K070123 · Dornier Medtech America, Inc. · OWB · Mar 9, 2007 · Radiology
Device Facts
Record ID
K070123
Device Name
TCA 5R / 5S
Applicant
Dornier Medtech America, Inc.
Product Code
OWB · Radiology
Decision Date
Mar 9, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The TCA 5R / 5S C-arm systems are designed for x-ray examinations, and in particular, for radioscopy, radiography, and diagnosis dedicated to: traumatology, pediatrics, simple interventional radiology, pacemaker implantation, operating room procedures, intensive care, and respiratory and skeleton procedures.
Device Story
TCA 5R / 5S are mobile C-arm fluoroscopic x-ray systems; consist of C-arm stand (x-ray tube, image intensifier, high-voltage generator) and monitor trolley (image processing system, 1-2 monitors). Used in clinical settings (OR, ICU, radiology) by healthcare professionals for real-time imaging during surgical and diagnostic procedures. Device captures x-ray signals, processes them via integrated system, and displays images on monitors to assist clinicians in visualization during interventions (e.g., pacemaker implantation, traumatology). Enables minimally invasive guidance; improves procedural accuracy; supports patient diagnosis and treatment.
Clinical Evidence
Bench testing only. Compliance with applicable performance standards confirmed via Declaration of Conformity.
Technological Characteristics
Mobile C-arm system; includes x-ray tube, image intensifier, high-voltage generator, and monitor trolley. Energy source: x-ray. Connectivity: integrated image processing system. Form factor: mobile unit with C-shaped arm for patient positioning.
Indications for Use
Indicated for x-ray examinations including radioscopy, radiography, and diagnosis in patients requiring traumatology, pediatric, interventional radiology, pacemaker implantation, OR, intensive care, respiratory, or skeletal procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K010123
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR 807.92.
| Submitter Name: | Dornier MedTech America, Inc. |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter Address: | 1155 Roberts Blvd.<br>Kennesaw, GA 30144 |
| Contact Person: | Tim Thomas, Vice President<br>Regulatory/ Quality/ Clinical<br>MAR 09 2007 |
| Phone Number: | 770-514-6163 |
| Fax Number: | 770-514-6288 |
| Establishment<br>Registration Number: | 1037955 |
| Date Prepared: | January 12, 2007 |
| Device Trade Name(s): | TCA 5R / 5S |
| Device Common Name: | Fluoroscopic Imaging System or Mobile C-Arm |
| Classification Name: | System, x-ray, fluoroscopic, image-intensified (JAA)<br>System, x-ray, mobile (IZL) |
| Device Classification: | II (21 CFR Sec. 892.1650 and 892.1720)<br>(Image-intensified fluoroscopic x-ray system / Mobile<br>x-ray system) |
| Predicate Device(s): | K992103 - Wuestec Medical Inc. (C-Quest) |
| | K002198 - Fluoroscan Imaging Systems (Hologic)<br>(FlouroScan Profile) |
| General Device<br>Description: | The TCA 5R / 5S are mobile C-arms specifically<br>designed for x-ray imaging |
| Intended Use: | The TCA 5R / 5S C-arm systems are designed for x-<br>ray examinations, and in particular, for radioscopy,<br>radiography, and diagnosis dedicated to:<br>traumatology, pediatrics, simple interventional<br>radiology, pacemaker implantation, operating room<br>procedures, intensive care, and respiratory and<br>skeleton procedures. |
| Technological<br>Characteristics: | The TCA 5R / 5S consists of two mobile units: a C-<br>Arm and a Monitor Trolley which holds one or two |
{1}------------------------------------------------
monitors, an image processing system and possible accessories. The C-Arm stand supports the highvoltage generator, x-ray controls, and the "C" shaped arm which supports the x-ray tube and the image intensifier. The C-Arm is designed to move in various directions to allow for proper positioning with the patient based on the procedure. Additional technological characteristics are provided in Section 10 (Device Description) and Section 11 (Substantial Equivalence)
The TCA 5R / 5S comply with the applicable performance standards listed in Section 8 Performance Data: (Declaration of Conformity)
> Based on a comparison to the predicate device determined to be substantially equivalent through the 510(k) premarket notification process, Dornier MedTech America, Inc., concludes that the TCA 5R / 5S are as safe, as effective, and performs as well as other legally marketed C-arm devices.
Conclusion:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized graphic of three curved lines that resemble a human figure in motion. The lines are thick and black, and they create a sense of movement and energy. The logo is simple and modern, and it is likely used to represent the Department of Health & Human Services.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. Tim Thomas Vice President, Quality / Regulatory / Clinical Dornier MedTech America, Inc. 1155 Roberts Blvd. KENNESAW GA 30144
MAR 0 9 2007
Re: K070123
Trade/Device Name: TCA 5R / 5S Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: JAA and IZL Dated: January 12, 2007 Received: January 17, 2007
Dear Mr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a black and white circular logo. The logo has the text "1906-2006" at the top, "FDA" in the middle, and "Centennial" at the bottom. There are three stars below the word "Centennial". The logo is surrounded by a dotted border.
Protecting and Promoting Public Health
{3}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known):
Device Name:
Indications for Use:
The TCA 5R / 5S C-arm systems are designed for x-ray examinations, and in particular, for radioscopy, radiography, and diagnosis dedicated to: traumatology, pediatrics, simple interventional radiology, pacemaker implantation, operating room procedures, intensive care, and respiratory and skeleton procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K070123
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.