The Ziehm Solo FD is intended for use in providing medical imaging for adults and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides contactless fluoroscopic image capture, temporarily storing, and display of digital subtraction of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholangiography, endoscopic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room fluoroscopy procedures. The visualization of such anatomical structures assists the clinical outcome. At the discretion of a physician, the device may be used for other imaging applications. This device does not support direct radiographic film exposures and is not intended for use in performing mammography.
Device Story
Mobile fluoroscopic C-arm system; utilizes mono-block high voltage generator and flat panel detector (SSXI/CMOS) for 2D medical imaging. Input: X-ray radiation passing through patient anatomy. Transformation: X-ray capture via flat panel detector; processing via Ziehm Adaptive Image Processing (ZAIP) for noise reduction; digital subtraction and cine loop storage. Output: High-resolution fluoroscopic images displayed on integrated monitors. Used in OR, clinic, or emergency settings by physicians/technicians. Provides real-time visualization of anatomical structures to assist in surgical/interventional guidance. Features include manual/motorized C-arm movement, touch screen control, and optional wireless footswitch. Connectivity via DICOM, USB, and DVD. Benefits include improved image quality for moving objects via variable pulse widths/frame rates and optimized surgeon positioning.
Clinical Evidence
Bench testing only. No clinical data. Verification and validation testing included internal functional specifications, software risk hazards, and non-clinical image comparisons against the Ziehm Vision² FD (K073346). Compliance demonstrated for DQE, resolution, dynamic range, beam alignment, dose rates, and frame rates per FDA guidance for solid-state X-ray imaging devices.
Technological Characteristics
Mobile C-arm with mono-block generator (stationary anode). Flat panel detector (SSXI/CMOS, 20.5cm x 20.5cm, 100µm pixel size, 2048x2048 matrix). 16-bit grayscale. ZAIP (Ziehm Adaptive Image Processing) for noise filtering. Connectivity: DICOM, USB, DVD, wireless/wired network. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, IEC 60601-2-54, IEC 60825-1, ISO 14971.
Indications for Use
Indicated for adults and pediatric populations requiring medical imaging via pulsed and continuous fluoroscopic imaging for diagnostic, interventional, and surgical procedures, including pediatric, cholangiography, endoscopic, urologic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room applications. Not for direct radiographic film exposures or mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 6, 2016
Ziehm Imaging Gmbh % Mr. Richard L. Westrich Director of Regulatory Affairs and Quality Assurance Ziehm Imaging, Inc. 6280 Hazeltine National Drive ORLANDO FL 32822
Re: K161976 Trade/Device Name: Ziehm Solo FD Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA, OXO Dated: July 15, 2016 Received: July 18, 2016
Dear Mr. Westrich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161976
Device Name Ziehm Solo FD
#### Indications for Use (Describe)
The Ziehm Solo FD is intended for use in providing medical imaging for adults and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides contactless fluoroscopic image capture, temporarily storing, and display of digital subtraction of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholangiography, endoscopic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room fluoroscopy procedures. The visualization of such anatomical structures assists the clinical outcome. At the discretion of a physician, the device may be used for other imaging applications. This device does not support direct radiographic film exposures and is not intended for use in performing mammography.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for ziehmimaging. The logo consists of a circular graphic to the left of the company name. The company name is written in a teal color and is in a sans-serif font.
U.S. Food and Drug Administration Center for Devices and Radiological Health Document Mail Center -W066-G609 10903 New Hampshire Avenue Silver Spring, MD 20993-0002
# Volume 005 510 (k) Summary
Sep 15, 2016
In accordance with 21 CFR §807.92 the following 510(k) summary information is provided:
| Submitter Address: | Ziehm Imaging GmbH<br>Donaustrasse 31<br>90451 Nuremberg<br>Germany<br>Phone: + 49.911.2172-219<br>Fax: +49.911.2172-390 |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person /<br>Agent: | Richard L. Westrich<br>Director of Requlatory Affairs and Quality Assurance<br>Ziehm Imaging, Inc.<br>Phone: (407) 615-8560 Ext 140<br>Fax: (470) 615-8561<br>Cell Phone: (909) 262 9930 |
| Secondary Contact Person: | Stefan Fiedler<br>Director QM/RA<br>Ziehm Imaging GmbH<br>Phone: + 49.911. 2172-219<br>Fax: +49.911.2172-390 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the text "ziehmimaging" on the right. The text is written in a sans-serif font and is a gradient of teal, with the left side being lighter and the right side being darker.
