K032046 · Philips Medical Systems, Inc. · OWB · Jul 17, 2003 · Radiology
Device Facts
Record ID
K032046
Device Name
PHILLIPS OMNIDIAGNOST ELEVA
Applicant
Philips Medical Systems, Inc.
Product Code
OWB · Radiology
Decision Date
Jul 17, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Philips OmniDiagnost Eleva is a diagnostic-imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.
Device Story
Multifunctional, universal, overtable X-ray system for radiographic, fluoroscopic, angiographic, and interventional applications. System comprises floor-mounted stand with integrated tilting patient support table, supported at one end for bilateral patient access. Configurable with Philips X-ray generators, digital spot film device, 38 cm multimode image intensifier, XTV imaging system, collimator, and glass or metal X-ray tubes. Used in clinical settings by healthcare professionals to acquire diagnostic images. Output displayed on TV monitors for clinical review and decision-making. Benefits include versatile patient positioning and integrated imaging capabilities for various diagnostic and interventional procedures.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Stationary X-ray system with integrated tilting patient support table. Components include 38 cm multimode image intensifier, XTV imaging system, collimator, and glass or metal X-ray tubes. Complies with 21 CFR Subchapter J (Radiological Health), UL 2601-1 safety standards, and ACR/NEMA DICOM standards for digital imaging communication.
Indications for Use
Indicated for diagnostic radiographic, fluoroscopic, angiographic, and interventional imaging applications in patients requiring such procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
JUL 17 2003
Image /page/0/Picture/1 description: The image shows the Philips logo, which consists of the word "PHILIPS" in bold, sans-serif font above a shield-shaped emblem. The emblem contains a stylized representation of radio waves or sound waves, with a circular shape and two horizontal wavy lines inside. There are also four stars or asterisks placed at the corners of the circular shape, adding a decorative element to the logo.
# Philips Medical Systems Nederland B.V.
The following information is be bmitted in accordance with the requirements of 21 CFR 807.92.
| Company Name:<br>Address: | Philips Medical Systems North America Company<br>22100 Bothell Everett Highway<br>P.O.Box 3003<br>Bothell, WA 98041-3003, USA |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Registration No .: | 1217116 |
| Contact Person:<br>Telephone No .: | Lynn Harmer<br>(425) 487-7312 |
| Date Prepared: | June 12, 2003 |
| Device (Trade) Name: | Philips OmniDiagnost Eleva |
| Classification Names: | Stationary X-ray system,<br>21CFR892.1680, Class II (code 90KPR) |
| | Image-intensified fluoroscopic X-ray system,<br>21CFR892.1650, Class II (code 90JAA) |
| | Angiographic X-ray system,<br>21CFR892.1600, Class II (code 90IZI) |
## Predicate Device:
The OmniDiagnost Eleva is substantially equivalent to the Philips OmniDiagnost manufactured by Philips Medical Systems. The OmniDiagnost system received a 510(k) substantially equivalent determination in K982993 on August 26, 1998.
## Device description:
The Philips OmniDiagnost Eleva is a multifunctional, universal, overtable X-ray system offering radiographic, fluoroscopic, angiographic, and interventional techniques in a wide variety of applications. The Philips OmniDiagnost Eleva consisting of a floor mounted stand with an integrated tilting patient support table is supported at only one end, allowing patient access from both sides.
As a fully integrated system, the Philips OmniDiagnost Eleva stand can be configured with X-ray generators from the Philips generator family and Philips digital spot film device. The system comes with a 38 cm multimode Image Intensifier, XTV imaging system, collimator for patient positioning without X-rays, Philips glass or metal X-ray tubes, and TV monitors.
{1}------------------------------------------------
510(k) Summary Philips OmniDiagnost Eleva Page 2 of 2
Image /page/1/Picture/1 description: The image shows the Philips logo, which consists of the word "PHILIPS" in bold, sans-serif font above a shield-like emblem. The emblem features a stylized representation of radio waves or sound waves emanating from a central point, with additional star-like symbols placed above and below the waves. The logo is presented in black and white.
Image /page/1/Picture/2 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are large and black, contrasting with the white background. The word is the primary focus of the image, taking up most of the space.
## Intended use:
K∅3 2∅4L
The Philips OmniDiagnost Eleva is a diagnostic-imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.
#### General Safety and Effectiveness:
The device and their labeling will comply with the applicable requirements of:
- 21 CFR, Subchapter J Radiological Health, parts 1020.30, 31, 32 .
- Underwriters Laboratories Standard for Safety UL 2601-1 and be classified by . Underwriters Laboratories.
- ACR/NEMA DICOM digital imaging communication standard. ●
#### Conclusion:
The Philips OmniDiagnost Eleva does not introduce any new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers the OmniDiagnost Eleva to be substantially equivalent with the predicate device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the department's name around the perimeter. In the center is a stylized depiction of an eagle with its wings spread, symbolizing the department's mission to protect and promote the health and well-being of the nation.
#### Public Health Service
JUL 17 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Philips Medical Systems North America Company % Mr. Marc Mouser Project Engineer/Program Manager Underwriters Laboratories, Inc. 2600 N.W. Lake Road CAMAS WA 98607
## Re: K032046
Trade/Device Name: Philips OmniDiagnost Eleva Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: 90 IZI, JAA, and KPR Dated: June 30, 2003 Received: July 2, 2003
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number if Known):______________________________________________________________________________________________________________________________________________________
Device Name: Philips OmniDiagnost Eleva
Indications for Use:
The Philips OmniDiagnost Eleva is a diagnostic imaging device intended for radiographic, fluoroscopic, angiographic, and interventional applications.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The Counter Use _________________________________________________________________________________________________________________________________________________________
David R. Seymon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number. K032046
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.