JAMIESON MODEL 2100 CINE FILM PROCESSOR AND KODAK CINEFLURE PROCESSOR MODEL 2100
Device Facts
| Record ID | K963334 |
|---|---|
| Device Name | JAMIESON MODEL 2100 CINE FILM PROCESSOR AND KODAK CINEFLURE PROCESSOR MODEL 2100 |
| Applicant | Jamieson Film Co. |
| Product Code | IXX · Radiology |
| Decision Date | Sep 11, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1900 |
| Device Class | Class 2 |
Intended Use
The Model 2100 is a leader card black-and-white 35 mm cine film processor, most commonly used for developing cineangiogram films.
Device Story
Leader card 35 mm cine film processor; processes cineangiogram films. System comprises four tanks: developer, fixer, wash, and dryer. Used in clinical environments for radiographic film development. Modifications include electrical/safety updates for CE Medical Device Directive compliance and minor transport assembly adjustments. Device automates film processing workflow; supports diagnostic imaging by preparing cineangiogram film for physician review.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Automatic radiographic film processor; 35 mm cine film format. Dimensions: 58.5" L x 17" W x 43" H. Four-tank system (developer, fixer, wash, dryer). Electrical/safety compliant with CE Medical Device Directive.
Indications for Use
Indicated for the development of 35 mm black-and-white cineangiogram film in a clinical setting.
Regulatory Classification
Identification
An automatic radiographic film processor is a device intended to be used to develop, fix, wash, and dry automatically and continuously film exposed for medical purposes.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- Jamieson JT/Kodak Ektacath (K913241/A)
Related Devices
- K981280 — CP 1000, CURIX HT 330, CLASSIC E.O.S., CURIX HT 530, MAMORAY HT 300, MAMORAY CMPACT PLUS, CURIX COMPACT PLUS, CURIX COMS · Bayer Corp., Agfa Div. · Apr 30, 1998
- K022154 — LEVEL 356 X-RAY FILM PROCESSOR, MODELS MD & MD-D · Fischer Industries, Inc. · Sep 30, 2002
- K023825 — FILM PROCESSOR, MODEL LIFE RAY PRO 2500 · S.A.I.E.P. S.R.L. · Feb 12, 2003
- K023613 — AX200 AUTOMATIC X-RAY FILM PROCESSOR, MODEL 6586.036 · Alphatek Corp. · Jan 23, 2003
- K024342 — RADIOGRAPHIC FILM PROCESSOR, MODELS XP400 AND XP1000 · Daito Co., Ltd. · Apr 23, 2003