| Device (Trade Name): | Ziehm Solo FD |
|-----------------------|---------------------------------------------|
| 510(k) Number | K161976 |
| Common /Usual Names: | Mobile Fluoroscopic C-Arm |
| Classification(s) | 21CFR 892.1650 |
| Classification Names: | Image-intensified fluoroscopic x-ray system |
| Device: | Interventional fluoroscopic x-ray system |
| Product Code: | OWB, OXO, JAA |
| Predicate Device: | Ziehm Solo (K092438) |
| Classification(s) | 21CFR 892.1650 |
| Classification Names: | Image-intensified fluoroscopic x-ray system |
| Device: | Interventional fluoroscopic x-ray system |
| Product Code: | OWB, OXO, JAA |
| Predicate Device: | Ziehm Vision² FD (K073346) |
| Classification(s) | 21CFR 892.1650 |
| Classification Names: | Image-intensified fluoroscopic x-ray system |
| Device: | Interventional fluoroscopic x-ray system |
| Product Code: | OWB, JAA, JAK |
General Description: The Ziehm Solo FD mobile fluoroscopy system is comprised of a mobile stand with a C-Profile shaped support with both a mono-block high voltage generator assembly and Flat Panel image receptor. These attach to either end of a C-Profile providing a fixed SID. The device performs 2D medical imaging using 4 axes of manual movement and one vertical axes of motorized movement. Additionally the device has the option of additional motorized axes of rotational movement for the (A-Axes C-profile motorization of the single A-axes provides the user an alternative for visualizing anatomical structures when performing certain procedures such as but not limited to working with lithotripsy procedures. A user touch screen provides for concise user selectable anatomical programs and X-ray technique control. Integrated highresolution flat panel display monitors directly mounted on the horizontal arm of the mobile stand providing the clinician with a precise angle for visualization of live
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a stylized globe with a human figure inside, followed by the text "ziehmimaging" in a teal color. The text is written in a sans-serif font and appears to be slightly faded.
fluoroscopic images of the patient's anatomy. This visualization helps to localize regions of pathology for surqical procedures. The mobile stand supports both a cable bound and optional wireless fluoroscopic footswitch. The Wireless footswitch operation allows for optimum positioning for the surgeon by removing the cable on the floor. The optional interface panel of the Ziehm Solo FD provides connection of peripheral devices such as external monitors, thermal video printers, and image storage devices (USB, DVD) and DICOM fixed wire and wireless network interfaces.
Indications for Use: The Ziehm Solo FD is intended for use in providing medical imaging for adults and pediatric populations, using pulsed and continuous fluoroscopic imaging. The device provides contactless fluoroscopic image capture, temporarily storing, and display of digital subtraction, and acquisition of cine loops during diagnostic, interventional and surgical procedures. Examples of clinical application may include pediatric, cholangiography, endoscopic, urologic, lithotripsy, orthopedic, neurologic, vascular, cardiac, angiographic, critical care, and emergency room fluoroscopy procedures. The visualization of such anatomical structures assists the clinician in the clinical outcome. At the discretion of
a physician, the device may be used for other imaging applications. This device does not support direct radiographic film exposures and is not intended for use in performing mammography.
Summary of Technological The comparisons of the predicate devices show the Characteristics: scientific and technology characteristics of the Ziehm Solo FD are substantial equivalence to that of the predicate devices.
