DR 800 with MUSICA Dynamic

K180589 · Agfa Healthcare N.V. · JAA · Apr 5, 2018 · Radiology

Device Facts

Record IDK180589
Device NameDR 800 with MUSICA Dynamic
ApplicantAgfa Healthcare N.V.
Product CodeJAA · Radiology
Decision DateApr 5, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesPediatric

Intended Use

The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: - Positioning fluoroscopy procedures - Gastro-intestinal examinations - Urogenital tract examinations - Angiography It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for manimography applications.

Device Story

DR 800 is a floor-mounted R/F X-ray system capturing digital images via flat-panel detectors. System includes tube, operator console, motorized tilting patient table, bucky, and optional wall stand/ceiling suspension. Operated by clinicians in clinical settings. Uses NX workstation with MUSICA Dynamic image processing to transform raw detector signals into diagnostic images. Output viewed by radiologists/physicians for clinical decision-making in fluoroscopy and radiography. Benefits include wide dynamic range, digital image capture replacing traditional image intensifiers, and optimized processing for various patient sizes/exams. Pediatric-specific features include grid-less FLFS, specific AEC values, and age-based protocol settings.

Clinical Evidence

Bench testing only. No clinical or animal trials performed. Validation included usability/functionality evaluations by radiographers, Full Leg Full Spine (FLFS) clinical validation, dose control validation using adult/pediatric phantoms, and image quality validation using anthropomorphic phantoms. Results confirmed performance met acceptance criteria, with image quality rated between good and excellent.

Technological Characteristics

Floor-mounted R/F X-ray system. Detectors: GOS or CsI scintillator flat-panel (17x17 in, 148 µm pixel size). Energy: 50-60 Hz, 100-240V. Connectivity: DICOM. Software: NX workstation with MUSICA Dynamic image processing. Sterilization: N/A. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-54, 21 CFR 1020.30-1020.32.

