K173314 · Nu Vasive, Incorporated · OWB · Nov 16, 2017 · Radiology
Device Facts
Record ID
K173314
Device Name
LessRay System
Applicant
Nu Vasive, Incorporated
Product Code
OWB · Radiology
Decision Date
Nov 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
LessRay® System is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
Device Story
LessRay is a software application interfacing with a fluoroscope via video cable. It captures, enhances, and displays fluoroscopic images on a computer monitor to improve clarity, contrast, and noise reduction, particularly for low-dose/pulsed imaging. The system includes C-arm tracking and instrument tracking capabilities. A tracker attached to an instrument allows the system to track its location relative to prior baseline x-rays, aiding the user in instrument positioning. The system provides visual cues to guide C-arm positioning. It is used in clinical settings by physicians and technicians. The device is passive, relying on the fluoroscope's video output without transmitting signals back to the fluoroscope. It serves as an auxiliary display to assist in surgical interventions, potentially allowing for lower radiation dose usage while maintaining diagnostic image quality.
Clinical Evidence
Bench testing only. Performance was demonstrated through verification of instrument tracking, alignment accuracy, image registration, and software validation. Electrical safety and EMC testing were performed per IEC 60601-1 and IEC 60601-1-2. No clinical data was provided.
Technological Characteristics
Software-based image processing system. Interfaces with fluoroscope via video cable. Hardware includes computer (Intel Core 2 Duo, NVIDIA Quadro 4000, 8GB RAM, 256GB HDD), frame grabber (Aver Media H339 or Elgato), and optical tracking hardware. Operates on Windows 8.1. Passive device; no signal transmission to fluoroscope. Image processing uses averaging algorithms for noise reduction and contrast/brightness enhancement.
Indications for Use
Indicated for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease. No specific patient population, age, or gender restrictions are defined.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
November 16, 2017
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
NuVasive, Incorporated % Manthan Damani Senior Regulatory Affairs Specialist 7475 Lusk Boulevard SAN DIEGO CA 92121
Re: K173314
Trade/Device Name: LessRay® System Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, LLZ, and JAA Dated: October 17, 2017 Received: October 19, 2017
Dear Manthan Damani:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
(1-800-638-2041 or 301-796-7100).
Sincerely,
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K173314
Device Name LessRay® System
Indications for Use (Describe)
LessRay® System is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left, with the word "NUVASIVE" in gray letters to the right of the leaf. The leaf shape is split into two parts, with the top part being purple and the bottom part being gray.
# 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
#### Submitted by: A.
Manthan J. Damani Senior Regulatory Affairs Specialist NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800
Date Prepared: October 17, 2017
#### B. Device Name
| Trade or Proprietary Name: | LessRay® System |
|----------------------------|---------------------------------------------|
| Common or Usual Name: | Image processing system |
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Device Class: | Class II |
| Classification: | 21 CFR § 892.1650 |
| Product Code: | OWB, LLZ, JAA |
#### Predicate Devices C.
The subject device is substantially equivalent to the predicate devices LessRay® with Tracking (K142243) and LessRay® with Enhanced Tracking (K170800).
#### D. Device Description
LessRay is a software application which can be interfaced to a fluoroscope with a video cable. The images produced by the fluoroscope are transmitted to a frame grabber in the computer running LessRay where the images are enhanced and then displayed. When used in connection with the low dose and/or pulse setting on the fluoroscope, the user can improve the quality (clarity, contrast, noise level, and usability-) of a noisy (low-quality) image. Using this system, much of the graininess of low radiation dose images can be eliminated. This allows for greater utility of low dose imaging.2 LessRay provides the additional feature of being able to interface LessRay with a tracking system in order to aid the C-arm technician in positioning the fluoroscope between the various views of the patient necessary for the intervention. LessRay with Tracking ensures that the fluoroscope is centered over the correct anatomy prior to taking any additional x-ray images.
LessRay System has additional capability of instrument tracking to aid the user in positioning an instrument using prior baseline x-rays. A tracker is attached to the instrument and as the instrument moves, the tracking system connected to LessRay tracks the location of the instrument. LessRay System uses this information to aid the user in positioning the instrument.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for NuVasive. The logo consists of a stylized leaf shape on the left, with the word "NUVASIVE" in capital letters to the right of the leaf. The leaf is colored purple on the top and gray on the bottom. The word "NUVASIVE" is gray.
