← Product Code [QKB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QKB) · K200323

# AutoContour (K200323)

_Radformation, Inc. · QKB · Oct 30, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QKB/K200323

## Device Facts

- **Applicant:** Radformation, Inc.
- **Product Code:** [QKB](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QKB.md)
- **Decision Date:** Oct 30, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Software as a Medical Device

## Intended Use

AutoContour is intended to assist radiation treatment planners in contouring and reviewing structures within medical images in preparation for radiation therapy treatment planning.

## Device Story

AutoContour assists radiation treatment planners in contouring medical images for radiation therapy. System comprises three components: Windows-based agent service monitoring network storage for new CT datasets; cloud-based automatic contouring service; and web-based application for manual registration, review, editing, and export of structure sets. Input consists of CT images; output is DICOM RTSTRUCT file. Device used in clinical settings by radiation oncology personnel. Machine learning algorithms generate initial contours; user reviews and edits output. Output facilitates radiation treatment planning by automating structure delineation, potentially improving workflow efficiency and consistency.

## Clinical Evidence

No clinical trials performed. Bench testing only. Verification and validation testing conducted per internal protocol to ensure software performance against requirements. Machine learning model validated on independent datasets separate from training/validation sets to demonstrate generalizability.

## Technological Characteristics

Software-only application; no energy delivered to patients. Components: Windows-based agent, cloud-based contouring service (Linux), and web-based application (Linux/modern browsers). Connectivity via TCP/IP. Machine learning-based contouring algorithm. Supports CT input for contouring; CT, MR, PET for registration. Outputs DICOM RTSTRUCT.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Mirada RTx ([K130393](/device/K130393.md))
- Workflow Box ([K181572](/device/K181572.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 30, 2020

Radformation, Inc. % Mr. Kurt Sysock Co-founder/CEO 335 Madison Avenue, 16th Floor NEW YORK NY 10017

Re: K200323

Trade/Device Name: AutoContour Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: QKB Dated: September 19, 2020 Received: September 22, 2020

Dear Mr. Sysock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K200323

Device Name AutoContour

Indications for Use (Describe)

AutoContour is intended to assist radiation treatment planners in contouring structures within medical images in preparation for radiation therapy treatment planning.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section 5. 510(k) Summary

{4}------------------------------------------------

This 510(k) Summary has been created per the requirements of the Safe Medical Device Act (SMDA) of 1990, and the content is provided in conformance with 21 CFR Part 807.92.

## 5.1. Submitter's Information

| Table 1 : Submitter's Information |                                                      |
|-----------------------------------|------------------------------------------------------|
| Submitter's Name:                 | Kurt Sysock                                          |
| Company:                          | Radformation, Inc.                                   |
| Address:                          | 335 Madison Avenue, 16th Floor<br>New York, NY 10017 |
| Contact Person:                   | Alan Nelson<br>Chief Science Officer, Radformation   |
| Phone:                            | 518-888-5727                                         |
| Fax:                              | ----------                                           |
| Email:                            | anelson@radformation.com                             |
| Date of Summary Preparation       | 9/16/2020                                            |

## 5.2. Device Information

| Table 2 : Device Information |                                                                                |
|------------------------------|--------------------------------------------------------------------------------|
| Trade Name:                  | AutoContour                                                                    |
| Common Name:                 | AutoContour, AutoContouring, AutoContour Agent,<br>AutoContour Web Application |
| Classification Name:         | Class II                                                                       |
| Classification:              | Picture archiving and communications system                                    |
| Regulation Number:           | 892.2050                                                                       |
| Product Code:                | QKB                                                                            |
| Classification Panel:        | Radiology                                                                      |

# 5.3. Predicate Devices Information

Mirada RTx (K130393) & Workflow Box (K181572)

{5}------------------------------------------------

## 5.4. Device Description

AutoContour consists of 3 main components:

- 1. An "agent" service designed to run on the Windows Operating System that is configured by the user to monitor a network storage location for new CT datasets that are to be automatically uploaded to:
- 2. A cloud-based AutoContour automatic contouring service that produces initial contours and
- 3. A web application accessed via web browser which allows the user to perform registration with other image sets as well as review, edit, and export the structure set containing the contours.

#### 5.5. Indications for Use

AutoContour is intended to assist radiation treatment planners in contouring and reviewing structures within medical images in preparation for radiation therapy treatment planning.

#### 5.6. Technological Characteristics

AutoContour (Subject Device) makes use of a Predicate Devices, Mirada RTx (K130393) & Workflow Box (K181572) for substantial equivalence comparison. Note that Workflow Box utilizes Mirada RTx as its predicate device.

