K163711 · Philips Medical Systems (Cleveland), Inc. · JAK · Apr 5, 2017 · Radiology
Device Facts
Record ID
K163711
Device Name
IQon Spectral CT
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
JAK · Radiology
Decision Date
Apr 5, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Intended Use
The IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. The IQon Spectral CT system acquires one CT dataset - composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number. This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures. The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
Device Story
IQon Spectral CT is a whole-body CT system; acquires dual-energy x-ray transmission data via rotating gantry and detector array; transforms data into cross-sectional images and material-specific information (attenuation, density, atomic number). Used in clinical settings by trained healthcare professionals; provides diagnostic visualization and analysis of anatomical/pathological structures. Includes low-dose lung cancer screening capability. System comprises scanner (gantry, patient couch, console), Spectral Reconstruction System, and Spectral CT Viewer. Output enables clinicians to assess chemical composition of tissues/contrast agents; aids in early detection of lung nodules. Modifications from predicate include minor spectral algorithm improvements (Constant Noise Suppression, Pre Decomposition DeNoise, Iterative Bone Correction) and updated labeling for lung screening.
Clinical Evidence
No clinical data required. Substantial equivalence demonstrated via non-clinical performance testing, comparative assessment of image quality, and technological characteristics against predicate and reference devices.
Technological Characteristics
Whole-body CT system; rotating gantry with x-ray tube and detector array. Spectral imaging via dual-energy acquisition. Software-based reconstruction and spectral analysis. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-44, ISO 14971. Cybersecurity and software documentation per FDA guidance.
Indications for Use
Indicated for use as a whole-body CT X-ray system for cross-sectional imaging, material analysis (attenuation, density, effective atomic number), and low-dose lung cancer screening for early detection of lung nodules in patients meeting established screening program/protocol criteria.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K203020 — Spectral CT · Philips Medical Systems Nederland B.V. · Feb 26, 2021
K212875 — Spectral CT on Rails · Philips Medical Systems Nederland B.V. · May 13, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines extending from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 5, 2017
Philips Medical Systems (Cleveland), Inc. % Mr. Steven Goldberg Regulatory Affairs Specialist 595 Miner Road CLEVELAND OH 44143
Re: K163711
Trade/Device Name: IQon Spectral CT Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: March 15, 2017 Received: March 17, 2017
Dear Mr. Goldberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K163711
Device Name IQon Spectral CT
#### Indications for Use (Describe)
The IQon Spectral CT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories. The IQon Spectral CT system acquires one CT dataset - composed of data from a higher-energy detected x-ray spectrum
and a lower- energy detected x-ray spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.
The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.
The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> |
|----------------------------------------------|-------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and the word is centered. The background is white, which makes the blue letters stand out.
## Section 5 510(k) Summary
{4}------------------------------------------------
### PHIL 1 PS
#### 510(k) Summary – K163711 Section 5:
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | March 14, 2017 | |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| Manufacturer: | Philips Medical Systems (Cleveland), Inc.<br>595 Miner Road<br>Cleveland, OH 44143, USA<br>Establishment Registration Number: 1525965 | |
| Primary Contact<br>Person: | Steven Goldberg<br>Regulatory Affairs Specialist<br>Phone: 972-(0)72-336-4218<br>E-mail: steven.goldberg@Philips.com | |
| Secondary<br>Contact Person: | Michael S. Preto<br>Senior Manager, Regulatory Affairs<br>Phone: 440-483-2295<br>Fax: 440-483-4918<br>E-mail: Michael.preto@Philips.com | |
| Device: | | |
| | Trade Name: | IQon Spectral CT |
| | Common name | Computed tomography x-ray system |
| | Classification Name: | Computed tomography x-ray system |
| | Classification<br>Regulation: | 21CFR 892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | II |
| | Primary Product<br>Code: | JAK |
| | Secondary Product<br>Code: | Not applicable |
| Primary Predicate<br>Device: | | |
| | Trade Name: | Philips IQon Spectral CT System |
| | Manufacturer: | Philips Medical Systems (Cleveland),<br>Inc. |
| | 510(k) Clearance: | K133674 (November 21, 2014) |
| | Classification Name: | Computed tomography x-ray system |
| | Classification<br>Regulation: | 21CFR §892.1750 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | JAK |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and aligned horizontally. The background is plain white, which makes the blue letters stand out.
| Reference Device: Trade Name: | Philips Multislice CT System with Low Dose CT Lung Cancer Screening |
|-------------------------------|---------------------------------------------------------------------|
| Manufacturer: | Philips |
| 510(k) Clearance: | K153444 (April 8, 2016) |
Device The IQon Spectral CT is a whole-body computed tomography Description: (CT) X-Ray System featuring a continuously rotating x-ray tube and detectors gantry and multi-slice capability. The acquired xray transmission data is reconstructed by computer into crosssectional images of the body taken at different angles and planes. This device also includes signal analysis and display equipment; patient and equipment supports; components; and accessories. The IQon Spectral CT includes the detector array previously described in K133674 "Philips IQon Spectral CT".
> The IQon Spectral CT consists of three main components - a scanner system that includes a rotating gantry, a movable patient couch, and an operator console for control and image reconstruction; a Spectral Reconstruction System; and a Spectral CT Viewer. On the gantry, the main active components are the x-ray high voltage (HV) power supply, the x-ray tube, and the detection system.