| Differences<br>Features/Technology | Subject Device<br>Ziehm Solo FD<br>(K161976) | Predicate Device<br>Ziehm Solo<br>(K092438) | Predicate Device<br>Ziehm Vision2 FD<br>(K073346) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Mobile fluoroscopic c-arm | Yes | Yes | Yes |
| Classification | 21CFR 892.1650 | 21CFR 892.1650 | 21CFR 892.1650 |
| Mobile stand user touch control<br>panel | Yes, same design | Yes, same design | Yes, same design |
| X-ray generator and tube<br>housing assembly monoblock<br>technology | Yes, same<br>monoblock design,<br>stationary anode<br>tube | Yes, same<br>monoblock design,<br>stationary anode<br>tube | Similar monoblock<br>design,<br>stationary anode<br>tube |
| KV Range 40-110 kV | Yes | Yes | Yes |
| KV Optional Range 40-120 kV | Yes | Yes | No |
| Max power output<br>2.4 kW has the ability to provide<br>higher mA at lower kV levels<br>allowing shorter pulse widths at<br>variable frame rates to increase<br>the image quality of moving<br>objects by reducing or<br>eliminating the movement during | 2.0 kW<br>Optional 2.4 kW | 2.0 kW<br>No | 2.0 kW<br>No |
| each single pulse.<br>Collimator/beam limiter shutters | Yes, same design<br>Asymmetric Slot | No, IRIS with Slot<br>Collimator for<br>circular image<br>receptor | Yes, same design<br>Asymmetric Slot |
| Pre Collimator | Square | Round | Square |
| Virtual collimation | Yes | Yes | Yes |
| X-ray Image Receptor | SSXI/CMOS<br>20.5 cm x 20.5 cm | I.I./cesium iodide<br>Ø 23 cm or Ø 31<br>cm | SSXI/Amorphous<br>silicon<br>20 cm x 20 cm |
| Flat Panel Detector (SSXI) | Yes | No | Yes |
| FPD FOV | Square FOV | Round FOV | Square FOV |
| Pixel Size | $100 \_\text{μm}$ | N/A | $194 \text{ μm}$ |
| Detector Matrix | 2,048 x 2,048 pixels | N/A | 1,024 x 1,024<br>pixels |
| Camera Matrix | NA | 1,024 x 1,024 pixels | NA |
| Grayscale: | 16 bit (65,536<br>shades of gray) | 12 bit (4,096<br>shades of gray) | 16 bit (65,536<br>shades of gray) |
| Dynamic range | 68 dB | N/A for image<br>intensifier devices | 94 dB |
| DQE | 70 % | 65 % | 77 % |
| System resolution | 4 lp/mm | 5.2-6.8 lp/mm | 2.4 lp/mm |
| ZAIP (Ziehm Adaptive Image<br>Processing) | Adaptive noise filter<br>adapts to the<br>physical<br>characteristics of<br>the input signal | No,<br>recursive noise<br>filtering | Adaptive noise<br>filter adapts to the<br>physical<br>characteristics of<br>the input signal |
| Removable anti-scatter grid | Yes | Yes | No |
| AERC Dose control system | Similar design and<br>control | Similar design and<br>control | Similar design<br>and control |
| Image Post Processing | Yes, same design | Yes, same design | Yes, same design |
| DICOM functionality including<br>RSDR | Yes, same design | Yes, same design | Yes, same design |
| TFT Flat Screen Displays | Yes | Yes | Yes |
| Orbital movement | 165°<br>Geometry with new<br>Flat Panel required | 135° | 135° |
| SID | 109 cm<br>Geometry with<br>new Flat Panel<br>requires larger<br>SID | 95 cm | 111 cm |
| Motorized Angulation | Yes, optional | Yes, optional | No |
| Option to dismantle the<br>device in several modules | Variant<br>"Portable" | Variant<br>"Portable" | No |
| Monitor Cart | Optional<br>"Viewing<br>Station" | Optional<br>"Viewing<br>Station" | Mandatory<br>Monitor Cart |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular icon on the left and the text "ziehmimaging" on the right. The icon features a stylized human figure inside a globe-like structure. The text is written in a sans-serif font, with the colors transitioning from a lighter teal to a darker teal.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ziehmimaging. The logo consists of a circular graphic with a stylized human figure inside, surrounded by orbital lines. To the right of the graphic is the text "ziehmimaging" in a teal, sans-serif font.
Summary of Non-Clinical Test Data:
Ziehm Solo FD is based on the direct modifications to cleared predicate devices Ziehm Solo (K092438) and Ziehm Vision2 FD (K073346); The design changes were completed in accordance with Ziehm Imaging GmbH Quality Management System Design Controls and Engineering, standards compliance, and Verification and Validation testing were successfully conducted on the Ziehm Solo FD. Further compliance testing for the modified device to all FDA requirements as stated in "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" as applicable including software requirements and software risk hazards was done.
Tests performed on the Ziehm Solo FD, demonstrated that the device is safe and effective, performs comparably to the predicate devices, and substantially equivalent to the predicate devices.
