EX3 STEPPER, CLASSIC STEPPER, MULTI-PURPOSE WORKSTATION STEPPER; MICRO-TOUCH, MICRO-TOUCH LP, MULIT-PURPOSE WORKSTATION,

K131161 · CIVCO Medical Instruments Co., Inc. · ITX · Sep 17, 2013 · Radiology

Device Facts

Record IDK131161
Device NameEX3 STEPPER, CLASSIC STEPPER, MULTI-PURPOSE WORKSTATION STEPPER; MICRO-TOUCH, MICRO-TOUCH LP, MULIT-PURPOSE WORKSTATION,
ApplicantCIVCO Medical Instruments Co., Inc.
Product CodeITX · Radiology
Decision DateSep 17, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2

Intended Use

Intended for use in ultrasonic procedures related to brachytherapy, transperineal template-guided biopsy, and/or fiducial marker placement (including volume determination of the prostate gland), and/or the application of radionuclide source(s) into the body during brachytherapy.

Device Story

System comprises Stepping Unit, Stabilizer, and Disposable Template Grid for prostate procedures. Stepping Unit holds/manipulates ultrasound probes; provides manual linear/rotational positioning. Stabilizer provides base/fixation for Stepping Unit on OR table or floor stand. Disposable Template Grid provides needle guidance via 5mm-spaced channels. EX3 Stepper variant includes encoders powered via USB-hub; firmware reports longitudinal/rotational position to external treatment planning software. Used in clinical settings by physicians for brachytherapy, biopsy, fiducial marker placement, and cryotherapy. Output (positional data) assists clinicians in accurate needle placement/therapy delivery, potentially improving procedural precision and patient outcomes.

Clinical Evidence

No clinical data. Substantial equivalence supported by literature review confirming safe/effective use in cryotherapy, biopsy, and fiducial marker placement, and non-clinical performance testing (verification of EX3 Stepper motion, encoder accuracy, and electrical safety per IEC 60601-1).

Technological Characteristics

Stepping Unit: manual linear/rotational positioning. Stabilizer: table/floor mount. Disposable Template Grid: ABS plastic, 5mm channel spacing, EtO sterilized. EX3 Stepper: USB-powered encoders, firmware-based position reporting. Electrical safety: IEC 60601-1.

