K100102 · Omega Medical Imaging, Inc. · OWB · Aug 31, 2010 · Radiology
Device Facts
Record ID
K100102
Device Name
CS-SERIES-FP
Applicant
Omega Medical Imaging, Inc.
Product Code
OWB · Radiology
Decision Date
Aug 31, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Omega Medical Imaging, Inc. CS-series-FP (SSXI) systems are intended for use in radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
Device Story
CS-series-FP is a fluoroscopy system utilizing a solid-state flat-panel detector (FPD) with a cesium iodide scintillator and amorphous silicon TFT panel. It replaces image intensifier/CCD-based detectors in existing CS-series systems. The system includes a floor-mounted C-arm (dual-plane adds ceiling-suspended C-arm) and a patient table. It captures X-ray signals, processes digital images via an acquisition workstation (MX-200-OMI), and displays them on medical-grade monochrome monitors. Features include pulsed fluoro, spot, and cine modes; edge enhancement; noise reduction with motion correction; and DICOM-compatible storage/networking. Used in clinical settings (e.g., interventional suites) by physicians and technicians. Output allows real-time visualization and review of diagnostic images, supporting clinical decision-making in cardiac and vascular interventions. Benefits include improved imaging performance over traditional intensifiers.
Clinical Evidence
Bench testing only. Performance was evaluated by comparing the CS-series-FP system against the existing Omega CS-series (image-intensifier/CCD-based) and by providing technical specifications and performance data for the Varian PaxScan 2020 flat-panel detector. No clinical data was required or provided.
Indicated for radiographic and fluoroscopic imaging in cardiac, vascular, general diagnostic, and interventional procedures. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# KIDDI02
Omega Medical Imaging, Inc.
## EXECTUTIVE 510(k) SUMMARY
AUG 31 2010
| Company Name: | Omega Medical Imaging, Inc |
|-----------------------|-------------------------------------------------------------------------------------------|
| Address: | 675 Hickman Circle<br>Sanford, FL 32771 |
| Telephone No: | 407-323-9400 |
| Registration No.: | 1052701 |
| Contact person: | James A. Princehorn |
| Date Prepared: | 22 July 2010 |
| Device (trade) name: | CS-series-FP (SSXI) Fluoroscopy system |
| Common/usual name: | Image Intensified Fluoroscopic X-ray System |
| Classification Name: | Solid State X-ray Imager, Class II, 90 MQB<br>Angiographic X-ray System, Class II, 90 IZI |
| Classification Panel: | Radiology |
| CFR Section: | 892.1650 and 892.1600 |
| Device Class: | Class II |
| Device Code: | 90 MQB, 90 IZI OWB, JAA |
### Predicate device(s):
- GE Innova 2000 (K022322) .
- Siemens Axiom Artis dFC (K021021) .
- Philips Allura Xper FD10 F (K020055) .
- . Toshiba Infinix CCi/FPD (K052884)
- . Omega Medical Imaging CS-series (K070834)
#### Device description:
- The Omega Medical Imaging, Inc. CS-series-FP (SSXI) systems incorporates a ◆ solid-state flat-panel detector (FPD) as an option to the cleared Omega Medical Imaging, Inc. CS-series fluoroscopy systems (K070834) in lieu of an image intensifier/CCD based imaging detector. The CS-series fluoroscopy single and dual plane x-ray imaging systems are configured with a floor mounted C-am and a patient table. The dual plane systems incorporate a ceiling suspended C-arm. The image detector utilizes a cesium iodide scintillator coupled to an amorphous silicon TFT panel. The caplured digital image is processed by the acquisition system which includes image processing, viewing functions, local storage, and DICOM compatibility.
