K201172 · Mx Imaging, Inc. · OWB · May 29, 2020 · Radiology
Device Facts
Record ID
K201172
Device Name
CFP-3131, CFP-2222
Applicant
Mx Imaging, Inc.
Product Code
OWB · Radiology
Decision Date
May 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
Intended for use in radiographic fluoroscopic applications including cardiar, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
Device Story
CMOS flat panel detectors designed as components for legacy surgical C-Arm systems; function as image intensifier replacements. Detectors capture X-ray signals using CMOS imaging sensors; provide real-time imaging for fluoroscopy/pulsed fluoroscopy and radiography. Interface with host X-ray systems via connectors and synchronize with X-ray generators via cables. Installed in standard image intensifier-based systems using mechanical adapters; provide plug-compatible outputs without requiring host system modifications. Calibration software provided for setup. Healthcare providers use output images for diagnostic and interventional guidance; enables continued use of legacy C-arm hardware with modern digital detector technology.
Clinical Evidence
No clinical data. Substantial equivalence determined via bench testing and performance evaluation.
Technological Characteristics
CMOS flat panel X-ray detectors. CFP-3131 (327x377 mm, 3096x3096 pixels); CFP-2222 (235x292 mm, 2170x2170 pixels). AC line powered. Standards: ANSI/AAMI ES60601-1:2005+C1:2009+A2:2020+A1:3012, IEC 60601-1-2:2007, 21 CFR 1020.30/31/32. Connectivity: Wired interface to host X-ray system and generator.
Indications for Use
Indicated for radiographic and fluoroscopic applications, including cardiac, vascular, general diagnostic, and interventional x-ray imaging.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
CS-series-FP with MX CFP 3131 or MX CFP 2222 Option (K171755)
Submission Summary (Full Text)
{0}------------------------------------------------
May 29, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an eagle emblem, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA" in a larger font size and a blue square behind it.
MX Imaging, Inc. % Mr. Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
Re: K201172
Trade/Device Name: CFP-3131, CFP-2222 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA Dated: April 29, 2020 Received: May 1, 2020
Dear Mr. Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201172
Device Name CFP-3131, CFP-2222.
Indications for Use (Describe)
Intended for use in radiographic fluoroscopic applications including cardiar, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text '510(k) Summary K201172' in bold, black font. The text is centered on the image and is underlined. The text appears to be a title or heading for a document or section of a document.
Image /page/3/Picture/1 description: The image shows a logo with the letters "MX" in a stylized font. The "M" is black, and the "X" is blue. There is a white oval shape in the upper right corner of the "X". The logo is simple and modern.
MX Imaging Inc. 2894 Columbia Street Torrance CA 90503-3808 US 310-381-3800
#### Administrative Information 1.
Reason for Submission: 510(k) Notification for CFP-3131, CFP-2222. Submitter: MX Imaging Inc. Submission contact person: John Ross john-ross@mximaging.com Contact telephone: 310-381-3800 May 27, 2020 Date prepared:
| 2. Identification: | CFP-3131, CFP-2222. |
|----------------------------|---------------------------------------------|
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1650 |
| Device Class: | Class II |
| Product Code: | OWB, JAA, |
#### 3. Substantially equivalent device:
| Trade Name: | CS-series-FP with MX CFP 3131 or MX CFP 2222 Option |
|----------------------------|-----------------------------------------------------|
| Manufacturer: | Omega Medical Imaging, LLC |
| 510(k) #: | K171755 |
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR §892.1650 |
| Device Class: | Class II |
| Product Code: | OWB, JAA, |
| | |
## 4. Device description: These CMOS X-ray detectors feature:
The MX Imaging CMOS Flat Panel Detectors are designed for medical x-ray imaging applications. The CMOS panels are real-time x-ray detectors. The Detectors use a state of the art CMOS Imaging sensor that is based on a special pixel architecture enabling the detectors to be used in high sensitivity applications such as x-ray fluoroscopy/pulsed fluoroscopy, as well as applications requiring a large dynamic x-ray range such as radiography serial and radiography applications.
The CMOS Flat Panel Detector is a component, and forms the part of a complete fluoroscopic x-ray system. The CMOS Detector interfaces to the Host X-Ray System through the associated interface connectors on the flat panel. The CMOS Flat Panel Detector is also synchronized to the X-Ray Generator via the interface cables.
The intended use for this device is for Image Intensifier replacement in legacy surgical C-Arm systems.
