K153416 · Arcscan, Inc. · IYO · Feb 2, 2016 · Radiology
Device Facts
Record ID
K153416
Device Name
ArcScan Insight 100
Applicant
Arcscan, Inc.
Product Code
IYO · Radiology
Decision Date
Feb 2, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1560
Device Class
Class 2
Intended Use
The ArcScan Insight 100 is indicated for use in adults to measure dimensions of components of the human eye. To provide tomographic, high-resolution ultrasound images of the anterior portion of the eye. It is also designed to measure these tissues and structures, such as anterior chamber depth, angle-to-angle width and sulcus-to-sulcus width. Measurements can be made of the cornea and its individual layers including the epithelium, stroma, and surgically induced surfaces. Measurement also may be made of pathological structures such as solid masses or cysts and it is therefore useful in evaluation and/or planning of refractive surgery and evaluation of pathologies of the anterior segment such as trauma, tumors, cysts, glaucoma and hypotony.
Device Story
Precision ultrasound device for imaging and biometry of anterior ocular structures. Uses 20-60 MHz transducer scanning in arc approximating anterior ocular surface curvature. Transforms ultrasonic pulsed echo signals into tomographic, high-resolution images. Used in clinical settings by eye care professionals. Provides quantitative measurements of corneal layers, anterior chamber depth, and pathological structures (masses/cysts). Output assists clinicians in refractive surgery planning and diagnosis of anterior segment pathologies. Benefits patients through precise anatomical assessment and improved surgical planning.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (General Safety) and IEC 60601-1-2 (EMC) demonstrated. Acoustic power output verified below Track 1 pre-amendment levels. Measurement performance validated against specifications for axial resolution, lateral resolution, thickness measurement, and lateral measurement.
Technological Characteristics
Ultrasonic pulsed echo imaging system; 20-60 MHz transducer (Blatek AT20573); arc-based scanning motion; SAL 10^-6 sterility; biocompatible patient contact surfaces; complies with IEC 60601-1 and IEC 60601-1-2 standards.
Indications for Use
Indicated for use in adults to measure dimensions of human eye components, including cornea layers (epithelium, stroma), anterior chamber depth, angle-to-angle width, and sulcus-to-sulcus width. Used for evaluation/planning of refractive surgery and assessment of anterior segment pathologies (trauma, tumors, cysts, glaucoma, hypotony).
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Artemis VHF Ultrasonic Arcscan System (Artemis 2) (K021333)
K070884 — POD B-SCAN · Starfish Product Engineering, Inc. · Apr 11, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 2, 2016
ArcScan, Inc. % Mr. John Mann Director RAQA Evergreen Research 433 Park Point Drive, Suite 140 GOLDEN CO 80401
Re: K153416
Trade/Device Name: ArcScan Insight 100 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: January 12, 2016 Received: January 13, 2016
Dear Mr. Mann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oolo
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K153416
Device Name ArcScan Insight 100
#### Indications for Use (Describe)
The ArcScan Insight 100 is indicated for use in adults to measure dimensions of the human eye. To provide tomographic, high-resolution ultrasound images of the eye. It is also designed to measure these tissues and structures, such as anterior chamber depth, angle-to-angle width and sulcus-to-sulcus width. Measurements can be made of the cornea and its individual layers including the epithelium, stroma, and surgically induced surfaces. Measurement also may be made of pathological structures such as solid masses or cysts and it is therefore useful in evaluation and/or planning of refractive surgery and evaluation of pathologies of the anterior segment such as trauma, tumors, cysts, glaucoma and hypotony.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Diagnostic Ultrasound Indications for Use
System: ArcScan Insight 100 Transducer: Blatek AT20573
Intended Use: The ArcScan Insight 100 is indicated for use in adults to measure dimensions of components of the human eye. To provide tomographic, high-resolution ultrasound images of the anterior portion of the eye. It is also designed to measure these tissues and structures, such as anterior chamber depth, angle width and sulcus-to-sulcus width. Measurements can be made of the cornea and its individual layers including the epithelium, stroma, and surgically induced surfaces. Measurement also may be made of pathological structures such as solid masses or cysts and it is therefore useful in evaluation and/or plaming of refractive surgery and evaluation of pathologies of the anterior segment such as trauma, turnors, cysts, glaucoma and hypotony.
