The SURGIVISIO medical device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during which the spine, pelvis or articulation structures are visualized.
Device Story
Mobile C-arm X-ray system providing 2D fluoroscopy and intraoperative 3D imaging. System comprises mobile C-arm (X-ray tube, flat panel detector, generator, collision avoidance) and mobile viewing workstation (dual monitors, computer, user interface). Operated by clinicians in surgical settings. Inputs: X-ray projections during manual or motor-driven C-arm rotation. Processing: Iterative 3D reconstruction algorithm transforms 2D projections into 3D volumes. Outputs: 2D fluoroscopic images and 3D anatomical reconstructions displayed on workstation monitors. Integrated Computer Aided Surgery (CAS) feature supports instrument positioning. Benefits: Intraoperative visualization of bony anatomy/metallic objects to guide surgical procedures; supports spine, pelvis, and articulation surgeries.
Clinical Evidence
No clinical tests were conducted. Evidence consists of bench testing: IEC 60601 series compliance, 2D image quality assessment, CBCT phantom testing (IEC 60601-2-44), distortion phantom concurrence study, iterative 3D reconstruction algorithm characterization, 3D image quality with breathing simulation, expert evaluation of sample clinical images, and radiation dose comparisons to literature.
Technological Characteristics
Mobile C-arm with flat panel detector (287mm x 265mm). Rotating anode X-ray tube (0.3–0.8 focal spot). 40–120 kV range. Motor-driven movement. Dual 22" tactile monitors. DICOM output. LAN connectivity. Iterative 3D reconstruction algorithm. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-43, 60601-2-54, and 60601-2-44.
Indications for Use
Indicated for surgical procedures requiring visualization of 2D or intraoperative 3D images of high x-ray attenuation structures (e.g., bone, metal), specifically spine, pelvis, or articulation structures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
March 26, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Surgivisio % Elodie Bouillet Quality and Regulatory Affairs Engineer Zone Mayencin II, Parc Equation - Bâtiment 1 2 Avenue de Vignate Gières, 38610 FRANCE
### Re: K202547
Trade/Device Name: Surgivisio system Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA Dated: February 23, 2021 Received: March 1, 2021
### Dear Elodie Bouillet:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K202547
Device Name Surgivisio system
#### Indications for Use (Describe)
The SURGIVISIO medical device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during which the spine, pelvis or articulation structures are visualized.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div style="display:flex; align-items:center;"><div style="margin-right:5px;">☑</div>Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|-----------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><div style="margin-right:5px;">☐</div>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K202547
# Submitter Information
| Submitter: | Surgivisio<br>Zone Mayencin II, Parc Equation – Bâtiment 1<br>2 avenue de Vignate<br>38610 Gières<br>France |
|------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Vincent Leré<br>Quality/Regulatory Affairs Director<br>Phone : 33-4-58 00 57 21<br>Email: vincent.lere@surgivisio.com |
| Date Summary Prepared: | July 31st, 2020 |
### Device Information
| Trade or proprietary name: | Surgivisio system |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or usual name: | Mobile Interventional Fluoroscopic X-ray System |
| Classification Name: | Interventional Fluoroscopic X-Ray System |
| Regulation Number: | 21 CFR 892.1650 - Image-intensified fluoroscopic x-ray system |
| Regulatory class: | II |
| Primary product code: | OWB |
| Legally marketed device to which equivalence is claimed: | K042793 - Arcadis Orbic (available with options 3D and 3D navigation interface) - Manufacturer: Siemens Medical Solutions, Inc |
| Device Description: | The Surgivisio system is a mobile x-ray system which provides 2D imaging and allows the generation of intraoperative 3D information of high contrast objects and anatomical structures.<br>The system consists of two mobile interconnected units: a mobile C-arm and a mobile viewing Workstation. These units are moved manually and are interconnected by a single cable that provides power and transfer of data and controls. |
{4}------------------------------------------------
The mobile C-arm comprises a high voltage generator, foot switches for radiation release, laser target devices, electronics cabinet, collision avoidance system, and a C-shaped structure mounting the X-ray tube assembly and the flat X-ray detector on distal ends of the 'C'.
The mobile viewing workstation comprises a computer, an image detector process unit, the main power supply, radiation indicator, dual viewing monitors and a user interface for patient management and image handling.
The system integrates a Computer Aided Surgery (CAS) feature that supports instruments positioning during surgical procedures.
- Indication for use: The SURGIVISIO medical device is intended to be used during surgical procedures in which the physician would benefit from the visualization of 2D medical imaging and/or intraoperatively generated 3D medical imaging of anatomical structures or objects with high x-ray attenuation such as bony anatomy or metallic objects. Such procedures include procedures during which the spine, pelvis or articulation structures are visualized.