{4}------------------------------------------------
While we have tested the device with the GE/OEC 9800 and 9900 series of C-Arms it would be appropriate to install them in any standard Image Intensifier based system with the proper mechanical adapter (Weight compensated) and signal plug adaptations. The system provides a "Plug Compatible" device with the same outputs as the target system. For these reasons no modification of the host system is needed or recommended. In the case of the GE/OEC 9800 the existing input/output connect to the provided detector connectors without modification. Calibration software and instructions are provided with the detector. The existing software and controls all work as before.
- 5. Indications for Use: Intended for use in radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
| Comparable<br>Properties | K171755 Omega CS-series-FP<br>with MX CFP 3131 or MX CFP<br>2222 Option<br>radiographic/fluoroscopy<br>system | CFP-3131, CFP-2222. K201172 | Comparison<br>Results |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Indications for<br>use | Intended for use in<br>radiographic/fluoroscopic<br>applications including cardiac,<br>vascular, general<br>radiographic/fluoroscopic<br>diagnostic, and interventional x-<br>ray imaging | Intended for use in<br>radiographic/fluoroscopic<br>applications including cardiac,<br>vascular, general<br>radiographic/fluoroscopic<br>diagnostic, and interventional<br>x-ray imaging | Identical |
| Digital X-Ray<br>Detectors | MX CFP 3131 and MX CFP 2222 | CFP-3131, CFP-2222. | SAME |
| Detector Sizes | CFP 3131 – 327 x 377 mm<br>CFP2222 – 235 x 292 mm | CFP-3131 – 327 x 377 mm<br>CFP-2222 – 235 x 292 mm | SAME |
| Pixel Resolutions | CFP-3131 3096x3096<br>CFP-2222 2170x2170,<br>(1x1 binning) | CFP-3131 3096x3096<br>CFP-2222 2170x2170,<br>(1x1 binning) | SAME |
| Power Source | AC Line | AC Line | SAME. |
| Standards | Same as below | See below | SAME |
## 6. Technological characteristics: Comparison Table
## 5. Non clinical testing: Testing was performed successfully according to the following standards:
| Standard<br>Developing<br>Organization | Standard Designation<br>Number And Date | Title of Standard |
|----------------------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| ANSI/AAMI | ANSI/AAMI ES60601-1:<br>2005 +C1:2009<br>+A2:2020 +A1:3012 | Medical Electrical Equipment Part 1: General Requirements<br>For Basic Safety And Essential Performance |
{5}------------------------------------------------
| Standard<br>Developing<br>Organization | Standard Designation<br>Number And Date | Title of Standard |
|----------------------------------------|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC | 60601-1-2:2007 | Medical Electrical Equipment Part 12: General Requirements<br>For Basic Safety And Essential Performance Collateral<br>Standard: Electromagnetic Disturbances Requirements And<br>Tests |
| Food and Drug<br>Administration | 21CFR 1020.30,<br>1020.31, and 1020.32<br>Current Version | Meets the applicable requirements of the US FDA Radiation<br>Safety Performance Standard |
Non-clinical testing was performed in accordance with the FDA Guidance Document Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices, Guidance for Industry and Food and Drug Administration Staff Document issued on: September 1, 2016. In recognition of possible cybersecurity threats to the software, we consulted this guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Druq Administration Staff October 2014. As a result, we updated our own internal standard operating procedures and added cybersecurity precautions to the software users' manuals. Also we have utilized FDA's pediatric guidance, "Pediatric Information for X-ray Imaging Device Premarket Notifications" in the preparation of our labeling documentation. Conforms to applicable portions of the US Radiation Safety Performance Standard, conformance with applicable portions of 21 CFR 1020.30, 1020.31, and 1020.32.
6. Clinical testing. Not required for a determination of substantial equivalence.
# 7. Substantial Equivalence Discussion.
When combined with compatible c-arm combination the CFP 3131 - CFP2222 - performs the same functions using the same technological methods to produce diagnostic fluoroscopic x-ray images. In all material aspects, the Omega and the MX Imaging systems are substantially equivalent to each other.
## 8. Substantial Equivalence Conclusion:
After analyzing bench test results, risk analysis, and clinical evaluation, it is the conclusion of MX Imaging LLC that the CFP 3131 - CFP2222 - is as safe and effective as the predicate device, has few technological differences, and has the same indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.