Browse hierarchy Radiology (RA) Subpart B — Diagnostic Devices 21 CFR 892.1550 Product Code IYN K102901 — HI VISION AVIUS DIAGNOSTIC ULTRASOUND SCANNER
HI VISION AVIUS DIAGNOSTIC ULTRASOUND SCANNER
K102901 · Hitachi Medical Systems America, Inc. · IYN · Oct 27, 2010 · Radiology
Device Facts
Record ID K102901
Device Name HI VISION AVIUS DIAGNOSTIC ULTRASOUND SCANNER
Applicant Hitachi Medical Systems America, Inc.
Product Code IYN · Radiology
Decision Date Oct 27, 2010
Decision SESE
Submission Type Special
Regulation 21 CFR 892.1550
Device Class Class 2
Attributes Pediatric
Indications for Use
The HI VISION Avius is intended for use by trained personnel (doctor, songrapher, etc.) for the diagnostic ultrasound evaluation of Abdominal, Cardiac, Intra-operative, Fetal, Pediatric, Small Organ, Peripheral vessel, Biopsy, Trans-rectal, Trans-vaginal, Musculoskeletal, Neonatal Cephalic, Adult Cephalic, Endoscopy, Intra-luminal, Gynecology, Urology and Laparoscopic clinical applications. The Modes of Operation of the HI VISION Avius are B mode, M mode, PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler), Color Doppler, Amplitude Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, and Real Time Tissue Elastography.
Device Story
The HI VISION Avius is a diagnostic ultrasound scanner used by trained clinicians (doctors, sonographers) in clinical settings. It utilizes ultrasound transducers to emit acoustic energy into the body and detect reflected echoes. A computer system processes these signals to generate images or Doppler data, which are displayed on a video monitor. The device supports multiple modes, including B-mode, M-mode, PW/CW Doppler, Color Doppler, 3D/4D imaging, and Real Time Tissue Elastography. Clinicians use the displayed images and Doppler data to perform diagnostic evaluations, guide biopsies, and assess fluid flow. The output assists in clinical decision-making by providing anatomical and physiological information, potentially improving diagnostic accuracy and patient management.
Clinical Evidence
Bench testing only. No clinical data provided. Validation included product specifications for the new transmitter and clinical safety/effectiveness documentation.
Technological Characteristics
Diagnostic ultrasound system utilizing pulsed Doppler imaging. Operates in 1-20 MHz frequency range. Includes various transducers for specific clinical applications. Features B, M, PW/CW Doppler, Color Doppler, 3D/4D, and Real Time Tissue Elastography modes. System architecture includes transducer, computer control/signal processing unit, and video display. Identical hardware and base software to predicate device (K093466).
Indications for Use
Indicated for diagnostic ultrasound evaluation of abdominal, cardiac, intra-operative, fetal, pediatric, small organ, peripheral vessel, biopsy, trans-rectal, trans-vaginal, musculoskeletal, neonatal/adult cephalic, endoscopy, intra-luminal, gynecological, urological, and laparoscopic applications in patients requiring diagnostic ultrasound imaging or fluid flow analysis.
Regulatory Classification
Identification An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
HITACHI HI VISION Preirus Diagnostic Ultrasound Scanner (K093466 )
Submission Summary (Full Text)
{0}------------------------------------------------
K/02901
OCT 2 7 2010
·
. . .
:
. . .
.....
# Section 1
.
、
・・
# 510(k) Summary
©2010 Hitachi Medical Systems America, Inc.
、
,
:
- (
: 上一篇:
·
・ ・・
1
เ
:
:
・・
:
{1}------------------------------------------------
OCT 2 7 2010
| Submitter: | Hitachi Medical Systems America, Inc.<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371<br>ph: (330) 425-1313<br>fax: (330) 963-0749 |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Douglas J. Thistlethwaite |
| Date: | September 27, 2010 |
## Device Name
Submitter Information
| Classification Name: | System, Imaging, Pulsed Doppler, Ultrasonic |
|-------------------------|--------------------------------------------------------------------|
| Classification Number: | 90-IYN |
| Trade/Proprietary Name: | HITACHI HI VISION Avius Diagnostic Ultrasound Scanner |
| Predicate Device(s): | HITACHI HI VISION Preirus Diagnostic Ultrasound Scanner<br>K093466 |
## Device Intended Use
The HI VISION Avius is intended for use by trained personnel (doctor, songrapher, etc.) for the diagnostic ultrasound evaluation of Abdominal, Cardiac, Intra-operative, Fetal, Pediatric, Small Organ, Peripheral vessel, Biopsy, Trans-rectal, Trans-vaginal, Musculoskeletal, Neonatal Cephalic, Adult Cephalic, Endoscopy, Intra-luminal, Gynecology, Urology and Laparoscopic clinical applications.