Tests included verification/validation testing to internal functional specifications (including software) and nonclinical image comparisons involving flat panel display images taken with the new device and the predicate device Ziehm Vision2 FD (K073346). Documentation
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the text "ziehmimaging" on the right. The graphic features a stylized human figure inside a globe-like structure. The text is in a teal color and appears to be a sans-serif font.
provided in this submission demonstrates compliance of the modified device Ziehm Solo FD to all FDA requirements stated in "A Guide for the Submission of Initial Reports on Diagnostic X-Ray Systems and Their Major Components" as applicable. This includes but is not limited to leakage radiation of diagnostic source assembly, peak tube potential (kV), tube current mA, fluoroscopic entrance exposure rates, and beam-limiting alignment to device image receptor. Further, this performance testing confirmed that the Ziehm Solo FD complies with 21 CFR 1020.30-32 Federal Performance Standards for X-Ray Fluoroscopic equipment and with relevant voluntary safety standards for Electrical Safety and Electromaqnetic Compatibility testing, specifically IEC standards listed in the table below. With regard to the flat panel detector (SSXI), test documentation provided in this submission demonstrates compliance of the modified device Ziehm Solo FD to "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices". This includes but is not limited to DQE, resolution, image quality of comparative sets of images, dynamic range, beam alignment, dose rates, frame rate, geometry, and resolution to predicate device.
The verification/validation activities successfully confirmed device requirements have been fulfilled, system functionality is consistent with the user needs, intended uses, and the Ziehm Solo FD device correctly performs as designed, and raises no new questions regarding either safety or effectiveness.
Therefore, when compared to the predicate devices the Ziehm Solo FD supports a determination of substantial equivalence to the predicate devices.
| FDA/CDRH<br>From 3626 (1/14) | "A Guide for the Submission of Initial Reports on Diagnostic X-<br>Ray Systems and Their Major Components" |
|------------------------------|------------------------------------------------------------------------------------------------------------|
| 21 CFR 1020.30-32 | Federal Performance Standard for Diagnostic X-ray Systems. |
| MDD 93/42/EEC | Annex II of the European Medical Devices Directive (MDD)<br>93/42/EEC. |
| EN ISO 13485 | Medical devices - Quality management systems -<br>Requirements for regulatory purposes |
#### Compliance Standards:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image features the logo of Ziehm Imaging. The logo consists of a circular graphic on the left, which contains a stylized human figure within interconnected lines, suggesting a global network or interconnectedness. To the right of the graphic is the text "ziehmimaging" in a teal-colored, serif typeface, with the letters slightly spaced apart.
| AAMI/ANSI ES60601-1 | Medical Electrical Equipment -- Part 1: General Requirements<br>for Basic Safety and Essential Performance (IEC 60601-<br>1:2005, mod)<br>Date: 2010<br>Conformance Standard #19-5 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | Medical Electrical Equipment, General Requirements for Safety<br>Edition 3.0 (+Corr.1+Corr.2), Date: 2005-12-15 |
| IEC 60601-1-2 | Medical Electrical Equipment, General Requirements for<br>Safety, Electromagnetic Compatibility<br>Edition 3.0, Date: 2007-03-30,<br>Conformance Standard #19-1 |
| IEC 60601-1-3 | Medical Electrical Equipment, Radiation Protection in<br>Diagnostic X-ray Equipment<br>Edition 2.0, Date: 2008-01-22<br>Conformance Standard #12-210 |
| IEC 60601-2-43 | Medical electrical equipment, Part 2-43: Particular<br>requirements for basic safety and essential performance of X-<br>ray equipment for interventional procedures<br>Edition 2.0, Date: 2010-03-25<br>Conformance Standard #12-202 |
| IEC 60601-2-54 | Medical electrical equipment, Part 2-54: Particular<br>requirements for the basic safety and essential performance<br>of X-ray equipment for radiography and radioscopy<br>Edition 1.0, Date: 2009-06-29<br>Conformance Standard #12-274 |
| IEC 60825-1 | Safety of laser products, Equipment Safety, requirements, and<br>user guide<br>Edition 2.0, Date: 2007-03-30<br>Conformance Standard #12-273 |
| ISO 14971 | Medical devices - Application of risk management to medical<br>devices<br>Edition 2.0, Date: 2007-03-01<br>Conformance Standard #5-40 |
- Ziehm Imaging GmbH considers the Ziehm Solo FD to be Conclusion: as safe, as effective, and performs substantially equivalent to the predicate device Ziehm Vision² FD (K073346) and Ziehm Solo (K092438) in accordance with its labeling.
End of 510(k) Summary
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Ziehm Imaging. The logo consists of a circular graphic on the left and the text "ziehmimaging" on the right. The circular graphic features a stylized human figure within a globe-like structure. The text "ziehmimaging" is written in a sans-serif font, with the colors transitioning from a lighter teal to a darker teal.
Richard A. Fiene
Richard L. Westrich Official Agent and Correspondent for Ziehm Imaging GmbH. Ziehm Imaging Inc, 6280 Hazeltine National Drive Orlando, Fl. 32822
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.