K213938 · Ge Healthcare Japan Corporation · JAK · Feb 4, 2022 · Radiology
Device Facts
Record ID
K213938
Device Name
Revolution Ascend
Applicant
Ge Healthcare Japan Corporation
Product Code
JAK · Radiology
Decision Date
Feb 4, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Pediatric
Intended Use
The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results. The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Device Story
Revolution Ascend is a head and whole-body CT system comprising a gantry, patient table, and operator console. It acquires x-ray transmission data at various angles/planes (Axial, Cine, Helical, Cardiac, Gated) and uses computer reconstruction to generate cross-sectional images. The system includes Deep Learning Iterative Reconstruction (DLIR) for image processing. Operated by clinicians in a clinical setting, the device provides images for diagnosis, therapy planning, and monitoring. The subject device introduces an alternative detector scintillator material (Merc40H) compared to the predicate's Merc40L. The system remains compliant with established safety and performance standards. Output is viewed by healthcare providers to inform clinical decision-making, potentially benefiting patients through improved diagnostic accuracy and treatment planning.
Clinical Evidence
Bench testing only. No clinical data. Verification and validation included risk management, software testing, and image quality (IQ) performance evaluation using standard phantoms. IQ testing performed per IEC 61223-3-5 Ed. 2. Performance of the subject device with the new detector was compared to the predicate device. Re-substantiation of DLIR performance was also conducted.
Technological Characteristics
75 cm gantry bore; 64-row detector; Performix 40 Plus X-Ray Tube with liquid metal rotor bearing; JEDI60DC high voltage generator (72 kW); Merc40H detector with backlit diode technology and GE low noise ASIC; DLIR (Deep Learning Iterative Reconstruction); compliant with IEC 60601-1 Ed. 3.1, IEC 60601-1-2 Ed 4.0, IEC 60601-1-3 Ed 2.1, IEC 60601-2-28 Ed 3.0, IEC 60601-2-44 Ed. 3.2, IEC 61223-3-5 Ed. 2.0, NEMA XR-25, XR-26, XR-28.
Indications for Use
Indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications in patients of all ages for diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K203169 — Revolution Ascend · Ge Healthcare Japan Corporation · Nov 20, 2020
K233749 — Revolution Ascend Sliding · Ge Hangwei Medical Systems Co., Ltd. · Feb 22, 2024
K052855 — STARSPEED SERIES CT SCANNER SYSTEMS · GE Healthcare · Oct 28, 2005
K033502 — LIGHTSPEED 6.0 CT SCANNER SYSTEM · Ge Medical Systems, LLC · Nov 20, 2003
K132222 — AQUILION ONE VISION, V6.0 · Toshibamedical Systems Corporation · Nov 7, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it.
February 4, 2022
GE Healthcare Japan Corporation % He Haibo Regulatory Affairs Leader 7-127, 4-Chome, Asahigaoka, Hino, Tokyo 191-8503 JAPAN
Re: K213938
Trade/Device Name: Revolution Ascend Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: December 15, 2021 Received: December 16, 2021
Dear He Haibo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined within a blue circle. The circle is surrounded by white, stylized water droplets or flourishes, giving it a dynamic and fluid appearance.
K213938
#### Section 4: Indications for Use Statement
Revolution Ascend
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K213938
Device Name Revolution Ascend
#### Indications for Use (Describe)
The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.
The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white water droplet shapes around the letters. The logo is simple and recognizable.
#### K213938 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided: Date: December 16th, 2021 Submitter: GE Healthcare Japan Corporation 7-127, Asahigaoka, 4-chome Hino-shi, Tokyo, 191-8503, Japan Primary Contact: He Haibo Regulatory Affairs Leader Phone: +86-010-5708-3413 Email: haibo.he1@ge.com Secondary Contacts: Helen Peng Sr Regulatory Affairs Director Phone: 262-4248222 Email: hong.peng@ge.com Laura Turner Sr Regulatory Affairs Leader Phone: 262-200-1044 Email: laura.turner@ge.com Proposed Device/ Revolution Ascend Device Trade Name: Device Classification: Class II Regulation Number/ 21 CFR 892.1750 Computed Tomography X-ray System / JAK Product Code:
#### Predicate Device Information
Device Name: Revolution Ascend
{5}------------------------------------------------
Image /page/5/Picture/13 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. There are decorative swirls or flourishes around the circle, giving it a classic and recognizable appearance.
| Manufacturer: | GE Healthcare Japan Corporation |
|-------------------------------------|--------------------------------------------------------|
| 510(k) Number: | K203169, Cleared on November 20, 2020 |
| Device Classification: | Class II |
| Regulation Number/<br>Product Code: | 21 CFR 892.1750 Computed Tomography X-ray System / JAK |
### Device Description
The Revolution Ascend is a head and whole-body CT system composed of a gantry, patient table, operator console with a host computer, power distribution unit, and interconnecting cables. The system also includes image acquisition and reconstruction hardware/software, general system software, accompanying documents, and associated accessories/interconnections. The system has a 75 cm gantry bore and 64-row detector.
Revolution Ascend generates cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions modes.
A design change has been made to the Revolution Ascend with an alternative detector scintillator material prompting this premarket notification. While this change is being made, the design and manufacturing is such that the system performance remains identical to its unmodified predicate. The proposed device carries over all the features, options and specifications of the predicate device, including the Deep Learning Iterative Recon (DLIR) cleared via K212067 without change.
