CS-series-FP with MX CFP 3131 or MX CFP 2222 Option Radiographic/Fluoroscopy System
K171755 · Omega Medical Imaging, LLC · OWB · Feb 9, 2018 · Radiology
Device Facts
Record ID
K171755
Device Name
CS-series-FP with MX CFP 3131 or MX CFP 2222 Option Radiographic/Fluoroscopy System
Applicant
Omega Medical Imaging, LLC
Product Code
OWB · Radiology
Decision Date
Feb 9, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Omega Medical Imaging, LLC CS-series-FP (SSXI) Systems are intended for use in Radiographic/fluroscopic applications including cardiac, vascular, general radiographic/fluroscopic diagnostic, and Interventional x-ray imaging.
Device Story
CS-series-FP is a radiographic/fluoroscopic x-ray system featuring a floor-mounted C-arm and patient table; dual-plane configurations include a ceiling-suspended C-arm. The system incorporates new MX CFP 3131 or MX CFP 2222 CMOS flat-panel detectors (FPD) using cesium iodide scintillators. The system captures x-ray signals, converting them to digital images via an acquisition system separate from the FPD. The acquisition system performs image processing, viewing, and local storage. Operated by clinicians in clinical settings, the system provides real-time imaging to assist in diagnostic and interventional procedures. The device is not internet-connected. Output is displayed for physician review to guide clinical decision-making during procedures, potentially improving diagnostic accuracy and procedural outcomes.
Clinical Evidence
Bench testing only. Performance compared to the existing Omega Medical Imaging CS-series-FP system using commercially available test objects, including low-contrast objects with varying absorbers, spatial and temporal resolution test objects, and dynamic range test objects.
Technological Characteristics
Solid-state CMOS flat-panel detector; cesium iodide scintillator; floor-mounted C-arm (with optional ceiling-suspended C-arm for dual-plane); DICOM compatible; non-networked (no browser). Complies with 21 CFR 1020.30/31/32, IEC 60601-1-2, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, IEC 60601-2-43, UL 60601-1, and CAN/USA C22.2 No.601.1-M90.
Indications for Use
Indicated for patients requiring radiographic/fluoroscopic diagnostic and interventional x-ray imaging, including cardiac and vascular procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
February 9, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The Department of Health and Human Services logo is a stylized representation of an eagle. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Omega Medical Imaging, LLC % Mr. John Newman Manager of Regulatory and Quality 675 Hickman Circle SANFORD FL 32771
Re: K171755
Trade/Device Name: CS-series-FP with MX CFP 3131 or MX CFP 2222 Option radiographic/fluoroscopy system Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA Dated: January 4, 2018 Received: January 10, 2018
Dear Mr. Newman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171755
Device Name
CS-series-FP with MX CFP 3131 or MX CFP 2222 Option Radiographic / Fluoroscopy System
Indications for Use (Describe)
The Omega Medical Imaging, LLC CS-series-FP (SSXI) Systems are intended for use in Radiographic/fluroscopic applications including cardiac, vascular, general radiographic/fluroscopic diagnostic, and Interventional x-ray imaging.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Omega Medical Imaging. The logo consists of a blue circular symbol on the left, resembling the Greek letter omega. To the right of the symbol, the words "OMEGA MEDICAL" are written in a gray sans-serif font, with the word "IMAGING" below it in a smaller font size and separated by horizontal lines.