The Modes of Operation of the HI VISION Preirus are B mode, M mode, PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler), Color Doppler, Amplitude Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, 4D Imaging, and Real Time Tissue Elastography.
## Device Description
## Function
An ultrasound system consists of the following:
- Ultrasound transducer(s) to generate the transmitted ultrasound energy and detect the ● reflected echoes
- � A computer system to control the transducer and analyze the signals resulting from the reflected echoes
- A video monitor with optional image recorder to display the computed image or derived . Doppler data
## Scientific Concepts
An acoustic wave is a mechanical perturbation of a medium which passes through a given medium at a fixed velocity. The rate at which the particles in the medium vibrate in the disturbance is the frequency of the wave, and is measure as cycles/second, or hertz (Hz). Frequencies above 20 kHz are not audible, and above this frequency, the acoustic energy is known as ultrasound. For the purposes of medical ultrasound imaging, frequencies in the range of 1-20 MHz are utilized.
{2}------------------------------------------------
The ultrasound waves comprising a beam travel in as straight line in homogeneous media. When an ultrasound wave reaches an interface between two media of different impedances, a portion of the beam energy may pass through the boundary (transmission), and a portion may be reflected. The direction of propagation of the transmitted beam is determined by the angle of incidence of the incident beam upon the boundary, and differences (if any) in the speed of sound in the two media. The direction of reflection is determined solely by the angle of incidence upon the boundary. The relative strength of the reflected wave depends upon the differences in the impedances between the two media. Reflection at a boundary between soft tissue and bone, as an example, involves a large impedance difference, and results in a relatively strong reflected echo. Reflection at a boundary between two soft tissue-types with a relatively small impedance difference, on the other hand, results in a relatively weak reflected echo.
The ultrasound transducer, when operating in pulsed mode, periodically emits an ultrasound burst at a predetermined rate described as the pulse repetition frequency (PRF). During the time duration that the transducer is not transmitting ultrasound energy, it may act as a received for the reflected ultrasound energy. Since the speed of propagation of ultrasound in tissues is estimated as 1540m/sec, the time elapsed between transmission of a pulse and receipt of an echo may be used to estimate the distance from the transducer to the tissue structure giving rise to the reflected echo. The relative strength of the reflected echo can be used to determine the brightness of display of the reflected echo or strength of derived Doppler signal.
## Physical and Performance Characteristics
The principle of operation of ultrasound imaging involves generation of an ultrasound wave with an electric signal applied to a transducer, direction of the resulting ultrasound wave into the tissue of the body, and reception and analysis of the echoes reflected back to the same or an adjacent transducer from the various tissues along the path of the ultrasound wave.
## Device Technological Characteristics
The technological characteristics of this device are identical to the primary predicate device. The control and image processing hardware and the base elements of the system software are identical to the predicate device. See Section 4 - Predicate Device Comparison.
## Conclusions
It is the opinion of Hitachi Medical Systems America. Inc. that HITACHI HI VISION Avius Diagnostic Ultrasound Scanner is substantially equivalent to the predicate devices. In addition, we have concluded that the subject system is substantially equivalent with respect to safety, effectiveness, and functionality to the HI VISION Preirus Diagnostic Ultrasound Scanner K093466 with the exception of the new transmitter. The Product Specification for new transmitter is included in Section 3 - General Device Description and the validation is included in Section 6 - Clinical Safety & Effectiveness.