The proposed device 's Intended Use and Indications for Use remain identical to those of the unmodified predicate device. Revolution Ascend with the modified detector remains compliant with IEC 60601-1 Ed. 3.1 and associated collateral and particular standards, NEMA XR25, XR26, XR28, and 21 CFR Subchapter J performance standards.
#### Intended Use
The system is intended to be used for head, whole body Computed Tomography applications.
#### Indications for Use
The system is intended to produce cross-sectional images of the body by computer reconstruction of Xray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment support, comport, components and accessories.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The frame has three teardrop-shaped elements evenly spaced around the circle. The logo is presented in a blue color.
This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.
The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
#### Technology
Revolution Ascend employs the same basic operating principles and fundamental technologies as the predicate device.
The table below summarizes the substantive feature/technological differences between the predicate device and the proposed device:
| Subsystem | Revolution Ascend<br>(Predicate Device, K203169) | Revolution Ascend<br>(Proposed Device) |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Gantry | Revolution Ascend Gantry<br>- Bore size: 75cm<br>- Physical Tilt (± 30°) | Revolution Ascend Gantry<br>- Bore size: 75cm<br>- Physical Tilt (± 30°) |
| | Performix 40 Plus X-Ray Tube<br>- Supports 40 mm beamwidth<br>- Liquid Metal rotor bearing | Performix 40 Plus X-Ray Tube<br>- Supports 40 mm beam width<br>- Liquid Metal rotor bearing |
| | JEDI60DC High Voltage Generator<br>- Peak Power: 72 kW | JEDI60DC High Voltage Generator<br>- Peak Power: 72 kW |
| | NGX Collimator (75cm bore)<br>- 40 mm max z-coverage | NGX Collimator (75cm bore)<br>- 40 mm max z-coverage |
| | Merc40L Detector<br>- Backlit Diode technology<br>- Chiclet Module design<br>- GE low noise ASIC technology<br>used for signal conversion | Merc40H Detector<br>(with alternative material)<br>- Backlit Diode technology<br>- Chiclet Module design<br>- GE low noise ASIC technology<br>used for signal conversion |
| Operator<br>Console | NIO Console:<br>- Host computer, keyboard,<br>- scan control unit, two<br>monitors. | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are four small white teardrop shapes around the edge of the circle, evenly spaced. The logo is simple and recognizable, and it is often used to represent the company's brand.
## GE Healthcare
# 510(k) Premarket Notification Submission for Revolution Ascend
| Subsystem | Revolution Ascend<br>(Predicate Device, K203169) | Revolution Ascend<br>(Proposed Device) |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| Deep Learning<br>Image<br>Reconstruction<br>(DLIR) | DLIR cleared with Revolution Ascend<br>(K212067). | Same |
| Standards | IEC 60601-1 Ed. 3.1<br>IEC 60601-1-2 Ed 4.0<br>IEC 60601-1-3 Ed 2.1<br>IEC 60601-2-28 Ed 3.0<br>IEC 60601-2-44 Ed. 3.2<br>IEC 61223-3-5 Ed. 1.0<br>NEMA XR-25<br>NEMA XR-26<br>NEMA XR-28 | IEC 60601-1 Ed. 3.1<br>IEC 60601-1-2 Ed 4.0<br>IEC 60601-1-3 Ed 2.1<br>IEC 60601-2-28 Ed 3.0<br>IEC 60601-2-44 Ed. 3.2<br>IEC 61223-3-5 Ed. 2.0<br>NEMA XR-25<br>NEMA XR-26<br>NEMA XR-28 |
The changes described above do not change the fundamental control mechanism, operating principle, energy type, and do not change the intended use from the predicate device Revolution Ascend.
#### Determination of Substantial Equivalence
The Revolution Ascend has completed testing and in compliance with AAMI/ANSI ES 60601-1 and IEC60601-1 Ed. 3.1 and its associated collateral and particular standards, 21 CFR Subchapter J, and NEMA standards XR 25, XR 26, and XR 28. The device has successfully completed engineering design V &V and bench testing in support of substantial equivalence between the subject device and predicate device. The following quality assurance measures were applied to the development of the system:
- Risk Analysis -
- -Required Reviews
- -Design Reviews
- -Testing on unit level (Module verification)
- -Integration testing (System verification)
- -Performance testing (Verification)
- -Safety testing (Verification)
- -Simulated use testing (Validation)
GE believes the Revolution Ascend CT system is of comparable type and substantially equivalent to our currently marketed system Revolution Ascend (K203169).
The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are in a stylized font, and there are four white swirls around the letters. The logo is simple and recognizable, and it is often used to represent the company.
#### Non-Clinical Testing
The verification and validation testing have been successfully completed as required by design control procedures under GE Healthcare's quality system. This includes risk management, software verification and validation testing as well as image quality performance evaluation using well established metrics and methods. IQ evaluation include:
- -General IQ Performance testing in accordance with IEC 61223-3-5 Ed. 2 to demonstrate the overall system performance in a standardized and referenceable manner.
- -Comparable IQ performance test using standard IQ, QA phantoms for typical conditions to demonstrate image quality equivalence of Revolution Ascend with Merc40H and predicate Revolution Ascend with Merc40L.
- -Re-substantiation of the imaging performance associated with the cleared DLIR(K212067) on the subject device Revolution Ascend.
Non-clinical bench test results demonstrated the subject device performs equivalently to the predicate device.
#### Substantial Equivalence Conclusion
Based on the conformance to standards, development under our quality system, and the engineering testing provided, GE Healthcare believes that the Revolution Ascend is as safe and effective, and performs in a substantially equivalent manner to the unmodified predicate device Revolution Ascend (K203169).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.