Indications for Use

Indicated for adult males with known or suspected prostate cancer.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K131161 Page 1 of 6 Image /page/0/Picture/1 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the word "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font. WWW.CIVCO.COM # Section 5 – 510(k) Summary SEP 17 2013 ## A. Submitter Information | | Submitter Name & Address: CIVCO Medical Instruments Co., Inc. d/b/a CIVCO Medica<br>Solutions<br>102 First Street South<br>Kalona, lowa 52247 | |-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Amanda Stahle, Regulatory Affairs Specialist<br>Telephone: 319-248-6628, Fax: 877-218-0324<br>amanda.stahle@civco.com | | Date Summary Prepared: | April 22, 2013 | | Trade Names: | EX3 Stepper™, Classic Stepper, Multi-Purpose<br>Workstation™ Stepper;<br>Micro-Touch™, Micro-Touch™ LP, Multi-Purpose<br>Workstation™, Multi-Purpose Workstation™ LP;<br>Disposable Template Grid<br>(These devices are marketed as part of the AccuCARE™<br>product line) | | Common Names: | Stepping Unit<br>Stabilizer<br>Disposable Template Grid | | Classification Names: | Transducer, Ultrasonic, Diagnostic<br>System, Applicator, Radionuclide, Manual<br>System, Applicator, Radionuclide, Remote-controlled | | Classification Numbers: | Class It under 21 CFR 892.1570<br>Class I under 21 CFR 892.5650<br>Class II under 21 CFR 892.5700 | | Review Panels: | Radiology | | Product Codes: | ITX, IWJ & JAQ | ## B. Predicate Devices The subject devices are substantially equivalent to the predicate devices included in the following 510(k)s: | Predicate Devices | | Mfg. | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|------------------------------------------------------------------------------------------------| | K972672: Brachystepper Stepping Unit, Brachystand Support and<br>Manual Adjustment Accessory<br>Note: The Brachystepper Needle Guide Template component in this<br>510(k) is not being used as a predicate. | | Barzell-Whitmore Maroon<br>Bells, Inc. (acquired by<br>CIVCO Medical<br>Instruments Co., Inc.) | | Corporate Headquarters | 102 First Street South, Kalona, IA 52247 USA | P: 319.248.6757 F: 319.248.6660 | | Coralville Office | 2301 Jones Blvd., Coralville, IA 52241 USA | P: 319.248.6757 F: 319.248.6660 | | Europe Office | Pasteurstraat 6, 2811 DX Reeuwijk, The Netherlands | P: +31(0) 182.394495 F: +31(0) 182.395014 | | Orange City Office | 1401 8th Street SE, Orange City, IA 51041 USA | P: 712.737.8688 F: 712.737.8654 | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the word "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font with wide spacing between the letters. The overall design is clean and professional. WWW.CIVCO.COM | Predicate Devices | Mfg. | |--------------------------------------------------|------------------------------------------------------------------------------------------------| | K981873: Brachystepper™ Disposable Template Grid | Barzell-Whitmore Maroon<br>Bells, Inc. (acquired by<br>CIVCO Medical<br>Instruments Co., Inc.) | The subject devices, with the exception of the EX3 Stepper, have been marketed for many years under K972672 and K981873. The purpose of this 510(k) is to modify the intended use of these devices to include cryotherapy, transperineal template-guided biopsy, and fiducial marker placement. These additional applications fall within the indications for use and the literature demonstrates that these devices have been safely and effectively used for many years in these applications. These changes do not raise new issues of safety and effectiveness because the system continues to be used for positioning in males with suspected or diagnosed prostate cancer. These changes do not impact or modify the therapy and CIVCO did not modify the design of these devices to enable use in these additional applications. The second purpose of this 510(k) is to provide a premarket notification for the EX3 Stepper. The EX3 Stepper is similar to the predicate in that it holds an ultrasound imaging probe and facilitates manual linear and rotational positioning of the probe. The EX3 Stepper differs from the predicate in that it offers continuous (free) longitudinal movement and contains encoder components that are powered via a USB-hub. Firmware in the encoders report position and this position is read by treatment planning software marketed by other companies. This position reporting function has been added to the EX3 intended use. Verification and validation testing confirmed that these changes in design do not raise new questions of safety and effectiveness. ## C. Device Description ## Stepping Unit The Stepping Unit holds an ultrasound imaging probe and facilitates manual linear and rotational positioning of the probe. The Stepping Unit consists of a Cradle designed to hold a specific