Indications for Use

Indicated for dynamic imaging (fluoroscopy/rapid sequence) and general radiography of all body parts in adult and pediatric patients. Procedures include positioning fluoroscopy, gastro-intestinal, urogenital, and angiography. Not indicated for mammography.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. Agfa HealthCare N.V. % Ms. Shaeann Cavanagh Regulatory Affairs Manager Agfa HealthCare Corporation 10 South Academy Street GREENVILLE. SC 29601 April 5, 2018 Re: K180589 Trade/Device Name: DR 800 with MUSICA Dynamic TM Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: JAA Dated: March 5, 2018 Received: March 6, 2018 Dear Ms. Cavanagh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) ### K180589 Device Name DR 800 with MUSICA Dynamic Indications for Use (Describe) The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: - · Positioning fluoroscopy procedures - · Gastro-intestinal examinations - · Urogenital tract examinations - · Angiography It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for manimography applications. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4 The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and mantain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PPAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(K) SUMMARY ### Agfa HealthCare DR 800 with MUSICA Dynamic #### I. SUBMITTER Agfa HealthCare N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Wim Govaerts, Prepared: March 5, 2018 Telephone: + 32 3444 6246 #### DEVICE II. Name of Device: DR 800 with MUSICA DynamicTM Common Name: System, X-Ray, Fluoroscopic, Image-Intensified Classification Name: Image-Intensified Fluoroscopic X-ray System Regulatory Classification: Class II, 21 CFR 892.1650 Product Code: JAA #### III. PREDICATE DEVICE This is a 510(k) for Agfa's DR 800 with MUSICA Dynamic, which is an image-intensified fluoroscopic x-ray system. It is substantially equivalent to General Medical Merate S.P.A.'s OPERA Swing (K140380). Name of Device: OPERA Swing Common Name: System, X-Ray, Fluoroscopic, Image-Intensified Classification Name: Image-Intensified Fluoroscopic X-ray System Regulatory Classification: Class II, 21 CFR 892.1650 Product Code: JAA This predicate device has not been subject to a design-related recall. #### IV. DEVICE DESCRIPTION Agfa HealthCare's DR 800 is an image-intensified fluoroscopic x-ray system (product code JAA) intended to capture images of the human body. The DR 800 is a floor-mounted R/F system that consists of a tube and operator console with a motorized tilting patient table and bucky with optional wall stand, FLFS overlay and ceiling suspension. The new device uses Agfa's NX workstation with MUSICA Dynamic™ image processing and flat-panel detectors for digital and wide dynamic range image capture. It is capable of replacing other direct radiography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes. {4}------------------------------------------------ This submission is to add the DR 800 with MUSICA Dynamic software to Agfa's radiography portfolio. The optional image processing allows users to conveniently select image processing settings for different patient sizes and examinations. The image processing algorithms in the new device are similar to those previously cleared and used in Aefa's radiography portfolio today which includes the DR 600 (K152639) and DR 400 (K141192). The addition of the dynamic image processing is identical to the predicate device (K140380). Principles of operation and technological characteristics of the new and predicate device are the same. The new device is virtually identical to predicate K140380 with the exception that it has Agfa's MUSICA Dynamic software instead of the predicate's ATS software. It uses the same flat panel detectors to capture and digitize the image. Differences in devices do not alter the intended diagnostic effect. Laboratory data and image quality evaluations conducted with independent radiologists confirm that performance is equivalent to the predicates. Configuration information for the flat-panel detectors can be found in the DR 14s (K161368) and DR 800. User Manuals. The DR 14s and RFFL4343 detectors can be integrated in an X-ray system that communicates to a workstation. The Service Manual details the possible configurations and integrations with the NX workstation and X-ray generator. All of Agfa HealthCare's DR X-ray systems (i.e. DX-D 100-K103597, DX-D 300-K103050, DX-D 600-K112670, DR 400-K141192, DR 600-K152639) will integrate with the detectors. The NX4.0 Service Manual, Chapter 4 and associated appendices addresses the installation and configuration with other system components. #### INDICATIONS FOR USE V. The DR 800 system is indicated for performing dynamic imaging examinations (fluoroscopy and/or rapid sequence) of the following anatomies/procedures: - Positioning fluoroscopy procedures - Gastro-intestinal examinations - Urogenital tract examinations - Angiography ● It is intended to replace fluoroscopic images obtained through image intensifier technology. In addition, the system is intended for project radiography of all body parts. The DR 800 is not intended for mammography applications. NOTE: The mammography applications embedded in the MUSICA software are for previously cleared CR imaging applications (K081963) and not intended for DR or fluoroscopic imaging. Furthermore, the additional mammography software is only available through additional license keys that must be purchased. These licenses are only available outside of the USA. {5}------------------------------------------------ #### VI. PEDIATRIC USE SUMMARY The DR 800 is intended for general populations, including adult and pediatric patients. Specific design features of the DR 800 for pediatric use include but not limited to using the FLFS application without the grid, specific AEC values, and NX protocol settings per pediatric age range. #### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICES Agfa HealthCare's DR 800 with MUSICA Dynamic and General Medical Merate S.P.A.'s OPERA Swing predicate