The SURGIVISIO device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during which the spine, pelvis or articulation structures are visualized.
Device Story
Mobile C-arm X-ray system providing 2D fluoroscopy and intraoperative 3D imaging; consists of mobile C-arm and viewing workstation. System uses X-ray tube and flat panel detector to capture images; integrates Computer Aided Surgery (CAS) for instrument positioning and stereotaxic guidance. New 3D Offset Imaging Set SPX1 accessory allows 10 cm offset of imaging area from standard configuration. Used in OR by surgeons/physicians; workstation displays images for intraoperative guidance. Output supports surgical decision-making and instrument placement. Benefits include improved visualization and navigation of anatomical structures during orthopedic/surgical procedures.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including design verification of the Offset Phantom SPX1, system-level accuracy verification, software verification, and packaging/shelf-life validation.
Technological Characteristics
Mobile C-arm with rotating anode X-ray tube (0.3-0.8 focal spot) and flat panel detector (287mm x 265mm). kV range 40-120; mA range 1-120. Connectivity via LAN/DICOM. Software compliant with IEC 62304. Accessories (phantoms/wedges) constructed of PETG and Tyvek 1073B. Integrated CAS interface.
Indications for Use
Indicated for surgical procedures requiring visualization of 2D or intraoperative 3D medical imaging of high x-ray attenuation structures (bony anatomy, metallic objects), specifically spine, pelvis, or articulation structures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
April 15, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Ecential Robotics % Mathilde Saulpic Quality Assurance & Regulatory Affairs Engineer Zone Mayencin II, Parc Equation - Bâtiment 1, 2 avenue de Vignate Gieres, 38610 FRANCE
Re: K220946
Trade/Device Name: SURGIVISIO Device Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA Dated: March 29, 2022 Received: April 1, 2022
Dear Mathilde Saulpic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220946
Device Name SURGIVISIO Device
#### Indications for Use (Describe)
The SURGIVISIO device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures during which the spine, pelvis or articulation structures are visualized.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="font-size:100%;"> <span style="font-family:Arial, sans-serif;"> </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## Submitter Information
| Submitter: | ECENTIAL ROBOTICS<br>2 avenue de Vignate<br>Zone Mayencin II, Parc Equation – Bâtiment 1<br>38610 Gières<br>France |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Mathilde SAULPIC<br>Quality Assurance & Regulatory Affairs Engineer<br>Phone : 0033 4 58 00 35 86<br>Email: mathilde.saulpic@ecential-robotics.com |
| Date Summary Prepared: | 08 April 2022 |
{4}------------------------------------------------
## Summary of the technological characteristics of the device compared to the predicate device
There have been no modifications on indication for use on SURGIVISIO Device since the previously approved submission K202547 & K220627.
This submission aims at adding a new Imaging set: 3D Offset Imaging Set SPX1 accessory, which enables to deport imaging zone regarding the already cleared 3D Imaging Set SPX1 accessory, giving user more flexibility by freeing space around the Patient reference SPX1 and the zone of interest. The new 3D Offset Imaging Set SPX1 accessory will not replace the already cleared 3D Imaging Set SPX1 accessory.
## Device Information
| Trade or proprietary name: | SURGIVISIO Device |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or usual name: | Mobile Interventional Fluoroscopic X-ray System |
| Classification Name: | Interventional Fluoroscopic X-Ray System |
| Regulation Number: | 21 CFR 892.1650 - Image-intensified fluoroscopic x-ray system |
| Regulatory class: | II |
| Primary product code: | OWB |
| Legally marketed device to which equivalence is claimed: | K202547 & K220627 - SURGIVISIO device - Manufacturer: ECENTIAL ROBOTICS (previously SURGIVISIO) |
| Device Description: | The SURGIVISIO Device is a mobile x-ray system which provides 2D imaging and allows the generation of intraoperative 3D information of high contrast objects and anatomical structures. The system consists of two mobile interconnected units: a mobile C-arm and a mobile viewing Workstation. These units are moved manually and are interconnected by a single cable that provides power and transfer of data and controls. The mobile C-Arm comprises a high voltage generator, foot switches for radiation release, laser target devices, electronics cabinet, collision avoidance system, and a C-shaped structure mounting the X-ray tube assembly and the flat X-ray detector on distal end of the 'C'. The mobile viewing workstation comprises a computer, an image detector process unit, the main power supply, radiation indicator, dual viewing monitors and a user interface for patient management and image handling. |
{5}------------------------------------------------
The system integrates a Computer Aided Surgery (CAS) feature that supports instruments positioning during surgical procedures.
The SPX1 Instruments are a set of dedicated accessories intended to be used in conjunction with the SURGIVISIO Device in order to enable stereotaxic guidance. More specifically, and when used with the 3D imaging, they support the registration of patient anatomy with the 3D reconstructed image. the Submissions K202547/ K220627 consider SPX1 instruments set with the standard 3D Imaging Set SPX1 accessory.
This submission aims at clearing a new imaging set accessory possibility: the 3D Offset Imaging Set SPX1 accessory, containing same patient reference and wedges as 3D imaging Set SPX1 accessory already 510(k) cleared, and an offset phantom (new instrument). The principle of use of the Offset Phantom SPX1 is identical to the already 510(k) cleared Phantom SPX1 (belonging to 3D Imaging set). However, the Offset Phantom SPX1 additional value is to offset the imaging area that will be navigated during surgery by 10 cm from the standard configuration. The 3D Imaging Set SPX1 accessory will be considered as the predicate for the 3D Offset Imaging Set SPX1 accessory.
- Indication for use: The SURGIVISIO device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during which the spine, pelvis or articulation structures are visualized. Indication for use is the same as the predicate.
- Technology comparison Same as the predicate. The technology is not impacted by the discussed device change, object of this Special 510(K) submission. This is an x-ray system that generates 2D and 3D images of anatomical structures. The 3D Offset Imaging Set accessory is used in combination with the imaging system in the same way than the 3D Imaging Set SPX1 accessory.
- Biocompatibility Same as the predicate. The discussed device change, object of this Special 510(K) submission does not require to re-perform Biocompatibility testing as the raw material, manufacturing processes are the same for 3D Imaging Set SPX1 accessory than for 3D Offset Imaging Set SPX1 accessory.
{6}------------------------------------------------
Electrical Safety Testing Same as the predicate. The discussed device change, object of this Special 510(K) submission does not impact Electrical Safety.
Performance Testing Same as the predicate.
Software Testing Same as the predicate.
> Software change to enable the identification of the Offset Phantom SPX1 were already taken into account in submission K220627. The clearance of the present Special 510(k) would enable to release the associated functionalities by clearing the associated required instrument.
SURGIVISIO software was verified and validated according to IEC 62304 Medical Device Software. Results of these validations and verifications confirm that the SURGIVISIO software complies with its specifications.
- Clinical images Same as the predicate. The discussed device change, object of this Special 510(K) submission does not impact clinical images.
- Conclusion Based upon comparison of devices and performance testing results, SURGIVISIO Device using 3D Offset Imaging Set SPX1 accessory is substantially equivalent to the predicate device.
{7}------------------------------------------------
| Devices | SURGIVISIO Device used<br>with 3D Offset Imaging<br>Set SPX1 accessory | SURGIVISIO Device<br>(K202547 & K220627)<br>used with 3D Imaging<br>Set SPX1 accessory |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use / Indication for use | The SURGIVISIO device is<br>intended to be used during<br>surgical procedures in<br>which the physician would<br>benefit from the<br>visualization of 2D medical<br>imaging and/or<br>intraoperatively generated<br>3D medical imaging of<br>anatomical structures or<br>objects with high x-ray<br>attenuation such as bony<br>anatomy or metallic<br>objects. Such procedures<br>include procedures during<br>which the spine, pelvis or<br>articulation structures are<br>visualized. | The SURGIVISIO device is<br>intended to be used during<br>surgical procedures in<br>which the physician would<br>benefit from the<br>visualization of 2D medical<br>imaging and/or<br>intraoperatively generated<br>3D medical imaging of<br>anatomical structures or<br>objects with high x-ray<br>attenuation such as bony<br>anatomy or metallic<br>objects. Such procedures<br>include procedures during<br>which the spine, pelvis or<br>articulation structures are<br>visualized. |
| Primary product code | OWB | OWB |
| Mechanical configuration | Mobile C-Arm | Mobile C-Arm |
| Movement range | Vertical: Up to 43.5cm<br>Horizontal: Up to 28cm<br>Orbital: 200°<br>Angulation: 180° | Vertical: Up to 43.5cm<br>Horizontal: Up to 28cm<br>Orbital: 200°<br>Angulation: 180° |
| Movement control | Motor-driven | Motor-driven |
| kV Range | 40 - 120 kV | 40 - 120 kV |
| mA Range | 1 – 120 mA | 1 – 120 mA |
| Pulse frequency | 1 - 12.5 fps | 1 - 12.5 fps |
| Detector technology | Flat panel, 287mm x 265mm | Flat panel, 287mm x 265mm |
| Image Matrix Size | 1560x1440 pixels<br>780x720 pixels | 1560x1440 pixels<br>780x720 pixels |
| X-ray tube technology | Rotating anode<br>0.3 - 0.8 focal spot | Rotating anode<br>0.3 - 0.8 focal spot |
| 2D Imaging | 2D Fluoroscopic | 2D Fluoroscopic |
| Pulsed Fluoroscopy | Yes | Yes |
| AERC system | Yes, kv/mA curve types | Yes, kv/mA curve types |
| 3D Imaging | Yes | Yes |
| Rotating movement for 3D<br>imaging | 180° | 180° |
| Devices | SURGIVISIO Device used<br>with 3D Offset Imaging<br>Set SPX1 accessory | SURGIVISIO Device<br>(K202547 & K220627)<br>used with 3D Imaging<br>Set SPX1 accessory |
| 3D Imaging characteristics | Cylindrical volume: 15 x<br>Ø13 cm<br>Elliptic cylindrical volume:<br>15 x Ø116 x Ø218 cm<br>Resolution 400 x 400 x 400<br>voxels | Cylindrical volume: 15 x<br>Ø13 cm<br>Elliptic cylindrical volume:<br>15 x Ø116 x Ø218 cm<br>Resolution 400 x 400 x 400<br>voxels |
| Monitor cart/Workstation | Yes | Yes |
| Screen displays | 2 Monitors 22" tactile<br>Screen Displays | 2 Monitors 22" tactile<br>Screen Displays |
| Image output format | DICOM | DICOM |
| LAN network connection | Yes | Yes |
| Computer Aided<br>Surgery<br>(CAS) interface | Yes, integrated | Yes, integrated |
| 3D Imaging set accessory<br>components | Offset Phantom SPX1 with<br>its ID card<br>Patient reference SPX1<br>with its ID card<br>Associated wedges | Phantom SPX1 with its ID<br>card<br>Patient reference SPX1<br>with its ID card<br>Associated wedges |
| 3D Imaging set accessory<br>components packaging | Offset Phantom SPX1 and<br>patient reference with<br>associated wedges<br>packaged in one unique<br>blister<br><br>Blister Material and<br>reference:<br>PETG & Tyvek 1073B | Phantom SPX1 and<br>patient reference<br>packaged in its own Blister<br>Patient reference SPX1<br>with associated wedges<br>packaged in its own Blister<br><br>Blister Material and<br>reference:<br>PETG & Tyvek 1073B |
| Software workflow | 3D Spine Universal<br>workflow | 3D Spine Universal<br>workflow |
{8}------------------------------------------------
{9}------------------------------------------------
## Performance Data
For further information about testing, standards, acceptance criteria and results, please refer to section Design Control Activities Summary.
Nonclinical tests: The following nonclinical tests were performed on the SURGIVISIO Device using 3D Offset Imaging Set SPX1 accessory to demonstrate substantial equivalence of safety and efficacy with the predicate device:
- -Design verification activities have been performed for the Offset Phantom SPX1
- -Accuracy verification has been conducted at the system level with the 3D Offset Imaging Set SPX1 accessory as intended
- -Software verification testing verifying the software requirements has been performed as intended and are already taken into account by the previous 510(k) cleared version (Submission K220627).
- Packaging process validation and verification activities, including Shelf-life validation, have been conducted as intended
Clinical tests: No clinical tests were conducted to demonstrate substantial equivalence.
## Conclusions drawn from Performance Data
The SURGIVISIO Device using 3D Offset Imaging Set SPX1 accessory is identical in indications for use and technological characteristics as the proposed predicate device. These aspects, along with the performance testing conducted, demonstrate the substantial equivalence to the SURGIVISIO Device (K202547/ K220627) and that the SURGIVISIO Device using 3D Offset Imaging Set SPX1 accessory does not raise different questions of safety and effectiveness when compared to this predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.