## 5.6.1. AutoContour vs. Mirada RTx (K130393) & Workflow Box (K181572)

RTx is "intended to display and visualize 2D & 3D multi-modal medical image data" and "supports the loading and saving of DICOM RT objects and allows the user to define, import, display, transform, store and export such objects including regions of interest structure and dose volumes to radiation therapy planning systems" (https://www.accessdata.fda.gov/cdrh docs/pdf13/K130393.pdf, accessed 1/20/2020).

Workflow Box "is a system designed to allow users to route DICOM-compliant data to and from automated processing components... [including] processing components for automatically contouring imaging data using … machine learning based algorithms."

(https://www.accessdata.fda.gov/cdrh docs/pdf18/K181572.pdf, accessed 1/20/2020)

| Table 3: Substantial Equivalence AutoContour vs. RTx & Workflow Box                 |                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                        |
|-------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic                                                                      | Subject Device:<br>AutoContour<br>Radformation                                                                                                                                                                                                                                                                                                                             | Predicate Device:<br>Workflow Box (K181572)                                                                                                                                                                                                                | Predicate Device: Mirada<br>RTx (K130393)                                                                                                                                                                                                                                                                                                              |
| Target<br>Population                                                                | Any patient type for<br>whom relevant modality<br>scan data is available.                                                                                                                                                                                                                                                                                                  | Any patient type for<br>whom relevant<br>modality scan data is                                                                                                                                                                                             | Any patient type for whom<br>relevant modality scan data<br>is available.                                                                                                                                                                                                                                                                              |
|                                                                                     | (Substantially<br>Equivalent)                                                                                                                                                                                                                                                                                                                                              | available.                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                        |
| Energy Used<br>and/or<br>Delivered                                                  | None - software only<br>application. The software<br>application does not<br>deliver or depend on<br>energy delivered to or<br>from patients<br>(Substantially<br>Equivalent)                                                                                                                                                                                              | None - software only<br>application. The<br>software application<br>does not deliver or<br>                                                                                                                                                                | None - software only<br>application. The software<br>application does not<br>deliver or depend on<br>energy delivered to or from<br>patients                                                                                                                                                                                                           |
| Intended users                                                                      | Trained radiation<br>oncology personnel<br>(Substantially<br>Equivalent)                                                                                                                                                                                                                                                                                                   | Designed to be used by<br>trained clinicians                                                                                                                                                                                                               | Designed to be used by<br>trained clinicians                                                                                                                                                                                                                                                                                                           |
| Design: Data<br>Visualisation/Gr<br>aphical User<br>Interface                       | Contains both an<br>automated processing<br>component and Data<br>Visualisation / Graphical<br>User Interface<br>(Substantially<br>Equivalent)                                                                                                                                                                                                                             | None - the proposed<br>device has no data<br>visualization<br>functionality. All data<br>processing is<br>automated and does<br>not require user<br>interaction. A control<br>interface is provided for<br>system administration<br>and configuration only | Yes.                                                                                                                                                                                                                                                                                                                                                   |
| Design: View<br>manipulation<br>and Volume<br>rendering                             | Window and level, pan,<br>zoom, cross-hairs, slice<br>navigation, fused views.<br>(Subject Device<br>functionality is a subset<br>of Predicate Devices)                                                                                                                                                                                                                    | None - Not applicable                                                                                                                                                                                                                                      | Window and level, pan,<br>zoom, cross-hairs, slice<br>navigation. Maximum or<br>minimum intensity projection<br>(MIP), volume rendering,<br>color rendering, surface<br>rendering, multi-planar<br>reconstruction (MPR), fused<br>views, gallery views.                                                                                                |
| Design: Image<br>registration                                                       | Manual Rigid<br>registration.<br>(Subject Device<br>functionality is a subset<br>of Predicate Devices)                                                                                                                                                                                                                                                                     | Registration for the<br>purposes of<br>replanning/re-contouring<br>and atlas based<br>contouring. The<br>algorithms used for<br>image registration are<br>the same for both<br>predicate and proposed<br>devices.                                          | Manual and Landmark<br>Rigid. Automatic multi-modal<br>rigid. Mono-modal and<br>multi-modal deformable<br>registration. Motion correction<br>in hybrid scans and gated<br>scans. Registration for the<br>purposes of<br>re-planning/recontouring and<br>atlas based contouring.                                                                        |
| Regions and<br>Volumes of<br>interest (ROI)                                         | Machine learning based<br>contouring and manual<br>ROI manipulation.<br>(Subject Device                                                                                                                                                                                                                                                                                    | Atlas Based contouring,<br>registration based<br>re-contouring, machine                                                                                                                                                                                    | 2D and 3D ROIs,<br>semi-automatic ROI<br>definition, isocontour ROIs<br>using threshold and                                                                                                                                                                                                                                                            |
