K151000 · Medtronic, Inc. · OWB · Aug 6, 2015 · Radiology
Device Facts
Record ID
K151000
Device Name
O-Arm O2 Imaging System
Applicant
Medtronic, Inc.
Product Code
OWB · Radiology
Decision Date
Aug 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The O-arm® O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs or greater and having an abdominal thickness greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The O-arm® O2 Imaging System is compatible with certain image guided surgery systems.
Device Story
Mobile x-ray system providing 2D fluoroscopic and 3D cone-beam CT imaging; utilizes 360-degree gantry rotation for 3D reconstruction. System comprises Image Acquisition System (IAS) with x-ray generator/detector and Mobile View Station (MVS) for image processing/viewing. Adds extended field-of-view (40cm) mode via half-fan scan acquisition. Used in clinical settings (e.g., OR) by physicians to visualize anatomic structures and metallic objects; compatible with image-guided surgery systems. Output assists clinicians in surgical planning and navigation; benefits include improved visualization of larger anatomic regions.
Clinical Evidence
Bench testing only. Includes AAMI/ANSI ES 60601-1, IEC 60601-1-2, 1-3, 2-28, and 2-43 compliance. Software/hardware verification and validation performed. Dosimetry report and image quality assessment conducted. Usability testing performed per FDA human factors guidance. Cadaver image pair study compared O-arm® O2 images against predicate O-arm® 1000 and reference Artis Zeego to evaluate clinical utility.
Technological Characteristics
Mobile cone-beam CT; 32kW RoHS-compliant generator; 40x30cm CsI flat-panel detector. Isocentric gantry motion. 2D fluoroscopy and 3D imaging (20cm full-fan, 40cm half-fan). Connectivity via cable between IAS and MVS. Power: 100/120/240 VAC, 50/60Hz. Internal battery for transport/gantry; UPS for MVS. Software-controlled imaging protocols.
Indications for Use
Indicated for adult and pediatric patients weighing ≥60 lbs with abdominal thickness >16cm requiring 2D fluoroscopic or 3D imaging of bony anatomy or metallic objects.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is an emblem that features a stylized image of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 6, 2015
Medtronic, Inc. Medtronic Navigation, Inc. (Littleton) % Mr. Rishi Sinha Prinicipal Regulatory Affairs Specialist 300 Foster Street LITTLETON MA 01460
Re: K151000
Trade/Device Name: O-arm® O2 Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO and JAA Dated: June 30, 2015 Received: July 2, 2015
Dear Mr. Sinha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) NA K151000
Device Name O-arm® O2 Imaging System
#### Indications for Use (Describe)
The O-arm® O2 Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging for adult and pediatric patients weighing 60 lbs or greater and having an abdominal thickness greater than 16cm, and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects.
The O-arm® O2 Imaging System is compatible with certain image guided surgery systems.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Submitter: | Medtronic Navigation, Inc. (Littleton)<br>300 Foster Street<br>Littleton, MA 01460 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Rishi Sinha (Primary)<br>Principal Regulatory Affairs Specialist<br>Phone: (269) 903-4373<br>Fax: (269) 353-5924<br>E-mail: rishi.k.sinha@medtronic.com<br><br>Paul Smolenski (Alternate)<br>Senior Regulatory Affairs Manager<br>Phone: (978)698-6065<br>Fax: (978)698-6090<br>E-mail: paul.d.smolenski@medtronic.com |
| Date Summary Prepared: | June 30th, 2015 |
| Device Trade Name: | Medtronic O-arm® O2 Imaging System |
| Common Name: | Interventional Fluoroscopic X-ray System |
| Device Classification: | Class II |
| Product Code: | Primary: OWB<br>Secondary: OXO, JAA |
| Classification Name: | 892.1650 - Image Intensified Fluoroscopic X-ray System, Mobile |
| Predicate Device: | K092564 - Medtronic O-arm® Imaging System |
| Device Description: | The O-arm® O2 Imaging System is a mobile x-ray system that provides<br>3D imaging as well as 2D fluoroscopic imaging. It was originally cleared<br>for market in 2005 via K050996. Additional submissions were made in<br>2006 (K060344) and 2009 (K092564). The device is classified under<br>primary product code OWB (secondary OXO, JAA) , ref 21 CFR<br>892.1650.<br><br>O-arm® O2 Imaging System, also referred to as “O-arm® O2”, adds an<br>extended field of view imaging mode that offers twice the lateral field of<br>view as the prior design to provide clinicians further visualization options<br>in larger anatomic regions and anatomical structures. It accomplishes<br>this task with essentially the same hardware design as described within.<br><br>The system consists of two parts: the O-arm® Image Acquisition System<br>(IAS), comprising of a x-ray generator, amorphous silicon flat panel x-ray<br>detector and the x-ray control user interface and the Mobile View Station<br>(MVS), comprising of the image processors, a user interface for image<br>and patient handling and viewing monitor.<br><br>The O-arm® O2 Imaging System consists of two main assemblies that<br>are used together during fluoroscopic imaging: |
| | • The Mobile View Station (MVS) |
| | The two units are interconnected by a single cable that provides power<br>and signal data. The O-arm® IAS has an internal battery pack that<br>provides power for motorized transportation and gantry positioning. In<br>addition the battery pack is used to power the X-ray tank. The MVS has<br>an internal UPS to support its function when mains power is<br>disconnected. |
| | The O-arm® operates off standard line voltage within the following<br>voltages:<br>• VAC 100, 120 or 240<br>• Frequency 60Hz; 50Hz<br>• Power Requirements 1440 VA |
| Indications for Use: | The O-arm® O2 Imaging System is a mobile x-ray system designed for<br>2D fluoroscopic and 3D imaging for adult and pediatric patients weighing<br>60lbs or greater and having an abdominal thickness greater than 16cm,<br>and is intended to be used where a physician benefits from 2D and 3D<br>information of anatomic structures and objects with high x-ray<br>attenuation such as bony anatomy and metallic objects. |