The purpose of this premarket notification is addition of a wired remote to the LessRay System and the integration of instrument tracking functionality previously cleared in K170800.
LessRay System is designed per recommendations provided in the following FDA guidance documents:
- o Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- Off-The-Shelf Software Use in Medical Devices ●
- o Content of Premarket Submissions for Management of Cybersecurity in Medical Devices
#### Indications for Use E.
LessRay® System is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
#### F. Comparison of Technological Characteristics with Predicate Device
As was established in this submission, the subject LessRay System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent to its predicate device through comparison in areas including design, labeling/intended use, and function.
| Specifi<br>cation/<br>Proper<br>ty | Predicate Device<br>LessRay with Tracking<br>(K142243) | Predicate Device<br>NuVasive LessRay with<br>Enhanced Tracking<br>(K170800) | Subject Device<br>LessRay System | Discus<br>sion |
|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Intende<br>d Use /<br>Indicati<br>ons for<br>Use | LessRay® with Tracking is<br>intended for use in any<br>application where a<br>fluoroscope is incorporated<br>to aid in diagnosis and<br>treatment of disease. | NuVasive® LessRay® with<br>Enhanced Tracking is<br>intended for use in any<br>application where a<br>fluoroscope is incorporated<br>to aid in diagnosis and<br>treatment of disease. | LessRay® System is<br>intended for use in any<br>application where a<br>fluoroscope is<br>incorporated to aid in<br>diagnosis and treatment<br>of disease. | Same |
| Device<br>Class | II | II | II | Same |
| Produc<br>t Code | OWB, JAA, LLZ | OWB, JAA, LLZ | OWB, JAA, LLZ | Same |
| Regula<br>tion<br>Numbe<br>r<br>(21CF<br>R) | §892.1650 | §892.1650 | §892.1650 | Same |
| Device<br>Classifi<br>cation<br>Name | Interventional Fluoroscopic<br>X-Ray System | Interventional Fluoroscopic<br>X-Ray System | Interventional<br>Fluoroscopic X-Ray<br>System | Same |
| Specifi<br>cation/<br>Proper<br>ty | Predicate Device<br>LessRay with Tracking<br>(K142243) | Predicate Device<br>NuVasive LessRay with<br>Enhanced Tracking<br>(K170800) | Subject Device<br>LessRay System | Discus<br>sion |
| Device<br>Functio<br>nalities<br>(image<br>acquisi<br>tion,<br>enhanc<br>ement,<br>and<br>display<br>) | Software based device<br>used to provide computer<br>display systems interfaced<br>to a fluoroscope through a<br>video cable. Images<br>produced by fluoroscope<br>are transmitted through a<br>cable to a frame capture<br>board in computer where<br>images are enhanced and<br>then displayed on the<br>monitor.<br><br>Enhanced images are<br>displayed on computer<br>monitor at the same time<br>that corresponding original<br>image is displayed on<br>fluoroscope monitor(s).<br><br>Uses software to control<br>frame capture,<br>intermediate image<br>manipulation, and image<br>display.<br><br>Delay between frame<br>acquisition and frame<br>display is less than the<br>time between frames, i.e.<br>33 msec.<br><br>Serves only as an image<br>display which is in<br>addition to the<br>fluoroscope's standard<br>image display device.<br>Device is passive, in that<br>the operation depends only<br>on the video output of the<br>fluoroscope, and it does<br>not transmit any signals or<br>images to the fluoroscope. | Software based device used<br>to provide computer<br>display systems interfaced<br>to a fluoroscope through a<br>video cable. Images<br>produced by fluoroscope<br>are transmitted through a<br>cable to a frame capture<br>board in computer where<br>images are enhanced and<br>then displayed on the<br>monitor.<br><br>Enhanced images are<br>displayed on computer<br>monitor at the same time<br>that corresponding original<br>image is displayed on<br>fluoroscope monitor(s).<br><br>Uses software to control<br>frame capture, intermediate<br>image manipulation, and<br>image display.<br><br>Delay between frame<br>acquisition and frame<br>display is less than the time<br>between frames, i.e. 33<br>msec.