|                                                                                     | functionality is a subset<br>of Predicate Devices)                                                                                                                                                                                                                                                                                                                         | learning based<br>contouring                                                                                                                                                                                                                               | percentage of maximum,<br>one-click seed-pointing<br>contouring, <b>manual ROI</b><br><b>manipulation</b> , <b>ROI</b><br>transformation, Atlas-based<br>contouring.                                                                                                                                                                                   |
| Design:<br>Region/volume<br>of interest<br>measurements<br>and size<br>measurements | None - not applicable<br>(Substantially<br>Equivalent)                                                                                                                                                                                                                                                                                                                     | None - not applicable                                                                                                                                                                                                                                      | Intensity, Hounsfield units,<br>activity and SUV<br>measurements including min,<br>max, mean, peak, standard<br>deviation, total glycolytic<br>activity, median, histogram,<br>max and mean ratio to<br>reference region. Gray for RT<br>Dose. Size measurements<br>include 2D and 3D<br>measurements including<br>rulers and volume, line<br>profile. |
| Design:<br>Region/Volume<br>Quantification                                          | None - not applicable<br>(Substantially<br>Equivalent)                                                                                                                                                                                                                                                                                                                     | None - not applicable                                                                                                                                                                                                                                      | Regions table with charting<br>supports analysis of<br>measurement over multiple<br>studies using standard<br>protocols such as RECIST,<br>PERCIST and WHO                                                                                                                                                                                             |
| Design:<br>Supported<br>modalities                                                  | CT input for contouring<br>or manual<br>registration/fusion.<br>MR, PET input for<br>manual<br>registration/fusion only.<br>DICOM RTSTRUCT for<br>output<br>(Minor differences)                                                                                                                                                                                            | <b>CT, MR, DICOM</b><br><b>RTSTRUCT</b> for image<br>processing. Any valid<br>DICOM data for data<br>routing                                                                                                                                               | Static and gated <b>CT and</b><br><b>PET</b> , and static MR, SPECT,<br>NM, DICOM RT                                                                                                                                                                                                                                                                   |
| Design:<br>Reporting and<br>data routing                                            | No built-in reporting,<br>supports exporting<br>DICOM RTSTRUCT file.<br>(Minor differences)                                                                                                                                                                                                                                                                                | Supports routing and<br>distribution of images to<br>other DICOM nodes<br>including to custom<br>executables determined<br>by the user.                                                                                                                    | Yes- Distribution of DICOM<br>compliant Images into other<br>DICOM compliant systems.<br>Built-in basic reporting                                                                                                                                                                                                                                      |
| Compatibility<br>with the<br>environment<br>and other<br>devices                    | Compatible with data<br>from any DICOM<br>compliant scanners for<br>the applicable modalities.<br>Agent Uploader<br>component compatible<br>with Microsoft Windows.                                                                                                                                                                                                        | Compatible with data<br>from any DICOM<br>compliant scanners for<br>the applicable<br>modalities. Compatible<br>with <b>Microsoft</b><br><b>Windows</b> . Integration<br>with Mirada DBx                                                                   | Compatible with data from<br>any DICOM compliant<br>scanners for the applicable<br>modalities. Compatible<br>with <b>Microsoft Windows</b> .<br>Integration with Mirada DBx<br>application launcher and data<br>browser.                                                                                                                               |
|                                                                                     | Cloud-based automatic<br>contouring service<br>compatible with Linux.<br><br>Web application Server<br>based application<br>compatible with Linux<br>with frontend compatible<br>with all modern web<br>browsers..<br>(Minor differences)                                                                                                                                  | application launcher and<br>data browser                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                        |
| Communication<br>s/Networking                                                       | TCP/IP<br>(Subject Device<br>functionality is a subset<br>of Predicate Devices)                                                                                                                                                                                                                                                                                            | TCP/IP and SCP                                                                                                                                                                                                                                             | TCP/IP and SCP                                                                                                                                                                                                                                                                                                                                         |
| Computer<br>platform &<br>operating<br>system                                       | Agent Uploader<br>supporting Microsoft<br>Windows 10 (64-bit) and<br>Microsoft Windows<br>Server 2016.<br><br>Cloud-based Server<br>based automatic<br>contouring application<br>compatible with Linux.<br><br>Web application Server<br>based application<br>compatible with Linux<br>with frontend compatible<br>with all modern web<br>browsers.<br>(Minor differences) | Server based application<br>supporting Microsoft<br>Windows 10 (64-bit) and<br>Microsoft Windows<br>Server 2016.                                                                                                                                           | Workstation and Server<br>based application supporting<br>Windows Server 2008 R2,<br>SP1 and Windows 7 (64-bit)                                                                                                                                                                                                                                        |
| Support for<br>radiation<br>treatment<br>planning                                   | AutoContour is intended<br>to assist radiation<br>treatment planners in<br>contouring and reviewing<br>structures within medical<br>images in preparation for<br>radiation therapy<br>treatment planning<br>(Subject Device intended<br>use is a subset of<br>predicate devices)                                                                                           | "Contours generated by<br>Workflow Box may be<br>used as an input to<br>clinical workflows<br>including, but not limited<br>to, radiation therapy<br>treatment planning"                                                                                   | "RTx supports the loading<br>and saving of DICOM RT<br>objects and allows the user to<br>define, import, display,<br>transform, store and export<br>such objects including<br>regions of interest structures<br>and dose volumes to<br>radiation therapy planning<br>systems."                                                                         |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Source for RTx & Workflow Box: (https://www.accessdata.fda.gov/cdrh_docs/pdf18/K181572.pdf, accessed 1/20/2020)

