CS-SERIES FLUOROSCOPY SYSTEM, CS-10 THROUGH CS-65 (11 CONFIGURATIONS)
K070834 · Omega Medical Imaging, Inc. · OWB · May 18, 2007 · Radiology
Device Facts
Record ID
K070834
Device Name
CS-SERIES FLUOROSCOPY SYSTEM, CS-10 THROUGH CS-65 (11 CONFIGURATIONS)
Applicant
Omega Medical Imaging, Inc.
Product Code
OWB · Radiology
Decision Date
May 18, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Omega Medical Imaging, Inc. CS-Series systems are intended for use in radiographic/fluoroscopic cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
Device Story
CS-Series is a floor-mounted C-Arm radiographic/fluoroscopy system; includes options for elevating or elevating/tilt patient tables; supports bi-plane operation via ceiling-suspended C-arm. System utilizes x-ray source and image intensifier/CCD camera for image acquisition. Operator controls motorized positioning of source/receptor. Digital image processing provides visualization for clinicians in interventional or diagnostic settings. Device assists in clinical decision-making by providing real-time or static radiographic/fluoroscopic images for cardiac, vascular, and general procedures.
Clinical Evidence
Bench testing only. The device complies with 21 CFR 1020.30, 1020.31, 1020.32, and international safety standards including IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, and IEC 60601-2-32, as well as UL 60601-1 and CAN/USA C22.2 No.601.1-M90.
Technological Characteristics
Floor-mounted C-Arm configuration; motorized positioning; x-ray source; image intensifier/CCD camera; digital image processing. Complies with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-7, 60601-2-28, 60601-2-32, UL 60601-1, and CAN/USA C22.2 No.601.1-M90.
Indications for Use
Indicated for patients requiring cardiac, vascular, or general radiographic/fluoroscopic diagnostic and interventional x-ray imaging.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Siemens AXIOM ARTIS U Angiography System (K040675)
Philips Integris H5000 Angiography System (K984545)
Omega Medical Imaging e-VIEW Fluoroscopy System (K062647)
Submission Summary (Full Text)
{0}------------------------------------------------
# OMEGA MEDICAL IMAGING, INC.
## 510(k) SUMMARY
MAY 1 8 2007
| Company Name: | Omega Medical Imaging, Inc |
|-----------------------|------------------------------------------------|
| Address: | 675 Hickman Circle<br>Sanford, FL 32771 |
| Telephone No: | 407-323-9400 |
| Registration No.: | 1052701 |
| Contact person: | James A. Princehorn |
| Date Prepared: | 1 February 2007 |
| Device (trade) name: | CS-Series radiographic/fluoroscopy system |
| Classification Name: | Angiographic X-ray System |
| Classification Panel: | Radiology |
| CFR Section: | 892.1650 and 892.1600 |
| Device Class: | Class II |
| Device Code: | JAA, IZI |
| Common/usual name: | System, X-Ray, Fluoroscopic, Image-Intensified |
### Predicate device(s):
- Siemens AXIOM ARTIS U Angiography System (K040675) .
- Philips Integris H5000 Angiography System (K984545) .
- Omega Medical Imaging e-VIEW Fluoroscopy System (K062647). .
### Device description:
- The Omega Medical Imaging, Inc. CS-Series systems are comprised of an x-ray source/image . receptor positioning device in a permanently floor mounted C-Arm configuration with options of either an elevating only or elevating/tilt patient table. The system may also be configure with a ceiling suspended "C" for bi-plane operation. The positioning of the source/receptor is achieved by motorized motions controlled by the operator. The imaging is achieved by way of an image intensifier/CCD camera with digital image processing.
### Intended use:
- The Omega Medical Imaging, Inc. CS-Series systems are intended for use in . radiographic/fluoroscopic cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
{1}------------------------------------------------
### Safety information:
- The Omega CS-Series systems will comply with the applicable requirements of 21 CFR . 1020.30, 21 CFR 1020.31, and 21 CFR 1020.32.
- The Omega CS-Series systems will comply with the international safety standards IEC 60601-. 1, IEC 60101-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, and IEC 60601-2-32 .
- The Omega CS-Series systems will comply with UL 60601-1 and CAN/USA C22.2 No.601.1-. M90
### Software/Firmware
Omega Medical Imaging, Inc. manufactures the mechanical sub-systems for the CS-series systems. The control of these mechanical sub-systems is achieved without involvement of software or firmware. Omega Medical Imaging, Inc. obtains certified components from FDA registered suppliers that contain software/firmware. Omega Medical Imaging does not modify any of the software/firmware obtained by the suppliers. Omega Medical Imaging, Inc. conducts Vendor Audits of these suppliers to assure that these vendored components are suitable for this application.
## Conclusion:
The Omega CS-Series systems do not introduce any new indications for use, nor does the use of the device result in any new potential hazard. Omega considers the CS-Series systems to be substantially equivalent with the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with three stylized, overlapping figures that resemble birds in flight or abstract human figures. The figures are depicted in black and are arranged diagonally, creating a sense of movement or progression. To the left of the figures, there is text arranged vertically along a curved path, but the text is too small to read.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. James A. Princehorn President Omega Medical Imaging, Inc. 675 Hickman Circle SANFORD FL 32771
MAY 1 8 2007
Re: K070834
Trade/Device Name: CS-Series Fluoroscopy System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: II Product Code: JAA and IZL Dated: March 22, 2007 Received: March 27, 2007
## Dear Mr. Princehorn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a black and white circular logo. The logo contains the text "1906-2006" at the top, followed by the letters "PA" in a stylized font. Below the letters, the word "Centennial" is written in a cursive font. Three stars are at the bottom of the logo.
Protecting and Promoting Public Health
{3}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Omega Medical Imaging, Inc.
K070834
Omega Medical Imaging, Inc. CS-series Radiographic/Fluoroscopic Systems
# Indications for Use:
The Omega Medical Imaging, Inc. CS-Series systems are intended for use in radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.
*Prescription Use* ✓
(Division Sign-Off)
Division of Reproductive, Abdominal, and
Radiological Devices
510(k) Number K070834
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.