ultrasound probe and a Carriage that moves longitudinally along the Stepping Unit. A Grid Platform connected at one end of the Stepper provides support for the Disposable Template Grid. The Stepping Unit connects to the Stabilizer. #### Stabilizer The Stabilizer provides a base for the Stepping Unit and offers fixation and support during insertion and final placement of the ultrasound imaging probe. On certain models, a fine tune mechanism is used to micro-adjust the probe or instrument to the ideal orientation. The Stabilizer mounts to an operating room table or is supported by a floor stand. #### Disposable Template Grid The Disposable Template Grid is a single-use, sterile grid that consists of rows and columns of holes (channels) spaced 5mm apart. These channels are labeled and provide placement of needles in predefined areas of the prostate. | Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | | 319.248.6757 | | 319.248.6660 | |------------------------|------------------------|-----------------------|-----------------|--|-------------------|--|-------------------| | Coralville Office | 2301 Jones Blvd. | Coralville, IA 52241 | USA | | 319.248.6757 | | 319.248.6660 | | Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | | +31(0) 182.394495 | | +31(0) 182.395014 | | Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | | 712.737.8688 | | 712.737.8654 | {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe-like graphic on the left, followed by the word "CIVCO" in large, bold letters. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in smaller letters. ## www.civco.com . The following part numbers are included in this 510(k): . 1 | Common<br>Name | Trade Name | Part No. | Device Description | |------------------|------------------------------------------|---------------------|----------------------------------------------------------------------------------| | Stepping<br>Unit | EX3<br>Stepper™ | 609-004 | EX3 Stepper for use with Best Sonalis<br>transducers | | | | 612-228 | EX3 Stepper for use with Acuson ER7B<br>transducers | | | | 614-119 | EX3 Stepper for use with Aloka UST-672-5/7.5<br>transducers | | | | 620-119 | EX3 Stepper for use with B-K Medical 8658<br>(8558), 8848 transducers | | | | 642-443 | EX3 Stepper for use with GE Healthcare ERB<br>transducers | | | | 644-081 | EX3 Stepper for use with Hitachi EUP-U533<br>transducers | | | | 676-177 | EX3 Stepper for use with Siemens Endo P-II<br>transducers | | | | 683-003 | EX3 Stepper for use with Terason 8B4S<br>transducers | | | Classic<br>Stepper | 614-092 | Classic Stepper for use with Aloka UST-672-5/7.5<br>transducers | | | | 644-064 | Classic Stepper for use with Hitachi EUP-U533<br>transducers | | | | 612-225 | Classic Stepper for use with Acuson ER7B<br>transducers | | | | 642-316 | Classic Stepper for use with GE Healthcare ERB<br>transducers | | | | 620-089 | Classic Stepper for use with B-K Medical 8658<br>(8558) transducers | | | | 676-114 | Classic Stepper for use with Siemens Endo-P II<br>transducers | | | | 609-001 | Classic Stepper for use with Best Sonalis<br>transducers | | | Multi-Purpose<br>Workstation™<br>Stepper | 614-098 | Multi-Purpose Workstation Stepper for use with<br>Aloka UST-672-5/7.5 | | | | 642-334 | Multi-Purpose Workstation Stepper for use with<br>GE Healthcare ERB | | | | 644-066 | Multi-Purpose Workstation Stepper for use with<br>Hitachi EUP-U533 | | | | 676-121 | Multi-Purpose Workstation Stepper for use with<br>Siemens Endo-P II | | | | 620-094 | Multi-Purpose Workstation Stepper for use with B-<br>K Medical 8658 (8558), 8808 | | | | 620-113 | Multi-Purpose Workstation Stepper for use with<br>BK Medical 8808e, 8818 | | | | 620-117 | Multi-Purpose Workstation Stepper for use with<br>BK Medical 8848 | | Stabilizer | Micro-Touch ™ | 610-911<br>610-911S | Micro-Touch (Dual-Sided Table Mount)<br>Micro-Touch Transportation Stand | | Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | 319.248.6757 | 319.248.6660 | |------------------------|------------------------|-----------------------|-----------------|-------------------|-------------------| | Coralville Office | 2301 Jones Blvd. | Coralville, IA 52241 | USA | 319.248.6757 | 319.248.6660 | | Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | +31(0) 182.394495 | +31(0) 182.395014 | | Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | 712.737.8688 | 712.737.8654 | {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the word "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller font size. The logo is black and white. | | | | WWW.CIVCO.COM | |--------------------------------|--------------------------------------|----------|----------------------------------------------------------------------------------| | | Micro-TouchTM<br>LP | 610-912 | Micro-Touch LP (Single-Sided Table Mount-right<br>side) | | | | 