech AB · May 5, 2004
- K972882 — APOLLO · Norland Corp. · Oct 31, 1997
- K964562 — DTX-200 BONE DENSITOMETER · Osteometer Meditech A/S · Aug 22, 1997