Browse hierarchy Radiology (RA) Subpart B — Diagnostic Devices 21 CFR 892.1550 Product Code IYN K181336 — HS50 / HS60 Diagnostic Ultrasound System
HS50 / HS60 Diagnostic Ultrasound System
K181336 · Samsung Medison Co., Ltd. · IYN · Dec 4, 2018 · Radiology
Device Facts
Record ID K181336
Device Name HS50 / HS60 Diagnostic Ultrasound System
Applicant Samsung Medison Co., Ltd.
Product Code IYN · Radiology
Decision Date Dec 4, 2018
Decision SESE
Submission Type Traditional
Regulation 21 CFR 892.1550
Device Class Class 2
Attributes Pediatric
Indications for Use
The HS50 / HS60 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Trans-esophageal(Cardiac), and peripheral vessel.
Device Story
The HS50/HS60 is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various linear, convex, phased, static, and volume probes (1.0–20.0 MHz) and displays it in multiple modes (B, M, Color M, Anatomical, Color/PW/CW Doppler, TDI, TDW, PD, ElastoScan, 3D/4D/XI STIC, Dual/Quad/Combined/Simultaneous/Zoom). The system uses digital multi-beam forming technology. It is operated by healthcare professionals in clinical settings to measure anatomical structures and provide analysis packages for diagnosis. The system displays real-time mechanical and thermal indices. Output is viewed on an integrated monitor to assist in clinical decision-making, potentially benefiting patients through high-resolution imaging and diagnostic analysis. It includes a Li-ion backup battery and battery assist function.
Clinical Evidence
No clinical studies were required to demonstrate substantial equivalence. Bench testing only, including acoustic output, software function, thermal, electrical, electromagnetic, and mechanical safety testing, was performed to demonstrate compliance with applicable standards.
Technological Characteristics
Mobile ultrasound system with digital multi-beam forming. Supports linear, convex, phased, static, and volume probes (1.0–20.0 MHz). Features include Li-ion backup battery, battery assist, and ECG interface. Safety standards: ANSI AAMI ES60601-1, IEC 60601-2-18, IEC 60601-1-2 (4th Ed), IEC 60601-2-37, NEMA UD 2, NEMA UD 3.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid analysis of the human body in patients including fetal/obstetrics, abdominal, gynecological, pediatric, small organ, neonatal/adult cephalic, trans-rectal, trans-vaginal, musculoskeletal, urological, cardiac (adult/pediatric/trans-esophageal), and peripheral vessel applications.
Regulatory Classification
Identification An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
HS50/HS60 Diagnostic Ultrasound System (K170493 )
RS85 Diagnostic Ultrasound System (K173204 )
WS80A Diagnostic Ultrasound System (K173513 )
HS40 Diagnostic Ultrasound System (K180409 )
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Samsung Medison Co., Ltd. Ji Yea Lee Regulatory Affairs Specialist 3366. Hanseo-ro. Nam-myeon. HONGCHEON-GUN, GANGWON-DO 25108 REPBLIC OF KOREA
Re: K181336
Trade/Device Name: HS50 / HS60 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: November 02, 2018 Received: November 05, 2018
Dear Ji Yea Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
December 4, 2018
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
### K181336
Device Name
HS50 / HS60 Diagnostic Ultrasound System
### Indications for Use (Describe)
The HS50 / HS60 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Trans-esophageal(Cardiac), and peripheral vessel.
### Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, bold, sans-serif font. The word is placed inside a blue, oval-shaped background that is tilted slightly to the right.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: HS50/HS60 Diagnostic Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | P | P | P | | Note 15 | Note 1 | Note 2,3,4, 7, 8, 9, 11 |
| | Abdominal (See Note 10) | P | P | P | P | Note 15 | Note 1 | Note 2,6, 7, 8, 9, 11, 12, 16 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | Note 15 | Note 1 | Note 2, 6, 7, 8, 9, 11 |
| | Small Organ (See Note 5) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11, 12 |
| | Neonatal Cephalic | P | P | P | P | Note 15 | Note 1 | Note 2, 7, 8, 9, 11 |
| | Adult Cephalic | P | P | P | P | Note 15 | Note 1 | Note 2, 7, 8, 9, 11 |
| | Trans-rectal | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 8, 9, 11, 12 |
| | Trans-vaginal | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 8, 9, 11, 12 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11, 12 |
| | Musculo-skel. (Superfic.) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11 |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 8, 9, 11, 12 |
| Cardiac | Cardiac Adult | P | P | P | P | Note 15 | Note 1 | Note 4, 7, 14, 17 |
| | Cardiac Pediatric | P | P | P | P | Note 15 | Note 1 | Note 4, 7, 14, 17 |
| | Trans-esophageal (Cardiac) | P | P | P | P | Note 15 | Note 1 | Note 4, 7, 14, 17 |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | P | Note 15 | Note 1 | Note 2, 6, 7, 9, 11 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, DuallQuad,
- B+C+CW, B+PD+CW, B+E
- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
- Note 15: Color Doppler includes Power (Amplitude) Doppler
- Note 16: Contrast Note 17: Strain+
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white letters on a blue background. The logo is tilted slightly to the right.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: LA3-14AD for use with HS60/ HS50
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|----------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | | | | | | | |
| | Abdominal (See Note 10) | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11, 12 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11, 12 |
| | Musculo-skel. (Superfic.) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11 |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | Note 15 | Note 1 | Note 2, 6, 7, 9, 11 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
#### Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M, B+C+CW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
Note 15: Color Doppler includes Power (Amplitude) Doppler
Note 16: Contrast
Note 17: Strain+
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed on a blue, oval-shaped background.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: CA1-7AD for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-----------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 9, 11 | |
| | Abdominal (See Note 10) | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 9, 11, 16 | |
| | Intra-operative (See Note 6) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 9, 11 | |
| | Musculo-skel. (Superfic.) | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 9, 11 | |
| | Intra-luminal | | | | | | | | |
| | Other (See Note 13) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate
Note 14: Tissue Doppler Imaging (TDI)
Note 15: Color Doppler includes Power (Amplitude) Doppler
Note 16: Contrast Note 17: Strain+
{6}------------------------------------------------
510(k) Premarket Notification - Traditional
Image /page/6/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The text is set against a blue, oval-shaped background. The logo is slightly tilted.
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: CF4-9 for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | P | P | P | | Note 15 | Note 1 | Notes 9, 11 |
| | Abdominal (See Note 10) | P | P | P | | Note 15 | Note 1 | Notes 9, 11 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | Note 15 | Note 1 | Notes 6, 9, 11 |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | P | P | P | | P | Note 1 | Notes 9, 11 |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | Note 15 | Note 1 | Notes 9, 11 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
- Note 15: Color Doppler includes Power (Amplitude) Doppler
- Note 16: Contrast Note 17: Strain+
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white letters on a blue, oval-shaped background. The logo is tilted slightly to the right. The logo is simple and recognizable.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: ER4-9 for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | |
|-----------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-------------------|
| General | Specific | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| (Track I only) | (Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 9 |
| | Abdominal (See Note 10) | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>&<br>Other | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 9, 12 |
| | Trans-vaginal | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 9, 12 |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) (See Note 13) | P | P | P | | Note 15 | Note 1 | Notes 2, 7, 9, 12 |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
#### Additional Comments:
Note 1: B+M, B+PW, B+CW, B+C, B+PD, B+Elastoscan, B+C+PW, B+DD+PW, B+DPD+PW, B+TD+PW, B+C+M, B+C+CW, Dual/Quad (B, B+C, B+PD, B+TD, B+DPD)
Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)
Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)
Note 10: Includes Renal, Gynecology/Pelvis
Note 11: Panoramic imaging
Note 12: ElastoScan
Note 13: Includes Urology/Prostate Note 14: Tissue Doppler Imaging (TDI)
Note 15: Color Doppler includes Power (Amplitude) Doppler
Note 16: Contrast
Note 17: Strain+
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the Samsung logo. The logo is in blue and features the word "SAMSUNG" in white, bold letters. The logo is tilted slightly to the right.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: LA3-16A for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|---------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal (See Note 3) | | | | | | | |