<br><br>Serves only as an image<br>display which is in addition<br>to the fluoroscope's<br>standard image display<br>device. Device is passive,<br>in that the operation<br>depends only on the video<br>output of the fluoroscope,<br>and it does not transmit any<br>signals or images to the<br>fluoroscope. | Software based device<br>used to provide<br>computer display<br>systems interfaced to a<br>fluoroscope through a<br>video cable. Images<br>produced by<br>fluoroscope are<br>transmitted through a<br>cable to a frame capture<br>board in computer<br>where images are<br>enhanced and then<br>displayed on the<br>monitor.<br><br>Enhanced images are<br>displayed on computer<br>monitor at the same<br>time that corresponding<br>original image is<br>displayed on<br>fluoroscope monitor(s).<br><br>Uses software to<br>control frame capture,<br>intermediate image<br>manipulation, and<br>image display.<br><br>Delay between frame<br>acquisition and frame<br>display is less than the<br>time between frames,<br>i.e. 33 msec.<br><br>Serves only as an<br>image display which is<br>in addition to the<br>fluoroscope's standard<br>image display device.<br>Device is passive, in<br>that the operation<br>depends only on the<br>video output of the<br>fluoroscope, and it<br>does not transmit any<br>signals or images to<br>the fluoroscope. | Same |
| | | Page 3 of 6 | | |
| Specifi<br>cation/<br>Proper<br>ty | Predicate Device<br>LessRay with Tracking<br>(K142243) | Predicate Device<br>NuVasive LessRay with<br>Enhanced Tracking<br>(K170800) | Subject Device<br>LessRay System | Discus<br>sion |
| Algorit<br>hms | - Image quality<br>improvement using<br>averaging algorithm<br>- Contrast and brightness<br>enhancement with<br>simultaneous reduction<br>of random noise | - Image quality<br>improvement using<br>averaging algorithm<br>- Contrast and brightness<br>enhancement with<br>simultaneous reduction of<br>random noise | - Image quality<br>improvement using<br>averaging algorithm<br>- Contrast and brightness<br>enhancement with<br>simultaneous reduction<br>of random noise | Same |
| C-arm<br>Trackin<br>g | - When tracking is<br>enabled, will<br>automatically choose the<br>Baseline when the<br>fluoroscope is near the<br>location and orientation<br>that the Baseline was<br>initially taken.<br>- When tracking is<br>enabled, requires<br>hardware components in<br>order to mount the off-<br>the-shelf tracking<br>hardware to the C-arm<br>and to the operating<br>table.<br>- When tracking is<br>enabled, requires the use<br>of an off-the-shelf<br>tracking system in order<br>to track the 6 DOF<br>location of the C-arm<br>relative to the operating<br>table.<br>- When tracking is<br>enabled, visual cues are<br>provided which help<br>guide the user in<br>positioning the C-arm<br>back to where a prior<br>Baseline was taken. | - When tracking is enabled,<br>will automatically choose<br>the Baseline when the<br>fluoroscope is near the<br>location and orientation<br>that the Baseline was<br>initially taken.<br>- When tracking is enabled,<br>requires hardware<br>components in order to<br>mount the off-the-shelf<br>tracking hardware to the<br>C-arm and to the<br>operating table.<br>- When tracking is enabled,<br>requires the use of an off-<br>the-shelf tracking system<br>in order to track the 6<br>DOF location of the C-<br>arm relative to the<br>operating table.<br>- When tracking is enabled,<br>visual cues are provided<br>which help guide the user<br>in positioning the C-arm<br>back to where a prior<br>Baseline was taken. | - When tracking is<br>enabled, will<br>automatically choose<br>the Baseline when the<br>fluoroscope is near the<br>location and<br>orientation that the<br>Baseline was initially<br>taken.<br>- When tracking is<br>enabled, requires<br>hardware components<br>in order to mount the<br>off-the-shelf tracking<br>hardware to the C-arm<br>and to the operating<br>table.<br>- When tracking is<br>enabled, requires the<br>use of an off-the-shelf<br>tracking system in<br>order to track the 6<br>DOF location of the<br>C-arm relative to the<br>operating table.<br>- When tracking is<br>enabled, visual cues<br>are provided which<br>help guide the user in<br>positioning the C-arm<br>back to where a prior<br>Baseline was taken. | Same |