> As shown in Table 3, almost all technological characteristics are either substantially equivalent or a subset of the predicate devices' technological characteristics.

{9}------------------------------------------------

#### 5.7. Discussion of differences

#### Subset of the Predicate Device

The comparison table above shows that several features of AutoContour are a subset of the features of the predicate device. This is because AutoContour's indication for use is narrower in scope than the predicate device in that it is intended only "to assist radiation treatment planners in contouring and reviewing structures within medical images in preparation for radiation therapy treatment planning". The design of AutoContour reflects that narrower scope and the features not shared between AutoContour are such that they can operate independently of the features that are not shared with the predicate devices, and therefore these differences do not create new questions regarding the safety and effectiveness of the device relative to the predicate devices.

#### Minor differences

The following minor differences exist, but do not represent any significant additional risks or decreased effectiveness for the device for its intended use:

- . Design: Supported modalities: AutoContour only supports CT image sets as inputs specifically for the contouring feature, while MR and PET image set inputs are only supported for registration purposes. In the predicate devices, all modalities are available for the workflows. This does not affect the safety and effectiveness of the device as it relates to its indications for use as the vast majority of radiation treatment planning is performed on CT images, and where MR or PET modalities are necessary for localization, AutoContour does support manual registration with those modalities which can be used to guide the contouring of the CT image set.
- . Design: Reporting and data routing: AutoContour does not include reporting features or direct data-routing with DICOM compliant systems instead AutoContour simply produces a DICOM compliant RT Structure Set file that the user may import into a DICOM compliant treatment planning system. This does not raise new questions regarding the safety and effectiveness of the device since the scope of AutoContour's indication for use is limited to making the contoured structures available for radiation treatment planning purposes and does not include managing or routing of the image data to various systems. The feature of managing or routing of the image data in this way is not a necessary component to achieve this goal, so not having this feature does not affect the safety or effectiveness of the device.
- Compatibility with the environment and other devices / Computer platform & operating system: AutoContour utilizes different computer platform & operating system for its 3 components, but this does not raise new questions regarding the safety and effectiveness of the device relative to the predicate device.

{10}------------------------------------------------

### 5.7. Performance Data

As with the Predicate Devices, no clinical trials were performed for AutoContour. Nonclinical tests were performed according to Radformation's AutoContour Complete Test Protocol and Report, which demonstrates that AutoContour performs as intended per its indications for use. Further tests were performed on independent datasets from those included in training and validation sets in order to validate the generalizability of the machine learning model.

Verification and validation tests were performed to ensure that the software works as intended and pass/fail criteria were used to verify requirements.

#### 5.8. Conclusion

AutoContour is deemed substantially equivalent to the Predicate Devices, Mirada RTx (K130393) & Workflow Box (K181572). Verification and Validation testing and Hazard Analysis demonstrate that AutoContour is as safe and effective as the Predicate Devices. The minor technological differences between AutoContour and the Predicate Devices with regard to the intended use do not raise any questions on the safety and effectiveness of the Subject Device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QKB/K200323](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/QKB/K200323)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