10-1
Section 10
{1}------------------------------------------------
The following tables indicate the technical specifications of the CS-series-FP option:
| 1. | Flat Panel Detector |
|----|---------------------|
|----|---------------------|
. •
| 1.1 Receptor Type | Amorphous Silicon |
|---------------------------|-------------------------------------------------------|
| 1.2 Conversion Screen | Cesium Iodide |
| 1.3 Pixel Area - Active | 19.8 cm X 19.8 cm |
| 1.4 Pixel Matrix - Active | 1024 X 1024 |
| 1.5 Pixel Pitch | 194 $\mu$ m |
| 1.6 Limiting Resolution | 2.58 lp/mm |
| 1.7 MTF, X-ray | $\geq$ 35% @ 1.30 lp/mm |
| 1.8 Energy Range | 40 - 150 kV |
| 1.9 Fill Factor | 70% |
| 1.10 Dynamic Range | 72 dB |
| 1.11 Contrast Ratio | Large Area (120mm): < 0.8%<br>Small Area (10mm): < 7% |
| 1.12 A/D Conversion | 14-bits |
### 2. Acquisition
| 2.1 Modes | Pulsed Fluoro, Spot, & Cine |
|---------------------------------|--------------------------------|
| 2.2 Pulsed Fluoro Rates | 3.75, 7.5, and 15 pps |
| 2.3 Spot Rates | Single & 2 fps |
| 2.4 Cine Rates | 15 & 30 fps |
| 2.5 Polarity | Positive & Negative |
| 2.6 Edge Enhancement | 5 level |
| 2.7 Noise Reduction | 6 level with motion correction |
| 2.8 Last Image Hold | Yes |
| 2.9 Fluoro Loops | Yes |
| 2.10 Horizontal & Vertical Flip | Yes |
## 3. Acquisition Workstation
| 3.1 Workstation Model | MX-200-OMI |
|--------------------------------|-----------------------------------------------------------|
| 3.2 Monitor, size | 2 X 18.1 inch, Medical Grade,<br>high contrast monochrome |
| 3.3 Grey Levels | |
| 3.4 Review (Fluoro Loops/Cine) | Forward, Reverse, Pause, and Speed |
| 3.5 Window/Level Adjust | Yes |
| 3.6 Annotation | Yes |
| 3.7 Image Invert | Yes |
| 3.8 Horizontal & Vertical Flip | Yes |
| 3.9 Zoom & Roam | Yes (2X) |
| 3.10 Image Storage | Minimum: 117,000 images |
| 3.11 Networking Capabilities | Ethernet 10/100 |
| 3.12 DICOM Store | Yes |
| 3.13 DICOM Print | Yes |
Omega Medical Imaging CS-series-FP 510(k) (revised 7 April 2010)
10-2
.
.
.
{2}------------------------------------------------
| 3.14 DICOM Worklist Query | Yes |
|---------------------------|----------------------|
| 3.15 DICOM Send | Yes |
| 3.16 Save to DVD/CD | With Embedded Viewer |
| 3.17 Windows Print | Yes |
| 3.18 Image Save | .JPEG & AVI |
## 4. Reference Monitor Identifiers [Review Station]
ﻌﺐ
| 4.1 Patient Name | 64 characters |
|------------------------|---------------------------------------|
| 4.2 Patient I.D. | 64 characters |
| 4.3 Date | 8 characters (05/08/03) |
| 4.4 Time | 4 characters (13::45) |
| 4.5 Window/Level | 0-4096 |
| 4.6 Hospital Name | 64 characters |
| 4.7 Physician Name | 64 characters |
| 4.8 Study I.D. | 16 characters |
| 4.9 Series/Image | 3 characters/3 characters |
| 4.10 Acquisition Mode | PF/DR/Cine |
| 4.11 Image Orientation | Indicated by "R" orientation |
| 4.12 Radiation | Radiation symbol during Live |
| 4.13 Image I.D. | e.g., LIH, Fluoro Loop, Spot, or Cine |
## 5. Live Monitor Identifiers [Image Area]
| 5.1 Patient Name | 64 characters |
|-------------------------|--------------------------------------------------------|
| 5.2 Patient I.D. | 64 characters |
| 5.3 Date | 8 characters (05/08/03) |
| 5.4 Time | 4 characters (13.:45) |
| 5.5 Window/Level | 0-4098 |
| 5.6 Hospital Name | 64 characters |
| 5.7 Physician Name | 64 characters |
| 5.8 Study I.D. | 16 characters |
| 5.9 Series/Image | 3 characters/3 characters |
| 5.10 Image Orientation | Indicated by "R" orientation |
| 5.11 Radiation Symbol | Radiation symbol during Live |
| 5.12 Image I.D. | e.g., FL LIH, FL Loops, Spot Replay, or Cine<br>Replay |
| 5.13 Detector Status | Ready/Not Ready |
| 5.14 Magnification Mode | Normal, MAG 1, MAG 2 |
| 5.15 Loop Replay | Play, Forward, Reverse, Speed, & Pause |
| | |
## GUI [Graphic User Interface on Reference Monitor]
| Patient/Exam Entry | |
|--------------------|-----------------------------------|
| 5.16 Directory | Opens 'Patient Directory' screen |
| 5.17 New | Opens 'Patient Data Entry' screen |
| 5.18 Close | Closes active patient |
10-3
Omega Medical Imaging CS-series-FP 510(k) (revised 7 April 2010)
{3}------------------------------------------------
| Acquisition | |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------|
| 5.19 NR Hi / NR Lo | Selects Noise Reduction Level |
| 5.20 L -> R | Live to Reference (monitor) transfer |
| 5.21 FL @ 15 | Right Click to invoke Fluoro Pulse Rate (15 f/s default, 7.5, & 3.75 f/s) |
| Note: Image Orientation available for Acquisition Mode [from Review ] | |
| Review | |
