K142126 · Philips Medical Systems Nederland B.V. · OWB · Dec 9, 2014 · Radiology
Device Facts
Record ID
K142126
Device Name
EP NAVIGATOR REL. 5.0
Applicant
Philips Medical Systems Nederland B.V.
Product Code
OWB · Radiology
Decision Date
Dec 9, 2014
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
EP navigator is intended to provide navigation support for cardiovascular devices for heart rhythm disorders, such as catheters and guidewires, by superimposing 3D cardiac anatomical image data, such as CT, MRI, or 3D rotational scan over intra-procedural X-ray images of the same anatomy.
Device Story
EP navigator Rel. 5.0 is a software medical device operating on a PC-based platform; used in clinical settings to assist physicians during cardiovascular procedures. Input data includes previously acquired 3D anatomical images (CT, MRI, or 3D rotational scans) and real-time intra-procedural X-ray images. The software segments 3D datasets, registers them with X-ray images, and superimposes the 3D anatomy and landmarks onto the live X-ray view. Features include visual marker placement (tag points), EndoView for internal volume visualization, manual 2D measurements, and anatomical-based viewing planes. It communicates with Philips X-ray systems to enable Automatic Position Control (APC). By providing a 3D roadmap of the patient's cardiac anatomy overlaid on live fluoroscopy, the device aids navigation of catheters and guidewires, potentially improving procedural accuracy and efficiency for heart rhythm disorder treatments.
Clinical Evidence
No clinical data; bench testing only. Verification and validation testing, including system verification and assessment of individual modifications, demonstrated compliance with IEC 62304, IEC 62366, ISO 14971, and NEMA PS 3.1-3.20 standards.
Technological Characteristics
PC-based software medical device. Integrates with Philips X-ray systems. Features: 3D image segmentation (CT, MRI, rotational scan), registration, and superimposition. Connectivity: DICOM compliant. Standards: IEC 62304 (software lifecycle), IEC 62366 (usability), ISO 14971 (risk management), NEMA PS 3.1-3.20 (DICOM).
Indications for Use
Indicated for patients with heart rhythm disorders requiring cardiovascular device navigation (e.g., catheters, guidewires) during interventional procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Philips Medical Systems Nederland B.V. % Ms. Liselotte Kornmann Regulatory Affairs Manager Veenpluis 4-6 Best, 5684 PC Netherlands
December 9, 2014
Re: K142126 Trade/Device Name: EP Navigator rel. 5.0 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, LLZ Dated: November 14, 2014 Received: November 17, 2014
Dear Ms. Kornmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K142126
Device Name
EP navigator
Indications for Use (Describe)
EP navigator is intended to provide navigation support for cardiovascular devices for heart rhythm disorders, such as catheters and guidewires, by superimposing 3D cardiac anatomical image data, such as CT, MRI, or 3D rotational scan over intra-procedural X-ray images of the same anatomy.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | July 31, 2014 | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Medical Systems Nederland B.V.<br>Veenpluis 4-6<br>5684 PC Best<br>The Netherlands<br>Establishment Registration Number: 3003768277 | |
| Contact Person: | Ms. Liselotte Kornmann, PhD<br>Regulatory Affairs Manager<br>Phone: +31 611621238<br>E-mail: Liselotte.Kornmann@philips.com | |
| Device: | Trade Name:<br>Common Name:<br>Device Class:<br>Classification Name:<br>Classification Regulation:<br>Classification Panel:<br>Primary Product Code:<br>Secondary Product Code: | EP navigator Rel. 5.0<br>EP navigator<br>Class II<br>Image-intensified fluoroscopic x-ray system<br>21CFR §892.1650<br>Radiology<br>OWB (Interventional x-ray system)<br>LLZ (system, image processing, radiological) |
| Primary1 Predicate<br>Device: | Trade Name:<br>Manufacturer:<br>510(k) Clearance:<br>Classification Regulation:<br>Classification Name:<br>Classification Panel:<br>Device Class:<br>Product Code: | Allura Xper FD series<br>Allura Xper OR Table series<br>Philips Medical Systems Nederland B.V.<br>K133292 (March 05, 2014)<br>21 CFR, Part 892.1650<br>Image-intensified fluoroscopic x-ray system<br>Radiology<br>Class II<br>OWB (primary), JAA (secondary) |
| Secondary Predicate<br>Devices #1: | Trade Name:<br>Manufacturer:<br>510(k) Clearance:<br>Classification Regulation:<br>Classification Name:<br>Classification Panel:<br>Device Class:<br>Product Code: | EP navigator Rel. 3.0<br>Philips Medical Systems Nederland B.V.<br>K101311 (Sep 30, 2010)<br>21 CFR, Part 892.2050<br>Picture Archiving and Communications System<br>Radiology<br>Class II<br>LLZ |
| Secondary Predicate<br>Devices #2: | Trade Name:<br>Manufacturer:<br>510(k) Clearance:<br>Classification Regulation:<br>Classification Name:<br>Classification Panel:<br>Device Class:<br>Product Code: | HeartNavigator Rel. 2.0<br>Philips Medical Systems Nederland B.V.<br>K140138 (June 10, 2014)<br>21 CFR, Part 892.1650<br>Image-intensified fluoroscopic x-ray system<br>Radiology<br>Class II<br>OWB, LLZ |
| Device Description: | EP navigator Rel. 5.0 is a software medical device that contains image<br>processing algorithms that are executed on a PC-based hardware platform.<br>EP navigator Rel. 5.0 can perform the following functions:<br>view previously acquired 3D anatomical image data in a slice viewer, such<br>●<br>as CT, MRI, or 3D rotational scan,<br>segment previously acquired 3D anatomical image data,<br>●<br>register the segmented 3D dataset and landmarks with intra-procedural X-ray<br>●<br>images,<br>superimpose the segmented 3D dataset and landmarks on intra-procedural X-<br>●<br>ray images of the same anatomy,<br>position visual markers such as tag points, and<br>●<br>visualize the inside of the 3D volume (EndoView).<br>● | |