610-912S | Micro-Touch Transportation Stand | | | | 610-913 | Micro-Touch LP (Single-Sided Table Mount-left<br>side) | | | | 610-913S | Micro-Touch Transportation Stand | | | Multi-Purpose<br>WorkstationTM | 610-974 | Multi-Purpose Workstation with adjustable floor<br>stand | | | Multi-Purpose<br>WorkstationTM<br>LP | 610-975 | Multi-Purpose Workstation LP with rail mount | | Disposable<br>Template<br>Grid | Disposable<br>Template Grid | 610-905 | Sterile 17GA Grid for use with AccuCARE<br>Positioning and Stabilizing equipment | | | | 610-906 | Sterile 18GA Grid for use with AccuCARE<br>Positioning and Stabilizing equipment | | | | 610-915 | Sterile 17GA Grid for use with B-K Medical UA-<br>1084 Stepper | | | | 610-916 | Sterile 18GA Grid for use with B-K Medical UA-<br>1084 Stepper | ### D. Indications for Use This device is indicated for use in adult males with known or suspected prostate cancer. #### E. Intended Use #### System Intended Use: Intended for use in ultrasonic procedures related to brachytherapy, transperineal template-guided biopsy, and/or fiducial marker placement (including volume determination of the prostate gland), and/or the application of radionuclide source(s) into the body during brachytherapy. #### Component-specific Intended Uses: Stepping Unit (Multi-Purpose Workstation Stepper, Classic Stepper): Holding and manipulating ultrasound imaging probes during prostate brachytherapy. transperineal template-guided biopsy, and/or fiducial marker placement procedures (including volume determination of the prostate gland), and/or the application of radionuclide source(s) into the body during brachytherapy. Stepping Unit (EX3 Stepper): Holding and manipulating ultrasound imaging probes, and reporting position, during prostate brachytherapy, transperineal templatequided biopsy, and/or fiducial marker placement procedures (including volume determination of the prostate gland), and/or the application of radionuclide source(s) into the body during brachytherapy. Stabilizer: Provides fixation, support and manipulation of transrectal ultrasound imaging probes during insertion and final placement. | Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | 319.248.6757 | 319.248.6660 | |------------------------|------------------------|-----------------------|-----------------|-------------------|-------------------| | Coralville Office | 2301 Jones Blvd. | Coralville, IA 52241 | USA | 319.248.6757 | 319.248.6660 | | Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | +31(0) 182.394495 | +31(0) 182.395014 | | Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | 712.737.8688 | 712.737.8654 | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, followed by the word "CIVCO" in a bold, sans-serif font. Below the word "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font. The logo is black and white. ### WWW.CIVCO.COM Disposable Template Grid: Accepting and guiding needles up to 1.3 mm (18 gauge) in diameter and providing coordinates as an aid to needle loading and positioning during radioactive seed implantation, cryotherapy, transperineal template-guided biopsy, and/or fiducial marker placement. ## F. Technological Characteristics ### Stepping Unit Both the subject and predicate Stepping Units are designed to hold and manipulate ultrasound imaging probes during prostate procedures and enable manual linear and rotational positioning of the probe. The subject and predicate Stepping Units consist of a Cradle designed to hold and rotate an ultrasound probe, a Carriage that moves longitudinally along the Stepping Unit, and a Grid Platform connected at one end of the Stepper to attach the Disposable Template Grid. The proposed EX3 Stepper differs from the predicate in that it offers continuous (free) longitudinal movement of the Carriage and contains encoder components that are powered via a USB-hub. Firmware in the encoders report position and this position is read by treatment planning software marketed by other companies. ## Stabilizer Both the subject and predicate Stabilizers are designed to serve as a base for the Stepping Unit. The subject and predicate Stabilizers offer fixation and support during insertion and final placement of the ultrasound imaging probe. Whereas the predicate device only mounted on a floor stand, several of the subject Stabilizers can be mounted to an operating table. The fine tune adjustment mechanism on the predicate device is incorporated in the Micro-Touch Stabilizers, but not the Multi-Purpose Workstation Stabilizers. ## Disposable Template Grid Both the subject and predicate Disposable Template Grids are single-use, sterile grids that consist of rows and columns of holes (channels) spaced 5mm apart. These channels are labeled and provide placement of needles in predefined areas of the prostate. The subject device is sterilized