device (K140380) are image-intensified fluoroscopic x-ray imaging devices. Product Code JAA. Agfa's DR 800 is substantially equivalent to the predicate device (K140380) in that it uses precisely the same technology to capture and transmit images. The DR 800 is a floormounted R/F system that consists of a tube and operator console with a motorized tilting patient table and bucky with optional wall stand, FLFS overlay and ceiling suspension. The new device uses Agfa's NX workstation with MUSICA Dynamic™ image processing and flat-panel detectors for digital and wide dynamic range image capture. It is capable of replacing other direct radiography, image intensifying tubes and TV cameras, including computed radiography systems with conventional or phosphorous film cassettes. Principles of operation and technological characteristics of the new and predicate device are the same. The new device is virtually identical to predicate K140380 with the exception that it has Agfa's MUSICA Dynamic software instead of the predicate's ATS software. It uses the same flat panel detectors to capture and digitize the image. Differences in devices do not alter the intended diagnostic effect. The optional image processing allows users to conveniently select image processing settings for different patient sizes and examinations. The image processing algorithms in the new device are similar to those previously cleared and used in Agfa's radiography portfolio today which includes the DR 600 (K152639) and DR 400 (K141192). The addition of the dynamic image processing is identical to the predicate device (K140380). Agfa's DR 800 has an Indications For Use statement virtually identical to the predicate device (K140380). Intended uses are the same. The devices have technological characteristics. The DR 800 indications for use is equivalent to the predicate (K140380) because both include the delineation of anatomical areas and imaging applications. The DR 800 includes the statement that the device is not indicated for mammography; however, the predicate device (K140380) Indications For Use statement does not indicate mammography in the identified list of imaging applications. Descriptive characteristics and performance data including image quality evaluations by qualified radiologists are adequate to ensure equivalence. Differences in devices do not alter the intended therapeutic/diagnostic effect. Table 1 on the next page summarizes the similarities and differences between the new device and predicate. {6}------------------------------------------------ | | DR 800 (New Device)<br>Agfa HealthCare | Opera Swing (PREDICATE-K140380)<br>General Medical Merate | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Communications | Same as predicate | DICOM | | Flat Panel Detectors | Same as predicate | RF FL4343, Pixium 2430 (DR10s),<br>Pixium 3543 (DR14s) | | Detector Material | Same as predicate | Gadolinium Oxysulfide (GOS) or<br>Cesium Iodide (CsI) scintillator | | Detector Sizes | Same as predicate | 17x17 in. | | Field Sizes | Same as predicate | 43 x 43 cm<br>30 x 30 cm<br>20 x 20 cm<br>15 x 15 cm | | Pixel Size | Same as predicate | 148 µm | | Dynamic Range | Same as predicate | 16 bit | | Matrix Size | Same as predicate | Static - 2880 x 2880<br>Dynamic - 1024 x 1024 | | Power Supply | Same as predicate | 50-60 Hz<br>100-240V auto ranging | | Operator Workstation | Agfa NX | Diagnostic Workstation | | Image processing | MUSICA Dynamic,<br>MUSICA2<br>MUSICA3/3+ | ATS Software | | Tabletop Features | Tilt +/- 90°<br>240 x 80 cm<br>265 kg maximum weight | Tilt +/- 90°<br>240 x 65 cm<br>227 kg maximum weight | | Generators | Same as predicate | Choice of three models:<br>50-80 KW | | Operating System | Same as predicate | Windows 7, 8, 8.1, 10 | | Display System | Separately cleared medical display (K051901) | Separately cleared medical display | | Indications for Use<br>Statements | The DR 800 systems is indicated for performing<br>dynamic imaging examinations (fluoroscopy and/or<br>rapid sequence) of the following<br>anatomies/procedures: Positioning fluoroscopy<br>procedures, Gastro-intestinal examinations,<br>Urogenital tract examinations, and Angiography. It<br>is intended to replace fluoroscopic images obtained<br>through intensifier technology. In addition, the<br>system is intended for project radiography of all<br>body parts. The DR 800 is not intended for<br>mammography applications. | The Opera Swing is indicated for<br>performing general radiography,<br>fluoroscopy and angiography<br>procedures/applications. The device is<br>intended for use in: Skeleton, Chest and<br>lungs, Pediatrics,<br>Emergency/traumatology,<br>Gastroenterology, Urology and<br>gynecology, Linear tomography, Digital<br>angiography and Stitching. | Table 1: Device Comparison Table {7}------------------------------------------------ #### VIII. PERFORMANCE DATA Laboratory testing and software testing (for a moderate level of concern device) using equivalent test protocols were evaluated by qualified individuals employed by the sponsor and independent radiologists to demonstrate that adequate design controls (according to 21 CFR 820.30) were in place. Verification and validation testing confirmed the device meets performance, safety, usability and security requirements. Pediatric indications were also taken into account. Results were verified and validated. No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device. No patient treatment was provided or withheld. ### Bench Testing Clinical image quality evaluations, performance/functionality and usability data, FLFS clinical validation and dose control evaluation data has been provided. No patient treatment was provided or withheld. No clinical or animal testing was performed in the development of the DR 800. - o Usability and functionality evaluations were conducted with qualified independent radiographers and internal experts. The results of these tests fell within the acceptance criteria for the DR 800 R/F X-ray system and some improvements will be implemented based on these results; the DR 800 supports a radiographic and fluoroscopic workflow including dynamic and static imaging, continuous and rapid sequence exams, calibration, and positioning. - Full Leg Full Spine (FLFS) clinical validation for the DR 800 R/F X-Ray system was ● conducted with a qualified internal radiographer. The FLFS clinical validation for the mount stitch grid, imaging ranges of a certain tolerance and transversal collimation, and medical ruler exposure fulfilled the acceptance criteria and passed the assessment with minor fails that will be solved. A comparison test and validation of FLFS landscape was conducted for NX Luna to compare the FLFS software with the current FLFS software already on the market. This comparison test was conducted with several qualified internal experts. The results of the FLFS comparison test for NX Luna concluded that the FLFS software is equal to or better than the current FLFS software currently on the market. The results of the FLFS landscape validation for the NX Luna concluded that the FLFS landscape functional design meets to user needs. - Dose control validation was conducted with a qualified internal expert to evaluate the level of dose control in adult and pediatric phantoms with pulsed and continuous fluoroscopy exams. The results fulfilled the acceptance criteria that none of the detector doses would measure higher than the DIN-norm or exceed the dose limit curve. - Image Quality Validation testing was conducted using anthropomorphic phantoms and evaluated by qualified independent radiographers and internal experts. The image quality validation included testing a full range of applications for the DR 800 R/F X-ray system with MUSICA Dynamic, MUSICA3 Abdomen+ and the R/F flat-panel detector compared to reference images using anonymized phantoms. The test results indicated that the pulsed {8}------------------------------------------------ and continuous fluoroscopy imaging of the DR 800 R/F X-ray system with MUSICA Dynamic was between good and excellent and passed the acceptance criteria. The test results showed MUSICA3 Abdomen+ images were suitable for diagnosis with an overall higher image quality. The test results proved clinical acceptability for static images made with the R/F flat-panel detector (FL4343). ### Software Verification and Validation Testing Verification and validation plans comprise of test protocols. The complete device has been certified and validated. During the final risk analysis meeting, the risk management team concluded that the medical risk is no greater than with conventional x-ray film previously released to the field. For the NX4.0 (NX Luna) there are a total of 274 risks in the broadly acceptable region and 26 risks in the ALARP region with only three of these risks identified. Zero risks were identified in the Not Acceptable Region. Therefore, the device is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk. The software risk assessed on solution level for the DR 800 and also includes separate risk assessments for the NX4.0 software and XRDI. The term "Level of Concern" means the level of risk that the software device is determined to be if the software were to fail. The Level of Concern for the DR 800 and NX4.0 has been determined to be moderate. ### Electrical Safety and Electromagnetic Compatibility (EMC) Testing: - IEC 60601-1: 2012 Medical Electrical Equipment: General Requirements for Safety and Essential ● Performance. - IEC 60601-1-2: 2007 Medical Electrical Equipment - Part 1-2: General Requirements for Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests. - IEC 60601-1-3: 2008 Medical Electrical Equipment Part 1-3: General Requirements for Safety and . Essential Performance - Collateral Standard: Radiation Protection in Diagnostic X-Ray Equipment - IEC 60601-2-54: 2009 Medical Electrical Equipment -- Part 2-54: Particular Requirements for the ● Basic Safety and Essential Performance of X-Ray Equipment for Radiography and Radioscopy. The DR 800 with MUSICA Dynamic is compliant to the FDA Subchapter J mandated performance standard 21 CFR 1020.30 – 1020.32. Agfa's in-house standard operating procedures were also used for the development of the device and software; these procedures conform to the following standards: - ISO 13485:2003 Medical Devices Quality Management Systems 0 - ISO 14971:2012 Application of Risk Management to Medical Devices - IEC 62304:2006 Medical Device Software Software life cycle processes ● - ISO 62366:2007 Medical Devices Application of Usability Engineering ● - ACR/NEMA PS3.1-3.20: 2011 Digital Imaging and Communications in Medicine (DICOM) ● {9}------------------------------------------------ ### Guidance Documents Agfa utilized the following guidance documents in the development of the DR 800 with MUSICA Dynamic™ - Guidance for the Content of Premarket Submissions for Software Contained in Medical ● Devices (May 2005) - Cybersecurity for Networked Medical Devices Containing Off-The-Shelf (OTS) Software (January 2005) - Off-the-Shelf Software Use in Medical Devices (September 1999) - Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (October 2014) - Guidance for Pediatric Information for X-ray Imaging Device Premarket Notifications ● (November 2017) ### Summary Based on the performance data as documented in the above testing, the DR 800 is found to have a safety and effectiveness profile that is similar to the predicate device. #### IX. CONCLUSIONS Agfa's DR 800 has indications for use that is consistent with those of the legally marketed predicate device (K140380). Intended uses are the same. Where technological characteristics differ lab tests concluded that the device is substantially equivalent to the predicate in that it does not alter the intended therapeutic/diagnostic effect. The new device and the OPERA Swing predicate device (K140380) are image-intensified fluoroscopic x-ray systems Product Code JAA. Agfa's DR 800 is substantially equivalent to the predicate device (K140380) in that it uses precisely the same technology to capture and transmit images. This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...