## Attachment
| Pressure of the contract of the comments of the comments of the comments of the first of the first of the first of the first of the first of the first and the first and the f<br>ATTACHMENT | POSITION |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Declaration of Conformity with Design Controls<br>Children and Children and Children and Children Children Children Children Children Children Children Children Children Children Children Children Children Children Children | |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three overlapping strokes forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Mr. Doug Thistlewaite Manager, Regulatory Affairs Hitachi Medical Systems America, Inc. 1959 Summit Commerce Park TWINSBURG OH 44087-2371
OCT 2 7 2010
Re: K102901
Trade/Device Name: HI VISION Avius Diagnostic Ultrasound Scanner Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: September 27, 2010 Received: September 30, 2010
Dear Mr. Thistlewaite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the HI VISION Avius Diagnostic Ultrasound Scanner, as described in your premarket notification:
Transducer Model Number
| EUP-B512 | EUP-CV524 | EUP-L54MA |
|-----------|-----------|------------------|
| EUP-B514 | EUP-CV714 | EUP-L65 |
| EUP-C511 | EUP-ES52E | EUP-L73S |
| EUP-C514 | EUP-ES52M | EUP-L74M |
| EUP-C516 | EUP-F334 | EUP-LV74 |
| EUP-C524 | EUP-F531 | EUP-053T |
| EUP-C532 | EUP-L52 | EUP-054J |
| EUP-C715 | EUP-L53 | EUP-OL334 |
| EUP-CC531 | EUP-L53L | EUP-R54AW-19, -3 |
{4}------------------------------------------------
#### EUP-VV731
- EUP-S50A EUP-S52 EUP-S70
### EUP-U533 EUP-V53W EUP-VV53 !
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.
Sincerely yours,
David G. Brown, Ph.D. Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
## Indications for Use
103901
OCT 2 7 2010
510(k) Number (if known):
## Device Name: HI VISION Avius Diagnostic Ultrasound Scanner
Indications For Use:
The HI VISION Avius is intended for use by trained personnel (doctor, songrapher, etc.) for the diagnostic ultrasound evaluation of Abdominal, Cardiac, Intra-operative, Fetal, Pediatric, Small Organ, Peripheral vessel, Biopsy, Trans-rectal, Trans-vaginal, Musculosketal, Neonatal Cephalic, Endoscopy, Intra-luminal, Gynecology, Urology and Laparoscopic clinical applications.
The Modes of Operation of the HI VISION Avius are B mode, M mode, PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler), Color Doppler, Amplitude Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, and Real Time Tissue Elastography.
X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-1
Page 1 of
Division of Radiolo
{6}------------------------------------------------
System:
-
## DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM HI VISION Avius
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|-----------------------------------------------------------------------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P |
| | Abdominal | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | Pb | Pb | Pb | | Pb | Pb | Pb |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | P | P | P | | P | P | P |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (Spec.) | Pd | Pd | Pd | | Pd | Pd | Pd |
| | Neonatal Cephalic | P | P | P | | P | P | P |
| Fetal Imaging<br>& Other | Adult Cephalic | P | P | P | P | P | P | P |
| | Trans-rectal | Ph | Ph | Ph | | Ph | Ph | Ph |
| | Trans-vaginal | Pf | Pf | Pf | | Pf | Pf | Pf |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | P | P | P | | P | P | P |
| | Musculo-skel. (Superfic.) | P | P | P | | P | P | P |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | P | P |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | P | P |
| | Trans-esophageal (card.) | Pg | Pg | Pg | P | Pg | Pg | Pg |
| | Other (spec.) | | | | | | | |
| Peripheral | Peripheral vessel | P | P | P | P | P | P | P |
| Vessel | Other (spec.) | | | | | | | |
| | N = new indication. P = previously cleared in K093466 | | | | | | | |
| | *Combination of each operating mode, B, M, PWD, CWD and Color Doppler. | | | | | | | |
| | **Amplitude Doppler, Harmonic Imaging, 3D Imaging, Real Time Tissue Elastography, Torapezoid, | | | | | | | |
| | Contrast, Oblique, High Compound imaging, Omni Directional M-mode, | | | | | | | |
| | Wide View, Tissue Doppler imaging | | | | | | | |
| Additional Comments: | | | | | | | | |
| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures | | | | | | | |
| | (including amniocentesis). | | | | | | | |
| Subscript "b": | Includes imaging of organs and structures exposed during surgery | | | | | | | |
| | (excluding neurosurgery and laparoscopic procedures). | | | | | | | |
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows: |
|--------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------|
·
| Subscript | Description |
|----------------|------------------------------------------------------------------------------------------------------------------------|
| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis). |
| Subscript "b": | Includes imaging of organs and structures exposed during surgery (excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis. |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy. |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy. |
| Subscript "g": | For pediatric patients. |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy. |
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
Radiological Devices
510(k) Number: K102901
.
{7}------------------------------------------------
System: Transducer
TALL PRO
STATE THE THE PERSONAL PARTY
Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P |
| | Abdominal | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Spec.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N = new indication. P = previously cleared in K093466
*Combination of each operating mode, B, M, PWD and Color Doppler.
**Amplitude Doppler, Harmonic Imaging, Contrast, High Compound imaging, Wide View
Additional Comments
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures Subscript "a": (including amniocentesis).