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General | Specific | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| (Track 1 Only) | (Tracks 1 & 3) | | | | | | | A-mode (P) |
| Ophthalmic | Ophthalmic | P | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under this appendix * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
{4}------------------------------------------------
# 510(k) Summary ArcScan Insight 100™
#### SUBMITTER INFORMATION
| Company name / address | ArcScan, Inc.<br>433 Park Point Drive<br>Golden CO 80401 |
|-------------------------------|------------------------------------------------------------------------------------------------------------------|
| 510(k) contact name / numbers | Andy Levien<br>Phone +1 877.363.7226<br>alevien@arcscan.com |
| Date summary prepared | 11/23/2015 |
| DEVICE IDENTIFICATION | |
| Trade names | ArcScan Insight 100TM |
| Common name | Ultrasound Bio Microscope |
| Classification name | Ultrasonic pulsed echo imaging system, 90-IYO,<br>892.1560<br>Diagnostic ultrasonic transducer, 90-ITX, 892.1570 |
| PREDICATE DEVICE | |
| Trade names | Artemis VHF Ultrasonic Arcscan System (Artemis<br>2) |
| 510(k) number | K021333 |
#### DEVICE DESCRIPTION
ArcScan Insight 100 is a precision ultrasound device for imaging and biometry of the eye. It uses a 20-60 MHz transducer that scans the eye in an arc whose curvature approximates the anterior ocular surfaces. Specifically, the ArcScan Insight 100 can measure the thickness of the cornea and its individual layers, the epithelium, stroma, and surgically induced surfaces. Measurements can also be made of the anatomic structures comprising the anterior of the eye such as anterior chamber depth, angle-to-angle width, and sulcus-to-sulcus width. Measurements can be made of pathologic structures such as solid masses and cysts.
{5}------------------------------------------------
### INDICATION FOR USE
The ArcScan Insight 100 is indicated for use in adults to measure dimensions of components of the human eye. To provide tomographic, high-resolution ultrasound images of the anterior portion of the eye. It is also designed to measure these tissues and structures, such as anterior chamber depth, angleto-angle width and sulcus-to-sulcus width. Measurements can be made of the cornea and its individual layers including the epithelium, stroma, and surgically induced surfaces. Measurement also may be made of pathological structures such as solid masses or cysts and it is therefore useful in evaluation and/or planning of refractive surgery and evaluation of pathologies of the anterior segment such as trauma, tumors, cysts, glaucoma and hypotony.
## TECHNOLOGICAL CHARACTERISTICS COMPARISON
Substantial Equivalence: The ArcScan Insight 100 is substantially equivalent to the classified device described in CFR 21 891.1560, "Ultrasonic Pulsed Echo Imaging System" and to other ultrasound systems that have been cleared by the 510(k) process, particularly the Artemis™ VHF Ultrasonic Arc-scan System cleared by ArcScan, Inc. (formerly Ultralink LLC) in K021333.
The 510(k) Substantial Equivalence Decision-making Process (detailed)from The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] was followed as described below:
- 1. These products have the same intended use, to produce high-resolution ultrasound images of the anterior portion of the eye, and to measure these tissues such as the thickness of the cornea and its individual layers, the epithelium stroma and surgically induced surfaces. They can also be used to measure pathologic structures such as solid masses or cysts.
- 2. The products have the same fundamental scientifictechnology.
- 3. The technological characteristics of this device are the same as those for the predicate device except for the method by which the larger radius of motion is achieved.
- 4. The acoustic output of this device is similar to that of predicate devices and is well below the preamendment levels described in the guidance document "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers."
- 5. The products have the same sterility assurance level of SAL10-6.
- 6. The products have the same biocompatibility for the patient contact surfaces.
### PERFORMANCE DATA
The ArcScan Insight 100 complies with IEC 60601-1, Ed. 3: Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance; Corrigendum 1(2006): Corrigendum 2(2007). The ArcScan Insight 100 complies with IEC 60601-1-2 Ed 3.0: Medical Electrical Equipment-Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility-Requirements and Tests.IEC 60601-1, General Electrical Safety.
The ArcScan Insight 100 Acoustic Power was demonstrated to be less than the preamendment levels for Track 1.
The ArcScan Insight 100 measurement performance capability was demonstrated to meet specifications for the following characteristics:
- Axial resolution
- Lateral resolution ●
- Thickness measurement
- Lateral measurement ●
{6}------------------------------------------------
# CONCLUSION
The ArcScan Insight 100 is substantially equivalent to the ArcScan Artemis 2.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.