| | The O-arm® O2 Imaging System is compatible with certain image guided<br>surgery systems. |
- The Image Acquisition System (IAS) •
{4}------------------------------------------------
| Substantial Equivalence: | O-arm® O2 is substantially equivalent to the following device: | |
|--------------------------|----------------------------------------------------------------|--|
| | · K092564 – O-arm® Imaging System | |
| | Subject Device | Predicate |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | O-arm® O2 Imaging System | O-Arm® 1000 Imaging System<br>(K092564) |
| Classification | Class 2 | Class 2 |
| Product Code | OXO; 892.1650 | OXO; 892.1650 |
| Indications for Use | The O-arm® O2 Imaging System is<br>a mobile x-ray system designed for<br>2D fluoroscopic and 3D imaging and<br>is intended to be used where a<br>physician benefits from 2D and 3D<br>information of anatomic structures<br>and objects with high x-ray<br>attenuation such as bony anatomy<br>and metallic objects.<br>The O-arm® Imaging System is<br>compatible with certain Image<br>Guided Surgery Systems. | The O-arm® Imaging System is a<br>mobile x-ray system designed for<br>2D fluoroscopic and 3D imaging<br>and is intended to be used where<br>a physician benefits from 2D and<br>3D information of anatomic<br>structures and objects with high<br>x-ray attenuation such as bony<br>anatomy and metallic objects.<br>The O-arm® Imaging System is<br>compatible with certain Image<br>Guided Surgery systems. |
## Table 1: O-arm® O2 Predicate Device Comparison Table
{5}------------------------------------------------
| | Subject Device | Predicate |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | O-arm® O2 Imaging System | O-Arm® 1000 Imaging System (K092564) |
| Cone Beam CT | The O-arm® O2 Imaging System is a mobile cone-beam x-ray system with isocentric motion options. It allows 3D image reconstruction using a 360 degree rotation of the x-ray source and detector within closed gantry. | The O-arm® Imaging System is a mobile cone-beam x-ray system with isocentric motion options. It allows 3D image reconstruction using a 360 degree rotation of the x-ray source and detector within closed gantry. |
| Detector Technology | 40 x 30 cm (RoHS compliant, Flat-Panel Detector using a Csl scintillation) | 40 x 30 cm (Flat-Panel Detector using a CsI scintillation) |
| Generator Technology | 32 kW, RoHS compliant generator with improved electrical interface. | 32kW Generator |
| 2D Imaging | 2D Fluoroscopic | 2D Fluoroscopic |
| 3D Imaging (20 cm FOV) | Full Fan (20cm FOV) scan acquisition | Full Fan (20cm FOV) scan acquisition |
| 3D Imaging Protocols (20 cm FOV) | Available presets:<br>1. Standard 3D<br>2. HD3D (High Definition)<br>3. Enhanced Cranial<br>4. Low Dose 3D | Available presets:<br>1. Standard 3D<br>2. HD3D (High Definition)<br>3. Enhanced Cranial |
| 3D Imaging (40 cm FOV) | Half-fan single scan acquisition | No 3D Imaging at 40cm FOV |
| 3D Imaging Protocols (40 cm FOV) | Available presets:<br>1. HD3D (high definition) equivalent to 750 projections | No 3D Imaging at 40cm FOV |
Performance Testing: Testing conducted demonstrates the product will perform as intended according to the outlined design requirements. The following testing was conducted on the O-arm O2 Imaging System device to establish substantial equivalence of the O-arm O2 Imaging System and verify that device will perform as intended meeting all of the design inputs.
- AAMI/ANSI ES 60601-1:2012 Medical Electrical Equipment -● Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2:2007 Medical Electrical Equipment Part 1-2: . General requirements for safety; Electromagnetic Compatibility -Requirements and Tests (2/2014)
- IEC 60601-1-3:2008 - Medical Electrical Equipment - Part 1-3:General Requirements for Basic Safety and Essential Performance – Collateral Standard: Radiation Protection in Diagnostic X-ray Equipment (1/2014)
- IEC 60601-2-28:2010 Medical electrical equipment part 2: . Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis (8/2013)
{6}------------------------------------------------
- . IEC 60601-2-43:2010 – Medical electrical equipment Part 2-43: Particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures (3/2010)
- Software Verification and Validation testing verifying the software requirements perform as intended
- . Hardware verification ensuring the hardware requirements identified for the system perform as intended.
- The Dose Setting Guidance for Extra-small Patient Size determines ● recommended dose settings (kVp/mA pairs) for an extra-small patient population to help the end user when imaging such patients.
- . The Image Quality Assessment of the O-arm® O2 System provides a quantitative image quality assessment of the O-arm® O2 system in comparison to the predicate O-arm® 1000 device.
- . The Dosimetry Report documents the dosimetry measurements for the various modes of the O-arm O2 System.
- . The Usability Testing was conducted according to the FDA quidance Applying Human Factors and Usability to Optimize Medical Device Design. Users conducted a series of imaging functions under simulated use conditions.
- The O-arm® Cadaver Image Pair Study evaluates the clinical utility ● of the images obtained using the O-arm® O2 Imaging System compared to the images obtained using the predicate O-arm® 1000 and the reference Artis Zeego device.
All performance testing was conducted to ensure the product meets all prescribed design inputs.
Conclusion: The O-arm® O2 system is similar in technological characteristics, imaging performance and indications for use as the predicate devices listed. These aspects, along with the functional testing conducted to the FDA recognized standards, demonstrate that O-arm® O2 does not raise new risks of safety and effectiveness when compared to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.