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis. |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy. |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy. |
| Subscript "g": | For pediatric patients. |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy. |
| (PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) | |
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
ﻟﻠ 10240 510(k) Number:
{8}------------------------------------------------
System: Transducer:
#### HI VISION Avius EUP-B514
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | P | P | P | P |
| | Abdominal | Pa | Pa | Pa | Pa | Pa | Pa | Pa |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Spec.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| Fetal Imaging<br>& Other | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows
N = new indication. P = previously cleared in K093466
*Combination of each operating mode, B, M, PWD and Color Doppler.
** Amplitude Doppler, Harmonic Imaging, Contrast, High Compound imaging, Wide View
Additional Comments:
| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis). |
|----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery |
| (excluding neurosurgery and laparoscopic procedures). | |
| Subscript "c": Includes thyroid, parathyroid, breast, scrotum, penis. | |
| Subscript "d": Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. | |
| Subscript "e": Includes imaging for guidance of transrectal biopsy. | |
| Subscript "f": Includes imaging for guidance of transvaginal biopsy. | |
| Subscript "g": For pediatric patients. | |
| Subscript "h": Includes imaging for guidance of transrectal biopsy. | |
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IE NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
• Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number: K102901
{9}------------------------------------------------
System: Transducer:
## EUP-C511
| Clinical Application | | Mode of Operation | | | | | | | |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | P | P | P | | P | P | P | |
| | Abdominal | Pa | Pa | Pa | | Pa | Pa | Pa | |
| | Intra-operative (Spec.) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | P | P | P | | P | P | P | |
| | Small Organ (Spec.) | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Cardiac | Cardiac Adult | P | P | P | | P | P | P | |
| | Cardiac Pediatric | P | P | P | | P | P | P | |
| | Trans-esophageal (card.) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | P | P | |
| | Other (spec.) | | | | | | | | |
Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows
N = new indication. P = previously cleared in K093466
*Combination of each operating mode, B, M, PWD and Color Doppler.
** Amplitude Doppler, Harmonic Imaging, Contrast, High Compound imaging,
Omni Directional M-mode, Wide View
Additional Comments:
Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures Subscript "a": (including amniocentesis).
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis. |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy. |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy. |
| Subscript "g": | For pediatric patients. |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy. |
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOT Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
{10}------------------------------------------------
System: Transducer:
EÜP-C514
| Clinical Application | | | | | | Mode of Operation | | | | |
|---------------------------|------------------------------|----|----|-----|-----|-------------------|----------------------|--------------------|--|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal | P | P | P | | P | P | P | | |
| | Abdominal | Pa | Pa | Pa | | Pa | Pa | Pa | | |
| | Intra-operative (Spec.) | | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | P | P | P | | P | P | P | | |
| | Small Organ (Spec.) | Pd | Pd | Pd | | Pd | Pd | Pd | | |
| | Neonatal Cephalic | | | | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | | |
| | Intra-luminal | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | | |
| | Other (spec.) | | | | | | | | | |
#### Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows:
N = new indication. P = previously cleared in K093466
*Combination of each operating mode, B, M, PWD and Color Doppler.
** Amplitude Doppler, Harmonic Imaging, Contrast, High Compound imaging,
Omni Directional Momode, Wide View
Additional Comments:
| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis). |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis. |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy. |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy. |
| Subscript "g": | For pediatric patients. |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy. |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number: K102901
{11}------------------------------------------------
System: HI VISION Avius
Transducer: EUP-C516
| Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows: | |
|--------------------------------------------------------------------------------------------------|-------------------|
| Clinical Application | Mode of Operation |
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | P | P | P |
| | Abdominal | Pa | Pa | Pa | | Pa | Pa | Pa |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P | P |
| | Small Organ (Spec.) | Pd | Pd | Pd | | Pd | Pd | Pd |
| | Neonatal Cephalic | | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
N = new indication. P = previously cleared in K093466
*Combination of each operating mode, B, M, PWD and Color Doppler.
** Amplitude Doppler, Harmonic Imaging, Contrast, High Compound imaging,
Omni Directional M-mode, Wide View
Additional Comments:
| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis). |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis. |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy. |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy. |
| Subscript "g": | For pediatric patients. |
Subscript "h": Includes imaging for guidance of transrectal biopsy.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number: K102901
{12}------------------------------------------------
System: HI VISION Avius Transducer: EUP-C524
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | P | P | P |
| | Abdominal | P | P | P | | P | P | P |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P | P |
| | Small Organ (Spec.) | Pc | Pc | Pc | | Pc | Pc | Pc |
| Fetal Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (spec.) | | | | | | | |
## Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows:
N = new indication. P = previously cleared in K093466
*Combination of each operating mode, B, M, PWD and Color Doppler.
