K072591 · Paieon, Inc. · IZI · Oct 10, 2007 · Radiology
Device Facts
Record ID
K072591
Device Name
CARDIOP-B SYSTEM, VERSION 2.1
Applicant
Paieon, Inc.
Product Code
IZI · Radiology
Decision Date
Oct 10, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1600
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The CardiOp-B System is a software system that assists in the evaluation of coronary lesions by enabling the creation of 3D images of coronary vessel segments based on two to three 2D angiography images obtained from single plane angiography. CardiOp-B provides quantitative information regarding the calculated dimensions of arterial segments based on the 3D image. CardiOp-B is intended for use in real-time in the catheterization lab and off-line for post-procedural analysis. It is intended for use by clinicians, technicians and research personnel.
Device Story
CardiOp-B is an add-on software system for conventional X-ray angiography systems; used in catheterization labs or off-line for post-procedural analysis. Input: two to three 2D angiography images from single-plane angiography. Processing: 3D reconstruction of coronary vessel segments; quantitative analysis of stenosed vessels, including bifurcation and side-branch analysis. Output: 3D images and quantitative cross-section measurements of arterial segments. Clinicians use output to evaluate coronary lesions and inform clinical decision-making regarding vessel stenosis. Benefits include improved visualization and objective dimensional data for cardiac assessment.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including software validation and phantom performance evaluation to verify the accuracy of 3D reconstruction and quantitative measurements for the modified features.
Technological Characteristics
Software-based 3D vessel analysis system; add-on to conventional X-ray angiography systems. Utilizes 3D reconstruction algorithms to process 2D angiography inputs. Operates in real-time or off-line environments.
Indications for Use
Indicated for clinicians, technicians, and research personnel to assist in the evaluation of coronary lesions by creating 3D images and providing quantitative dimensional analysis of coronary vessel segments using 2-3 2D single-plane angiography images.
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K07259/
Image /page/0/Picture/1 description: The image shows the word "Paieon" in a stylized font. Below the word "Paieon" is the phrase "Realizing Cardiac Imaging". The text is black and the background is white.
510(k) SUMMARY
### 510(k) SUMMARY—CardiOp-B System
OCT 』0 2007
| Submitter Name: | Paieon Inc. | | |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Submitter Address: | 747 Third Ave., 4th floor New York, NY 10017-2803 | | |
| Contact Person: | Ravit Barkama, MD | | |
| Phone Number: | +972 3 915 0000 | | |
| Fax Number: | +972 3 901 2324 | | |
| Date Prepared: | Sept 12, 2007 | | |
| Device Trade Name: | The CardiOp-B System | | |
| Device Common Name: | 3D Vessel Analysis System | | |
| Classification Name: | 3D Vessel Analysis System | | |
| Predicate Devices: | CardiOp-B cleared for marketing under K030139, QCA-CMS cleared<br>under K993763 and CAAS QCA 3D cleared under K063344 | | |
| Device Description: | The CardiOp-B System is an image acquisition and processing software<br>system designed as an add-on to conventional X-ray angiography<br>systems. The CardiOp-B system improves the output of coronary<br>angiography by presenting a three-dimensional reconstruction of the<br>stenosed vessel as well as quantitative cross-section information. The<br>modifications to the CardiOp-B System which are the subject of this<br>Special 510(K) include:<br>Bifurcation and side-branch analysis | | |
{1}------------------------------------------------
Realizing Cardiac Imaging
### 510(k) SUMMARY
| Intended Use: | The CardiOp-B System is a software system that assists in the<br>evaluation of coronary lesions by enabling the creation of 3D images of<br>coronary vessel segments based on two to three 2D angiography<br>images obtained from single plane angiography. CardiOp-B provides<br>quantitative information regarding the calculated dimensions of arterial<br>segments based on the 3D image. CardiOp-B is intended for use in<br>real-time in the catheterization lab and off-line for post-procedural<br>analysis. It is intended for use by clinicians, technicians and research<br>personnel |
|----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Characteristics<br>Compared to Predicate Device: | The technological characteristics, e.g., overall design, principle of<br>action, mode of operation, performance characteristics, etc., and the<br>intended use of the Cardio-Op B system are substantially equivalent to<br>the predicate devices cited above. |
| Performance Data: | Applicable performance testing was performed to evaluate the<br>modifications to the CardiOp-B system. Testing included software<br>validation and phantom performance evaluation. The test results were<br>found to be acceptable as required by the respective test plans and<br>protocols, demonstrating that the modified device performs according<br>to its specifications. The test results did not raise new safety or<br>effectiveness issues. |
| Conclusion: | The testing reported in this 510(K) establishes the modified CardiOp-B<br>is substantially equivalent to the predicate devices and is safe and<br>effective for its intended use. |
.
:
: : : : :
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular logo with the words "Department of Health & Human Services USA" written around the perimeter. Inside the circle is a stylized symbol consisting of three curved lines that resemble a bird in flight or a stylized human figure. The logo is black and white and appears to be of low resolution.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
OCT 1 0 2007
Paieon, Inc. % Ravit Barkama, M.D. Medical Director Paieon Medical Ltd. 23 Hamelacha St., P.O.B 11355 Rosh Haavin 48091 ISRAEL
Re: K072591
Trade/Device Name: The CardiOp-B System Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: II Product Code: IZI
Dated: September 12, 2007 Received: September 14, 2007
#### Dear Dr. Barkama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows a circular emblem with the text "1906-2006" at the top. Below the text is a large, stylized "PA" and the word "Centennial" underneath. There are three stars at the bottom of the emblem. The emblem is surrounded by a dotted border.
Promoting and Promoting Public Health.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Paieon
Realizing Cardiac Imaging
INDICATIONS FOR USE
# Indications for Use
510(k) Number (if known): K072591
Device Name: The CardiOp-B System
Indications for Use:
The CardiOp-B System is a software system that assists in the evaluation of coronary lesions by enabling the creation of 3D images of coronary vessel segments based on two to three 2D angiography images obtained from single plane angiography. CardiOp-B provides quantitative information regarding the calculated dimensions of arterial segments based on the 3D image. CardiOp-B is intended for use in real-time in the catheterization lab and off-line for post-procedural analysis. It is intended for use by clinicians, technicians and research personnel.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ami M. M.
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Page _________________________________________________________________________________________________________________________________________________________________________ of
(Posted November 13, 2003)
Sept 2007