| Indications for Use: | EP navigator is intended to provide navigation support for cardiovascular devices<br>for heart rhythm disorders, such as catheters and guidewires, by superimposing<br>3D cardiac anatomical image data, such as CT, MRI, or 3D rotational scan over<br>intra-procedural X-ray images of the same anatomy. | |
| Fundamental<br>Scientific<br>Technology: | EP navigator Rel. 5.0 is intended to be used in combination with the primary<br>predicate device Allura X-ray system (K133292). While the secondary predicate<br>device EP navigator Rel. 3.0 (K101311) and EP navigator Rel. 5.0 share the<br>same technological characteristics, EP navigator Rel. 5.0 offers the following<br>enhancements to existing functions: (1) Support of the import and automatic<br>segmentation of MR datasets, (2) Reduced angular range of 3D EP rotational<br>scan and support of the automatic segmentation of a 3D EP rotational scan with<br>reduced angular angle, (3) Manual 2D measurements, (4) Anatomical-based<br>viewing planes, and (5) Communication with Philips X-ray system, including<br>Automatic Position Control (APC). The modifications represented by<br>EP navigator Rel. 5.0 do not affect the safety and effectiveness of the device. | |
<sup>1</sup> Note: the EP navigator Rel. 5.0 software medical device is an accessory to the currently marketed and primary predicate Allura Xper FD series and Allura Xper OR Table series K133292. The Allura Xper FD series and Allura Xper OR Table series will be abbreviated as Allura X-ray system in this 510(k) Premarket notification. Substantial equivalence demonstration to the primary predicate is not warranted as they are devices with which EP navigator Rel. 5.0 is to be used. Therefore, substantial equivalence is demonstrated to the secondary predicate devices.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters against a white background. The font is sans-serif and the letters are evenly spaced. The word is horizontally oriented and centered in the image.
Therefore, EP navigator Rel. 5.0 is substantially equivalent to the currently marketed and secondary predicate EP navigator Rel. 3.0 (K101311) and HeartNavigator Rel. 2.0 (K140138) in terms of fundamental scientific technology.
Summary of Nonclinical Performance Data:
Non-clinical performance testing has been performed on EP navigator Rel. 5.0 and demonstrates compliance with the following International and FDArecognized consensus standards and FDA guidance document:
- IEC 62304 Medical device software Software life cycle processes (Ed. 1.0, . 2006).
- IEC 62366 Medical devices - Application of usability engineering to medical devices (Ed. 1.0, 2007),
- ISO 14971 Medical devices Application of risk management to medical ● devices (Ed. 2.0, corrected version, 2007),
- NEMA PS 3.1-3.20 Digital Imaging and Communications in Medicine (DICOM) Set (2011), and
- . Guidance for Industry and FDA Staff - Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005, document number 337).
Verification and validation tests have been performed to address intended use, the technical claims, requirement specifications and the risk management results. The modifications of the EP navigator Rel. 5.0 that are required to implement the enhancements are within the design controls and specifications; a system verification test was performed to ensure that the modifications are properly introduced; verification and validation testing was conducted to ensure the proper introduction of the individual modifications listed; sample clinical images as well as conformance to IEC standards and guidance documents were provided. All of these components and tests were used to support substantial equivalence of the subject device.
The test results in this 510(k) premarket notification demonstrate that EP navigator Rel. 5.0 complies with the aforementioned international and FDArecognized consensus standards and FDA guidance document, meets the acceptance criteria, and is adequate for its intended use. Additionally, all risks are sufficiently mitigated, no new risks are introduced and the overall residual risk is acceptable.
Therefore, EP navigator Rel. 5.0 is substantially equivalent to the currently marketed and secondary predicate devices EP navigator Rel. 3.0 (K101311) and HeartNavigator Rel. 2.0 (K140138) in terms of safety and effectiveness.
Summary of Clinical EP navigator Rel. 5.0 did not require clinical studies to demonstrate substantial Data: equivalence to the currently marketed and secondary predicate devices EP navigator Rel. 3.0 (K101311) and HeartNavigator Rel. 2.0 (K140138).
The EP navigator Rel. 5.0 software medical device is substantially equivalent to Substantial Equivalence the currently marketed and secondary predicate devices EP navigator Rel. 3.0 Conclusion: (K101311) and HeartNavigator Rel. 2.0 (K140138) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness. Additionally, substantial equivalence was demonstrated with nonclinical performance testing, which complied with the requirements specified in
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and aligned horizontally. The font is sans-serif and the background is white.
the international and FDA-recognized consensus standards. The non-clinical performance tests provided in this 510(k) premarket notification demonstrates that the proposed EP navigator Rel. 5.0 is as safe and effective as its predicate devices without raising any new safety and/or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.