via EtO gas, whereas the predicate device was sterilized by gamma irradiation. Also, the subject device is made of ABS while the predicate device was constructed from polycarbonate. The subject device is labeled with a 3 year expiration date. ## G. Literature Review , Non-Clinical Performance Testing, and Conclusions A literature review was conducted to support the modification of the intended use to include cryotherapy, transperineal template-guided biopsy, and fiducial marker placement. The literature demonstrates that these devices have been safely and effectively used in cryotherapy, transperineal template-guided biopsy, and fiducial marker placement. No design changes were made to enable use of the devices in these additional applications. | Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | | <span style="font-family: sans-serif;">Ⓟ</span> 319.248.6757 | | <span style="font-family: sans-serif;">Ⓕ</span> 319.248.6660 | |------------------------|------------------------|-----------------------|-----------------|--|-------------------------------------------------------------------|--|-------------------------------------------------------------------| | Coralville Office | 2301 Jones Blvd. | Coralville, IA 52241 | USA | | <span style="font-family: sans-serif;">Ⓟ</span> 319.248.6757 | | <span style="font-family: sans-serif;">Ⓕ</span> 319.248.6660 | | Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | | <span style="font-family: sans-serif;">Ⓟ</span> +31(0) 182.394495 | | <span style="font-family: sans-serif;">Ⓕ</span> +31(0) 182.395014 | | Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | | <span style="font-family: sans-serif;">Ⓟ</span> 712.737.8688 | | <span style="font-family: sans-serif;">Ⓕ</span> 712.737.8654 | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for CIVCO Medical Solutions. The logo features a stylized globe on the left, with curved lines forming a grid pattern. To the right of the globe is the word "CIVCO" in bold, sans-serif font. Below "CIVCO" are the words "MEDICAL SOLUTIONS" in a smaller, sans-serif font. K13 1161 Page 6 of 6 ## www.civco.com Additionally, non-clinical performance testing was conducted on the EX3 Stepper including verification testing on the following characteristics: - Stepping Unit is able to provide continuous (free) longitudinal movement ● - Encoders correctly report longitudinal and rotational motion of Stepping Unit . - Stepping Unit meets electrical safety requirements of IEC 60601-1 and IEC . 60601-1-1 Validation was also conducted to confirm the device remains safe and effective for its intended use and included testing with treatment planning software. All testing confirmed that the EX3 Stepper is substantially equivalent to the predicate device in regards to safety and effectiveness and the new design features have not diminished the safety and effectiveness of the device. | Corporate Headquarters | 102 First Street South | Kalona, IA 52247 | USA | 319.248.6757 | 319.248.6660 | |------------------------|------------------------|-----------------------|-----------------|-------------------|-------------------| | Coralville Office | 2301 Jones Blvd. | Coralville, IA 52241 | USA | 319.248.6757 | 319.248.6660 | | Europe Office | Pasteurstraat 6 | 2811 DX Reeuwijk | The Netherlands | +31(0) 182.394495 | +31(0) 182.395014 | | Orange City Office | 1401 8th Street SE | Orange City, IA 51041 | USA | 712.737.8688 | 712.737.8654 | {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract human figures connected by flowing lines, representing the department's mission to protect the health of all Americans. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 17, 2013 CIVCO Medical Instruments Co., Inc. d/b/a CIVCO Medical Solutions % Ms. Amanda Stahle Regulatory Affairs Specialist 102 First Street South KALONA IA 52247 Re: K131161 Trade/Device Name: Ex3 Stepper" Classic Stepper, Multi-Purpose Workstation " Stepper, Micro-Touch", Micro-Touch" LP, Multi-Purpose Workstation ", Multi-Purpose Workstation™ LP; and Disposable Template Grid Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: ITX, IWJ, JAQ Dated: August 28, 2013 Received: August 29, 2013 Dear Ms. Stahle: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {7}------------------------------------------------ Page 2-Ms. Stahle forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, for Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications for Use 510(k) Number (if known): K131161 Device Name: Stepping Unit Stabilizer Disposable Template Grid Indications for Use: This device is indicated for use in adult males with known or suspected prostate cancer. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Sm27) (Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health K131161 510(k)_ Page 1 of
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...