| 5.22 Edge (enhancement) | Increments from 0, 1, 2, 3, & 4 |
| 5.23 Win./Lev. | Window/Level adjustment |
| 5.24 Split | Invokes 4 on 1 Image display |
| 5.25 Reference - Live | Toggles image between the Reference and Live monitors |
| 5.26 Zoom | Up to 2X zoom with panning |
| 5.27 Polarity | Video white on black or black on white |
| 5.28 Collimate | Invokes electronic shutters |
| 5.29 R [image orientation] | Toggles thru 1 of 4 positions |
| Additional Functions | |
| 5.30 Send | Opens 'DICOM server node selection' |
| 5.31 Print | Sends Image to a Windows compatible printer |
| 5.32 Save | Saves current image to Hard Drive |
| 5.33Text | Invokes the Annotation and Measurement function |
- System Components: This option includes the solid-state x-ray imager (FPD) and . image processor, along with associated mounting hardware and electrical/electronic components. These components are installed in lieu of the image intensifier/CCD camera and its image processor.
- . Materials: The solid-state x-ray imager (FPD) construction and materials comply with UL2601-1, IEC60601-1, CSA 22.2 No. 601.1-M90, and is CE Marked. The imager is classified by Underwriters Laboratories.
- Available Configurations: The CS-series-FP is available in single plane and dual . plane configurations.
#### Intended use:
- . The Omega Medical Imaging, Inc. CS-Series-FP (SSXI) systems are intended for use in radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
#### Comparison with Predicate Devices:
- It is the opinion of Omega Medical Imaging, Inc. that the CS-series-FP is essentially ● equivalent to the Toshiba Infinix CCi/FPD system (K052884), the GE Innova 2000 (K993037), Philips Alura Xper FD10 F (K020055), and the Siemens Axiom Artis dFC (K021021).
{4}------------------------------------------------
## DISCUSSION:
Review of the line items in the Comparison Table (Section 10.1) Indicate that all items compared are either similar or the same as the predicate devices. The Flat-Panel Detector manufactured by Varian Medical Systems is utilized in similar FDA cleared fluoroscopic systems and has been in production for several years. All of the major operational features are either similar or the same as the predicate devices and therefore should offer essentially the same performance to the user as the predicate devices.
### Non-Clinical Performance:
- Included in this report is detailed data comparing performance with the existing . Omega Medical Imaging, Inc. CS-series systems utilizing an image-intensifier/CCD based image acquisition system. Also included is detailed specifications and performance of the Varian PaxScan 2020 flat-panel detector.
#### Safety information:
- . The Omega CS-series-FP (SSXI) systems will comply with the applicable requirements of 21 CFR 1020.30, 21 CFR 1020.31, and 21 CFR 1020.32.
- The Omega CS-series-FP systems will comply with the international safety . standards IEC 60601-1, IEC 60101-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, and IEC 60601-2-43.
#### CONCLUSION:
The Omega CS-series-FP (SSXI) systems do not introduce any new indications for use, nor does the use of the device result in any new potential hazard. Omega considers the CSseries-FP (SSXI) systems with the flat-panel detector option to be substantially equivalent with the predicate devices.
Section 10
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. James Princehorn President Omega Medical Imaging, Inc. 675 Hickman Circle SANFORD FL 32771
JUL 30 2012
Re: K100102
Trade/Device Name: CS-series-FP Fluoroscopy System with optional Solid State X-ray Imaging Device (Flat-Panel Detector)
Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB. JAA. and MQB Dated: July 23, 2010 Received: July 29, 2010
Dear Mr. Princehorn:
This letter corrects our substantially equivalent letter of August 31, 2010.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of
{6}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Janine M. Morri Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): _
Device Name: CS-series-FP (SSXI)
Indications for Use:
17
The Omega Medical Imaging, Inc. CS-series-FP (SSXI) systems are intended for use in radiographic/fluoroscopic application including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
Prescription Use V (Part 21 CFR 801 Subpart D)
WAS AFF Comments on the State Charm Change Charm Ch
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ・・
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K100162
Page _1_ of _1
AUG 3 1 2010
\$\neg\$
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.