** Amplitude Doppler, Harmonic Imaging, High Compound imaging, Omni Directional Mrmode, Wide View
Additional Comments:
| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis). |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis. |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy. |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy. |
| Subscript "g": | For pediatric patients. |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy. |
THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED! WKITE RETOW gli DU INC Concurrence of CDRH Office of Device Evaluation (ODE)
(Division Sign-Off) 2 Division of Reproductive, Abdominal, ENT, and Radiological Devices
{13}------------------------------------------------
System: HI VISION Avius EUP-C532
Transducer:
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | Pa | Pa | Pa | | Pa | Pa | Pa |
| | Intra-operative (Spec.) | Pb | Pb | Pb | | Pb | Pb | Pb |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P | P |
| | Small Organ (Spec.) | Pd | Pd | Pd | | Pd | Pd | Pd |
| | Neonatal Cephalic | P | P | P | | P | P | P |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | P | P |
| | Other (spec.) | | | | | | | |
Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human oody as follows
N = new indication. P = previously cleared in K093466
*Combination of each operating mode, B, M, PWD and Color Doppler.
** Amplitude Doppler, Harmonic Imaging, Real Time Tissue Elastography, High Compound imaging, Wide View
Additional Comments:
| Subscript "a": | Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures<br>(including amniocentesis). |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis. |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy. |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy. |
| Subscript "g": | For pediatric patients. |
Subscript "h": Includes imaging for guidance of transrectal biopsy.
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDBH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number: K102901
Prescription Use (Per 21 CFR 801.109)
{14}------------------------------------------------
System: HI VISION Avius Transducer: EUP-C715
| Clinical Application | | Mode of Operation | | | | | | | |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | P | P | P | | P | P | P | |
| | Abdominal | Pa | Pa | Pa | | Pa | Pa | Pa | |
| | Intra-operative (Spec.) | | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | P | P | P | | P | P | P | |
| | Small Organ (Spec.) | Pd | Pd | Pd | | Pd | Pd | Pd | |
| Fetal Imaging<br>& Other | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| | Cardiac Adult | | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | | |
| | Other (spec.) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | | |
| | Other (spec.) | | | | | | | | |
#### Intended use: Diagnostic ultrasound imaging or fluid flow analysis if the human body as follows:
N = new indication. P = previously cleared in K093466
* Combination of each operating mode, B, M, PWD and Color Doppler.
** Amplitude Doppler, Harmonic Imaging, Contrast, High Compound imaging,
Omni Directional M-mode, Wide View
Additional Comments:
Subscript "a". Includes imaging for guidance of percutaneous biopsy of abdominal organs and structures (including amniocentesis).
| Subscript "b": | Includes imaging of organs and structures exposed during surgery<br>(excluding neurosurgery and laparoscopic procedures). |
|----------------|---------------------------------------------------------------------------------------------------------------------------|
| Subscript "c": | Includes thyroid, parathyroid, breast, scrotum, penis. |
| Subscript "d": | Includes thyroid, parathyroid, breast, scrotum, penis and imaging for guidance of biopsy. |
| Subscript "e": | Includes imaging for guidance of transrectal biopsy. |
| Subscript "f": | Includes imaging for guidance of transvaginal biopsy. |
| Subscript "g": | For pediatric patients. |
| Subscript "h": | Includes imaging for guidance of transrectal biopsy. |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number: K102901
Prescription Use (Per 21 CFR 801.109)
{15}------------------------------------------------
System: Transducer:
EUP-CC531
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------|-------------------|----|-----|-----|------------------|----------------------|--------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined*<br>(Spec.) | Other**<br>(Spec.) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | P | P | P | | P | P | P |
| | Abdominal | | | | | | | |
| | Intra-operative (Spec.) | | | | | | | |
| | Intra-operative (Neuro.) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Spec.) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | Pe | Pe | Pe | | Pe | Pe | Pe |
| | Trans-vaginal | Pf | Pf | Pf | | Pf | Pf | Pf |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skel. (Convent.) | | | | | | | |
| | Musculo-skel. (Superfic.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (spec.) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esophageal (card.) | | | | | | | |
| | Other (spec.) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel |…