StentBoost Mobile is intended for use as a vascular x-ray interventional application. StentBoost Mobile provides high image quality visualization of stents in relation to vessels. StentBoost Mobile assists in the treatment of endovascular diseases by visualizing the placement and deployment of stents.
Device Story
StentBoost Mobile is a software application (iApp) designed for vascular x-ray interventional procedures. It processes 2D image sequences transmitted from a digital X-ray imaging system to provide high-resolution visualization of stents relative to vessels. The device operates on Philips Zenition series mobile surgery interventional X-ray systems. It is controlled by physicians in the exam room via a mobile view station, stand user interface, or touchscreen module. By enhancing stent visualization, the device assists clinicians in the accurate placement and deployment of stents during endovascular interventions, potentially improving procedural outcomes.
Clinical Evidence
No clinical data. Substantial equivalence was demonstrated through non-clinical performance testing, including software verification and validation, compliance with international consensus standards (IEC 62304, IEC 62366-1, IEC 82304-1, ISO 14971, ISO 15223-1), and assessment of safety and effectiveness.
Technological Characteristics
Software-based image processing application. Operates on Philips Zenition series mobile surgery interventional X-ray systems. Connectivity via digital X-ray imaging system integration. Complies with IEC 62304 (software life cycle), IEC 62366-1 (usability), IEC 82304-1 (health software safety), ISO 14971 (risk management), and ISO 15223-1 (symbols).
Indications for Use
Indicated for use as a vascular x-ray interventional application to assist in the treatment of endovascular diseases by visualizing stent placement and deployment.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
April 5, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
Philips India Limited % Shruti Sancheti Regulatory Approbation Officer Plot No. B-79, MIDC, Phase-II, Chakan, Taluka-Khed, Village - Savardari Pune, Maharashtra 410501 INDIA
## Re: K203623
Trade/Device Name: StentBoost Mobile Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, LLZ Dated: March 2, 2021 Received: March 8, 2021
## Dear Shruti Sancheti:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203623
Device Name StentBoost Mobile
Indications for Use (Describe)
StentBoost Mobile is intended for use as a vascular x-ray interventional application.
StentBoost Mobile provides high image quality visualization of stents in relation to vessels.
StentBoost Mobile assists in the treatment of endovascular diseases by visualizing the placement and deployment of stents.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:flex; align-items:center;"> <input checked="true" type="checkbox"/> <span>For Investigative Testing/STP Validation Study</span> </div> | <div style="display:flex; align-items:center;"> <input type="checkbox"/> <span>For In-Situ Testing/STP Validation Study</span> </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K203623 510(k) Summary
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR \$807.92.
| 21 CFR §807.92. | Date Prepared: | December 8, 2020 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Manufacturer: | Philips India Limited<br>Plot No. B-79, MIDC, Phase-II, Chakan<br>Taluka - Khed, Village - Savardari<br>District: Pune<br>Maharashtra 410501<br>India<br>Establishment Registration Number: 3010685285 | |
| Primary Contact<br>Person: | Shruti Sancheti<br>Regulatory Approbation Officer<br>Phone: +91 9518546814<br>E-mail: shruti.sancheti@philips.com | |
| Secondary Contact<br>Person: | Neena Sonavane<br>Senior RA Manager, Q&R R&C Affairs - IGT<br>Phone: +91 8446020023<br>E-mail: neena.sonavane@philips.com | |
| Device: | Trade Name: | StentBoost Mobile |
| | Classification Name: | Interventional fluoroscopic x-ray system |
| | Classification Regulation: | 21CFR §892.1650 |
| | Classification Panel: | Radiology |
| | Regulatory Class: | Class II |
| | Product Code: | Primary Code: OWB<br>Subsequent Codes: LLZ |
| Predicate Device: | Trade Name: | StentBoost Rel. 4 |
| | Manufacturer: | Philips Medical Systems Nederland B.V. |
| | 510(k) Clearance: | K132305 (October 30, 2013) |
| | Classification Name: | Image intensified fluoroscopic x-ray system |
| | Classification Regulation: | 21CFR §892.1650 |
| | Classification Panel: | Radiology |
| | Device Class: | Class II |
| | Product Code: | Primary Code: OWB<br>Subsequent Code: LLZ |
{4}------------------------------------------------
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
StentBoost Mobile is a software product (iApp) intended to provide a high-Device description: resolution visualization of stents in vessels. It supports the physician in placing and deploying stents. Indications for Use: StentBoost Mobile is intended for use as a vascular x-ray interventional application. StentBoost Mobile provides high image quality visualization of stents in relation to vessels. StentBoost Mobile assists in the treatment of endovascular diseases by visualizing the placement and deployment of stents. The indications for use of StentBoost Mobile is a subset of StentBoost Rel. 4, as StentBoost Mobile is intended for use as a vascular application only, and, StentBoost Rel. 4 is used for vascular or cardiovascular applications. Based on the information provided above. StentBoost Mobile is considered substantially equivalent to the currently marketed and predicate device StentBoost Rel. 4 in terms of Indications for Use. Comparison of The fundamental scientific technology provided in the subject device StentBoost Technological Mobile and predicate device StentBoost Rel. 4 is same based on the following: characteristics with . Image processing algorithm: the Predicate Device: . Execution on and connection to a software hosting platform/ workstation; Transmission of 2D image sequence from a digital X-ray imaging system . to the software hosting platform; Display of image series on reference monitor screen. ● The differences between StentBoost Mobile and the predicate device are as follows: . Predicate device StentBoost Rel. 4 runs on the independent hosting software functionality platform of the currently marketed Philips Interventional Workspot (K121296), whereas, the subject device StentBoost Mobile runs on compatible platform, which is available on Philips Zenition series mobile surgery interventional X-ray systems (K183101, K183040). Change in the user interface and visualization components between the . predicate and the subject device. StentBoost Rel. 4 can be controlled from both the control room and exam . room. In the exam room. StentBoost Rel. 4 can be operated via the touchscreen module of the Allura X-ray system. On the contrary, StentBoost Mobile can be controlled only from the compatible Zenition series mobile surgery systems, via the Mobile view station, Stand user interface and/or Touch screen module (option).
The differences between the proposed StentBoost Mobile and the predicate device StentBoost rel. 4 do not raise any new questions regarding safety or
{5}------------------------------------------------
effectiveness. Based on the information provided in this 510(k) submission, the proposed StentBoost Mobile is considered substantially equivalent to the currently marketed predicate StentBoost rel. 4 in terms of fundamental scientific technology.
#### Performance Data: The non-clinical performance testing has been performed on StentBoost Mobile and demonstrates compliance with the following FDA recognized consensus standards:
- . IEC 62304 Medical device software – Software life cycle processes (Edition 1.1, 2015-06). FDA/CDRH recognition number 13-79.
- . IEC 62366-1 Medical devices - Part 1: Application of usability engineering to medical devices (Edition 1.0, 2015-02). FDA/CDRH recognition number 5-114.
- IEC 82304-1 Health software Part 1: General requirements for product . safety (Edition 1.0 2016-10), FDA/CDRH recognition number 13-97.
- . ISO 14971 Medical devices – Application of risk management to medical devices (Edition 2.0, corrected 2007). FDA/CDRH recognition number 5-40.
- . ISO 15223-1 Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements (Third Edition, 2016-11-01). FDA/CDRH recognition number 5-117.
Software verification testing of the functional requirements as well as performance and safety has been performed to verify that all the requirements of System Requirements Specification as well as the safety risk control measures from the Detailed Risk Management Matrix and the Privacy and Security requirements and mitigations have been implemented. Results demonstrated that all executed verification tests were passed.
Non-clinical validation testing has been performed to validate that StentBoost Mobile conforms to the intended use, claims, user needs, effectiveness of safety measures and instructions for use. All these tests were used to support substantial equivalence of the subject device and demonstrate that StentBoost Mobile:
- . complies with the aforementioned international and FDA-recognized consensus standards, and
- . meets the acceptance criteria and is adequate for its intended use.
Therefore, StentBoost Mobile is substantially equivalent to the predicate device StentBoost Rel. 4 in terms of safety and effectiveness.
{6}------------------------------------------------
The proposed StentBoost Mobile did not require clinical study since substantial Clinical Performance Data: equivalence to the currently marketed and predicate device StentBoost was demonstrated with the following attributes:
- Indication for use; ●
- . Technological characteristics;
- Non-clinical performance testing; and ●
- Safety and effectiveness. .
StentBoost Mobile is substantially equivalent to the currently marketed predicate Conclusion: device StentBoost Rel. 4 (K132305) in terms of indications for use, fundamental scientific technology (image processing algorithm), and safety and effectiveness. Additionally, substantial equivalence was demonstrated by non-clinical performance tests provided in this 510(k) premarket notification. These tests demonstrate that StentBoost Mobile complies with the requirements specified in the international and FDA-recognized consensus standards and is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.