# Special 510(k) SUMMARY
| Company Name: | Omega Medical Imaging, LLC |
|-----------------------|----------------------------------------------------------------------------------------|
| Address: | 675 Hickman Circle, Sanford, Florida 32771 |
| Telephone No: | 407-323-9400 |
| Registration No: | 1052701 |
| Contact person: | Brian J. Fleming |
| Date Prepared: | 05/26/2017 |
| Device (trade) name: | CS-series-FP with MX CFP 3131 or MX CFP 2222 Option<br>radiographic/fluoroscopy system |
| Common/usual name: | Fluoroscopic/Radiographic X-ray system |
| Classification Name: | Solid State X-ray Imager, Class II, 90 MQB |
| Classification Panel: | Radiology |
| CFR section: | 892.1650 |
| Device Class: | Class II |
| Device Code: | 90 OWB / JAA |
## Predicate Device(s):
- Omega Medical Imaging CS-series-FP (K121293)
## Device description:
The Omega Medical Imaging, LLC, CS-series-FP systems currently incorporate a 19.8cm x 19.8cm or 29.8cm x 29.8cm solid-state flat-panel detector (FPD). This 510(k) submission adds a slightly larger 21.7cm x 21.7cm and 30.3cm x 30.3cm solid-state CMOS flat-panel detector as an additional option. The CS-series-FP fluoroscopy single and dual plane x-ray imaging systems are configured with a floor mounted c-arm and patient table. The dual plane systems incorporate a ceiling suspended C-arm into the system. The MX CFP 3131 flat-panel image detector utilizes a cesium iodide scintillator coupled to an amorphous silicon TFT panel. The captured digital image is processed by the acquisition system (separate from the Flat Panel Detector) which includes image processing, viewing functions, local storage, and DICOM compatibility. The Image Processor does not have the capability to connect to the internet as there is no browser.
Subject Device Flat Panel Detectors MX CFP 2222 / 3131 are to be used only with the Omega CSseries-FP Fluoroscopic Systems
{4}------------------------------------------------
### Indications for use:
- The Omega Medical Imaging, LLC CS-series-FP with Optional MX CFP 3131 and MX CFP 2222 are intended for use in radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
- The intended use of the modified device, as described in the labeling, has not changed as a result of the modification for this filling.
#### Comparison with Predicate Devices:
- · The Omega Medical Imaging MX CFP 2222 and 3131 option the CS-series-FP system utilizes the same technology as the above-mentioned predicate device. The input scintillator is Cesium lodide coupled to a CMOS-light sensitive imaging component. The image processing is achieved with conventional computer based image processing system that is separate from the Flat Panel Detector. The MX CFP 2222 and 3131 Flat Panel Detectors are manufactured by the same company as the Image Processor. It is the opinion of Omega Medical Imaging that the CS-series-FP with the MX CFP 2222 and MX CFP 3131 option is equivalent to the existing Omega Medical Imaging cleared CS-series-FP system.
#### Substantial Equivalence:
- · SE was determined on Bench performance testing. Included in this report is detailed data comparing performance with the existing Omega Medical Imaging CS-series-FP system utilizing the 29.8cm x 29.8cm x 19.8cm x 19.8cm format FPD image acquisition systems. The tests that were performed utilized commercially available Test Object that include lowcontrast objects with varying absorbers simulating different patient sizes, spatial and temporal resolution test objects, and dynamic range test objects.
- Guidance for this submission of 510(k) for (SSXID) Solid State X-ray Imaging Devices issued on: September 1, 2016 was used to establish substantial equivalence
#### Safety information:
- The Omega CS-series-FP with the MX CFP 2222 and MX CFP 3131 option systems comply with the applicable requirements of 21 CFR 1020.30, 21 CFR 1020.31, and 21 CFR 1020.32.
- The Omega CS-series-FP with the MX CFP 2222 and MX CFP 3131 option systems comply with the international safety standards IEC 60601-1-2, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, and IEC 60601-2-43.
- · The Omega CS-series-FP with the MX CFP 2121 and MX CFP 3131 option systems comply with UL 60601-1 and CAN/USA C22.2 No.601.1-M90
{5}------------------------------------------------
# Conclusion:
The Omega CS-series-FP with the MX CFP 2222 and MX CFP 3131 options do not introduce any new indications for use, nor does the use of the device result in any new potential hazard. Omega considers the Omega CS-series-FP with the MX CFP 2222 and MX CFP 3131 Flat panel detectors to be substantially equivalent with the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.