| | Abdominal (See Note 10) | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | P | P | P | | Note 15 | Note 1 | Note 2,7,9,11,12 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | Note 15 | Note 1 | Note 2,7,9,11,12 |
| | Musculo-skel. (Superfic.) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11 |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | Note 15 | Note 1 | Note 2, 6, 7, 9, 11 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
- Note 15: Color Doppler includes Power (Amplitude) Doppler
- Note 16: Contrast Note 17: Strain+
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the Samsung logo. The logo is a blue oval with the word "SAMSUNG" written in white letters inside. The logo is simple and recognizable. The font is sans-serif and the letters are evenly spaced.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: LA3-16AI for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|--------------------------|------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|--|
| General | Specific | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | |
| (Track I only) | (Tracks I & III) | | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal (See Note 3) | | | | | | | | |
| | Abdominal | | | | | | | | |
| | Intra-operative (See Note 6) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | Note 15 | Note 1 | Note 7, 9, 11 | |
| | Musculo-skel. (Superfic.) | P | P | P | | Note 15 | Note 1 | Note 7, 9, 11 | |
| | Intra-luminal | | | | | | | | |
| | Other (See Note 13) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | Note 15 | Note 1 | Note 6, 7, 9, 11 | |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
- Note 15: Color Doppler includes Power (Amplitude) Doppler
- Note 16: Contrast Note 17: Strain+
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the Samsung logo. The logo is in blue and white. The word "SAMSUNG" is written in white letters on a blue background. The background is an oval shape.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: LA2-9A for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|---------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | | | | | | | |
| | Abdominal (See Note 10) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11 |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11 |
| | Musculo-skel. (Superfic.) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 9, 11 |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | Note 15 | Note 1 | Note 2, 6, 7, 9, 11 |
| | Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
- Note 15: Color Doppler includes Power (Amplitude) Doppler
- Note 16: Contrast Note 17: Strain+
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, set against a blue, oval-shaped background. The oval is tilted slightly, giving the logo a dynamic appearance.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: PA3-8B for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | | | | | | | | |
| | Abdominal (See Note 10) | P | P | P | P | Note 15 | Note 1 | Note 7 | |
| | Intra-operative (See Note 6) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | P | P | P | P | Note 15 | Note 1 | Note 6, 7 | |
| | Small Organ (See Note 5) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | P | P | P | P | Note 15 | Note 1 | Note 7 | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (See Note 13) | | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7, 14,17 | |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7, 14, 17 | |
| | Trans-esophageal (Cardiac) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
- Note 15: Color Doppler includes Power (Amplitude) Doppler
- Note 16: Contrast Note 17: Strain+
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, bold letters. The word is set against a blue, oval-shaped background.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: PE2-4 for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal/Obstetrics (See Note 3) | | | | | | | | |
| | Abdominal (See Note 10) | P | P | P | P | Note 15 | Note 1 | Note 7 | |
| | Intra-operative (See Note 6) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| Fetal Imaging<br>& Other | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | P | P | P | P | Note 15 | Note 1 | Note 7 | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (See Note 13) | | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | Note 1 | Note 4, 7, 14,17 | |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | Note 1 | Note 4, 7, 14, 17 | |
| | Trans-esophageal (Cardiac) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (spec.) | | | | | | | | |
N= new indication; P= previously cleared by FDA K170493; E= added under Appendix E
#### Additional Comments:
Color Doppler includes Power (Amplitude) Doppler
Note 1: B+M, B+PW, B+C, B+PD, B+TD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E
Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: Tissue Doppler Imaging (TDI)
- Note 15: Color Doppler includes Power (Amplitude) Doppler
- Note 16: Contrast Note 17: Strain+
Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed on a blue, oval-shaped background.
510(k) Premarket Notification - Traditional
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE STATEMENT
510(k) No.:
Device Name: CV1-8AD for use with HS50/HS60
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Mode of Operation (*includes simultaneous B-mode) | | | | | | |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|---------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal Imaging<br>& Other | Fetal/Obstetrics (See Note 3) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 8, 9, 11 |
| | Abdominal (See Note 10) | P | P | P | | Note 15 | Note 1 | Note 2, 7, 8, 9, 11 |
| | Intra-operative (See Note 6) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (See Note 5) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Cardiac) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (See Note 13) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (Cardiac) | | | | | | | |
| | Other (spec.) | | | | | | |…