### Summary of the technological characteristics of the device compared to the predicate device
The Surgivisio system is substantially equivalent to the Arcadis Orbic (available with options 3D and 3D navigation interface) - K042793
| Devices | Surgivisio system | The Arcadis Orbic with options<br>3D and 3D navigation interface<br>(K042793) |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use /<br>Indication for<br>use | The SURGIVISIO medical device is<br>intended to be used during surgical<br>procedures in which the physician<br>would benefit from the visualization<br>of 2D medical imaging and/or<br>intraoperatively generated 3D<br>medical imaging of anatomical<br>structures or objects with high x-ray<br>attenuation such as bony anatomy<br>or metallic objects. Such<br>procedures include procedures<br>during which the spine, pelvis or<br>articulation structures are<br>visualized. | The Arcadis Orbic is a mobile X-ray<br>system designed to provide<br>fluoroscopic and digital spot-film<br>imaging of the patient during<br>surgical and interventional<br>procedures. Clinical application<br>may include, but are not limited to,<br>cholangiography, endoscopic,<br>urologic, pain management,<br>orthopedic, neurologic, vascular,<br>cardiac, critical care and<br>emergency room procedures.<br>The Arcadis Orbic 3D option<br>provides 3D imaging and is<br>intended to be used whenever the<br>physician benefits form<br>intraoperatively-generated 3D<br>information of high contrast objects<br>and anatomical structures. |
| Device | Surgivisio system | The Arcadis Orbic with options 3D and 3D navigation interface (K042793) |
| Primary product code | OWB | OWB |
| Mechanical configuration | Mobile C-Arm | Mobile C-Arm |
| Movement range | Vertical: Up to 43.5cm<br>Horizontal: Up to 28cm<br>Orbital: 200°<br>Angulation: 180° | Vertical: Up to 40cm<br>Horizontal: Up to 20cm<br>Orbital: 190°<br>Angulation: 190° |
| Movement control | Motor-driven | Manual<br>Motor-driven |
| kV Range | 40 – 120 kV | 40-110 kV |
| mA Range | 1 – 120 mA | 0.2 – 15.2 mA, pulsed up to 23 mA<br>mAs up to 150 |
| Pulse frequency | 1 – 12.5 fps | Up to 15 fps |
| Detector technology | Flat panel, 287mm x 265mm | X-ray Image intensifier Ø23cm |
| Image Matrix Size | 1560x1440 pixels<br>780x720 pixels | 1024x1024 pixels |
| X-ray tube technology | Rotating anode<br>0.3 – 0.8 focal spot | Stationary anode<br>0.6 focal spot |
| 2D Imaging | 2D Fluoroscopic | 2D Fluoroscopic |
| Pulsed Fluoroscopy | Yes | Yes |
| AERC system | Yes, kv/mA curve types | Yes, kv/mA curve types |
| 3D Imaging | Yes | Yes |
| Rotating movement for 3D imaging | 180° | [120° - 190°] |
| 3D Imaging characteristics | Cylindrical volume: 15 x Ø13 cm<br>Elliptic cylindrical volume: 15 x Ø16 x Ø218 cm<br>Resolution 400 x 400 x 400 voxels | Cubic volume: 11 x 11 x 11 cm |
| Monitor cart/Workstation | Yes | Yes |
| Screen displays | 2 Monitors 22" tactile Screen Displays | 2 Monitors 19" Screen Displays |
| Image output format | DICOM | DICOM |
| LAN network connection | Yes | Yes |
| Computer Aided Surgery (CAS) interface | Yes, integrated | Yes, provides the 3D navigation interface that enables interoperability with navigation |
{5}------------------------------------------------
{6}------------------------------------------------
# Performance Data
#### The following nonclinical tests were performed on the Surgivisio system Nonclinical tests: to demonstrate substantial equivalence of safety and efficacy with the predicate device:
- IEC 60601-1 (Edition 3.1): Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 (Edition 4.0): Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-3 (Edition 2.1): Medical electrical equipment Part 1-3 : General requirements for basic safety and essential performance - Collateral Standard : Radiation protection in diagnostic X-ray equipment
- IEC 60601-2-43 (Edition 2.1): Medical electrical equipment Part 2-43: Particular requirements for basic safety and essential performance of X-ray equipment for interventional procedures
- IEC 60601-2-54 (Edition 1.1): Medical electrical equipment Part 2-54: Particular i requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy [Including: Amendment 2 (2018)]
- 2D Image Quality Assessment of the Surgivisio system provides a quantitative image । quality assessment of subject device in comparison to the predicate device
- । Basic CBCT bench testing using phantoms and based on standard IEC 60601-2-44 (Edition 3.2) Medical electrical equipment - Part 2-44: Particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography. to demonstrate the effectiveness and safety of the subject device
- । Concurrence study using a distortion phantom to demonstrate the ability of the subject device to provide 3D images equivalent to those of the predicate device
- Quality characterization of the iterative 3D reconstruction algorithm to demonstrate the effectiveness of the 3D imaqing feature
- 3D image quality assessment with breathing simulation to demonstrate the i effectiveness of the 3D imaging feature
- Sample clinical images evaluated by a qualified expert to support the intended use of the 2D and 3D imaging features
- । Comparison of radiation doses associated with clinical image data provided by the subject device to literature
- Software Verification testing verifying the software requirements perform as intended -
Clinical tests: No clinical tests were conducted to demonstrate substantial equivalence.
{7}------------------------------------------------
# Conclusions drawn from Performance Data
The Surgivisio system is similar in indications for use and technological characteristics as the proposed predicate device. These aspects, along with the performance testing conducted, demonstrate the substantial equivalence to the Arcadis Orbic system (K042793) and that the Surgivisio system does not raise different questions of safety and effectiveness when compared to this predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.