| Trackin<br>g<br>options | Electromagnetic or optical | Electromagnetic or optical | Optical | Same |
| Specifi<br>cation/<br>Proper<br>ty | Predicate Device<br>LessRay with Tracking<br>(K142243) | Predicate Device<br>NuVasive LessRay with<br>Enhanced Tracking<br>(K170800) | Subject Device<br>LessRay System | Discus<br>sion |
| Compa<br>tible<br>hardwa<br>re<br>platfor<br>ms | Any computer that meets<br>the following minimum<br>specifications:<br>CPU: Intel Core 2 Duo<br>GPU: NVIDIA Quadro<br>4000<br>RAM: 8 GB<br>HDD: 256 GB<br>Frame Grabber: Aver Media<br>H339 or Elgato<br>Operating System:<br>Windows 7 or 8.1 | Any computer that meets the<br>following minimum<br>specifications:<br>CPU: Intel Core 2 Duo<br>GPU: NVIDIA Quadro 4000<br>RAM: 8 GB<br>HDD: 256 GB<br>Frame Grabber: Aver Media<br>H339 or Elgato<br>Operating System: Windows<br>8.1 | LessRay computer<br>specifications:<br>CPU: Intel Core 2 Duo<br>GPU: NVIDIA Quadro<br>4000<br>RAM: 8 GB<br>HDD: 256 GB<br>Frame Grabber: Aver<br>Media H339 or Elgato<br>Operating System:<br>Windows 8.1 | Same |
| Instru<br>ment<br>Trackin<br>g | No | Yes- NuVasive LessRay<br>with Enhanced Tracking has<br>additional capability of<br>instrument tracking to aid the<br>user in positioning an<br>instrument using prior<br>baseline x-rays. | Yes- LessRay System has<br>additional capability of<br>instrument tracking to aid<br>the user in positioning an<br>instrument using prior<br>baseline x-rays. | Same |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left, with the word "NUVASIVE" in a sans-serif font to the right. The leaf shape is divided into two parts, with the top part in purple and the bottom part in gray. The word "NUVASIVE" is in gray.
# Traditional 510(k) Submission NuVasive® LessRay® System
{6}------------------------------------------------
# Traditional 510(k) Submission NuVasive® LessRay® System
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a stylized leaf shape on the left and the word "NUVASIVE" on the right. The leaf shape is divided into two parts, with the top part being purple and the bottom part being gray. The word "NUVASIVE" is written in a simple, sans-serif font and is also gray.
# Traditional 510(k) Submission Nu Vasive® LessRay® System
#### G. Performance Data
Nonclinical testing was performed to demonstrate that the subject LessRay System is substantially equivalent to the predicate device. The following testing was performed:
- Verification of Instrument Tracking to confirm that subject device allows the user to position the instrument back to where it was located when the desired image is taken.
- Verification of Alignment Accuracy to confirm that the alignment algorithm performs ● according to specifications.
- Verification of Image Registration Performance to confirm that the software performs ● image registration according to specifications.
- Software Validation to validate the functionality and user operations of LessRay System.
- Electrical Safety and Electromagnetic Compatibility (EMC) testing of LessRay ● System per IEC 60601-1 and IEC 60601-1-2 respectively.
The results demonstrate that the subject LessRay System is substantially equivalent to the predicate.
#### H. Conclusions
The subject LessRay System has been shown to be substantially equivalent to legally marketed predicate devices for their intended use.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left and the word "NUVASIVE" on the right. The leaf shape is split into two colors, with the top portion being purple and the bottom portion being gray. The word "NUVASIVE" is written in gray, sans-serif font.
1 As evaluated by a human observer in a side by side visual comparison of 30 image pairs with and without LessRay processing.
2 In clinical practice, the amount of improvement achieved when a Pulsed and/or Low Dose image is processed with LessRay is dependent on the clinical task, patient size, anatomical location, and clinical practice. The dose should be set at a level to which the physician is able to achieve the adequate image quality needed for the particular clinical task. A consultation with a radiologist and a physicist may aid in determining the appropriate dose settings.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.