> In addition to the above components and the software operating them, the system includes workstation hardware and software for data acquisition; and image display, manipulation, storage, and filming, as well as post-processing into views other than the original axial images. Patient supports (positioning aids) are used to position the patient.
### Device Modifications
This 510(k) addresses the minor modifications that were implemented in the currently marketed and predicate device, Philips IQon Spectral CT, cleared by FDA via 510(k) number K133674, Nov. 21, 2014. These minor changes include clarifications, materials, components, defect labeling fixes/improvements, and minor improvements to some spectral algorithms, e.g., Constant Noise Suppression, Pre Decomposition DeNoise, parameters tuning, and Iterative Bone Correction that were implemented since clearance of the currently marketed and predicate device. Philips IQon Spectral CT and do not change the indications for use of the device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
This 510k also adds Low Dose CT Lung Cancer Screening in the Indications for Use, cleared by FDA for the reference device, Philips Multislice CT System with Low Dose CT Lung Cancer Screening via K153444, April 8, 2016.
The IQon Spectral CT is a Computed Tomography X-Ray Indications for Use: System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. This device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
> The IQon Spectral CT system acquires one CT dataset composed of data from a higher-energy detected x-ray spectrum and a lower- energy detected x-ray spectrum. The two spectra may be used to analyze the differences in the energy dependence of the attenuation coefficient of different materials. This allows for the generation of images at energies selected from the available spectrum and to provide information about the chemical composition of the body materials and/or contrast agents. Additionally, materials analysis provides for the quantification and graphical display of attenuation, material density, and effective atomic number.
> This information may be used by a trained healthcare professional as a diagnostic tool for the visualization and analysis of anatomical and pathological structures.
> The system is also intended to be used for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*.
> The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
> *Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
Technological The main components (detection system, the reconstruction Characteristics algorithm, and the x-ray system) that are used in the proposed IQon Spectral CT System have the same fundamental design characteristics and are based on the same technologies as the
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
currently marketed and predicate device, Philips IQon Spectral CT.
The design, indications for use, and the fundamental scientific technology provided with the proposed IQon Spectral CT are identical to the currently marketed and predicate device, Philips IQon Spectral CT.
Summary of Non-Non-clinical performance testing has been performed on the Clinical IQon Spectral CT System and demonstrates compliance with Performance Data: the following international and FDA-recognized consensus standards:
- International and FDA-recognized consensus ● standards:
- o IEC 60601-1:2005+A1:2012 (FDA reorganization number :19-4)
- IEC 60601-1-2:2007 (FDA reorganization O number :19-1)
- IEC 60601-1-3 Edition 2.0:2008 (FDA o reorganization number :12-210)
- IEC 60601-1-6:2010 (FDA reorganization O number :5-85)
- IEC 60601-2-44:2009 (FDA reorganization O number :12-257)
- ISO 14971 Application of risk management to o medical devices (2007) (FDA/CDRH reorganization number :5-40)
- Device specific guidance documents: Guidance for . Industry and FDA Staff – Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005, document number 337)
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (issued October 2, 2014).
Non-Clinical verification and or validation tests have been performed with regards to the intended use, the technical claims, the requirement specifications and the risk management results. The addition of lung cancer screening to the IFU statement was supported with a comparative assessment of image quality and technological characteristics to the devices cleared for this use in K153444.
Clinical verification and or validation test results Non demonstrate that the IQon Spectral CT System:
{8}------------------------------------------------
# PHILIPS
| • | Complies with the aforementioned international and<br>FDA-recognized consensus standards, Guidance for<br>Industry and FDA Staff – Guidance for the Content of<br>Premarket Submissions for Software Contained in<br>Medical Devices" (issued May 11, 2005, document<br>number 337), and Content of Premarket Submissions<br>for Management of Cybersecurity in Medical Devices<br>(issued October 2, 2014). |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • | Meets the acceptance criteria and is adequate for its<br>intended use. |
| | Therefore, the IQon Spectral CT System is substantially<br>equivalent to the currently marketed and predicate device,<br>Philips IQon Spectral CT (K133674, Nov. 21, 2014) in terms of<br>safety and effectiveness. |
| Summary of<br>Clinical<br>Performance Data: | The IQon Spectral CT System did not require a clinical study<br>because substantial equivalence to the currently marketed and<br>predicate device, Philips IQon Spectral CT, was demonstrated<br>with the following attributes:<br>• Indication for use;<br>• Technological characteristics;<br>• Non-clinical performance testing;<br>• Labeling; and<br>• Safety and effectiveness. |
| Substantial<br>Equivalence<br>Summary: | The proposed IQon Spectral CT System is substantially<br>equivalent to the currently marketed and predicate device,<br>Philips IQon Spectral CT (K133674), in terms of design<br>features, fundamental scientific technology, indications for<br>use, and safety and effectiveness. Also, the Low Dose CT<br>Lung Cancer Screening provided with the IQon Spectral CT<br>System is substantially equivalent to the reference device,<br>Philips Multislice CT System with Low Dose CT Lung Cancer<br>Screening (K153444, April 8, 2016). Additionally, substantial<br>equivalence was demonstrated with non-clinical performance<br>(verification and validation) tests, which complied with the<br>requirements specified in the aforementioned international and<br>FDA-recognized consensus standards. The results of these<br>tests demonstrate that the IQon Spectral CT System met the<br>acceptance criteria and is adequate for its intended use. |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.